Health Canada
www.hc-sc.gc.ca
Home
> Drugs & Health Products
> Compliance & Enforcement
> Recalls
> Recalls
Institutional links
-
Back to
-
Explore...
-
Proactive Disclosure
Medical Device Recall Listings - Sorted by Recall Posting Date
(From July 2005 - September 2005)
- Recall Posting Date
- 2005-09-19
- Trade Name
- A) ACL Futur A/ACLl Advance
B) ACL Top
- Manufacturer
- Instrumentation Laboratory Co.
- Recall Start Date
- 2005-05-19
- Recall Number
- 28468
- Hazard Classification
- Type II
- Model or Catalog #
- A) 280000, 999001
B) N/A
- Lot or Serial #
- A)Serial number starting with 05041164, Serial number with 0505354.
B) N/A
- Reason for Recall
- Patient prothrombin time (PT) may report low for individuals on anticoagulent therapy.
- Recall Posting Date
- 2005-09-19
- Trade Name
- CoolGard 3000
- Manufacturer
- Alsius Corporation
- Recall Start Date
- 2005-08-02
- Recall Number
- 28831
- Hazard Classification
- Type I
- Model or Catalog #
- N
- Lot or Serial #
- 465131
- Reason for Recall
- Coolgard 3000 units manufactured before April 12th, 2005, may overheat their coolant if left in a warm room, for prolonged periods in standby mode. At 44 c. Alarm will trigger and system latches.
- Recall Posting Date
- 2005-09-19
- Trade Name
- Galileo Immuno Haematology System
- Manufacturer
- Immucor Inc.
- Recall Start Date
- 2005-06-30
- Recall Number
- 28664
- Hazard Classification
- Type II
- Model or Catalog #
- 64999
- Lot or Serial #
- N
- Reason for Recall
- The immediate spin crossmatch (IS_XM) assay performed on the Galileo may not give the expected incompatible result.
- Recall Posting Date
- 2005-09-19
- Trade Name
- Gemini GXL 6
- Manufacturer
- Philips Medical Systems (Cleveland), Inc.
- Recall Start Date
- 2005-08-25
- Recall Number
- 28819
- Hazard Classification
- Type II
- Model or Catalog #
- 4535 679 72021
- Lot or Serial #
- 4004
- Reason for Recall
- An anomaly was identified in the gemini ct image reconstruction subsystem of the Gemini GXL Pet
- Recall Posting Date
- 2005-09-19
- Trade Name
- Pyramidal Tip Obturator
- Manufacturer
- Keir Surgical Ltd.
- Recall Start Date
- 2005-08-09
- Recall Number
- 28751
- Hazard Classification
- Type I
- Model or Catalog #
- KS81-1055
- Lot or Serial #
- KS81-1055
- Reason for Recall
- Distal tip of obturator, lot# 10123, separated from the proximal shaft. A recent affected device was found after initial company recall done on April 2002.
- Recall Posting Date
- 2005-09-19
- Trade Name
- Synchron CRP
- Manufacturer
- Beckman Coulter Inc.
- Recall Start Date
- 2005-07-18
- Recall Number
- 28832
- Hazard Classification
- Type II
- Model or Catalog #
- N
- Lot or Serial #
- N
- Reason for Recall
- Synchron CRP turbidimetric assay may not be sensitive or specific enough as a sole diagnostic marker of infection in neonates.
- Recall Posting Date
- 2005-09-05
- Trade Name
- 3M Microfoam Surgical Tape
- Manufacturer
- 3M Health Care
- Recall Start Date
- 2005-08-08
- Recall Number
- 28764
- Hazard Classification
- Type III
- Model or Catalog #
- 1528-1, 1528-2, 1528-3, 1528-4
- Lot or Serial #
- 2010-06 (HE, HR, HS, HU, IM, IN)
- Reason for Recall
- The tape may not perform as intended in some situations as portions of the tape may not appear to have adhesive or the adhesive may separate from the foam backing.
- Recall Posting Date
- 2005-09-05
- Trade Name
- A) Access Immunoassay Sys: Dil-AFP
B) Access Immunoassay Sys: Dil-AFP
- Manufacturer
- Beckman Coulter Inc.
- Recall Start Date
- 2005-07-21
- Recall Number
- 28789
- Hazard Classification
- Type III
- Model or Catalog #
- A) 33211
B) 33210
- Lot or Serial #
- A) All lots
B) All lots
- Reason for Recall
- Potential splashing may cause erroneous results.
- Recall Posting Date
- 2005-09-05
- Trade Name
- A) Alto VR625 Cardioverter Defibrillator
B) Alto DR614 Cardioverter Defibrillator
C) Alto 2 DR624 Cardioverter Defibrillator
- Manufacturer
- ELA Medical, S.A.
- Recall Start Date
- 2005-07-20
- Recall Number
- 28746
- Hazard Classification
- Type I
- Model or Catalog #
- A) N/A
B) 614
C) N/A
- Lot or Serial #
- A) 304XI002, 304XI108, 402XI001, 404XI049, 413XI027, 413XI035
B) 222XF028, 222XF035, 222XF085, 222XF079, 222XF033
C) 314XJ284, 238XJ029, 403XJ002, 403XJ092, 409XJ014, 409XJ036, 410XJ019, 410XJ024, 410XJ025, 410XJ031, 410XJ042
- Reason for Recall
- Premature battery depletion and
- Recall Posting Date
- 2005-09-05
- Trade Name
- A) Arcadis Varic
B) Siremobil Compact
- Manufacturer
- Siemens Aktiengesellschaft
- Recall Start Date
- 2005-08-08
- Recall Number
- 28754
- Hazard Classification
- Type III
- Model or Catalog #
- A) 8080017
B) 3780629
- Lot or Serial #
- A) N/A
B) 10051, 10243, 10256, 1639
- Reason for Recall
- The screws fixing the laser diode out windows can become loosened.
- Recall Posting Date
- 2005-09-05
- Trade Name
- A) Immage Low Conc. IgA (IGALC) Reagent
B) Immage Low Conc. IgM (IGALC) Reagent
- Manufacturer
- Beckman Coulter Inc.
- Recall Start Date
- 2005-04-17
- Recall Number
- 28610
- Hazard Classification
- Type III
- Model or Catalog #
- A) 447460
B) 447470
- Lot or Serial #
- A) Tous les lots
B) Tous les lots
- Reason for Recall
- Reagents do not meet performance claim.
- Recall Posting Date
- 2005-09-05
- Trade Name
- A) Magnetom Avanto System
B) Manetom Espree
- Manufacturer
- Siemens Aktiengesellschaft
- Recall Start Date
- 2005-08-18
- Recall Number
- 28779
- Hazard Classification
- Type II
- Model or Catalog #
- A) 7391167
B) 7727121
- Lot or Serial #
- A) N/A
B) 25285, 25490, 25340, 25043, 30048, 25039, 25541, 25234, 25462, 25173, 25089, 25526, 25110, 25066, 25137, 25486, 25049, 25097, 25311, 25168, 25333, 25555, 25244
- Reason for Recall
- A misalignment of the patient table may occur between the patient table assembly and the opening to the magnet bore.
- Recall Posting Date
- 2005-09-05
- Trade Name
- A) Plum A+ Infusion Pumps
B) Plum A+3 Infusion Pumps
- Manufacturer
- Abbott Laboratories
- Recall Start Date
- 2005-06-02
- Recall Number
- 28438
- Hazard Classification
- Type I
- Model or Catalog #
- A) N/A
B) N/A
- Lot or Serial #
- A) Numerus lot numbers
B) Numerus lot numbers
- Reason for Recall
- If battery service warning and E320 messages are not responded to appropriately, the pump may shut off during battery operation without providing adequate response time.
- Recall Posting Date
- 2005-09-05
- Trade Name
- A) Simplastin Excel
B) Simplastin L
C) Simplastin - Excel S
- Manufacturer
- Biomerieux, Inc.
- Recall Start Date
- 2005-08-08
- Recall Number
- 28780
- Hazard Classification
- Type I
- Model or Catalog #
- A) N/A
B) N/A
C) N/A
- Lot or Serial #
- A) 161711, 161929, 161897
B) 161750
C) 161720, 161760, 161761, 161783, 161811
- Reason for Recall
- The ISI values published in the package insert of Simplastin Excel products are incorrect for the devices: fibrometer (mechanical), option, XM and MTX (photo optic).
- Recall Posting Date
- 2005-09-05
- Trade Name
- Abbott Corzyme
- Manufacturer
- Abbott Laboratories Diagnostic Division
- Recall Start Date
- 2005-04-01
- Recall Number
- 28747
- Hazard Classification
- Type II
- Model or Catalog #
- 9977-24
- Lot or Serial #
- 18488M100, 15893M100, 16792M200, 10527M200, 23326M100, 19804M100
- Reason for Recall
- A study to compare the panels with a reference serum showed a difference in pei units
- Recall Posting Date
- 2005-09-05
- Trade Name
- Access Immunoassay Unconj Estriol Calib
- Manufacturer
- Beckman Coulter Inc.
- Recall Start Date
- 2005-07-14
- Recall Number
- 28643
- Hazard Classification
- Type III
- Model or Catalog #
- 33575
- Lot or Serial #
- 416534
- Reason for Recall
- Does not meet expiration claim of 12 months.
- Recall Posting Date
- 2005-09-05
- Trade Name
- Cell-DYN 4000 System
- Manufacturer
- Abbott Laboratories Diagnostic Division
- Recall Start Date
- 2005-03-17
- Recall Number
- 28750
- Hazard Classification
- Type II
- Model or Catalog #
- 01H01-01, 01H01-03, 01H02-01, 01H02-03, 01H03-01
- Lot or Serial #
- 30268AA, 30271AA, 30455AA, 17500AA96, 30272AA, 30670AA, 287AA, 30266AA, 30403AA, 30653AA, 30654AA, 30697AA, 30416AA, 30488AA, 30665AA, 30672AA, 30615AA, 30709AA , 30712AA, 30686AA, 30683AA, 30723AA.
- Reason for Recall
- The optical platelet algorithm does not correctly gate the plto for the calibrator and the plus control and causes calibration problems and control plto results to fall outside the lower limit of assay range.
- Recall Posting Date
- 2005-09-05
- Trade Name
- I-Stat G3+ Cartridges
- Manufacturer
- I-Stat Corporation
- Recall Start Date
- 2005-08-05
- Recall Number
- 28778
- Hazard Classification
- Type III
- Model or Catalog #
- N
- Lot or Serial #
- L05048, L05088
- Reason for Recall
- Certain lots of I-stat G3+ cartridges may exhibit higher than expected rates of star-outs (***) in place of oxygen results.
- Recall Posting Date
- 2005-09-05
- Trade Name
- LH 1500 Series Lab Automation System
- Manufacturer
- Beckman Coulter Inc.
- Recall Start Date
- 2005-06-23
- Recall Number
- 28510
- Hazard Classification
- Type III
- Model or Catalog #
- 6605528
- Lot or Serial #
- N
- Reason for Recall
- Warnings in manual conflict with sections in the manual. Potential for mechanical and voltage hazards during maintenance.
- Recall Posting Date
- 2005-09-05
- Trade Name
- One Touch UltraSmart Meter
- Manufacturer
- Lifescan Inc.
- Recall Start Date
- 2005-06-15
- Recall Number
- 28489
- Hazard Classification
- Type III
- Model or Catalog #
- 011-555, 020-720
- Lot or Serial #
- All lots.
- Reason for Recall
- CORRECTION TO RECALL # 28489 PREVIOUSLY POSTED ON 2005-08-22.
Labelling incorrectly states that the 'snack' message will appear on .Meter between 1.1mmol/l to 3.9 mmol/l, labelling should state between 1mmol/l to 3.3 mmol/l.
- Recall Posting Date
- 2005-09-05
- Trade Name
- Paragon CZE 2000 Software Version 2.0
- Manufacturer
- Beckman Coulter Inc.
- Recall Start Date
- 2005-07-27
- Recall Number
- 28734
- Hazard Classification
- Type III
- Model or Catalog #
- A09744
- Lot or Serial #
- N
- Reason for Recall
- Potential for incorrect results to be displayed and printed out.
- Recall Posting Date
- 2005-09-05
- Trade Name
- Prisma Continuous Replacement System
- Manufacturer
- Gambro Dasco S.P.A.
- Recall Start Date
- 2005-08-16
- Recall Number
- 28774
- Hazard Classification
- Type I
- Model or Catalog #
- N
- Lot or Serial #
- N
- Reason for Recall
- Serious injuries or deaths have resulted because of excessive ultrafiltration. This can occur when the cause of the "incorrect weight change detected" alarm is not identified and removed by the user.
- Recall Posting Date
- 2005-09-05
- Trade Name
- Remisol 2000 Data Manager System
- Manufacturer
- Beckman Coulter Inc.
- Recall Start Date
- 2005-06-27
- Recall Number
- 28505
- Hazard Classification
- Type III
- Model or Catalog #
- A27343
- Lot or Serial #
- N
- Reason for Recall
- Results may upload with an incorrect or blank sample ID.
- Recall Posting Date
- 2005-09-05
- Trade Name
- Scissors Insert for Single Use
- Manufacturer
- Richard Wolf GMBH
- Recall Start Date
- 2005-04-29
- Recall Number
- 28206
- Hazard Classification
- Type II
- Model or Catalog #
- 8392.0403, 8393.0403, 8394.0403
- Lot or Serial #
- 953046, 9532217, 953486
- Reason for Recall
- Damage to the insulation part may arise by unintentional "jamming" of the scissor on the existing edges in the access instrument when being introduced or pulled out. Insulation material may come off and fall.
- Recall Posting Date
- 2005-09-05
- Trade Name
- StemCXP Software Version 1.0
- Manufacturer
- Beckman Coulter Inc.
- Recall Start Date
- 2005-06-09
- Recall Number
- 28471
- Hazard Classification
- Type III
- Model or Catalog #
- 628843
- Lot or Serial #
- N
- Reason for Recall
- 7HPCSA and 7HPCSA control protocols may not invoke the algorithm under certain conditions.
- Recall Posting Date
- 2005-09-05
- Trade Name
- Synchron Systems Amphetamine Reagent
- Manufacturer
- Beckman Coulter Inc.
- Recall Start Date
- 2005-07-07
- Recall Number
- 28607
- Hazard Classification
- Type III
- Model or Catalog #
- 475000
- Lot or Serial #
- M501225, M503380
- Reason for Recall
- These lots have different than expected cross reactivities. May cause false positive amphetamine results in patients taking medications containing ephedrine.
- Recall Posting Date
- 2005-09-05
- Trade Name
- Testpack Plus HCG Combo with OBC
- Manufacturer
- Abbott Laboratories Diagnostic Division
- Recall Start Date
- 2005-03-22
- Recall Number
- 28752
- Hazard Classification
- Type II
- Model or Catalog #
- 7B34-16
- Lot or Serial #
- 16006M200, 19199M300
- Reason for Recall
- Manufacturer has received user complaints reporting occurrences of false positive results or weak positive colour on the patient bar for specimens where the results disagree with other methods.
- Recall Posting Date
- 2005-08-22
- Trade Name
- A) Axiom Artis BA
B) Axiom Artis DFC
C) Axiom Artis FA ALPHA
D) Axiom Artis MP
E) Axiom Artis TA
- Manufacturer
- Siemens Aktiengesellschaft
- Recall Start Date
- 2005-07-26
- Recall Number
- 28718
- Hazard Classification
- Type III
- Model or Catalog #
- A) 5904656
B) 7412807, 7727717
C) 5904441
D) 5904466
E) 7007755
- Lot or Serial #
- A) N/A
B) 35195, 35261, 35219, 35298, 40033, 25006, 25076, 20070, 20215, 28064, 32070
C) N/A
D) N/A
E) N/A
- Reason for Recall
- The live display of the Axiom Artis in the examination room may sporadically stop working.
- Recall Posting Date
- 2005-08-22
- Trade Name
- A) CA1500 Instrument
B) CA540 Instrument
C) CA560 Instrument
- Manufacturer
- Sysmex Corporation
- Recall Start Date
- 2005-07-21
- Recall Number
- 28685
- Hazard Classification
- Type II
- Model or Catalog #
- A) CA-1500WBCR
B) B4260-540
C) B4260-560
- Lot or Serial #
- A) A2399, A2483, A2400, A3358, A2401, A2727, A1888, A3295, A3276, A3632, A2620, A2739, A1825, A2270, A2753, A2522, A1556, A3905, A2480, A4001, A2636, A3546, A2320, A1465, A1151, A2646, A4047, A4050, A3548, A2713, A2269, A2271, A3634, A1646, A2182, A2170, A3521, A4028, A2171, A3284, A1105, A1581, A2427, A2257, A2632, A2621, A1644, A2296.
B) A3616, A2858, A3359, A2530, A2578, A2353, A2536, A3250, A1307, A2384, A3637, A1449, A2993, A2954, A3729, A2352, A3406, A3249, A2005, A2996, A2246, A2854, A2308, A3273, A3274, A3687, A3360, A2841, A1452, A3482, A2859, A3407, A3403, A2474, A2351, A2953, A1309, A2027, A2248, A3636, A2997, A3602, A3477, A3600, A3599, A3345, A2861, A2309, A2995, A2532, A1308, A2531, A3446. A3417, A3548, 2026, A2759, A2003, A1454, A3479, A3297, A3321, A3320, A2952, A3361, A3362, A3394, A3478, A3447, A2004B, A2994, A3404, A1456, A3409, A2862, A2863, A2857, A2860, A3452, A3319, A3344, A3401, A3483, A3450, A3475, A2490, A1450, A2395, A2840, A2955, A3405, A3296, A2951, A3311, A3318, A3448, A3662, A3772, A1457, A3476, A3667, A3783, A3927, A2337.
C) A1323, A1413, A1088, A1821, A1336, A1633, A1276, A1315, A1102. A1850, A1858, A1105, A1510, A1395, A1870, A1866, A1856, A1846, A1869, A1474, A1103, A1113, A1839, A1108, A1290, A1861, A1265, A1278, A1672, A1630, A1198, A1770.
- Reason for Recall
- Manufacturer has found an issue when using the akima method of calculating the standard curve. A change to the standard curve method for this assay is needed.
- Recall Posting Date
- 2005-08-22
- Trade Name
- A) Homechoice Automated PD System
B) Homechoice Pro Automated PD System
- Manufacturer
- Baxter Healthcare Corporation
- Recall Start Date
- 2005-07-05
- Recall Number
- 28726
- Hazard Classification
- Type II
- Model or Catalog #
- A) 5C4471
B) 5C8310, 5C8310P, 5C8310PE
- Lot or Serial #
- A) All serial numbers.
B) All serial numbers.
- Reason for Recall
- Remote risk of electrical shock from screws used to secure the power entry module to the back of the panel, this module is the electrical connector where the power cord plugs into the unit.
- Recall Posting Date
- 2005-08-22
- Trade Name
- ACMI Snap & Peel Introducer Sheaths
- Manufacturer
- ACMI Corporation
- Recall Start Date
- 2005-06-30
- Recall Number
- 28738
- Hazard Classification
- Type III
- Model or Catalog #
- 60810BX, 61012BX
- Lot or Serial #
- 1297324D, 1312704E, 1321584E, 1325384E, 1339994F, 1344104G, 1389384K, 1394004L, 1400154M, 1419225A, 1297374D, 1312724E, 1316954E, 1321644E, 1325394E, 1336944F, 1339964F, 1344094G, 1355274H, 1350354H, 1359054H, 1364234J, 1367404J, 1372814J, 1383294K, 1412604M, 1419235A, 1427565A, 1430855B, 1441195B
- Reason for Recall
- ACMI received a report from B. Braun, manufacturer of device, that there were cracks in the hub
- Recall Posting Date
- 2005-08-22
- Trade Name
- AMVISC
- Manufacturer
- Anika Therapeutics Inc.
- Recall Start Date
- 2005-07-25
- Recall Number
- 28723
- Hazard Classification
- Type II
- Model or Catalog #
- 59051, 59081
- Lot or Serial #
- B050120B, B050510B
- Reason for Recall
- Some cannulae packaged with AMVISC and AMVISC Plus Visoelastic may have an incomplete pouch seal.
- Recall Posting Date
- 2005-08-22
- Trade Name
- Architect System Software V2.10 & 2.11
- Manufacturer
- Abbott Laboratories Diagnostic Division
- Recall Start Date
- 2005-06-23
- Recall Number
- 28679
- Hazard Classification
- Type III
- Model or Catalog #
- 1G06-01, 7-99077-01, 99077-103
- Lot or Serial #
- All serial numbers.
- Reason for Recall
- The software calculation utilizes 17ul rather than 12 ul for the alkaline washing solution dispensed, the displayed value for volume remaining will be less than the actual amount in the bottle.
- Recall Posting Date
- 2005-08-22
- Trade Name
- Argus Epic Camera
- Manufacturer
- ADAC Laboratories, Inc.
- Recall Start Date
- 2005-07-13
- Recall Number
- 28730
- Hazard Classification
- Type II
- Model or Catalog #
- 2145-3007A
- Lot or Serial #
- A0105217, 01548B, 03338B, 00000000009701075, 00964B, 00227BC, 01922B, 01920B, 01921B, 00000000009901158, 00000000009901157, 00709B, 00000000009908167, 00000000009801107, 01525B, A03030282, 01332B, 03041B, 01333B, 03094B, 01310B, 03368B
- Reason for Recall
- Fracture was discovered on an argus fork which holds the detector.
- Recall Posting Date
- 2005-08-22
- Trade Name
- D-TRON Plus Adapter
- Manufacturer
- Disetronic Medical Systems AG
- Recall Start Date
- 2005-07-16
- Recall Number
- 28683
- Hazard Classification
- Type I
- Model or Catalog #
- 3000803
- Lot or Serial #
- N/A
- Reason for Recall
- Incorrect function may occur within 15 min of replacing the adapter and priming the set. The drop in pressure may result in the delivery of up to 1.8 units of insulin which affects users 13 years and under.
- Recall Posting Date
- 2005-08-22
- Trade Name
- Leveen Coaccess Electrode
- Manufacturer
- Radiotherapeutics Corporation
- Recall Start Date
- 2005-08-02
- Recall Number
- 28745
- Hazard Classification
- Type II
- Model or Catalog #
- 26-216, 26-217
- Lot or Serial #
- 7108067, 6556464, 6826039, 6666258, 6556462, 7356251, 31804, 7021736, 7091462, 21403R, 6558336, 6538341, 6556258, 6538336, 7257089, 7481332
- Reason for Recall
- Manufacturer has received reports of extended ablation times, lack of roll-off, and
- Recall Posting Date
- 2005-08-22
- Trade Name
- One Touch UltraSmart Meter
- Manufacturer
- Lifescan Inc.
- Recall Start Date
- 2005-06-15
- Recall Number
- 28489
- Hazard Classification
- Type III
- Model or Catalog #
- 011-555, 020-720
- Lot or Serial #
- All lots.
- Reason for Recall
- Glucose meter does not prompt user to take a snack at glucose levels over the entire range of 1.1MMOL
- Recall Posting Date
- 2005-08-22
- Trade Name
- STAAR Surgical STAARVISC II
- Manufacturer
- STAAR Surgical Company
- Recall Start Date
- 2005-07-18
- Recall Number
- 28728
- Hazard Classification
- Type II
- Model or Catalog #
- N
- Lot or Serial #
- B041118C
- Reason for Recall
- The cannula package has not been validated to a sterility assurance level (sal) of 0.000001 and therefore the sterility can not be assured.
- Recall Posting Date
- 2005-08-22
- Trade Name
- STAGO ATA R System
- Manufacturer
- Diagnostica Stago
- Recall Start Date
- 2005-06-08
- Recall Number
- 28662
- Hazard Classification
- Type III
- Model or Catalog #
- 57160
- Lot or Serial #
- All serial numbers.
- Reason for Recall
- Manufacturer determined there is potential defect relating to change of disposable cuvette roll change, filled cuvettes may wait on the conveyer belt prolonging contact time with plasma & intermediate regent.
- Recall Posting Date
- 2005-08-08
- Trade Name
- A) AXSYM Total B-HCG Reagent
B) IMX Total B-HCG Reagent
- Manufacturer
- Abbott Laboratories Diagnostic Division
- Recall Start Date
- 2005-04-05
- Recall Number
- 28629
- Hazard Classification
- Type III
- Model or Catalog #
- A) 1A06-22
B) 7A59-22
- Lot or Serial #
- A) 26148Q100, 26278Q100, 16381Q100
B) 25459Q100
- Reason for Recall
- Tests have shown that protein aggregates are present in the specimen diluent and causes calibration errors and a shift in control values and no valid patient results could be generated.
- Recall Posting Date
- 2005-08-08
- Trade Name
- A) Brilliance CT 16 Slice
B) Brilliance CT 16 Power
C) Brilliance CT 40 Channel
- Manufacturer
- Philips Medical Systems (Cleveland), Inc.
- Recall Start Date
- 2005-07-18
- Recall Number
- 28640
- Hazard Classification
- Type II
- Model or Catalog #
- A) 453567078851
B) 453567005721
C) 455011004011
- Lot or Serial #
- A) 5074
B) 6110
C) 9081, 6072
- Reason for Recall
- Couch positions that are reported on the gantry control panel are correct during a CCT procedure, but the couch positions that are displayed on the monitor may be incorrect.
- Recall Posting Date
- 2005-08-08
- Trade Name
- A) Colleague 3 Volumetric Infusion Pump
B) Colleague Volumetric Infusion Pump
- Manufacturer
- Baxter Healthcare Corporation
- Recall Start Date
- 2005-07-21
- Recall Number
- 28660
- Hazard Classification
- Type I
- Model or Catalog #
- A) 2M8153, 2M8163, DNM8153
B) 2M8151, 2M8161, DNM8151
- Lot or Serial #
- A) All Serial Numbers
B) All Serial Numbers
- Reason for Recall
- Design problem with the clocking circuit that can disrupt internal communications causing the pumps to shut down with certain error codes and disrupt critical medication infusions. Death has been involved.
- Recall Posting Date
- 2005-08-08
- Trade Name
- A) Contak Renewal 4 AVT
B) Contak Renewal 4 AVT HE
C) Ventak Prizm AVT
D) Vitality AVT AICD
- Manufacturer
- Cardiac Pacemakers Inc. A Wholly Owned Subsidiary of Guidant
- Recall Start Date
- 2005-07-22
- Recall Number
- 28665
- Hazard Classification
- Type II
- Model or Catalog #
- A) M170, M175
B) M177
C) 1900
D) A155
- Lot or Serial #
- A) N/A
B) N/A
C) N/A
D) N/A
- Reason for Recall
- One of the original recommendations to recall #28528 - programming atrial tachy episode data storage to 0% - can cause latching in a subset of AVT devices that have previously stored atrial episode data.
- Recall Posting Date
- 2005-08-08
- Trade Name
- A) Contak TR CHFD
B) CPI Discovery DR Pacemaker
C) CPI Discovery SR Pacemaker
D) CPI Meridian DDD Pacemaker
E) CPI Meridian DR Pacemaker
F) CPI Meridian SR Pacemaker
G) CPI Meridian SSI Pacemaker
H) CPI Pulsar DDD, Model 970 Pacemaker
I) CPI Pulsar DDD, Model 972 Pacemaker
J) CPI Pulsar DR Pacemaker
K) CPI Pulsar Max DR Pacemaker
L) CPI Pulsar Max SR, Model 1170 Pacemaker
M) CPI Pulsar Max SR, Model 1171 Pacemaker
N) CPI Pulsar SSI Pacemaker
O) CPI Pulsar VDD Pacemaker
P) Discovery II DDD
Q) Discovery II DR
R) Discovery II SR
S) Discovery II SSI
T) Pulsar Max II Pacemaker DDDR
U) Pulsar Max II Pacemaker SSIR
W) CPI Pulsar SR Pacemaker
- Manufacturer
- Cardiac Pacemakers Inc. A Wholly Owned Subsidiary of Guidant
- Recall Start Date
- 2005-07-19
- Recall Number
- 28650
- Hazard Classification
- Type I
- Model or Catalog #
- A) 1241
B) 1273, 1274, 1275
C) 1174, 1175
D) 976
E) 1276
F) 1176
G) 476
H) 970
I) 972
J) 1272
K) 1270
L) 1170
M) 1171
N) 470
O) 870
P) 981
Q) 1283, 1284, 1285, 1286
R) 1184, 1186, 1187
S) 481
T) 1280
U) 1180, 1181
W) 1172
- Lot or Serial #
- A) N/A
B) N/A
C) N/A
D) N/A
E) N/A
F) N/A
G) N/A
H) N/A
I) N/A
J) N/A
K) N/A
L) N/A
M) N/A
N) N/A
O) N/A
P) N/A
Q) N/A
R) N/A
S) N/A
T) N/A
U) N/A
W) N/A
- Reason for Recall
- Hermetic sealing component may experience a gradual degradation, resulting in a higher than normal moisture content within the pacemaker Case late in the device's service life.
- Recall Posting Date
- 2005-08-08
- Trade Name
- Access Immuno Sys Unconj Estriol Calib
- Manufacturer
- Beckman Coulter, Inc.
- Recall Start Date
- 2005-07-14
- Recall Number
- 28643
- Hazard Classification
- Type III
- Model or Catalog #
- 33575
- Lot or Serial #
- 416534
- Reason for Recall
- Device does not meet expiration claim of 12 months.
- Recall Posting Date
- 2005-08-08
- Trade Name
- Axym Anti-TPO Reagent
- Manufacturer
- Abbott Laboratories Diagnostic Division
- Recall Start Date
- 2005-06-28
- Recall Number
- 28675
- Hazard Classification
- Type III
- Model or Catalog #
- 5F57-20
- Lot or Serial #
- 22827HN00
- Reason for Recall
- Manufacturer has identified a manufacturing error at a contract facility in which a different reagent was mistakenly packed in its place for a subset of this reagent lot.
- Recall Posting Date
- 2005-08-08
- Trade Name
- BD DTXPlus Disposable Transducer Kit
- Manufacturer
- Becton Dickinson
- Recall Start Date
- 2005-06-29
- Recall Number
- 28624
- Hazard Classification
- Type II
- Model or Catalog #
- 682000, 682034
- Lot or Serial #
- 407032, 407104
- Reason for Recall
- Device (stopcock tap) may easily detach from the stopcock and cause leakage of the iv fluids and
- Recall Posting Date
- 2005-08-08
- Trade Name
- Biograph Duo LSO
- Manufacturer
- CTI Pet Systems Inc.
- Recall Start Date
- 2005-07-12
- Recall Number
- 28625
- Hazard Classification
- Type III
- Model or Catalog #
- 8415957
- Lot or Serial #
- 5408-0301009
- Reason for Recall
- Under certain circumstances the system will continue to scan after an IRS function failure occurs. The system will scan only until the end of the planned sequence and then will shut off.
- Recall Posting Date
- 2005-08-08
- Trade Name
- Cell DYN 1800 System
- Manufacturer
- Abbott Laboratories Diagnostic Division
- Recall Start Date
- 2005-05-18
- Recall Number
- 28635
- Hazard Classification
- Type III
- Model or Catalog #
- 07H77-01
- Lot or Serial #
- 19017AY
- Reason for Recall
- The HP6122 printer will print one page correctly but may have problems printing more than one page from the data log or the Qc log when "print data log" and "print Qc log" commands are given.
- Recall Posting Date
- 2005-08-08
- Trade Name
- Clotting Factor 8 Deficient Plasma
- Manufacturer
- Dade Behring Marburg GMBH
- Recall Start Date
- 2005-07-01
- Recall Number
- 28642
- Hazard Classification
- Type II
- Model or Catalog #
- OTXW17
- Lot or Serial #
- 503848A, 503849E, 503850C, 503856B, 503858A, 503860D,
- Reason for Recall
- Manufacturer has determined that these lots do not demonstrate acceptable performance and patient samples may exhibit falsely high factor viii results, could lead to misdiagnosis.
- Recall Posting Date
- 2005-08-08
- Trade Name
- Dell AC Adapter
- Manufacturer
- Sensormedics Corporation
- Recall Start Date
- 2005-03-11
- Recall Number
- 28669
- Hazard Classification
- Type III
- Model or Catalog #
- 22536, 22538
- Lot or Serial #
- N
- Reason for Recall
- Dell recalled the adapter units due to a risk of overheating which could lead to fire and electrical shock. The adapters are used in conjunction with laptops-peripheral components of vmax and sleep systems.
- Recall Posting Date
- 2005-08-08
- Trade Name
- D-TRON Plus Adapter
- Manufacturer
- Disetronic Medical Systems AG
- Recall Start Date
- 2005-07-06
- Recall Number
- 28686
- Hazard Classification
- Type II
- Model or Catalog #
- 3000803
- Lot or Serial #
- part # 3000803
- Reason for Recall
- Small percentage of adapters may drop in pressure which may result in the delivery of up to 1.8 units of insulin. This may occur within 15 minutes of replacing the adapter and priming the set.
- Recall Posting Date
- 2005-08-08
- Trade Name
- Heartstring II Proximal Seal System
- Manufacturer
- Guidant Corporation Cardiac Surgery
- Recall Start Date
- 2005-07-07
- Recall Number
- 28623
- Hazard Classification
- Type II
- Model or Catalog #
- N
- Lot or Serial #
- N
- Reason for Recall
- Failure modes include incomplete or no Aortotomy with use of device on unaltered tissue, or incomplete Aortotomy when device is used on altered tissue (e.g. Cardioplegia hole or Aortotomy incision).
- Recall Posting Date
- 2005-08-08
- Trade Name
- MISYS Laboratory System Blood Bank
- Manufacturer
- MISYS Healthcare Systems Inc.
- Recall Start Date
- 2005-05-10
- Recall Number
- 28492
- Hazard Classification
- Type III
- Model or Catalog #
- N
- Lot or Serial #
- Sites using software ver. 6.1
- Reason for Recall
- CORRECTION TO RECALL # 28492 PREVIOUSLY POSTED ON 2005-07-11.
Information entered into the comment field may be lost when resulting the blood type for the first time via a blood bank instrument interface. The software problem will be corrected with the installation of a software patch.
- Recall Posting Date
- 2005-08-08
- Trade Name
- Multileaf Collimater (MLC)+Software
- Manufacturer
- Varian Medical Systems, Inc.
- Recall Start Date
- 2005-07-01
- Recall Number
- 28682
- Hazard Classification
- Type II
- Model or Catalog #
- N
- Lot or Serial #
- H520003-H520943
H530012-H534223
H540381-H549004
- Reason for Recall
- Potential mistreatment may occur because initial MLC shape may be an open field and not match planned MLC field when executing dynamic MLC treatment types.
- Recall Posting Date
- 2005-08-08
- Trade Name
- Prismaflex Software 1.07.3 KJ
- Manufacturer
- Gambro Dasco S.P.A.
- Recall Start Date
- 2005-05-27
- Recall Number
- 28506
- Hazard Classification
- Type II
- Model or Catalog #
- 6023014700
- Lot or Serial #
- All serial numbers
- Reason for Recall
- Pumps may stop while green light is on with no audible alarm. No hemodynamic problems but access eventually clot as well as lost blood in set.
- Recall Posting Date
- 2005-08-08
- Trade Name
- Synchron Clinical Sys. Alcohol Reagent
- Manufacturer
- Beckman Coulter, Inc.
- Recall Start Date
- 2005-05-17
- Recall Number
- 28371
- Hazard Classification
- Type III
- Model or Catalog #
- 91DIC
- Lot or Serial #
- UPDATED RECALL # 28371 PREVIOUSLY POSTED ON 2005-06-27.
M405028, M409024, M407255, M412090, M503108
- Reason for Recall
- Reagent is past expiry date thus will not produce results since it cannot calibrate successfully.
- Recall Posting Date
- 2005-08-08
- Trade Name
- TDX
- Manufacturer
- Abbott Laboratories Diagnostic Division
- Recall Start Date
- 2005-04-05
- Recall Number
- 28653
- Hazard Classification
- Type III
- Model or Catalog #
- 9797-02, 9797-12
- Lot or Serial #
- 18229Q100, 18230Q100
- Reason for Recall
- Manufacturer has discovered a change in performance in which patient results are depressed 5-8%, elevated values when running a calibration curve.
- Recall Posting Date
- 2005-07-25
- Trade Name
- A) Contak Renewal 2
B)Contak Renewal
- Manufacturer
- Cardiac Pacemakers Inc. A Wholly Owned Subsidiary of Guidant
- Recall Start Date
- 2005-06-21
- Recall Number
- 28519
- Hazard Classification
- Type II
- Model or Catalog #
- A) H135
B) H155
- Lot or Serial #
- A) N/A
B) N/A
- Reason for Recall
- Deterioration in a wire insulator within the lead connector block, could cause a short circuit and loss of device function due to diversion of therapy energy away from the heart and into device circuitry.
- Recall Posting Date
- 2005-07-25
- Trade Name
- A) Contak Renewal 4 AVT
B) Contak Renewal 4 AVT HE
C) Contak Renewal 4
D) Contak Renewal 4 HE
- Manufacturer
- Cardiac Pacemakers Inc. A Wholly Owned Subsidiary of Guidant
- Recall Start Date
- 2005-06-24
- Recall Number
- 28531
- Hazard Classification
- Type II
- Model or Catalog #
- A) M170, M175
B) M177, M179
C) H190, H195
D) H197, H199
- Lot or Serial #
- A) N/A
B) N/A
C) N/A
D) N/A
- Reason for Recall
- Component failure may limit available tachyarrhythmia therapy. The magnetic switch might stick in the closed position.
- Recall Posting Date
- 2005-07-25
- Trade Name
- A) Contak Renewal 4 AVT
B) Contak Renewal 4 AVT HE
C) Ventak Prizm AVT
D) Vitality AVT
- Manufacturer
- Cardiac Pacemakers Inc. A Wholly Owned Subsidiary of Guidant
- Recall Start Date
- 2005-06-21
- Recall Number
- 28528
- Hazard Classification
- Type II
- Model or Catalog #
- A) M170, M175
B) M177, M179
C) 1900
D) A135, A155
- Lot or Serial #
- A) N/A
B) N/A
C) N/A
D) N/A
- Reason for Recall
- Random memory error causes functional "latching" that limits available therapy. Latching can only occur when storing tachy detection
- Recall Posting Date
- 2005-07-25
- Trade Name
- Axiom Multistar
- Manufacturer
- Siemens Aktiengesellschaft
- Recall Start Date
- 2005-06-17
- Recall Number
- 28539
- Hazard Classification
- Type III
- Model or Catalog #
- 3772501, 3773004
- Lot or Serial #
- 1027, 1094, 1142, 1163, 1229, 1290, 1795, 1818, 1102, 1112, 1113, 1129
- Reason for Recall
- A system error may occur which results in unintended movement during examination.
- Recall Posting Date
- 2005-07-25
- Trade Name
- Bd Unopette - RBC Osmotic Fragility Determination for Manual Methods
- Manufacturer
- Becton Dickinson
- Recall Start Date
- 2005-06-27
- Recall Number
- 28536
- Hazard Classification
- Type II
- Model or Catalog #
- 365830
- Lot or Serial #
- 5014328
- Reason for Recall
- Lot may have a problem that could cause an increased susceptibility for lysis of red cells from normal subjects. Yields uninterpretable results and causing the test results to not be generated.
- Recall Posting Date
- 2005-07-25
- Trade Name
- Behring Nephelometer Pro Spec Analyzer
- Manufacturer
- Dade Behring Marburg GMBH
- Recall Start Date
- 2005-06-24
- Recall Number
- 28589
- Hazard Classification
- Type III
- Model or Catalog #
- 400252.1003
- Lot or Serial #
- 420926, 341237, 241226, 141169, 331087, 490129, 100215, 410745, 410737, 120753, 310622, 410720, 300392, 310624, 310667, 210615, 210582, 210579, 210613A, 110540, 210577, 210545, 110539, 400476,
- Reason for Recall
- Manufacturer has received reports that flags "l" and "h" indicating results below and above the reference range may not be correct for formula assays.
- Recall Posting Date
- 2005-07-25
- Trade Name
- Bio-intact PTH (I-84) Assay
- Manufacturer
- Nichols Institute Diagnostics
- Recall Start Date
- 2005-03-25
- Recall Number
- 28457
- Hazard Classification
- Type II
- Model or Catalog #
- 62-7040
- Lot or Serial #
- 62-402622
- Reason for Recall
- Internal testing indicates that the assay does not meet the following performance specifications stated in the directional insert functional sensitivity, reproducibility, parallelism, recover and interfaces.
- Recall Posting Date
- 2005-07-25
- Trade Name
- Microstrep Plus 1
- Manufacturer
- Dade Behring, Inc.
- Recall Start Date
- 2005-06-23
- Recall Number
- 28563
- Hazard Classification
- Type III
- Model or Catalog #
- B1027-201
- Lot or Serial #
- 2005-08-06, 2006-01-14, 2006-01-25, 2006-02-24
- Reason for Recall
- There is a potential for out of range (high) quality control Clindamycin results for ATC'S. pneumoniae 49619 and
- Recall Posting Date
- 2005-07-25
- Trade Name
- Quicklyte Sensor Cartridge
- Manufacturer
- Dade Behring, Inc.
- Recall Start Date
- 2005-06-17
- Recall Number
- 28555
- Hazard Classification
- Type III
- Model or Catalog #
- S600
- Lot or Serial #
- 5CD813, 5CD816, 5DD817, 5DD8120
- Reason for Recall
- Manufacturer has confirmed that urine chloride test results in patients, qc and proficiency survey samples may be elevated by 10-15%.
- Recall Posting Date
- 2005-07-25
- Trade Name
- Set Radical Handheld Pulse Oximeter
- Manufacturer
- Masimo Corporation
- Recall Start Date
- 2005-06-13
- Recall Number
- 28558
- Hazard Classification
- Type II
- Model or Catalog #
- N
- Lot or Serial #
- Less than or equal to 102999 - P3027 through P3148
- Reason for Recall
- The internal speaker could fail to produce an audio alarm during an alarm condition. The affected units were manufactured prior to December 11, 2001.
- Recall Posting Date
- 2005-07-25
- Trade Name
- Sonoline Antares Diagnostic Ultrasound
- Manufacturer
- Siemens Medical Solutions USA, Inc.
- Recall Start Date
- 2005-06-05
- Recall Number
- 28540
- Hazard Classification
- Type II
- Model or Catalog #
- 5936518
- Lot or Serial #
- 112607, 112593, 112609, 112611, 112530, 112595, 112612, 112687, 112610, 10077, 10145, 10280, 111223, 111233, 111234, 112579, 10438, 111197, 112736, 10068, 111843, 111922, 111923, 111639, 111643, 111660, 111666, 10928, 11092, 11096, 112658, 112666, 10838, 10925, 112237, 10134, 112267, 112070, 112073, 10940, 111166
- Reason for Recall
- Error occurs after performing a labeled measurement on a frozen waveform utilizing a 2d image and an angle-corrected pulse wave doppler waveform.
- Recall Posting Date
- 2005-07-25
- Trade Name
- Swiss Lithoclast Ultrasound Handpiece
- Manufacturer
- Electro Medical Systems (EMS) SA
- Recall Start Date
- 2005-06-17
- Recall Number
- 28544
- Hazard Classification
- Type II
- Model or Catalog #
- FR-093
- Lot or Serial #
- 20124, 20172, 20268, 29342, 20372, 20403, 20406
- Reason for Recall
- Manufacturer has become aware that handpieces have twisted cables and there is a small possibility that they may not function properly or the function may diminish over time.
- Recall Posting Date
- 2005-07-11
- Trade Name
- A) Axiom Aristos FX
B) Axiom Aristos MX
- Manufacturer
- Siemens Aktiengesellschaft
- Recall Start Date
- 2005-06-14
- Recall Number
- 28493
- Hazard Classification
- Type III
- Model or Catalog #
- A) 7414803
B) 5895003
- Lot or Serial #
- A) 1298, 1120, 1129, 1131, 1292, 1288, 1086, 1236, 1006, 1243, 1018, 1311, 1315, 1326, 1340, 1354, 1153, 1443, 1337, 1173, 1190, 1244, 1072,
1273, 1175, 1124, 1134, 1272, 1333, 1145, 1386,1065, 1328, 1162, 1128, 1119, 1176.
B) N/A
- Reason for Recall
- A system lock-up may occur when repeating patient registrations under a specific workflow.
- Recall Posting Date
- 2005-07-11
- Trade Name
- A) Homechoice Pro PD Aut. System-Software
B) PBO Enal Link
- Manufacturer
- Baxter Healthcare Corporation
- Recall Start Date
- 2005-06-23
- Recall Number
- 28534
- Hazard Classification
- Type II
- Model or Catalog #
- A) R5C4532, R5C4534
B) 5M5590
- Lot or Serial #
- A) Renal Software Suite
B) Renal Software Suite
- Reason for Recall
- This is a followup to the recall of April 2004 regarding an anomaly that may occur when changing the name of a medication, Baxter is now providing a cd with a software upgrade.
- Recall Posting Date
- 2005-07-11
- Trade Name
- A) Model 6060 Homerun Vol. Infusion Pump
B) Model 6060 Homerun Vol. Infusion Pump
- Manufacturer
- Baxter Healthcare Corporation
- Recall Start Date
- 2005-06-22
- Recall Number
- 28512
- Hazard Classification
- Type II
- Model or Catalog #
- A) 2M9832, 2M9832R, 60600040IR
B) 60600043, D2M9832
- Lot or Serial #
- A) All Serial Numbers
B) All Serial Numbers
- Reason for Recall
- Manufacturer is issuing a notice to all users on two issues 1)specific keystrokes sequence can result in over-infusion, 2)to safeguard the security codes that allow programming pump parameters.
- Recall Posting Date
- 2005-07-11
- Trade Name
- Access Immunoassay Prolactin Calibrator
- Manufacturer
- Beckman Coulter, Inc.
- Recall Start Date
- 2005-03-24
- Recall Number
- 27979
- Hazard Classification
- Type III
- Model or Catalog #
- N
- Lot or Serial #
- 419474
- Reason for Recall
- This calibrator lot will not support the open vial storage recommendations of the 2 months at 2 to 10 degrees c. This issue could effect patient results.
- Recall Posting Date
- 2005-07-11
- Trade Name
- Compression Hip Screw System-Lag Screw
- Manufacturer
- Smith & Nephew, Inc. Orthopaedic Division
- Recall Start Date
- 2005-04-11
- Recall Number
- 28190
- Hazard Classification
- Type III
- Model or Catalog #
- 12-1108
- Lot or Serial #
- 04LT84854
- Reason for Recall
- Chart stick label and inner tyvek pouch label state that the product is a hex screwdriver. The outside label correctly identifies the product.
- Recall Posting Date
- 2005-07-11
- Trade Name
- Configuration Transfer Cable
- Manufacturer
- Baxter Healthcare Corporation
- Recall Start Date
- 2005-03-14
- Recall Number
- 28488
- Hazard Classification
- Type III
- Model or Catalog #
- 2M8155
- Lot or Serial #
- 50604
- Reason for Recall
- Baxter has been provided and distributed cables with incorrect dimensions on the thumbscrews that attach to the colleague pump, the thumbscrews were not manufactured to baxter specifications.
- Recall Posting Date
- 2005-07-11
- Trade Name
- Coulter 5C Cell Control
- Manufacturer
- Beckman Coulter, Inc.
- Recall Start Date
- 2005-03-09
- Recall Number
- 27862
- Hazard Classification
- Type III
- Model or Catalog #
- N
- Lot or Serial #
- 111220K-111224K for PN 624506, 102866K-102880K for PN 7547001, 102785K-102794K for PN 7547002, 869400F-869900F for PN 7547011, 100128K-100152K for PN 7547116, 112198K-112101K for PN 7547124, 100348K-100351K for PN 7547199
- Reason for Recall
- The recovery of the MCV parameter for the abnormal II level of coulter 5c cell control may recover values above the upper expected range limit.
- Recall Posting Date
- 2005-07-11
- Trade Name
- DIM RCRP Rev C-Reactive PRTN Flex 120TS
- Manufacturer
- Dade Behring, Inc.
- Recall Start Date
- 2005-05-25
- Recall Number
- 28487
- Hazard Classification
- Type III
- Model or Catalog #
- DF34
- Lot or Serial #
- DH5341, EG5300
- Reason for Recall
- Manufacturer has determined a stability issue with the particle reagent and may result in inability to calibrate, flagged results, imprecision of inaccuracy, QC results shift high or low, error message.
- Recall Posting Date
- 2005-07-11
- Trade Name
- Disposable Tourniquet Cuffs
- Manufacturer
- Zimmer Orthopaedic Surgical Products
- Recall Start Date
- 2005-03-10
- Recall Number
- 28499
- Hazard Classification
- Type II
- Model or Catalog #
- 60707000100, 60707000300, 60707000400, 60707000500, 60707000600
- Lot or Serial #
- 60154121, 60171346, 60162824, 60171348, 60180450, 60171347, 60173808, 60194151, 60194152, 60188441, 60184634, 60202399, 60205580, 60208192, 60176164
- Reason for Recall
- Some units manufactured with a lower separation strength for the port-to-tube connection. The port
- Recall Posting Date
- 2005-07-11
- Trade Name
- Immulite CEA
- Manufacturer
- Euro
- Recall Start Date
- 2005-05-25
- Recall Number
- 28484
- Hazard Classification
- Type II
- Model or Catalog #
- LKCE1
- Lot or Serial #
- 302, 303
- Reason for Recall
- Manufacturer has determined an adverse trend in stability which may result in significant decrease in adjustment slopes and a positive shift of 20-30% at low conc., 20% at higher conc. of patient results.
- Recall Posting Date
- 2005-07-11
- Trade Name
- ITHA52 Healing Abutments
- Manufacturer
- Implant Innovations, Inc.
- Recall Start Date
- 2005-05-18
- Recall Number
- 28369
- Hazard Classification
- Type III
- Model or Catalog #
- ITHA52
- Lot or Serial #
- 330918
- Reason for Recall
- The colour code on the label to identify the implant platform diameter was incorrect. The label has a yellow stripe instead of a blue stripe.
- Recall Posting Date
- 2005-07-11
- Trade Name
- KyphX HV-R Bone Cement
- Manufacturer
- Kyphon Inc.
- Recall Start Date
- 2005-03-14
- Recall Number
- 27927
- Hazard Classification
- Type III
- Model or Catalog #
- N
- Lot or Serial #
- EL08304
- Reason for Recall
- US version of the instructions for use accompanied product sent to Canada.
- Recall Posting Date
- 2005-07-11
- Trade Name
- Magnetom Avanto System
- Manufacturer
- Siemens Aktiengesellschaft
- Recall Start Date
- 2005-06-21
- Recall Number
- 28493
- Hazard Classification
- Type II
- Model or Catalog #
- 7391167
- Lot or Serial #
- 25285, 25340, 25043, 25039, 25234, 25173, 25089, 25110, 25066, 25137, 25049, 25097, 25311, 25168, 25333, 25224
- Reason for Recall
- For the safety of the service technicians, the acc power connection X102 need to be checked and labelled.
- Recall Posting Date
- 2005-07-11
- Trade Name
- MISYS Laboratory System
- Manufacturer
- MISYS Healthcare Systems Inc.
- Recall Start Date
- 2005-05-13
- Recall Number
- 28491
- Hazard Classification
- Type I
- Model or Catalog #
- SQI-MIC-SY
- Lot or Serial #
- LARS versions 5.3 or later
- Reason for Recall
- In rare situations, modification of previously reported specimen results using the lars application, may modify other previously filed results on the same specimen.
- Recall Posting Date
- 2005-07-11
- Trade Name
- MISYS Laboratory System Blood Bank
- Manufacturer
- MISYS Healthcare Systems Inc.
- Recall Start Date
- 2005-05-10
- Recall Number
- 28492
- Hazard Classification
- Type III
- Model or Catalog #
- N
- Lot or Serial #
- Sites using software ver. 6.1
- Reason for Recall
- When a previously reported microbiology organism result is deleted on a selected patient, reported susceptibilities will not stay with associated organisms and one or more susceptibilities may be deleted.
- Recall Posting Date
- 2005-07-11
- Trade Name
- Purevision Toric Tinted Contact Lenses
- Manufacturer
- Bausch & Lomb Inc.
- Recall Start Date
- 2005-06-13
- Recall Number
- 28497
- Hazard Classification
- Type II
- Model or Catalog #
- 738
- Lot or Serial #
- W58405708, W58405713
- Reason for Recall
- Portion of two Purevision Toric product lots did not complete the required cycle time in the extraction and hydration process. As a result, residual extractable in the lenses exceeded specification.
- Recall Posting Date
- 2005-07-11
- Trade Name
- Six-Part Bar Code Labels, Sarstedt Tubes
- Manufacturer
- Roche Molecular Systems, Inc.
- Recall Start Date
- 2005-05-17
- Recall Number
- 28326
- Hazard Classification
- Type I
- Model or Catalog #
- N
- Lot or Serial #
- 28868, 029213
- Reason for Recall
- Rolls of six-part bar code labels, roche M
- Recall Posting Date
- 2005-07-11
- Trade Name
- Viewforum Software
- Manufacturer
- Philips Medical Systems Nederland B.V.
- Recall Start Date
- 2005-05-27
- Recall Number
- 28501
- Hazard Classification
- Type II
- Model or Catalog #
- M1M1171
- Lot or Serial #
- 5V9371J, 00F1F8B16DA, 00F1F9385CE
- Reason for Recall
- The "automated analysis of vascular obstruction" tool in the Viewforum 2003 software gives unreliable calibration and analysis results when using unprocessed images acquired on the Philips Allura FD10.