Health Canada
Symbol of the Government of Canada

Institutional links

Drugs and Health Products

Medical Device Recall Listings - Sorted by Manufacturer

(From July 2005 - September 2005)

Manufacturer
3M Health Care
Recall Posting Date
2005-09-05
Trade Name
3M Microfoam Surgical Tape
Recall Start Date
2005-08-08
Recall Number
28764
Hazard Classification
Type III
Model or Catalog #
1528-1, 1528-2, 1528-3, 1528-4
Lot or Serial #
2010-06 (HE, HR, HS, HU, IM, IN)
Reason for Recall
The tape may not perform as intended in some situations as portions of the tape may not appear to have adhesive or the adhesive may separate from the foam backing.
Manufacturer
Abbott Laboratories
Recall Posting Date
2005-09-05
Trade Name
A) Plum A+ Infusion Pumps

B) Plum A+3 Infusion Pumps
Recall Start Date
2005-06-02
Recall Number
28438
Hazard Classification
Type I
Model or Catalog #
A) N/A

B) N/A
Lot or Serial #
A) Numerus lot numbers.
B) Numerus lot numbers.
Reason for Recall
If battery service warning and E320 messages are not responded to appropriately, the pump may shut off during battery operation without providing adequate response time.
Manufacturer
Abbott Laboratories Diagnostic Division
Recall Posting Date
2005-08-22
Trade Name
Architect System Software V2.10 & 2.11
Recall Start Date
2005-06-23
Recall Number
28679
Hazard Classification
Type III
Model or Catalog #
1G06-01, 7-99077-01, 99077-103
Lot or Serial #
All serial numbers.
Reason for Recall
The software calculation utilizes 17ul rather than 12 ul for the alkaline washing solution dispensed, the displayed value for volume remaining will be less than the actual amount in the bottle.
Manufacturer
Abbott Laboratories Diagnostic Division
Recall Posting Date
2005-08-08
Trade Name
A) AXSYM Total B-HCG Reagent

B) IMX Total B-HCG Reagent
Recall Start Date
2005-04-05
Recall Number
28629
Hazard Classification
Type III
Model or Catalog #
A) 1A06-22
B) 7A59-22
Lot or Serial #
A) 26148Q100, 26278Q100, 16381Q100
B) 25459Q100
Reason for Recall
Tests have shown that protein aggregates are present in the specimen diluent and causes calibration errors and a shift in control values and no valid patient results could be generated.
Manufacturer
Abbott Laboratories Diagnostic Division
Recall Posting Date
2005-08-08
Trade Name
Axym Anti-TPO Reagent
Recall Start Date
2005-06-28
Recall Number
28675
Hazard Classification
Type III
Model or Catalog #
5F57-20
Lot or Serial #
22827HN00
Reason for Recall
Manufacturer has identified a manufacturing error at a contract facility in which a different reagent was mistakenly packed in its place for a subset of this reagent lot.
Manufacturer
Abbott Laboratories Diagnostic Division
Recall Posting Date
2005-08-08
Trade Name
Cell DYN 1800 System
Recall Start Date
2005-05-18
Recall Number
28635
Hazard Classification
Type III
Model or Catalog #
07H77-01
Lot or Serial #
19017AY
Reason for Recall
The HP6122 printer will print one page correctly but may have problems printing more than one page from the data log or the Qc log when "print data log" and "print Qc log" commands are given.
Manufacturer
Abbott Laboratories Diagnostic Division
Recall Posting Date
2005-08-08
Trade Name
TDX
Recall Start Date
2005-04-05
Recall Number
28653
Hazard Classification
Type III
Model or Catalog #
9797-02, 9797-12
Lot or Serial #
18229Q100, 18230Q100
Reason for Recall
Manufacturer has discovered a change in performance in which patient results are depressed 5-8%, elevated values when running a calibration curve.
Manufacturer
Abbott Laboratories Diagnostic Division
Recall Posting Date
2005-09-05
Trade Name
Abbott Corzyme
Recall Start Date
2005-04-01
Recall Number
28747
Hazard Classification
Type II
Model or Catalog #
9977-24
Lot or Serial #
18488M100, 15893M100, 16792M200, 10527M200, 23326M100, 19804M100
Reason for Recall
A study to compare the panels with a reference serum showed a difference in pei units
Manufacturer
Abbott Laboratories Diagnostic Division
Recall Posting Date
2005-09-05
Trade Name
Cell-DYN 4000 System
Recall Start Date
2005-03-17
Recall Number
28750
Hazard Classification
Type II
Model or Catalog #
01H01-01, 01H01-03, 01H02-01, 01H02-03, 01H03-01
Lot or Serial #
30268AA, 30271AA, 30455AA, 17500AA96, 30272AA, 30670AA, 287AA, 30266AA, 30403AA, 30653AA, 30654AA, 30697AA, 30416AA, 30488AA, 30665AA, 30672AA, 30615AA, 30709AA , 30712AA, 30686AA, 30683AA, 30723AA.
Reason for Recall
The optical platelet algorithm does not correctly gate the plto for the calibrator and the plus control and causes calibration problems and control plto results to fall outside the lower limit of assay range.
Manufacturer
Abbott Laboratories Diagnostic Division
Recall Posting Date
2005-09-05
Trade Name
Testpack Plus HCG Combo with OBC
Recall Start Date
2005-03-22
Recall Number
28752
Hazard Classification
Type II
Model or Catalog #
7B34-16
Lot or Serial #
16006M200, 19199M300
Reason for Recall
Manufacturer has received user complaints reporting occurrences of false positive results or weak positive colour on the patient bar for specimens where the results disagree with other methods.
Manufacturer
ACMI Corporation
Recall Posting Date
2005-08-22
Trade Name
ACMI Snap & Peel Introducer Sheaths
Recall Start Date
2005-06-30
Recall Number
28738
Hazard Classification
Type III
Model or Catalog #
60810BX, 61012BX
Lot or Serial #
1297324D, 1312704E, 1321584E, 1325384E, 1339994F, 1344104G, 1389384K, 1394004L, 1400154M, 1419225A, 1297374D, 1312724E, 1316954E, 1321644E, 1325394E, 1336944F, 1339964F, 1344094G, 1355274H, 1350354H, 1359054H, 1364234J, 1367404J, 1372814J, 1383294K, 1412604M, 1419235A, 1427565A, 1430855B, 1441195B
Reason for Recall
ACMI received a report from B. Braun, manufacturer of device, that there were cracks in the hub
Manufacturer
ADAC Laboratories, Inc.
Recall Posting Date
2005-08-22
Trade Name
Argus Epic Camera
Recall Start Date
2005-07-13
Recall Number
28730
Hazard Classification
Type II
Model or Catalog #
2145-3007A
Lot or Serial #
A0105217, 01548B, 03338B, 00000000009701075, 00964B, 00227BC, 01922B, 01920B, 01921B, 00000000009901158, 00000000009901157, 00709B, 00000000009908167, 00000000009801107, 01525B, A03030282, 01332B, 03041B, 01333B, 03094B, 01310B, 03368B
Reason for Recall
Fracture was discovered on an argus fork which holds the detector.
Manufacturer
Alsius Corporation
Recall Posting Date
2005-09-19
Trade Name
CoolGard 3000
Recall Start Date
2005-08-02
Recall Number
28831
Hazard Classification
Type I
Model or Catalog #
N
Lot or Serial #
465131
Reason for Recall
Coolgard 3000 units manufactured before April 12th, 2005, may overheat their coolant if left in a warm room, for prolonged periods in standby mode. At 44 c. Alarm will trigger and system latches.
Manufacturer
Anika Therapeutics Inc.
Recall Posting Date
2005-08-22
Trade Name
AMVISC
Recall Start Date
2005-07-25
Recall Number
28723
Hazard Classification
Type II
Model or Catalog #
59051, 59081
Lot or Serial #
B050120B, B050510B
Reason for Recall
Some cannulae packaged with AMVISC and AMVISC Plus Visoelastic may have an incomplete pouch seal.
Manufacturer
Bausch & Lomb Inc.
Recall Posting Date
2005-07-11
Trade Name
Purevision Toric Tinted Contact Lenses
Recall Start Date
2005-06-13
Recall Number
28497
Hazard Classification
Type II
Model or Catalog #
738
Lot or Serial #
W58405708, W58405713
Reason for Recall
Portion of two Purevision Toric product lots did not complete the required cycle time in the extraction and hydration process. As a result, residual extractable in the lenses exceeded specification.
Manufacturer
Baxter Healthcare Corporation
Recall Posting Date
2005-08-22
Trade Name
A) Homechoice Automated PD System

B) Homechoice Pro Automated PD System
Recall Start Date
2005-07-05
Recall Number
28726
Hazard Classification
Type II
Model or Catalog #
A) 5C4471

B) 5C8310, 5C8310P, 5C8310PE
Lot or Serial #
A) All serial numbers.
B) All serial numbers.
Reason for Recall
Remote risk of electrical shock from screws used to secure the power entry module to the back of the panel, this module is the electrical connector where the power cord plugs into the unit.
Manufacturer
Baxter Healthcare Corporation
Recall Posting Date
2005-08-08
Trade Name
A) Colleague 3 Volumetric Infusion Pump

B) Colleague Volumetric Infusion Pump
Recall Start Date
2005-07-21
Recall Number
28660
Hazard Classification
Type I
Model or Catalog #
A) 2M8153, 2M8163, DNM8153

B) 2M8151, 2M8161, DNM8151
Lot or Serial #
A) All Serial Numbers
B) All Serial Numbers
Reason for Recall
Design problem with the clocking circuit that can disrupt internal communications causing the pumps to shut down with certain error codes and disrupt critical medication infusions. Death has been involved.
Manufacturer
Baxter Healthcare Corporation
Recall Posting Date
2005-07-11
Trade Name
A) Homechoice Pro PD Aut. System-Software

B) PBO Enal Link
Recall Start Date
2005-06-23
Recall Number
28534
Hazard Classification
Type II
Model or Catalog #
A) R5C4532, R5C4534
B) 5M5590
Lot or Serial #
A) Renal Software Suite
B) Renal Software Suite
Reason for Recall
This is a followup to the recall of April 2004 regarding an anomaly that may occur when changing the name of a medication, Baxter is now providing a cd with a software upgrade.
Manufacturer
Baxter Healthcare Corporation
Recall Posting Date
2005-07-11
Trade Name
A) Model 6060 Homerun Vol. Infusion Pump

B) Model 6060 Homerun Vol. Infusion Pump
Recall Start Date
2005-06-22
Recall Number
28512
Hazard Classification
Type II
Model or Catalog #
A) 2M9832, 2M9832R, 60600040IR
B) 60600043, D2M9832
Lot or Serial #
A) All Serial Numbers
B) All Serial Numbers
Reason for Recall
Manufacturer is issuing a notice to all users on two issues 1)specific keystrokes sequence can result in over-infusion, 2)to safeguard the security codes that allow programming pump parameters.
Manufacturer
Baxter Healthcare Corporation
Recall Posting Date
2005-07-11
Trade Name
Configuration Transfer Cable
Recall Start Date
2005-03-14
Recall Number
28488
Hazard Classification
Type III
Model or Catalog #
2M8155
Lot or Serial #
50604
Reason for Recall
Baxter has been provided and distributed cables with incorrect dimensions on the thumbscrews that attach to the colleague pump, the thumbscrews were not manufactured to baxter specifications.
Manufacturer
Beckman Coulter Inc.
Recall Posting Date
2005-09-05
Trade Name
StemCXP Software Version 1.0
Recall Start Date
2005-06-09
Recall Number
28471
Hazard Classification
Type III
Model or Catalog #
628843
Lot or Serial #
N
Reason for Recall
7HPCSA and 7HPCSA control protocols may not invoke the algorithm under certain conditions.
Manufacturer
Beckman Coulter Inc.
Recall Posting Date
2005-09-05
Trade Name
LH 1500 Series Lab Automation System
Recall Start Date
2005-06-23
Recall Number
28510
Hazard Classification
Type III
Model or Catalog #
6605528
Lot or Serial #
N
Reason for Recall
Warnings in manual conflict with sections in the manual. Potential for mechanical and voltage hazards during maintenance.
Manufacturer
Beckman Coulter Inc.
Recall Posting Date
2005-09-05
Trade Name
Synchron Systems Amphetamine Reagent
Recall Start Date
2005-07-07
Recall Number
28607
Hazard Classification
Type III
Model or Catalog #
475000
Lot or Serial #
M501225, M503380
Reason for Recall
These lots have different than expected cross reactivities. May cause false positive amphetamine results in patients taking medications containing ephedrine.
Manufacturer
Beckman Coulter Inc.
Recall Posting Date
2005-09-05
Trade Name
A) Immage Low Conc. IgA (IGALC) Reagent

B) Immage Low Conc. IgM (IGALC) Reagent
Recall Start Date
2005-04-17
Recall Number
28610
Hazard Classification
Type III
Model or Catalog #
A) 447460
B) 447470
Lot or Serial #
A) All lots
B) All Lots
Reason for Recall
Reagents do not meet performance claim.
Manufacturer
Beckman Coulter Inc.
Recall Posting Date
2005-09-05
Trade Name
Access Immunoassay Unconj Estriol Calib
Recall Start Date
2005-07-14
Recall Number
28643
Hazard Classification
Type III
Model or Catalog #
33575
Lot or Serial #
416534
Reason for Recall
Does not meet expiration claim of 12 months.
Manufacturer
Beckman Coulter Inc.
Recall Posting Date
2005-09-05
Trade Name
Paragon CZE 2000 Software Version 2.0
Recall Start Date
2005-07-27
Recall Number
28734
Hazard Classification
Type III
Model or Catalog #
A09744
Lot or Serial #
N
Reason for Recall
Potential for incorrect results to be displayed and printed out.
Manufacturer
Beckman Coulter Inc.
Recall Posting Date
2005-09-05
Trade Name
Remisol 2000 Data Manager System
Recall Start Date
2005-06-27
Recall Number
28505
Hazard Classification
Type III
Model or Catalog #
A27343
Lot or Serial #
N
Reason for Recall
Results may upload with an incorrect or blank sample ID.
Manufacturer
Beckman Coulter Inc.
Recall Posting Date
2005-09-05
Trade Name
A) Access Immunoassay Sys: Dil-AFP

B) Access Immunoassay Sys: Dil-AFP
Recall Start Date
2005-07-21
Recall Number
28789
Hazard Classification
Type III
Model or Catalog #
A) 33211
B) 33210
Lot or Serial #
A) All lots.

B) All lots
Reason for Recall
Potential splashing may cause erroneous results.
Manufacturer
Beckman Coulter Inc.
Recall Posting Date
2005-09-19
Trade Name
Synchron CRP
Recall Start Date
2005-07-18
Recall Number
28832
Hazard Classification
Type II
Model or Catalog #
N
Lot or Serial #
N
Reason for Recall
Synchron CRP turbidimetric assay may not be sensitive or specific enough as a sole diagnostic marker of infection in neonates.
Manufacturer
Beckman Coulter, Inc.
Recall Posting Date
2005-08-08
Trade Name
Access Immuno Sys Unconj Estriol Calib
Recall Start Date
2005-07-14
Recall Number
28643
Hazard Classification
Type III
Model or Catalog #
33575
Lot or Serial #
416534
Reason for Recall
Device does not meet expiration claim of 12 months.
Manufacturer
Beckman Coulter, Inc.
Recall Posting Date
2005-08-08
Trade Name
Synchron Clinical Sys. Alcohol Reagent
Recall Start Date
2005-05-17
Recall Number
28371
Hazard Classification
Type III
Model or Catalog #
91DIC
Lot or Serial #
UPDATED RECALL # 28371 PREVIOUSLY POSTED ON 2005-06-27.
M405028, M409024, M407255, M412090, M503108
Reason for Recall
Reagent is past expiry date thus will not produce results since it cannot calibrate successfully.
Manufacturer
Beckman Coulter, Inc.
Recall Posting Date
2005-07-11
Trade Name
Access Immunoassay Prolactin Calibrator
Recall Start Date
2005-03-24
Recall Number
27979
Hazard Classification
Type III
Model or Catalog #
N
Lot or Serial #
419474
Reason for Recall
This calibrator lot will not support the open vial storage recommendations of the 2 months at 2 to 10 degrees c. This issue could effect patient results.
Manufacturer
Beckman Coulter, Inc.
Recall Posting Date
2005-07-11
Trade Name
Coulter 5C Cell Control
Recall Start Date
2005-03-09
Recall Number
27862
Hazard Classification
Type III
Model or Catalog #
N
Lot or Serial #
111220K-111224K for PN 624506, 102866K-102880K for PN 7547001, 102785K-102794K for PN 7547002, 869400F-869900F for PN 7547011, 100128K-100152K for PN 7547116, 112198K-112101K for PN 7547124, 100348K-100351K for PN 7547199
Reason for Recall
The recovery of the MCV parameter for the abnormal II level of coulter 5c cell control may recover values above the upper expected range limit.
Manufacturer
Becton Dickinson
Recall Posting Date
2005-08-08
Trade Name
BD DTXPlus Disposable Transducer Kit
Recall Start Date
2005-06-29
Recall Number
28624
Hazard Classification
Type II
Model or Catalog #
682000, 682034
Lot or Serial #
407032, 407104
Reason for Recall
Device (stopcock tap) may easily detach from the stopcock and cause leakage of the iv fluids and
Manufacturer
Becton Dickinson
Recall Posting Date
2005-07-25
Trade Name
Bd Unopette - RBC Osmotic Fragility Determination for Manual Methods
Recall Start Date
2005-06-27
Recall Number
28536
Hazard Classification
Type II
Model or Catalog #
365830
Lot or Serial #
5014328
Reason for Recall
Lot may have a problem that could cause an increased susceptibility for lysis of red cells from normal subjects. Yields uninterpretable results and causing the test results to not be generated.
Manufacturer
Biomerieux, Inc.
Recall Posting Date
2005-09-05
Trade Name
A) Simplastin Excel

B) Simplastin L

C) Simplastin - Excel S
Recall Start Date
2005-08-08
Recall Number
28780
Hazard Classification
Type I
Model or Catalog #
A) N/A
B) N/A
C) N/A
Lot or Serial #
A) 161711, 161929, 161897
B) 161750
C) 161720, 161760, 161761, 161783, 161811
Reason for Recall
The ISI values published in the package insert of Simplastin Excel products are incorrect for the devices: fibrometer (mechanical), option, XM and MTX (photo optic).
Manufacturer
Cardiac Pacemakers Inc. A Wholly Owned Subsidiary of Guidant
Recall Posting Date
2005-08-08
Trade Name
A) Contak Renewal 4 AVT

B) Contak Renewal 4 AVT HE

C) Ventak Prizm AVT

D) Vitality AVT AICD
Recall Start Date
2005-07-22
Recall Number
28665
Hazard Classification
Type II
Model or Catalog #
A) M170, M175

B) M177

C) 1900

D) A155
Lot or Serial #
A) N/A

B) N/A

C) N/A

D) N/A
Reason for Recall
One of the original recommendations to recall #28528 - programming atrial tachy episode data storage to 0% - can cause latching in a subset of AVT devices that have previously stored atrial episode data.
Manufacturer
Cardiac Pacemakers Inc. A Wholly Owned Subsidiary of Guidant
Recall Posting Date
2005-08-08
Trade Name
A) Contak TR CHFD

B) CPI Discovery DR Pacemaker

C) CPI Discovery SR Pacemaker

D) CPI Meridian DDD Pacemaker

E) CPI Meridian DR Pacemaker

F) CPI Meridian SR Pacemaker

G) CPI Meridian SSI Pacemaker

H) CPI Pulsar DDD, Model 970 Pacemaker

I) CPI Pulsar DDD, Model 972 Pacemaker

J) CPI Pulsar DR Pacemaker

K) CPI Pulsar Max DR Pacemaker

L) CPI Pulsar Max SR, Model 1170 Pacemaker

M) CPI Pulsar Max SR, Model 1171 Pacemaker

N) CPI Pulsar SSI Pacemaker

O) CPI Pulsar VDD Pacemaker

P) Discovery II DDD

Q) Discovery II DR

R) Discovery II SR

S) Discovery II SSI

T) Pulsar Max II Pacemaker DDDR

U) Pulsar Max II Pacemaker SSIR

W) CPI Pulsar SR Pacemaker
Recall Start Date
2005-07-19
Recall Number
28650
Hazard Classification
Type I
Model or Catalog #
A) 1241

B) 1273, 1274, 1275

C) 1174, 1175

D) 976

E) 1276

F) 1176

G) 476

H) 970

I) 972

J) 1272

K) 1270

L) 1170

M) 1171

N) 470

O) 870

P) 981

Q) 1283, 1284, 1285, 1286

R) 1184, 1186, 1187

S) 481

T) 1280

U) 1180, 1181

W) 1172
Lot or Serial #
A) N/A

B) N/A

C) N/A

D) N/A

E) N/A

F) N/A

G) N/A

H) N/A

I) N/A

J) N/A

K) N/A

L) N/A

M) N/A

N) N/A

O) N/A

P) N/A

Q) N/A

R) N/A

S) N/A

T) N/A

U) N/A

W) N/A
Reason for Recall
Hermetic sealing component may experience a gradual degradation, resulting in a higher than normal moisture content within the pacemaker Case late in the device's service life.
Manufacturer
Cardiac Pacemakers Inc. A Wholly Owned Subsidiary of Guidant
Recall Posting Date
2005-07-25
Trade Name
A) Contak Renewal 2

B)Contak Renewal
Recall Start Date
2005-06-21
Recall Number
28519
Hazard Classification
Type II
Model or Catalog #
A) H135

B) H155
Lot or Serial #
A) N/A

B) N/A
Reason for Recall
Deterioration in a wire insulator within the lead connector block, could cause a short circuit and loss of device function due to diversion of therapy energy away from the heart and into device circuitry.
Manufacturer
Cardiac Pacemakers Inc. A Wholly Owned Subsidiary of Guidant
Recall Posting Date
2005-07-25
Trade Name
A) Contak Renewal 4 AVT

B) Contak Renewal 4 AVT HE

C) Contak Renewal 4

D) Contak Renewal 4 HE
Recall Start Date
2005-06-24
Recall Number
28531
Hazard Classification
Type II
Model or Catalog #
A) M170, M175

B) M177, M179

C) H190, H195

D) H197, H199
Lot or Serial #
A) N/A

B) N/A

C) N/A

D) N/A
Reason for Recall
Component failure may limit available tachyarrhythmia therapy. The magnetic switch might stick in the closed position.
Manufacturer
Cardiac Pacemakers Inc. A Wholly Owned Subsidiary of Guidant
Recall Posting Date
2005-07-25
Trade Name
A) Contak Renewal 4 AVT

B) Contak Renewal 4 AVT HE

C) Ventak Prizm AVT

D) Vitality AVT
Recall Start Date
2005-06-21
Recall Number
28528
Hazard Classification
Type II
Model or Catalog #
A) M170, M175

B) M177, M179

C) 1900

D) A135, A155
Lot or Serial #
A) N/A

B) N/A

C) N/A

D) N/A
Reason for Recall
Random memory error causes functional "latching" that limits available therapy. Latching can only occur when storing tachy detection
Manufacturer
CTI Pet Systems Inc.
Recall Posting Date
2005-08-08
Trade Name
Biograph Duo LSO
Recall Start Date
2005-07-12
Recall Number
28625
Hazard Classification
Type III
Model or Catalog #
8415957
Lot or Serial #
5408-0301009
Reason for Recall
Under certain circumstances the system will continue to scan after an IRS function failure occurs. The system will scan only until the end of the planned sequence and then will shut off.
Manufacturer
Dade Behring Marburg GMBH
Recall Posting Date
2005-08-08
Trade Name
Clotting Factor 8 Deficient Plasma
Recall Start Date
2005-07-01
Recall Number
28642
Hazard Classification
Type II
Model or Catalog #
OTXW17
Lot or Serial #
503848A, 503849E, 503850C, 503856B, 503858A, 503860D,
Reason for Recall
Manufacturer has determined that these lots do not demonstrate acceptable performance and patient samples may exhibit falsely high factor viii results, could lead to misdiagnosis.
Manufacturer
Dade Behring Marburg GMBH
Recall Posting Date
2005-07-25
Trade Name
Behring Nephelometer Pro Spec Analyzer
Recall Start Date
2005-06-24
Recall Number
28589
Hazard Classification
Type III
Model or Catalog #
400252.1003
Lot or Serial #
420926, 341237, 241226, 141169, 331087, 490129, 100215, 410745, 410737, 120753, 310622, 410720, 300392, 310624, 310667, 210615, 210582, 210579, 210613A, 110540, 210577, 210545, 110539, 400476,
Reason for Recall
Manufacturer has received reports that flags "l" and "h" indicating results below and above the reference range may not be correct for formula assays.
Manufacturer
Dade Behring, Inc.
Recall Posting Date
2005-07-25
Trade Name
Microstrep Plus 1
Recall Start Date
2005-06-23
Recall Number
28563
Hazard Classification
Type III
Model or Catalog #
B1027-201
Lot or Serial #
2005-08-06, 2006-01-14, 2006-01-25, 2006-02-24
Reason for Recall
There is a potential for out of range (high) quality control Clindamycin results for ATC'S. pneumoniae 49619 and
Manufacturer
Dade Behring, Inc.
Recall Posting Date
2005-07-25
Trade Name
Quicklyte Sensor Cartridge
Recall Start Date
2005-06-17
Recall Number
28555
Hazard Classification
Type III
Model or Catalog #
S600
Lot or Serial #
5CD813, 5CD816, 5DD817, 5DD8120
Reason for Recall
Manufacturer has confirmed that urine chloride test results in patients, qc and proficiency survey samples may be elevated by 10-15%.
Manufacturer
Dade Behring, Inc.
Recall Posting Date
2005-07-11
Trade Name
DIM RCRP Rev C-Reactive PRTN Flex 120TS
Recall Start Date
2005-05-25
Recall Number
28487
Hazard Classification
Type III
Model or Catalog #
DF34
Lot or Serial #
DH5341, EG5300
Reason for Recall
Manufacturer has determined a stability issue with the particle reagent and may result in inability to calibrate, flagged results, imprecision of inaccuracy, QC results shift high or low, error message.
Manufacturer
Diagnostica Stago
Recall Posting Date
2005-08-22
Trade Name
STAGO ATA R System
Recall Start Date
2005-06-08
Recall Number
28662
Hazard Classification
Type III
Model or Catalog #
57160
Lot or Serial #
All serial numbers.
Reason for Recall
Manufacturer determined there is potential defect relating to change of disposable cuvette roll change, filled cuvettes may wait on the conveyer belt prolonging contact time with plasma & intermediate regent.
Manufacturer
Disetronic Medical Systems AG
Recall Posting Date
2005-08-22
Trade Name
D-TRON Plus Adapter
Recall Start Date
2005-07-16
Recall Number
28683
Hazard Classification
Type I
Model or Catalog #
3000803

Lot or Serial #
N/A

Reason for Recall
Incorrect function may occur within 15 min of replacing the adapter and priming the set. The drop in pressure may result in the delivery of up to 1.8 units of insulin which affects users 13 years and under.
Manufacturer
Disetronic Medical Systems AG
Recall Posting Date
2005-08-08
Trade Name
D-TRON Plus Adapter
Recall Start Date
2005-07-06
Recall Number
28686
Hazard Classification
Type II
Model or Catalog #
3000803
Lot or Serial #
part # 3000803
Reason for Recall
Small percentage of adapters may drop in pressure which may result in the delivery of up to 1.8 units of insulin. This may occur within 15 minutes of replacing the adapter and priming the set.
Manufacturer
ELA Medical, S.A.
Recall Posting Date
2005-09-05
Trade Name
A) Alto VR625 Cardioverter Defibrillator

B) Alto DR614 Cardioverter Defibrillator

C) Alto 2 DR624 Cardioverter Defibrillator
Recall Start Date
2005-07-20
Recall Number
28746
Hazard Classification
Type I
Model or Catalog #
A) N/A

B) 614

C) N/A
Lot or Serial #
A) 304XI002, 304XI108, 402XI001, 404XI049, 413XI027, 413XI035

B) 222XF028, 222XF035, 222XF085, 222XF079, 222XF033

C) 314XJ284, 238XJ029, 403XJ002, 403XJ092, 409XJ014, 409XJ036, 410XJ019, 410XJ024, 410XJ025, 410XJ031, 410XJ042
Reason for Recall
Premature battery depletion and
Manufacturer
Electro Medical Systems (EMS) SA
Recall Posting Date
2005-07-25
Trade Name
Swiss Lithoclast Ultrasound Handpiece
Recall Start Date
2005-06-17
Recall Number
28544
Hazard Classification
Type II
Model or Catalog #
FR-093
Lot or Serial #
20124, 20172, 20268, 29342, 20372, 20403, 20406
Reason for Recall
Manufacturer has become aware that handpieces have twisted cables and there is a small possibility that they may not function properly or the function may diminish over time.
Manufacturer
Euro
Recall Posting Date
2005-07-11
Trade Name
Immulite CEA
Recall Start Date
2005-05-25
Recall Number
28484
Hazard Classification
Type II
Model or Catalog #
LKCE1
Lot or Serial #
302, 303
Reason for Recall
Manufacturer has determined an adverse trend in stability which may result in significant decrease in adjustment slopes and a positive shift of 20-30% at low conc., 20% at higher conc. of patient results.
Manufacturer
Gambro Dasco S.P.A.
Recall Posting Date
2005-08-08
Trade Name
Prismaflex Software 1.07.3 KJ
Recall Start Date
2005-05-27
Recall Number
28506
Hazard Classification
Type II
Model or Catalog #
6023014700
Lot or Serial #
All serial numbers
Reason for Recall
Pumps may stop while green light is on with no audible alarm. No hemodynamic problems but access eventually clot as well as lost blood in set.
Manufacturer
Gambro Dasco S.P.A.
Recall Posting Date
2005-09-05
Trade Name
Prisma Continuous Replacement System
Recall Start Date
2005-08-16
Recall Number
28774
Hazard Classification
Type I
Model or Catalog #
N
Lot or Serial #
N
Reason for Recall
Serious injuries or deaths have resulted because of excessive ultrafiltration. This can occur when the cause of the "incorrect weight change detected" alarm is not identified and removed by the user.
Manufacturer
Guidant Corporation Cardiac Surgery
Recall Posting Date
2005-08-08
Trade Name
Heartstring II Proximal Seal System
Recall Start Date
2005-07-07
Recall Number
28623
Hazard Classification
Type II
Model or Catalog #
N
Lot or Serial #
N
Reason for Recall
Failure modes include incomplete or no Aortotomy with use of device on unaltered tissue, or incomplete Aortotomy when device is used on altered tissue (e.g. Cardioplegia hole or Aortotomy incision).
Manufacturer
Immucor Inc.
Recall Posting Date
2005-09-19
Trade Name
Galileo Immuno Haematology System
Recall Start Date
2005-06-30
Recall Number
28664
Hazard Classification
Type II
Model or Catalog #
64999
Lot or Serial #
N
Reason for Recall
The immediate spin crossmatch (IS_XM) assay performed on the Galileo may not give the expected incompatible result.
Manufacturer
Implant Innovations, Inc.
Recall Posting Date
2005-07-11
Trade Name
ITHA52 Healing Abutments
Recall Start Date
2005-05-18
Recall Number
28369
Hazard Classification
Type III
Model or Catalog #
ITHA52
Lot or Serial #
330918
Reason for Recall
The colour code on the label to identify the implant platform diameter was incorrect. The label has a yellow stripe instead of a blue stripe.
Manufacturer
Instrumentation Laboratory Co.
Recall Posting Date
2005-09-19
Trade Name
A) ACL Futur A/ACLl Advance

B) ACL Top
Recall Start Date
2005-05-19
Recall Number
28468
Hazard Classification
Type II
Model or Catalog #
A) 280000, 999001

B) N/A
Lot or Serial #
A)Serial number starting with 05041164, Serial number with 0505354.

B) N/A
Reason for Recall
Patient prothrombin time (PT) may report low for individuals on anticoagulent therapy.
Manufacturer
I-Stat Corporation
Recall Posting Date
2005-09-05
Trade Name
I-Stat G3+ Cartridges
Recall Start Date
2005-08-05
Recall Number
28778
Hazard Classification
Type III
Model or Catalog #
N
Lot or Serial #
L05048, L05088
Reason for Recall
Certain lots of I-stat G3+ cartridges may exhibit higher than expected rates of star-outs (***) in place of oxygen results.
Manufacturer
Keir Surgical Ltd.
Recall Posting Date
2005-09-19
Trade Name
Pyramidal Tip Obturator
Recall Start Date
2005-08-09
Recall Number
28751
Hazard Classification
Type I
Model or Catalog #
KS81-1055
Lot or Serial #
KS81-1055
Reason for Recall
Distal tip of obturator, lot# 10123, separated from the proximal shaft. A recent affected device was found after initial company recall done on April 2002.
Manufacturer
Kyphon Inc.
Recall Posting Date
2005-07-11
Trade Name
KyphX HV-R Bone Cement
Recall Start Date
2005-03-14
Recall Number
27927
Hazard Classification
Type III
Model or Catalog #
N
Lot or Serial #
EL08304
Reason for Recall
US version of the instructions for use accompanied product sent to Canada.
Manufacturer
Lifescan Inc.
Recall Posting Date
2005-08-22
Trade Name
One Touch UltraSmart Meter
Recall Start Date
2005-06-15
Recall Number
28489
Hazard Classification
Type III
Model or Catalog #
011-555, 020-720
Lot or Serial #
All lots.
Reason for Recall
Glucose meter does not prompt user to take a snack at glucose levels over the entire range of 1.1MMOL
Manufacturer
Lifescan Inc.
Recall Posting Date
2005-09-05
Trade Name
One Touch UltraSmart Meter
Recall Start Date
2005-06-15
Recall Number
28489
Hazard Classification
Type III
Model or Catalog #
011-555, 020-720
Lot or Serial #
All lots.
Reason for Recall
CORRECTION TO RECALL # 28489 PREVIOUSLY POSTED ON 2005-08-22.

Labelling incorrectly states that the 'snack' message will appear on .Meter between 1.1mmol/l to 3.9 mmol/l, labelling should state between 1mmol/l to 3.3 mmol/l.
Manufacturer
Masimo Corporation
Recall Posting Date
2005-07-25
Trade Name
Set Radical Handheld Pulse Oximeter
Recall Start Date
2005-06-13
Recall Number
28558
Hazard Classification
Type II
Model or Catalog #
N
Lot or Serial #
Less than or equal to 102999 - P3027 through P3148
Reason for Recall
The internal speaker could fail to produce an audio alarm during an alarm condition. The affected units were manufactured prior to December 11, 2001.
Manufacturer
MISYS Healthcare Systems Inc.
Recall Posting Date
2005-08-08
Trade Name
MISYS Laboratory System Blood Bank
Recall Start Date
2005-05-10
Recall Number
28492
Hazard Classification
Type III
Model or Catalog #
N
Lot or Serial #
Sites using software ver. 6.1
Reason for Recall
CORRECTION TO RECALL # 28492 PREVIOUSLY POSTED ON 2005-07-11.

Information entered into the comment field may be lost when resulting the blood type for the first time via a blood bank instrument interface. The software problem will be corrected with the installation of a software patch.
Manufacturer
MISYS Healthcare Systems Inc.
Recall Posting Date
2005-07-11
Trade Name
MISYS Laboratory System
Recall Start Date
2005-05-13
Recall Number
28491
Hazard Classification
Type I
Model or Catalog #
SQI-MIC-SY
Lot or Serial #
LARS versions 5.3 or later
Reason for Recall
In rare situations, modification of previously reported specimen results using the lars application, may modify other previously filed results on the same specimen.
Manufacturer
MISYS Healthcare Systems Inc.
Recall Posting Date
2005-07-11
Trade Name
MISYS Laboratory System Blood Bank
Recall Start Date
2005-05-10
Recall Number
28492
Hazard Classification
Type III
Model or Catalog #
N
Lot or Serial #
Sites using software ver. 6.1
Reason for Recall
When a previously reported microbiology organism result is deleted on a selected patient, reported susceptibilities will not stay with associated organisms and one or more susceptibilities may be deleted.
Manufacturer
Nichols Institute Diagnostics
Recall Posting Date
2005-07-25
Trade Name
Bio-intact PTH (I-84) Assay
Recall Start Date
2005-03-25
Recall Number
28457
Hazard Classification
Type II
Model or Catalog #
62-7040
Lot or Serial #
62-402622
Reason for Recall
Internal testing indicates that the assay does not meet the following performance specifications stated in the directional insert functional sensitivity, reproducibility, parallelism, recover and interfaces.
Manufacturer
Philips Medical Systems (Cleveland), Inc.
Recall Posting Date
2005-08-08
Trade Name
A) Brilliance CT 16 Slice

B) Brilliance CT 16 Power

C) Brilliance CT 40 Channel
Recall Start Date
2005-07-18
Recall Number
28640
Hazard Classification
Type II
Model or Catalog #
A) 453567078851

B) 453567005721

C) 455011004011
Lot or Serial #
A) 5074

B) 6110

C) 9081, 6072
Reason for Recall
Couch positions that are reported on the gantry control panel are correct during a CCT procedure, but the couch positions that are displayed on the monitor may be incorrect.
Manufacturer
Philips Medical Systems (Cleveland), Inc.
Recall Posting Date
2005-09-19
Trade Name
Gemini GXL 6
Recall Start Date
2005-08-25
Recall Number
28819
Hazard Classification
Type II
Model or Catalog #
4535 679 72021
Lot or Serial #
4004
Reason for Recall
An anomaly was identified in the gemini ct image reconstruction subsystem of the Gemini GXL Pet
Manufacturer
Philips Medical Systems Nederland B.V.
Recall Posting Date
2005-07-11
Trade Name
Viewforum Software
Recall Start Date
2005-05-27
Recall Number
28501
Hazard Classification
Type II
Model or Catalog #
M1M1171
Lot or Serial #
5V9371J, 00F1F8B16DA, 00F1F9385CE
Reason for Recall
The "automated analysis of vascular obstruction" tool in the Viewforum 2003 software gives unreliable calibration and analysis results when using unprocessed images acquired on the Philips Allura FD10.
Manufacturer
Radiotherapeutics Corporation
Recall Posting Date
2005-08-22
Trade Name
Leveen Coaccess Electrode
Recall Start Date
2005-08-02
Recall Number
28745
Hazard Classification
Type II
Model or Catalog #
26-216, 26-217
Lot or Serial #
7108067, 6556464, 6826039, 6666258, 6556462, 7356251, 31804, 7021736, 7091462, 21403R, 6558336, 6538341, 6556258, 6538336, 7257089, 7481332
Reason for Recall
Manufacturer has received reports of extended ablation times, lack of roll-off, and
Manufacturer
Richard Wolf GMBH
Recall Posting Date
2005-09-05
Trade Name
Scissors Insert for Single Use
Recall Start Date
2005-04-29
Recall Number
28206
Hazard Classification
Type II
Model or Catalog #
8392.0403, 8393.0403, 8394.0403
Lot or Serial #
953046, 9532217, 953486
Reason for Recall
Damage to the insulation part may arise by unintentional "jamming" of the scissor on the existing edges in the access instrument when being introduced or pulled out. Insulation material may come off and fall.
Manufacturer
Roche Molecular Systems, Inc.
Recall Posting Date
2005-07-11
Trade Name
Six-Part Bar Code Labels, Sarstedt Tubes
Recall Start Date
2005-05-17
Recall Number
28326
Hazard Classification
Type I
Model or Catalog #
N
Lot or Serial #
28868, 029213
Reason for Recall
Rolls of six-part bar code labels, roche M
Manufacturer
Sensormedics Corporation
Recall Posting Date
2005-08-08
Trade Name
Dell AC Adapter
Recall Start Date
2005-03-11
Recall Number
28669
Hazard Classification
Type III
Model or Catalog #
22536, 22538
Lot or Serial #
N
Reason for Recall
Dell recalled the adapter units due to a risk of overheating which could lead to fire and electrical shock. The adapters are used in conjunction with laptops-peripheral components of vmax and sleep systems.
Manufacturer
Siemens Aktiengesellschaft
Recall Posting Date
2005-08-22
Trade Name
A) Axiom Artis BA

B) Axiom Artis DFC

C) Axiom Artis FA ALPHA

D) Axiom Artis MP

E) Axiom Artis TA
Recall Start Date
2005-07-26
Recall Number
28718
Hazard Classification
Type III
Model or Catalog #
A) 5904656

B) 7412807, 7727717

C) 5904441

D) 5904466

E) 7007755
Lot or Serial #
A) N/A

B) 35195, 35261, 35219, 35298, 40033, 25006, 25076, 20070, 20215, 28064, 32070

C) N/A

D) N/A

E) N/A
Reason for Recall
The live display of the Axiom Artis in the examination room may sporadically stop working.
Manufacturer
Siemens Aktiengesellschaft
Recall Posting Date
2005-07-25
Trade Name
Axiom Multistar
Recall Start Date
2005-06-17
Recall Number
28539
Hazard Classification
Type III
Model or Catalog #
3772501, 3773004
Lot or Serial #
1027, 1094, 1142, 1163, 1229, 1290, 1795, 1818, 1102, 1112, 1113, 1129
Reason for Recall
A system error may occur which results in unintended movement during examination.
Manufacturer
Siemens Aktiengesellschaft
Recall Posting Date
2005-07-11
Trade Name
A) Axiom Aristos FX

B) Axiom Aristos MX
Recall Start Date
2005-06-14
Recall Number
28493
Hazard Classification
Type III
Model or Catalog #
A) 7414803

B) 5895003
Lot or Serial #
A) 1298, 1120, 1129, 1131, 1292, 1288, 1086, 1236, 1006, 1243, 1018, 1311, 1315, 1326, 1340, 1354, 1153, 1443, 1337, 1173, 1190, 1244, 1072,
1273, 1175, 1124, 1134, 1272, 1333, 1145, 1386,1065, 1328, 1162, 1128, 1119, 1176.

B) N/A
Reason for Recall
A system lock-up may occur when repeating patient registrations under a specific workflow.
Manufacturer
Siemens Aktiengesellschaft
Recall Posting Date
2005-07-11
Trade Name
Magnetom Avanto System
Recall Start Date
2005-06-21
Recall Number
28493
Hazard Classification
Type II
Model or Catalog #
7391167
Lot or Serial #
25285, 25340, 25043, 25039, 25234, 25173, 25089, 25110, 25066, 25137, 25049, 25097, 25311, 25168, 25333, 25224
Reason for Recall
For the safety of the service technicians, the acc power connection X102 need to be checked and labelled.
Manufacturer
Siemens Aktiengesellschaft
Recall Posting Date
2005-09-05
Trade Name
A) Arcadis Varic

B) Siremobil Compact
Recall Start Date
2005-08-08
Recall Number
28754
Hazard Classification
Type III
Model or Catalog #
A) 8080017

B) 3780629
Lot or Serial #
A) N/A

B) 10051, 10243, 10256, 1639
Reason for Recall
The screws fixing the laser diode out windows can become loosened.
Manufacturer
Siemens Aktiengesellschaft
Recall Posting Date
2005-09-05
Trade Name
A) Magnetom Avanto System

B) Manetom Espree
Recall Start Date
2005-08-18
Recall Number
28779
Hazard Classification
Type II
Model or Catalog #
A) 7391167

B) 7727121
Lot or Serial #
A) N/A


B) 25285, 25490, 25340, 25043, 30048, 25039, 25541, 25234, 25462, 25173, 25089, 25526, 25110, 25066, 25137, 25486, 25049, 25097, 25311, 25168, 25333, 25555, 25244
Reason for Recall
A misalignment of the patient table may occur between the patient table assembly and the opening to the magnet bore.
Manufacturer
Siemens Medical Solutions USA, Inc.
Recall Posting Date
2005-07-25
Trade Name
Sonoline Antares Diagnostic Ultrasound
Recall Start Date
2005-06-05
Recall Number
28540
Hazard Classification
Type II
Model or Catalog #
5936518
Lot or Serial #
112607, 112593, 112609, 112611, 112530, 112595, 112612, 112687, 112610, 10077, 10145, 10280, 111223, 111233, 111234, 112579, 10438, 111197, 112736, 10068, 111843, 111922, 111923, 111639, 111643, 111660, 111666, 10928, 11092, 11096, 112658, 112666, 10838, 10925, 112237, 10134, 112267, 112070, 112073, 10940, 111166
Reason for Recall
Error occurs after performing a labeled measurement on a frozen waveform utilizing a 2d image and an angle-corrected pulse wave doppler waveform.
Manufacturer
Smith & Nephew, Inc. Orthopaedic Division
Recall Posting Date
2005-07-11
Trade Name
Compression Hip Screw System-Lag Screw
Recall Start Date
2005-04-11
Recall Number
28190
Hazard Classification
Type III
Model or Catalog #
12-1108
Lot or Serial #
04LT84854
Reason for Recall
Chart stick label and inner tyvek pouch label state that the product is a hex screwdriver. The outside label correctly identifies the product.
Manufacturer
STAAR Surgical Company
Recall Posting Date
2005-08-22
Trade Name
STAAR Surgical STAARVISC II
Recall Start Date
2005-07-18
Recall Number
28728
Hazard Classification
Type II
Model or Catalog #
N
Lot or Serial #
B041118C
Reason for Recall
The cannula package has not been validated to a sterility assurance level (sal) of 0.000001 and therefore the sterility can not be assured.
Manufacturer
Sysmex Corporation
Recall Posting Date
2005-08-22
Trade Name
A) CA1500 Instrument

B) CA540 Instrument

C) CA560 Instrument
Recall Start Date
2005-07-21
Recall Number
28685
Hazard Classification
Type II
Model or Catalog #
A) CA-1500WBCR

B) B4260-540

C) B4260-560
Lot or Serial #
A) A2399, A2483, A2400, A3358, A2401, A2727, A1888, A3295, A3276, A3632, A2620, A2739, A1825, A2270, A2753, A2522, A1556, A3905, A2480, A4001, A2636, A3546, A2320, A1465, A1151, A2646, A4047, A4050, A3548, A2713, A2269, A2271, A3634, A1646, A2182, A2170, A3521, A4028, A2171, A3284, A1105, A1581, A2427, A2257, A2632, A2621, A1644, A2296.

B) A3616, A2858, A3359, A2530, A2578, A2353, A2536, A3250, A1307, A2384, A3637, A1449, A2993, A2954, A3729, A2352, A3406, A3249, A2005, A2996, A2246, A2854, A2308, A3273, A3274, A3687, A3360, A2841, A1452, A3482, A2859, A3407, A3403, A2474, A2351, A2953, A1309, A2027, A2248, A3636, A2997, A3602, A3477, A3600, A3599, A3345, A2861, A2309, A2995, A2532, A1308, A2531, A3446. A3417, A3548, 2026, A2759, A2003, A1454, A3479, A3297, A3321, A3320, A2952, A3361, A3362, A3394, A3478, A3447, A2004B, A2994, A3404, A1456, A3409, A2862, A2863, A2857, A2860, A3452, A3319, A3344, A3401, A3483, A3450, A3475, A2490, A1450, A2395, A2840, A2955, A3405, A3296, A2951, A3311, A3318, A3448, A3662, A3772, A1457, A3476, A3667, A3783, A3927, A2337.

C) A1323, A1413, A1088, A1821, A1336, A1633, A1276, A1315, A1102. A1850, A1858, A1105, A1510, A1395, A1870, A1866, A1856, A1846, A1869, A1474, A1103, A1113, A1839, A1108, A1290, A1861, A1265, A1278, A1672, A1630, A1198, A1770.
Reason for Recall
Manufacturer has found an issue when using the akima method of calculating the standard curve. A change to the standard curve method for this assay is needed.
Manufacturer
Varian Medical Systems, Inc.
Recall Posting Date
2005-08-08
Trade Name
Multileaf Collimater (MLC)+Software
Recall Start Date
2005-07-01
Recall Number
28682
Hazard Classification
Type II
Model or Catalog #
N
Lot or Serial #
H520003-H520943
H530012-H534223
H540381-H549004
Reason for Recall
Potential mistreatment may occur because initial MLC shape may be an open field and not match planned MLC field when executing dynamic MLC treatment types.
Manufacturer
Zimmer Orthopaedic Surgical Products
Recall Posting Date
2005-07-11
Trade Name
Disposable Tourniquet Cuffs
Recall Start Date
2005-03-10
Recall Number
28499
Hazard Classification
Type II
Model or Catalog #
60707000100, 60707000300, 60707000400, 60707000500, 60707000600
Lot or Serial #
60154121, 60171346, 60162824, 60171348, 60180450, 60171347, 60173808, 60194151, 60194152, 60188441, 60184634, 60202399, 60205580, 60208192, 60176164
Reason for Recall
Some units manufactured with a lower separation strength for the port-to-tube connection. The port