Health Canada
www.hc-sc.gc.ca
Home
> Drugs & Health Products
> Compliance & Enforcement
> Recalls
> Recalls
Institutional links
-
Back to
-
Explore...
-
Proactive Disclosure
Medical Device Recall Listings - Sorted by Manufacturer
(From July 2010 - September 2010)
- Manufacturer
- Abbott Laboratories Diagnostic Division
- Recall Posting Date
- 2010-08-30
- Trade Name
- A) Cell-Dyn 4000 Vent Needle and Accessory Kit
B) Cell-Dyn Sapphire Vent Head Assembly and Accessory Kit
- Recall Start Date
- 2010-08-06
- Recall Number
- 56047
- Hazard Classification
- Type III
- Model or Catalog #
- A) 01H01-01
B) 08H00-01
- Lot or Serial #
- A) N/A
B) N/A
- Reason for Recall
- During manufacturing, the aspiration probe was hitting the top portion of the vent head assembly and would not go all the way down. Upon further evaluation, it was found that the vent needle was not manufactured correctly by the supplier.
- Manufacturer
- Abbott Laboratories Diagnostic Division
- Recall Posting Date
- 2010-09-27
- Trade Name
- A) Cell-Dyn 3700 Haematology System-SL Analyser
B) Cell-Dyn 3200SL System
C) Cell-Dyn 3700 Haematology System-CS Analyser
- Recall Start Date
- 2010-08-31
- Recall Number
- 56461
- Hazard Classification
- Type III
- Model or Catalog #
- A) 02H31-01
B) 4H60-01
C) 02H30-01
- Lot or Serial #
- A) All serial numbers
B) All serial numbers
C) All serial numbers
- Reason for Recall
- Cell-Dyn 3700 customers had reported occurrences of visible fire and smoke from the analyzer. In addition, a Cell-Dyn 3200 customer reported smoke from the analyzer. Returned power supply units were analysed and were identified to contain the incorrect fuse for the electrical configuration at the customer site.
- Manufacturer
- Abbott Point of Care Inc.
- Recall Posting Date
- 2010-08-30
- Trade Name
- i-Stat Level 2 Control Value Assignment Sheet (VAS)
- Recall Start Date
- 2010-08-02
- Recall Number
- 55982
- Hazard Classification
- Type III
- Model or Catalog #
- 06F13-01
- Lot or Serial #
- B09342, B09154
- Reason for Recall
- Abbott has determined that the i-Stat level 2 Control Value Assignment Sheets contain the incorrect range for PCO2, blood urea nitrogen (BUN) and glucose for the i-Stat cartridges. Abbott has made the decision to update the affected value assignment sheets to ensure the performance of their products.
- Manufacturer
- Accuray, Inc.
- Recall Posting Date
- 2010-08-02
- Trade Name
- CyberKnife Robotic Radiosurgery System - Base CyberKnife
- Recall Start Date
- 2010-07-06
- Recall Number
- 55783
- Hazard Classification
- Type II
- Model or Catalog #
- 028000, 030000, 032000
- Lot or Serial #
- C0231, C0202
- Reason for Recall
- When the cosmetic linac cover (the 'nose' cover) is reattached to the linac after removal, the rotating latches that secure the cover may not be properly engaged with the connection pins on the linac. If the latches are not fully engaged, the linac cover can become loose and may unexpectedly drop off the linac. If a patient treatment is initiated while the cover is loose, there is potential for it to contact the patient and lead to injury depending on the position and height of the linac at that time.
- Manufacturer
- Acumed, LLC
- Recall Posting Date
- 2010-07-05
- Trade Name
- Trial Gage
- Recall Start Date
- 2010-05-11
- Recall Number
- 54806
- Hazard Classification
- Type III
- Model or Catalog #
- TR-TG01
- Lot or Serial #
- 211365, 214346, 220365, 222678, 224183
- Reason for Recall
- These parts are being recalled as they were manufactured from a different grade of aluminum than required. The part therefore has an increased risk of breaking which could occur during a surgery.
- Manufacturer
- Advanced Neuromodulation Systems, Inc.
- Recall Posting Date
- 2010-07-19
- Trade Name
- A) Genesis IPG Neurostimulator - Lamitrode (Surgical) Leads
B) Lamitrode S-4 Lead Kits
C) Lamitrode S-8 Lead Kits
D) Lamitrode C Series Leads
E) Lamitrode Neurostimulation Lead Kit
F) Lamitrode S-Series Neurostimulation Lead Kit
G) Lamitrode Tripole 8 Lead Kit
H) Lamitrode Tripole 8C Lead Kit
I) Lamitrode Tripole 16C Lead Kit
J) Lamitrode Exclaim Lead Kit
K) Lamitrode Tripole 16 Lead Kit
- Recall Start Date
- 2010-06-18
- Recall Number
- 55435
- Hazard Classification
- Type III
- Model or Catalog #
- A) 3240, 3244, 3280
B) 3243, 3246
C) 3283, 3286
D) 3245
E) 3254, 3255, 3262, 3263, 3264, 3265
F) 3266, 3267, 3268, 3269
G) 3208
H) 3210
I) 3214
J) 3224
K) 3219
- Lot or Serial #
- A) N/A
B) N/A
C) N/A
D) N/A
E) N/A
F) N/A
G) N/A
H) N/A
I) N/A
J) N/A
K) N/A
- Reason for Recall
- St. Jude Medical (SJM) has identified the potential for the rare occurrence of neurological deficit adverse events such as autonomic symptoms, pain (manifesting itself as chest or abdominal pain) or paralysis associated with implanting the paddle leads listed above.
- Manufacturer
- AGFA Healthcare N.V.
- Recall Posting Date
- 2010-08-16
- Trade Name
- A) NX8000 workstations used in combination with 19'' Barco monitor and mounting stand
B) NX8200 workstations used in combination with 19'' Barco monitor and mounting stand
- Recall Start Date
- 2010-06-30
- Recall Number
- 55730
- Hazard Classification
- Type II
- Model or Catalog #
- A) E5LHV, E5LJX
B) X.0.8200
- Lot or Serial #
- A) N/A
B) N/A
- Reason for Recall
- Attempt to reposition the monitor by pulling or pushing the monitor housing will cause strain on the monitor. Depending on the extent of stress and after an extended period of time the monitor may fail at or near the mounting bracket pins.
- Manufacturer
- American Medical Systems
- Recall Posting Date
- 2010-09-27
- Trade Name
- GreenLight HPS fibers, P/N 0010-2090
- Recall Start Date
- 2010-01-29
- Recall Number
- 53499
- Hazard Classification
- Type III
- Model or Catalog #
- 0010-2090A, 0010-2090B, 0010-2090C,
0010-2090D
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- The laser fibres had intermittent packaging and labelling issues. Three scenarios: HPS 0010-2090 as PV2079, HPS0010-2090 as PV 0010-2090, and product insert may be missing.
- Manufacturer
- Amo Manufacturing USA, LLC
- Recall Posting Date
- 2010-09-13
- Trade Name
- Wavescan Wavefront System and Offline Programming Module (OPM)
- Recall Start Date
- 2010-08-04
- Recall Number
- 56048
- Hazard Classification
- Type III
- Model or Catalog #
- HS-1
- Lot or Serial #
- N/A
- Reason for Recall
- An incorrect patient treatment table may be generated by a coding error within the Wavescan Wavefront system software. The field correction is in regard to the software upgrade of Wavescan Wavefront system and offline programming module (OPM), which is used to determine patient treatment.
- Manufacturer
- Arrow International, Inc.
- Recall Posting Date
- 2010-08-16
- Trade Name
- Central Venous Catheterization Kit
- Recall Start Date
- 2010-07-23
- Recall Number
- 55893
- Hazard Classification
- Type II
- Model or Catalog #
- CA-22702
- Lot or Serial #
- RF0033779
- Reason for Recall
- Arrow has received complaints regarding the introducer needle being "flimsy" and being difficult to insert into the blood vessel. It was determined that some finished products lots had incorrect cannula with a blunt bevel.
- Manufacturer
- B. Braun Medical Inc.
- Recall Posting Date
- 2010-08-02
- Trade Name
- AddEase Binary Connector
- Recall Start Date
- 2010-07-06
- Recall Number
- 55733
- Hazard Classification
- Type I
- Model or Catalog #
- N7990, N7993
- Lot or Serial #
- 60984562, 61019650, 61031167, 61057867
- Reason for Recall
- B.Braun pab bags demonstrate coring when used with AddEase connectors and third party needles. The incidence rate appears to be higher with the use of AddEase connectors.
- Manufacturer
- B. Braun Medical Inc.
- Recall Posting Date
- 2010-08-16
- Trade Name
- Addease Binary Connector
- Recall Start Date
- 2010-07-22
- Recall Number
- 55733
- Hazard Classification
- Type I
- Model or Catalog #
- N7990, N7993
- Lot or Serial #
- 60984562, 61019650, 61031167, 61057867
- Reason for Recall
- B.Braun pab bags demonstrate coring when used with Addease connectors and third party needles. The incidence rate appears to be higher with the use of Addease connectors.
- Manufacturer
- Baxter Healthcare Corporation
- Recall Posting Date
- 2010-09-27
- Trade Name
- A) Micro-Volume Extension Sets with Filter
B) Micro-Volume Extension Set
- Recall Start Date
- 2010-09-07
- Recall Number
- 56507
- Hazard Classification
- Type III
- Model or Catalog #
- A) 2N3350
B) 2N3347
- Lot or Serial #
- A) UR07I25092, UR08B04109, UR08B09058, UR09E12013, UR09F04083, UR192179X, UR242057, UR253450, UR261362
B) UE103267
- Reason for Recall
- Baxter has identified a labelling issue affecting all lots of extension sets 2N3347 and 2N3350, whereby the product configuration and label illustration have the slide clamp above the filter, but the written instructions tell the user to close the clamp below filter when set is not in use. The current clamp configuration could result in the unintended fluid administration of approximately 0.2ml.
- Manufacturer
- Beckman Coulter, Inc.
- Recall Posting Date
- 2010-07-05
- Trade Name
- A) P3, P4 Plates for the PK7200 Automated Microplate System-Microplate Reader
B) P3, P4 Plates for the PK7200 Automated Microplate System-Analyzer Class 3
C) P3, P4 Plates for the PK7300 Automated Microplate System-Instrument Class 3
D) P3, P4 Plates for the PK7300 Automated Microplated System-Instrument
- Recall Start Date
- 2010-04-26
- Recall Number
- 55300
- Hazard Classification
- Type II
- Model or Catalog #
- A) PK7200
B) PK7200
C) PK7300
D) PK7300
- Lot or Serial #
- A) P3-91A,92A,93A,95D,96D, P4-93B,94B,99C
B) P3-91A,92A,93A,95D,96D, P4-93B,94B,99C
C) P3-91A,92A,93A,95D,96D, P4-93B,94B,99C
D) P3-91A,92A,93A,95D,96D, P4-93B,94B,99C
- Reason for Recall
- Beckman Coulter has confirmed the presence of deformed well terraces in the PK Microplate wells may cause fold over or slip down of positive reaction patterns in the PK7200 and PK7300 Automated Microplate System.
- Manufacturer
- Beckman Coulter, Inc.
- Recall Posting Date
- 2010-07-19
- Trade Name
- Access Thyroglobulin Antibody II Reagent
- Recall Start Date
- 2010-05-17
- Recall Number
- 55500
- Hazard Classification
- Type II
- Model or Catalog #
- A32898
- Lot or Serial #
- 916271, 917073, 918433
- Reason for Recall
- Beckman Coulter has confirmed that the three lots of Access Thyroglobulin Antibody II Reagent may produce significantly different results in a small subset of patient samples when compared to previously released reagent lots.
- Manufacturer
- Beckman Coulter, Inc.
- Recall Posting Date
- 2010-08-02
- Trade Name
- A) Unicel DXI 600 Access Immunoassay System - Instrument Class 2
B) Unicel DXI 600 Access Immunoassay System - Instrument Class 3
C) Unicel DXI 800 Access Immunoassay - Analyzer-Class 3
D) Unicel DXI 800 Access Immunoassay System - Analyzer - Class II
- Recall Start Date
- 2010-07-05
- Recall Number
- 55603
- Hazard Classification
- Type II
- Model or Catalog #
- A) A30260
B) A30260
C) 973100
D) 973100
- Lot or Serial #
- A) All lots
B) All lots
C) All lots
D) All lots
- Reason for Recall
- Beckman Coulter has confirmed the following issue with the software version 4.3 being used in the Unicel DXI 600 and DXI 800 Access Immunoassay Systems. If a laboratory changed a default sample type setting on the tests screen while running Unicel DXI System software version 4.3, the new setting would not be saved to the system database. The changed sample type setting would reset to the previously saved setting when power was interrupted to the system, or if an operator rebooted the system.
- Manufacturer
- Beckman Coulter, Inc.
- Recall Posting Date
- 2010-08-30
- Trade Name
- Sex Hormone Binding Globulin (SHBG) Reagent
- Recall Start Date
- 2010-07-26
- Recall Number
- 55983
- Hazard Classification
- Type II
- Model or Catalog #
- A48617
- Lot or Serial #
- 009046, 010202, 829586, 909837, 914557, 915628, 917613, 919184
- Reason for Recall
- The instruction for use (IFU) for access Sex Hormone Binding Globulin (SHBG) does not specify the units of measure to be used when calculating the free androgen index (FAI %). Access SHBG (nmol/l) values and access testosterone (ng/ml) values are reported in different default units. The testosterone value (when reported in ng/ml) requires conversion to nmol/l before the FAI % can be calculated..
- Manufacturer
- Beckman Coulter, Inc.
- Recall Posting Date
- 2010-08-30
- Trade Name
- Synchron System(s) Creatine Reagent
- Recall Start Date
- 2010-08-03
- Recall Number
- 55984
- Hazard Classification
- Type II
- Model or Catalog #
- 472525
- Lot or Serial #
- M911428, M911519
- Reason for Recall
- Beckman Coulter Inc. has confirmed reports that black/brown particulates were observed in some bottles of the creatinine alkaline buffer in modular Creatinine Reagent Kit lots M911428 and M911519.
- Manufacturer
- Beckman Coulter, Inc.
- Recall Posting Date
- 2010-08-30
- Trade Name
- Synchron Systems Glucose Reagent (GLU)
- Recall Start Date
- 2010-08-09
- Recall Number
- 56165
- Hazard Classification
- Type II
- Model or Catalog #
- 442640
- Lot or Serial #
- All non-expired lots
- Reason for Recall
- Synchron Systems Glucose Reagent (GLU) when used in conjunction with Unicel DXC and Synchron LX instruments may give falsely high results for moderately hemolysed samples.
- Manufacturer
- Beckman Coulter, Inc.
- Recall Posting Date
- 2010-09-13
- Trade Name
- Navios Flow Cytometer
- Recall Start Date
- 2010-06-21
- Recall Number
- 55631
- Hazard Classification
- Type III
- Model or Catalog #
- A52101, A52102, A52103
- Lot or Serial #
- AN29103-AS15099, AN06024-AS03008, AM51001-AS15102
- Reason for Recall
- Beckman Coulter has confirmed an issue with the Navios Flow Cytometer with version 1.0 software: red and violet laser related errors are not displayed in the event of a laser failure. The blue laser is not impacted by this issue.
- Manufacturer
- Beckman Coulter, Inc.
- Recall Posting Date
- 2010-09-13
- Trade Name
- Synchron Systems Enzymatic CO2 (CO2E) Reagent
- Recall Start Date
- 2010-07-12
- Recall Number
- 55665
- Hazard Classification
- Type III
- Model or Catalog #
- A60290
- Lot or Serial #
- M907574
- Reason for Recall
- Blank rate high (SH) instrument flag may be generated when less than 30 tests left in the CO2E cartridge and recalibration attempts may fail or may not be maintained.
- Manufacturer
- Beckman Coulter, Inc.
- Recall Posting Date
- 2010-09-13
- Trade Name
- A) Access System - Immunoassay Aanalyzers Class 2
B) Unicel DXC 600I Synchron Access Clinical Analyzer
C) Unicel DXI 600 Access Immunoassay System - Instrument Class 2
D) Unicel DXI 600 Access Immunoassay System - Instrument Class 3
E) Unicel DXC 880I Synchron Access Clinical Analyzer
F) Access Immunoassay Analyzers - Class 3
G) Access Immunoassay System - Analyzer Class 3
H) Access System Access 2 Analyzer - Class 2
I) Unicel DXI 800 Access Immunoassay System - Analyzer - Class III
J) Unicel DXI 800 Access Immunoassay System - Analyzer - Class II
K) Synchron LXI 725
L) Unicel DXC 660I Synchron Access Clinical Analyzer
M) Unicel DXC 680I Synchron Access Clinical Analyzer
N) Unicel DXC 860I Synchron Access Clinical Analyzer
- Recall Start Date
- 2010-08-09
- Recall Number
- 56259
- Hazard Classification
- Type II
- Model or Catalog #
- A) 81600
B) 4767
C) A30260
D) A30260
E) 4768
F) 81600
G) 81600N
H) 81600N
I) 973100
J) 973100
K) 476501
L) 4771
M) 4772
N) 4773
- Lot or Serial #
- A) All lots
B) All lots
C) All lots
D) All lots
E) All lots
F) All lots
G) All lots
H) All lots
I) All lots
J) All lots
K) All lots
L) All lots
M) All lots
N) All lots
- Reason for Recall
- If a reagent pack is not loaded onto Access and DXI instruments in accordance with the instrument operator's guide, it is possible for the instrument to pipette insufficient or incorrect reagents and generate erroneous quality control or patient results.
- Manufacturer
- Beckman Coulter, Inc.
- Recall Posting Date
- 2010-09-13
- Trade Name
- SYNCHRON System(s) Immunoglobulin A (Ig-A) Reagent
- Recall Start Date
- 2010-06-28
- Recall Number
- 55480
- Hazard Classification
- Type II
- Model or Catalog #
- 467920
- Lot or Serial #
- M812164, M902345
- Reason for Recall
- Ig-A Reagent lots M812164 and M902345 contain an unidentified cross reacting antibody that may yield falsely high Ig-A results in survey samples for immunoglobulins which are formulated with Ig-M monoclonal antibodies.
- Manufacturer
- Beckman Coulter, Inc.
- Recall Posting Date
- 2010-09-27
- Trade Name
- A) AU5400 Clinical Chemistry Analyzer
B) AU2700 Clinical Chemistry Analyzer
C) AU400/AU400E Clinical Chemistry Analyzers
D) AU600/AU640/AU640E Clinical Chemistry Analyzers
E) AU680 Clinical Chemistry Analyzer
F) AU480 Clinical Chemistry Analyzer
- Recall Start Date
- 2010-07-12
- Recall Number
- 56411
- Hazard Classification
- Type I
- Model or Catalog #
- A) AU5400
B) AU2700
C) AU400, AU400E
D) AU600, AU640, AU640E
E) AU680
F) AU480
- Lot or Serial #
- A) All serial numbers
B) All serial numbers
C) All serial numbers
D) All serial numbers
E) All serial numbers
F) All serial numbers
- Reason for Recall
- Company has received reports of incorrect results generated by analyzers as a result of cuvette overflow. Current system software unable to complete real-time monitoring of reaction profiles to identify impact of cuvette overflow on results, yielding incorrect results without any flags or alarms to alert operator.
- Manufacturer
- Becton Dickinson and Company
- Recall Posting Date
- 2010-09-13
- Trade Name
- BD Gaspak EZ Large Incubation Container
- Recall Start Date
- 2010-08-17
- Recall Number
- 56228
- Hazard Classification
- Type III
- Model or Catalog #
- 260672
- Lot or Serial #
- 7110573
- Reason for Recall
- Product may not maintain air tight environment due to a manufacturing defect. The defect is a missing clear plastic post on the inside of the container lid that is found in the middle position.
- Manufacturer
- Becton Dickinson and Company
- Recall Posting Date
- 2010-09-27
- Trade Name
- BBL Staphyloslide Latex Test Kit
- Recall Start Date
- 2010-07-19
- Recall Number
- 56555
- Hazard Classification
- Type III
- Model or Catalog #
- 240952, 240953
- Lot or Serial #
- 16605410, 16640910, 16679310, 16778710, 16781210, 16781510,
16821684
- Reason for Recall
- Kits may exhibit false agglutination with non-staphylococcus aureus colonies as well as the negative control reagent. Quality control should be performed each day the kit is used. The kit should not be used if the reactions with the control organisms are incorrect.
- Manufacturer
- Becton Dickinson Diagnostic Instrument Systems
- Recall Posting Date
- 2010-09-13
- Trade Name
- Bactec FX
- Recall Start Date
- 2010-08-12
- Recall Number
- 56174
- Hazard Classification
- Type III
- Model or Catalog #
- 441385
- Lot or Serial #
- FT001 to FT0906, FT0922
- Reason for Recall
- Patient sample database on the Bactec FX are sometimes found to be unusable upon rebooting the instrument. The reboot proceeds normally until it tries to load the database file, at which point the instrument reboots and then repeats this cycle indefinitely or customers see a message on the display that says "please insert a bootable device". The system will continue to agitate and incubate the samples.
- Manufacturer
- Becton Dickinson Infusion Therapy Systems Inc.
- Recall Posting Date
- 2010-08-30
- Trade Name
- A) BD Q-Syte Extension Set
B) MPS Acadia Extension Set With Rightbore
C) MPS Acadia Specialty Tri-Extension Set
D) Dual Port BD Nexiva Closed IV Catheter
Updated recall #51661 previously posted on 2009-12-07.
- Recall Start Date
- 2009-11-02
- Recall Number
- 51661
- Hazard Classification
- Type I
- Model or Catalog #
- A) 385102
B) 385150, 385151
C) 385164
D) 383530, 383531
- Lot or Serial #
- A) 8308330
B) A1967, A2399
C) A1951
D) 8238450, 8345951, 9013867
- Reason for Recall
- Potential risk of end-users inadvertently taking a BD Q-Syte device from BD Nexiva product or package, and placing it on a central venous catheter.
- Manufacturer
- Biomerieux SA
- Recall Posting Date
- 2010-08-16
- Trade Name
- Vidas System Rub IGG II Assay
- Recall Start Date
- 2010-07-21
- Recall Number
- 55865
- Hazard Classification
- Type II
- Model or Catalog #
- 30221
- Lot or Serial #
- 837810801
- Reason for Recall
- One or the two pouch of lot #837810801 may contain out-of-specification cones that were affected by a manufacturing incident.
- Manufacturer
- Biomerieux, Inc.
- Recall Posting Date
- 2010-08-30
- Trade Name
- A) Vitek 2 Compact 30 System - Test Cards
B) Vitek 2 Compact 60 System - Test Cards
C) Vitek 2 System - Test Cards
- Recall Start Date
- 2010-08-02
- Recall Number
- 55961
- Hazard Classification
- Type II
- Model or Catalog #
- A) 22225
B) 22225
C) 22225
- Lot or Serial #
- A) 345158910
B) 345158910
C) 345158910
- Reason for Recall
- There are small slits on the backsides of Vitek 2 AST-N085 lot 345158910 foil pouches near the seam separating the desiccant from the main body of the pouch.
- Manufacturer
- Biomerieux, Inc. Portland
- Recall Posting Date
- 2010-08-16
- Trade Name
- O-F Dextrose Tubed Media
- Recall Start Date
- 2010-06-08
- Recall Number
- 55201
- Hazard Classification
- Type III
- Model or Catalog #
- T7141
- Lot or Serial #
- 313295-1
- Reason for Recall
- O-F Dextrose labelling information on the tubes incorrectly described the product as O-F Staph.
- Manufacturer
- Biomet Microfixation, Inc.
- Recall Posting Date
- 2010-09-13
- Trade Name
- Rees Insulated Forceps
- Recall Start Date
- 2010-06-30
- Recall Number
- 55527
- Hazard Classification
- Type II
- Model or Catalog #
- 02-0461
- Lot or Serial #
- 042308D08, 112108D08, 031309C09, 041009D09, 042310B10, 021910B10, 092707I07, 103607J07, 032607C07, 041607D07
- Reason for Recall
- Cracking or peeling off of the insulation of the forceps, causing burns to the patients.
- Manufacturer
- Biomet Sports Medicine Inc.
- Recall Posting Date
- 2010-07-05
- Trade Name
- Artificial Ligament Fixation Device
- Recall Start Date
- 2010-03-31
- Recall Number
- 53910
- Hazard Classification
- Type II
- Model or Catalog #
- 904755
- Lot or Serial #
- 050400, 089710, 648470, 192250, 228610
- Reason for Recall
- The ziploop extended toggleloc device may contain an out of specification saddle length (too long), which could result in insufficient graft being introduced into the surgical tunnel.
- Manufacturer
- Bio-Rad
- Recall Posting Date
- 2010-07-05
- Trade Name
- Strep B Select
- Recall Start Date
- 2010-06-08
- Recall Number
- 55233
- Hazard Classification
- Type II
- Model or Catalog #
- 63750
- Lot or Serial #
- 0D0039
- Reason for Recall
- The lot 0D0039 gives an abnormal rate of blue colonies that were not further confirmed as GBS (group B streptococcus). On this particular lot, some enterococci could produce blue colonies instead of the pink to violet colonies that are expected for these organisms.
- Manufacturer
- Bio-Rad Laboratories Deeside
- Recall Posting Date
- 2010-08-30
- Trade Name
- IN2IT - A1C Test Cartridges
- Recall Start Date
- 2010-08-10
- Recall Number
- 56054
- Hazard Classification
- Type II
- Model or Catalog #
- 2810001EX
- Lot or Serial #
- All lots
- Reason for Recall
- Bio-Rad notifies in2it users that when performing in2it testing at altitudes above 1 000 meters (3 280 feet) they may obtain high results. This problem can occur on all lots of A1C test cartridges.
- Manufacturer
- Boston Scientific Corporation
- Recall Posting Date
- 2010-07-19
- Trade Name
- A) Synchro 2 Soft Guidewire
B) Synchro 2 Standard Guidewire
C) Synchro PVS 1300 Neuro Guidewire
- Recall Start Date
- 2010-06-07
- Recall Number
- 55223
- Hazard Classification
- Type II
- Model or Catalog #
- A) M00326010, M00326310
B) M00326410, M00326510
C) M00313010, M00313020, M00313310
- Lot or Serial #
- A) N/A
B) N/A
C) N/A
- Reason for Recall
- Product may exhibit a PTFE coating issue that could impact patient safety, more specifically, Boston Scientific Corporation has identified scrapes on the coating during production of the affected lots, which could lead to flaking.
- Manufacturer
- Boston Scientific Corporation
- Recall Posting Date
- 2010-08-30
- Trade Name
- A) Sterling Monorail PTA Balloon
B) Sterling Over-The-Wire PTA Balloon
- Recall Start Date
- 2010-05-11
- Recall Number
- 54721
- Hazard Classification
- Type III
- Model or Catalog #
- A) H74939135152010, H74939135202010, H74939135203010, H74939135204010, H74939135252010, H74939135253010, H74939135254010, H74939135302010, H74939135303010, H74939135304010, H74939135352010, H74939135354010, H74939135402010, H74939135403010, H74939135404010
B) H74939134152010, H74939134202010, H74939134203010, H74939134204010, H74939134252010, H74939134253010, H74939134254010, H74939134302010, H74939134303010, H74939134304010, H74939134352010, H74939134353010, H74939134354010, H74939134402010, H74939134403010, H74939134404010
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
- Reason for Recall
- Directions for use for the affected lots do not include certain language previously requested.
- Manufacturer
- Boston Scientific Corporation
- Recall Posting Date
- 2010-09-13
- Trade Name
- Ultraflex Precision Colonic Stent
- Recall Start Date
- 2010-07-16
- Recall Number
- 55796
- Hazard Classification
- Type II
- Model or Catalog #
- M00557360
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- Boston Scientific has identified an issue that may potentially result in the failure to deploy the stent due to the deployment suture breaking.
- Manufacturer
- Brainlab AG
- Recall Posting Date
- 2010-09-13
- Trade Name
- A) Kolibri Navigation Station 2.0 with Mobile Camera Stand 2.0
B) Kolibri Mobile Camera Stand 2.0
- Recall Start Date
- 2010-08-18
- Recall Number
- 56243
- Hazard Classification
- Type II
- Model or Catalog #
- A) 19506C
B) 19801A
- Lot or Serial #
- A) 000000005042210001, 000000005042310001, 000000003934009001, 000000002185609001, 000000004809910001, 000000003509109001, 000000004367509001
B) N/A
- Reason for Recall
- For the Spectra camera interfaces assembled within the Kolibri Mobile Camera Stand, a long-term reliable fixation of the camera unit cannot be guaranteed due to a manufacturing error.
- Manufacturer
- Brentwood Medical Technology, Corp. DBA Midmark Diagnostics Group
- Recall Posting Date
- 2010-07-19
- Trade Name
- IQmanager Diagnostic Workstation
- Recall Start Date
- 2010-05-26
- Recall Number
- 55107
- Hazard Classification
- Type III
- Model or Catalog #
- 4-100-1200
- Lot or Serial #
- Versions 8.3, 8.3.1, 8.3.2
- Reason for Recall
- When the user access the QT dialogue box to manually accept or edit the QT interval, the software recalculates the QTC value using Hodges equation for all patients rather than for only patients 16 years and older.
- Manufacturer
- Cardiac Pacemakers Incorporated
- Recall Posting Date
- 2010-09-13
- Trade Name
- A) Contak Renewal 3 RF CRT-D
B) Contak Renewal 3 RF HE CRT-D
C) Contak Renewal 4 CRT-D
D) Contak Renewal 4 HE CRT-D
- Recall Start Date
- 2010-07-23
- Recall Number
- 55829
- Hazard Classification
- Type III
- Model or Catalog #
- A) H215
B) H217
C) H190
D) H-197, H-199
- Lot or Serial #
- A) N/A
B) N/A
C) N/A
D) N/A
- Reason for Recall
- Application of a magnet (typically in a clinic hospital environment) may cause a magnetic reed switch located within the device to become permanently stuck in a closed position and prevent delivery of programmed tachy therapy.
- Manufacturer
- Cardinal Health
- Recall Posting Date
- 2010-08-30
- Trade Name
- A) Sterile Craniotomy Pack
B) Sterile Ear Pack
- Recall Start Date
- 2010-08-04
- Recall Number
- 56051
- Hazard Classification
- Type II
- Model or Catalog #
- A) SNE30CRXXX
B) SEN30EAXXX, SEN30MEXXX
- Lot or Serial #
- A) 887069, 905131, 909364, 919774, 926904, 933512, 953095, 885710, 890855, 894055, 900358, 918868, 933456, 940554, 945788
B) 887885, 908312, 912899, 916104, 916587, 890561, 906447, 935731, 953954
- Reason for Recall
- The possibility of reduced adhesion between the patties and the blue radio-opaque x-ray stripe. This may result in the x-ray stripe becoming detached from the pattie during use and remaining in surgical site.
- Manufacturer
- Cardinal Health
- Recall Posting Date
- 2010-09-27
- Trade Name
- Medi-Vac Non Conductive Suction Tubing
- Recall Start Date
- 2010-08-18
- Recall Number
- 56305
- Hazard Classification
- Type II
- Model or Catalog #
- PN510, PN512, PN56A, PN610, PN612, PN66A, PN710, PN712, PN76A
- Lot or Serial #
- All Lot Numbers Prior to Y10KXXXX (10K is August 2010)
- Reason for Recall
- The recall has been initiated due to the possibility that packaging seals on the referenced lot numbers may be compromised.
- Manufacturer
- CaridianBCT, Inc.
- Recall Posting Date
- 2010-08-02
- Trade Name
- A) COBE Spectra White Blood Cell Sets
B) COBE Spectra Therapeutic Plasma Exchange Sets
- Recall Start Date
- 2010-07-16
- Recall Number
- 55784
- Hazard Classification
- Type II
- Model or Catalog #
- A) 70620
B) 70600
- Lot or Serial #
- A) 01S15239
B) 01S15238
- Reason for Recall
- Potential leaking at the return luer connection on two manufacturing lots of COBE Spectra White Blood Cell Disposable Tubing Sets.
- Manufacturer
- CaridianBCT, Inc.
- Recall Posting Date
- 2010-08-30
- Trade Name
- COBE Spectra White Blood Cell Sets
Updated recall # 55784 previously posted 2010-08-02. / Mise
jour du retrait # 55784 afflich le 2010-08-02.
- Recall Start Date
- 2010-07-16
- Recall Number
- 55784
- Hazard Classification
- Type II
- Model or Catalog #
- 70620
- Lot or Serial #
- 01S15239
- Reason for Recall
- Potential leaking at the return luer connection on two manufacturing lots of COBE Spectra White Blood Cell Disposable Tubing Sets.Note that COBE Spectra Therapeutic Plasma Exchange Sets has been removed from this recall.
- Manufacturer
- CaridianBCT, Inc.
- Recall Posting Date
- 2010-09-27
- Trade Name
- Cobe Spectra Apheresis System wbc
- Recall Start Date
- 2010-07-16
- Recall Number
- 55784
- Hazard Classification
- Type II
- Model or Catalog #
- 70620
- Lot or Serial #
- 01S15239, 01S15238
- Reason for Recall
- Leak at the return luer connection on 2 MFR lots of Cobe Spectra white blood cell disposable tubing sets. Leak is observed as a very small fluid leak originating between the male Luer on the return line and the attached customer needle or as an observation of air in the line when blood warmer tubing is attached to the male Luer.
- Manufacturer
- Ciba Vision Corporation
- Recall Posting Date
- 2010-09-13
- Trade Name
- Softperm
- Recall Start Date
- 2010-07-03
- Recall Number
- 55841
- Hazard Classification
- Type II
- Model or Catalog #
- Softperm
- Lot or Serial #
- >100 Numbers, Contact Manufacturer
- Reason for Recall
- Out-of-specification in a product stability study resulting from a change in the vial stopper.
- Manufacturer
- Codman & Shurtleff Inc.
- Recall Posting Date
- 2010-08-16
- Trade Name
- Codman Surgical Patties, X-Ray Detectable
- Recall Start Date
- 2010-07-22
- Recall Number
- 55895
- Hazard Classification
- Type II
- Model or Catalog #
- 80-1396, 80-1399
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- Codman Surgical Patties, which are labelled with a radio-opaque marker, and were distributed by Codman between May 2009 and May 2010, may have reduced adhesion between the patty and the blue radio-opaque x-ray stripe due to a recently discovered processing problem. This reduced adhesion may result in the x-ray stripe becoming detached from the patty during use and remaining in the surgical site inadvertently.
- Manufacturer
- Cook Urological Inc.
- Recall Posting Date
- 2010-08-02
- Trade Name
- Modified Novy Cornual Cannualtion Set
- Recall Start Date
- 2010-01-09
- Recall Number
- 52817
- Hazard Classification
- Type II
- Model or Catalog #
- J-NCS-503570
- Lot or Serial #
- U1799467
- Reason for Recall
- The curved tip of the introducing catheter has the potential to separate during use. If the affected product has already been used successfully, there is not additional patient risk.
- Manufacturer
- Cook, Inc.
- Recall Posting Date
- 2010-09-13
- Trade Name
- A) Ciaglia Blue Rhino Percutaneous
B) Ciaglia Blue Dolphin Percutaneous
- Recall Start Date
- 2010-04-23
- Recall Number
- 54482
- Hazard Classification
- Type I
- Model or Catalog #
- A) C-PTIS-100-HC-PERC6, C-PTIS-100-HC-PERC6-CA, C-PTIS-100-HC-PERC8, C-PTIS-100-HC-PERC8-CA
B) C-PTBS-2600-PERC6, C-PTBS-2800-PERC8, C-PTBSY-2600-PERC6, C-PTBSY-2800-PERC8
- Lot or Serial #
- A) G12403, G12566, G13165, G36037, G13166, G12523, G36038, G12565, G53178, G53169, G53170
B) G36116, G36118, G51971, G36117, G49507, G51972
- Reason for Recall
- Certain cuffed Shiley tracheostomy tubes, in which the cuff does not hold air as a result of leaks in the pilot balloon inflation assembly.
- Manufacturer
- Coopersurgical Inc.
- Recall Posting Date
- 2010-08-16
- Trade Name
- A) INCA Infant Nasal Cannulae Assembly for CPAP
B) INCA Infant Nasal Cannulae Assembly for CPAP Replacement Set
- Recall Start Date
- 2010-07-27
- Recall Number
- 55932
- Hazard Classification
- Type III
- Model or Catalog #
- A) 44-2707, 44-2709, 44-2710, 44-2712, 44-2715
B) 44-0707, 44-0709, 44-0710, 44-0712, 44-0715
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
- Reason for Recall
- As a result of Coopersurgical's voluntary recall for the Infant Nasal Cannulae (CPAP)(INCA) assembly the classification of the device was reviewed by Health Canada. Subsequently, Coopersurgical is recalling the INCA since it was sold in Canada without a license.
- Manufacturer
- Depuy International Ltd.
- Recall Posting Date
- 2010-09-27
- Trade Name
- A) Depuy ASR Cup System - Heads
B) ASR XL Acetabular Cup System - Heads
C) ASR XL Acetabular Cup - Sleeve Adaptors
D) Depuy ASR Acetabular Cup System - Shells
E) ASR 300 Spiked Acetabular Shells
- Recall Start Date
- 2010-08-25
- Recall Number
- 56326
- Hazard Classification
- Type II
- Model or Catalog #
- A) 9998-03-239, 9998-03-441, 9998-03-643, 9998-03-845, 9998-03-946, 9998-04-047, 9998-04-249, 9998-04-451, 9998-04-653, 9998-04-855, 9998-05-057, 9998-05-359, 9998-05-561, 9998-05-763
B) 9998-90-139, 9998-90-141, 9998-90-143, 9998-90-145, 9998-90-146, 9998-90-147, 9998-90-149, 9998-90-151, 9998-90-153, 9998-90-155, 9998-90-157, 9998-90-159, 9998-90-161, 9998-90-163
C) 9998-00-102, 9998-00-105, 9998-00-108, 9998-00-200, 9998-00-203, 9998-00-206, 9998-00-209, 9998-00-300, 9998-00-303, 9998-00-313
D) 9998-03-944, 9998-04-146, 9998-04-348, 9998-04-550, 9998-04-652, 9998-04-754, 9998-04-956, 9998-05-158, 9998-05-360, 9998-05-562, 9998-05-764, 9998-05-966, 9998-06-168, 9998-06-370
E) 999830744, 999830746, 999830748, 999830750, 999830752, 999830754, 999830756, 999830758, 999830760, 999830762, 999830764, 999830766, 999830768, 999830770
- Lot or Serial #
- A) >500 Numbers, Contact Manufacturer
B) >100 Numbers, Contact Manufacturer
C) >500 Numbers, Contact Manufacturer
D) >500 Numbers, Contact Manufacturer
E) >100 Numbers, Contact Manufacturer
- Reason for Recall
- Depuy issued a field safety notice in March 2010 (recall number 53553) after receiving data that demonstrated the ASR system had a higher than expected revision rate at three years when used with smaller head sizes. New data now shows a higher than expected revision rate at five years, across the entire size range. The manufacturer is therefore issuing a voluntary recall of all ASR products.
- Manufacturer
- Depuy Mitek
- Recall Posting Date
- 2010-07-05
- Trade Name
- A) VAPR S90 Suction Electrode (Integrated)
B) VAPR S50 Knee Electrode (Integrated)
C) VAPR Premiere90 Electrode (Integrated)
D) VAPR Premiere50 Electrode (Integrated)
- Recall Start Date
- 2010-06-09
- Recall Number
- 55222
- Hazard Classification
- Type II
- Model or Catalog #
- A) 225370
B) 227355
C) 227204
D) 227504
- Lot or Serial #
- A) >100 Numbers, Contact Manufacturer
B) 805107, 904050, 904051, 904056, 904061
C) >10 Numbers, Contact Manufacturer
D) 906032
- Reason for Recall
- Some of the products within the recall list did not meet the required packaging specifications for pouch seal width dimensions, which could potentially lead to a breach in sterility. The recall applies to all non-expired inventory with the exception of inventory that is designated with round green stickers.
- Manufacturer
- Depuy Orthopaedics, Inc.
- Recall Posting Date
- 2010-08-02
- Trade Name
- Summit Broach Handle
- Recall Start Date
- 2010-07-14
- Recall Number
- 55756
- Hazard Classification
- Type III
- Model or Catalog #
- 2750-00-000
- Lot or Serial #
- A1109, A1209, A0210, A0310
- Reason for Recall
- Some of Depuy's straight broach handles are experiencing early failure due to a weakened handle locking mechanism (leaf spring).
- Manufacturer
- Diasorin Inc.
- Recall Posting Date
- 2010-08-30
- Trade Name
- Liason Analyzer
- Recall Start Date
- 2010-08-10
- Recall Number
- 56139
- Hazard Classification
- Type II
- Model or Catalog #
- 15970
- Lot or Serial #
- All Serial Numbers
- Reason for Recall
- During system initialization when pipettors are washing, if sample or reagent area flap are opened, the shaker rod may stop moving. This will affect the suspension of magnetic particles resulting in low RLU values for calibrators, controls & patient samples giving unexpected positive or negative results or incorrect dose.
- Manufacturer
- Draeger Medical AG & Co. KG
- Recall Posting Date
- 2010-08-02
- Trade Name
- NBP Cuffs NEO Sizes 1 to 5
- Recall Start Date
- 2010-06-29
- Recall Number
- 55447
- Hazard Classification
- Type I
- Model or Catalog #
- MP00901, MP00902, MP00903, MP00904, MP00905
- Lot or Serial #
- Devices Delivered Between 12-01-2010 and 12-06-2010.
- Reason for Recall
- Cases where measurements with the affected neonatal blood pressure cuffs deviate from the actual values. This deviation can be up to 50%, meaning the measured and displayed value can be 50% higher than the actual one.
- Manufacturer
- Draeger Medical AG & Co. KG
- Recall Posting Date
- 2010-09-13
- Trade Name
- Oxylog 3000
- Recall Start Date
- 2010-08-16
- Recall Number
- 56204
- Hazard Classification
- Type I
- Model or Catalog #
- 2M86955
- Lot or Serial #
- SRZB-0059, SRZA-0128, SRYK-0134, SRYK-0133, SRYK-0132, SRYK-0131, SRYK-0069, SRYK-0068, SRYB-0074, SRYA-0195, SRYA-0014
- Reason for Recall
- Patients have been inadequately ventilated because the dead space of the breathing circuit was not completely considered by the user when setting the ventilation parameters. The information for use (IFU) for the Oxylog 3000 devices delivered from September 2008 contains additional warnings and cautions. A safety notice with the IFU amendment will be sent out to users who own affected Oxylog 3000. No patient injury has been reported in relation to that dead space issue.
- Manufacturer
- Draeger Medical AG & Co. KG
- Recall Posting Date
- 2010-09-13
- Trade Name
- Fabius GS Anesthesia System
- Recall Start Date
- 2010-08-25
- Recall Number
- 56304
- Hazard Classification
- Type II
- Model or Catalog #
- 8604700
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- The Fabius GS or Fabius Tiro delivered between June 2005 and June 2006 could have been equipped with a power cord of the manufacturer Electri-Cord Manufacturing Corporation (ECM). The US Food and Drug Administration (FDA) has associated a risk of sparks, fire, device failures or electrical shocks with these power cords. Although Draeger Medical has not received any reports of failures of power cords manufactured by ECM, Draeger Medical will replace the possible affected power cords.
- Manufacturer
- Draeger Medical Systems, Inc.
- Recall Posting Date
- 2010-07-19
- Trade Name
- A) Gamma XXL
B) Delta Monitor
- Recall Start Date
- 2010-05-24
- Recall Number
- 55089
- Hazard Classification
- Type II
- Model or Catalog #
- A) MS18852
B) MS18597
- Lot or Serial #
- A) 5399463952, 5399468849, 5399469251, 5399494054, 5399498745, 5399532156, 5399532753, 5399533850, 5399662354, 5399664949, 6000446771, 6000796170
B) >10 Numbers, Contact Manufacturer
- Reason for Recall
- Investigation determined that the monitor's fixed keys may become inoperative or activate without user interaction due to a contamination present in the fixed key panel material. This is a concern because the worst case scenario of this malfunction could be that the monitor may discharge to a patient automatically.
- Manufacturer
- Elekta Business Area Software Systems ATA Computerized Medical Systems Inc.
- Recall Posting Date
- 2010-07-05
- Trade Name
- Focal Software
- Recall Start Date
- 2010-06-10
- Recall Number
- 55234
- Hazard Classification
- Type II
- Model or Catalog #
- N/A
- Lot or Serial #
- Release 4.40.00 & Up
- Reason for Recall
- The couch position can be moved (X or Y direction) between scans of the same CT study. When images are imported, the software is not checking the Dicom image position (patient) tag & is not aligning the image sets. A shift between image sets will exist of the same amount that the CT couch was moved.
- Manufacturer
- Excelsior Medical Corporation
- Recall Posting Date
- 2010-09-27
- Trade Name
- Prefilled Flush Syringe
- Recall Start Date
- 2010-09-02
- Recall Number
- 56438
- Hazard Classification
- Type II
- Model or Catalog #
- S-5
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- During a routine test for the syringe integrity, a possible integrity failure was discovered.
- Manufacturer
- Focus Medical, LLC
- Recall Posting Date
- 2010-09-27
- Trade Name
- Naturalase 1064 Laser
- Recall Start Date
- 2010-09-01
- Recall Number
- 55902
- Hazard Classification
- Type II
- Model or Catalog #
- LT100
- Lot or Serial #
- N/A
- Reason for Recall
- The recall is being conducted because the licence to market and sell these systems in Canada had expired at the time the units were sold
- Manufacturer
- Fujirebio Diagnostics, Inc.
- Recall Posting Date
- 2010-08-16
- Trade Name
- Mesomark Kits
- Recall Start Date
- 2010-07-08
- Recall Number
- 55955
- Hazard Classification
- Type II
- Model or Catalog #
- 801-900
- Lot or Serial #
- 57K04510, 57K04409 (Already Expired)
- Reason for Recall
- Certain lots of Mesomark Kits may generate inaccurate absorbance values due to the presence of a residue on the underside of some microtiter wells that can be detected visually.
- Manufacturer
- GE Healthcare
- Recall Posting Date
- 2010-07-05
- Trade Name
- A) Prospeed Scanner System
B) SYTEC6000
- Recall Start Date
- 2009-03-26
- Recall Number
- 47109
- Hazard Classification
- Type II
- Model or Catalog #
- A) B7970YZ
B) P9183AH
- Lot or Serial #
- A) 298920YMO & 455302YMO
B) 121615YM9
- Reason for Recall
- The x-ray leakage from the diagnostic source assembly (tube+collimator) of the Zj/Qj families of computed tomography systems exceeded 21 CFR regulatory requirement and IEC standard.
- Manufacturer
- GE Healthcare IT
- Recall Posting Date
- 2010-07-19
- Trade Name
- Centricity Ultra Laboratory Information System
- Recall Start Date
- 2010-06-15
- Recall Number
- 55072
- Hazard Classification
- Type II
- Model or Catalog #
- CORE
- Lot or Serial #
- All Software Versions
- Reason for Recall
- Under specific conditions, the laboratory system may report incorrect results such as truncating a value after a comma is encountered. For example, a result of 1,234 would be reported as 1.0.
- Manufacturer
- GE Healthcare Japan Corporation
- Recall Posting Date
- 2010-08-16
- Trade Name
- HiSpeed DX/I LX/I CT System
- Recall Start Date
- 2010-07-26
- Recall Number
- 55695
- Hazard Classification
- Type III
- Model or Catalog #
- B7510ZZ
- Lot or Serial #
- 00000532977YM6, 00000715131YM9
- Reason for Recall
- Computed tomography dose index (CTDI) values on dose display & in operator manual are incorrect for scan settings over 3mm slice thickness. The provided CTDI value may be more than 20% higher than actual CTDI value.
- Manufacturer
- GE Medical Systems Information Technologies, Inc.
- Recall Posting Date
- 2010-09-27
- Trade Name
- A) Transport Pro V2 Patient Monitor
B) Transport Pro Patient Monitor (PDM Compatible)
- Recall Start Date
- 2010-08-30
- Recall Number
- 56100
- Hazard Classification
- Type II
- Model or Catalog #
- A) TRANS-AC-HXXX
B) TRANS-BC-HXXX
- Lot or Serial #
- A) AAD05360553GA, AAD06031134GA, AAD06131474GA, AAD06352093GA, AAD06352095GA, AAD05350491GA
B) >10 Numbers, Contact Manufacturer
- Reason for Recall
- Transport Pro stops communication with Carescape patient data modules after 414 days of continuous run time, resulting in the loss of the displayed parameter info. (Eg: waveforms, numerical data, alarms).
- Manufacturer
- GE Medical Systems Kretztechnik GMBH & CO OHG
- Recall Posting Date
- 2010-07-05
- Trade Name
- A) Voluson E8 Ultrasound System Main Unit
B) Voluson E8 Expert Ultrasound System
Updated reall # 51227 previously posted on 2009/11/09
- Recall Start Date
- 2009-10-21
- Recall Number
- 51227
- Hazard Classification
- Type II
- Model or Catalog #
- A) H48651PT
B) H48661N
- Lot or Serial #
- A) S/N D00229
B) S/N D04842, D05589 & D50233
- Reason for Recall
- When using combination of 2D-compound resolution imaging (CRI) and a pulsed wave Doppler single gate (PW) with steered Doppler angle, the displayed velocity scale is incorrect, which could delay treatment.
- Manufacturer
- GE Medical Systems SCS
- Recall Posting Date
- 2010-07-05
- Trade Name
- Innova 2000 Cardiovascular System
- Recall Start Date
- 2010-06-09
- Recall Number
- 54810
- Hazard Classification
- Type II
- Model or Catalog #
- S1872LC
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- Source to image distance (SID) failure could occur, which may result in loss of x-ray generation. An error message may or may not be displayed, depending on certain conditions.
- Manufacturer
- GE Medical Systems Ultrasound & Primary Care Diagnostics LLC
- Recall Posting Date
- 2010-09-27
- Trade Name
- Logiq E9 Ultrasound System-Sector Transducer
- Recall Start Date
- 2010-09-02
- Recall Number
- 56175
- Hazard Classification
- Type III
- Model or Catalog #
- H4908SC
- Lot or Serial #
- 94900496
- Reason for Recall
- Rating plate of S1-5 probe identifies name and model number of S4-10 probe. Correct S1-5 name appears on transducer handle and packaging.
- Manufacturer
- Gyrus Acmi, Inc.
- Recall Posting Date
- 2010-08-30
- Trade Name
- Pleatman Sac Tissue Removal System
- Recall Start Date
- 2010-08-04
- Recall Number
- 56025
- Hazard Classification
- Type II
- Model or Catalog #
- 004942-903
- Lot or Serial #
- 186540FC
- Reason for Recall
- Trays used to package the Pleatman Sac product had cracks and/or dimples on them. Further investigation revealed that the dimples could form cracks when the tray is manipulated. This could lead to a breach in the sterile barrier which could compromise sterility of the product.
- Manufacturer
- Gyrus ACMI, Inc.
- Recall Posting Date
- 2010-08-16
- Trade Name
- Dissector PlasmaKnife (DPK)
- Recall Start Date
- 2010-07-15
- Recall Number
- 55925
- Hazard Classification
- Type II
- Model or Catalog #
- 7035-3005
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- A non conformance report (NCR) was initiated at Gyrus ACMI facility because a qualification run of fluid electrodes failed International Safe Transit Association (ISTA) testing due to a failure of the seal on the sterile packaging. A breach in the sterile barrier would compromise the sterility of the product.
- Manufacturer
- Haemonetics Corporation
- Recall Posting Date
- 2010-07-05
- Trade Name
- Plasma Bowl, 0625B-00
- Recall Start Date
- 2010-06-10
- Recall Number
- 55256
- Hazard Classification
- Type II
- Model or Catalog #
- 0625B-00
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- There is a defect in a very small number of plasma bowls in which a portion of the wall of the bowl may be too thin. There is a potential that a defective bowl may leak or fail within the centrifuge chamber during a plasma collection procedure. If a leak or failure were to occur, the PCS2 system is designed to detect the problem and safely terminate the collection procedure. However, it is possible that blood could escape from the device and spray into the surrounding environment and onto surrounding objects or people.
- Manufacturer
- Heine Optotechnik GMBH & Co. KG
- Recall Posting Date
- 2010-08-16
- Trade Name
- XP Laryngoscope Blabe Mac 3
- Recall Start Date
- 2010-07-08
- Recall Number
- 55909
- Hazard Classification
- Type III
- Model or Catalog #
- F-000.22.763
- Lot or Serial #
- 62117
- Reason for Recall
- There is a tolerance issue on a certain production batch from April 2010. The result is that the heel of the blade may be too wide to be inserted into the head of laryngoscope handles manufactured to the "green standard".
- Manufacturer
- Hill-Rom, Inc.
- Recall Posting Date
- 2010-07-05
- Trade Name
- Hill-Rom 100 Low Bed
- Recall Start Date
- 2010-05-03
- Recall Number
- 54548
- Hazard Classification
- Type II
- Model or Catalog #
- P3930
- Lot or Serial #
- N/A
- Reason for Recall
- Potential for siderails used on the Hill-Rom 100 Low Bed to operate incorrectly (false latching, unintentional lowering) or become inoperate.
- Manufacturer
- Hill-Rom, Inc.
- Recall Posting Date
- 2010-07-05
- Trade Name
- A) Hill-Rom 1000 Beds
B) Care Assist P1170
- Recall Start Date
- 2009-07-13
- Recall Number
- 54567
- Hazard Classification
- Type II
- Model or Catalog #
- A) P1600
B) P1170
- Lot or Serial #
- A) N/A
B) N/A
- Reason for Recall
- There is a potential for a shearing action between the siderail arm and the bracket used to secure the siderail to the bed.
- Manufacturer
- Hill-Rom, Inc.
- Recall Posting Date
- 2010-07-19
- Trade Name
- A) Hill-Rom 1000 Beds
B) Careassist Hospital Bed
- Recall Start Date
- 2010-06-10
- Recall Number
- 55218
- Hazard Classification
- Type II
- Model or Catalog #
- A) P1160
B) P1170
- Lot or Serial #
- A) N/A
B) N/A
- Reason for Recall
- Failure of the plastic housing used to support the trapeze handle from the patient helper.
- Manufacturer
- Hologic, Inc.
- Recall Posting Date
- 2010-09-13
- Trade Name
- A) Atec Breast Biopsy & Excision System Sapphire
B) Atec Breast Biopsy & Excision System Emerald
C) Atec Breast Biopsy & Excision System Pearl
- Recall Start Date
- 2010-08-09
- Recall Number
- 56247
- Hazard Classification
- Type II
- Model or Catalog #
- A) N/A
B) N/A
C) N/A
- Lot or Serial #
- A) 003016
B) 003016
C) 003016
- Reason for Recall
- The aperture of the needle may be a 20mm size, versus the 12mm size indicated on the label.
- Manufacturer
- Hospira Inc.
- Recall Posting Date
- 2010-08-30
- Trade Name
- Gemstar Pump Set (110 Yellow)(AMB.INFU)
- Recall Start Date
- 2010-08-10
- Recall Number
- 56130
- Hazard Classification
- Type II
- Model or Catalog #
- 13261, 13273
- Lot or Serial #
- 740545H, 680295H, 770878H
- Reason for Recall
- Under delivery reports occurring during clinical use under low rate settings (less than 10ml/hr).
- Manufacturer
- Howmedica Osteonics Corporation
- Recall Posting Date
- 2010-08-30
- Trade Name
- Osteonics Series 7000 Total Knee Standard Tibial Tray
- Recall Start Date
- 2010-08-04
- Recall Number
- 56012
- Hazard Classification
- Type II
- Model or Catalog #
- 7115-0007
- Lot or Serial #
- MJK68N
- Reason for Recall
- Stryker Howmedica Osteonics has become aware that there was a cross mix between the Osteonics Series 7000 Total Knee Standard Tibial Tray and the Triathlon Total Knee System Primary Cemented Tibial Base-Plate. The packaging was labelled as a Series 7000 Total Knee Standard Tibial Tray, while the package actually contained a Triathlon Total Knee System Primary Cemented Tibial Base-Plate.
- Manufacturer
- Immucor Inc.
- Recall Posting Date
- 2010-07-19
- Trade Name
- Anti-FY Blood Grouping Reagent
- Recall Start Date
- 2010-06-16
- Recall Number
- 55381
- Hazard Classification
- Type III
- Model or Catalog #
- 0007594
- Lot or Serial #
- 613007, 613007-1
- Reason for Recall
- Lots 613007 and 613007-1 demonstrate weaker reactivity than previous lots of Anti-FY.
- Manufacturer
- Immucor Inc.
- Recall Posting Date
- 2010-08-16
- Trade Name
- Panoscreen I, II and III
- Recall Start Date
- 2010-07-16
- Recall Number
- 55873
- Hazard Classification
- Type I
- Model or Catalog #
- 0002377
- Lot or Serial #
- 24223
- Reason for Recall
- Panoscreen I, II and III lot 24223 does not include an S positive donor cell.
- Manufacturer
- Immucor Inc.
- Recall Posting Date
- 2010-08-16
- Trade Name
- Galileo Blood Bank Analyzer
- Recall Start Date
- 2010-07-28
- Recall Number
- 55949
- Hazard Classification
- Type II
- Model or Catalog #
- 0064999
- Lot or Serial #
- N/A
- Reason for Recall
- If an unrecognized rack is removed, and a sample is than removed or switched in the rack, followed by reinserting the rack into the lane, it is possible for the sample ID that originally occupied the position to be assigned to the newly loaded sample or the empty position. Under these conditions, it is possible that results will be associated with the wrong sample.
- Manufacturer
- Immucor Inc.
- Recall Posting Date
- 2010-09-13
- Trade Name
- Galileo Echo Blood Bank Analyzer
- Recall Start Date
- 2010-08-12
- Recall Number
- 56238
- Hazard Classification
- Type II
- Model or Catalog #
- 0087000
- Lot or Serial #
- N/A
- Reason for Recall
- When a sample rack is partially removed after testing is complete and another rack is inserted into an adjacent position, it is possible for the partially removed rack to be inadvertently loaded ahead of the second rack. In this instance, the second rack's sample ID's will be incorrectly assigned to the original rack.
- Manufacturer
- Immucor Inc.
- Recall Posting Date
- 2010-09-27
- Trade Name
- Panocell-20
- Recall Start Date
- 2010-08-30
- Recall Number
- 56415
- Hazard Classification
- Type II
- Model or Catalog #
- 0005020
- Lot or Serial #
- 29304
- Reason for Recall
- Cell 5 (donor B3907) of lot 29304 is demonstrating weak positive reactions in direct antiglobulin testing (DAT). Tests performed with this cell may produce unexpected positive reactions at the anti-human globulin (AHG) phase of testing.
- Manufacturer
- Immunotech A.S.
- Recall Posting Date
- 2010-08-02
- Trade Name
- Estrone Radioimmunoassay (RIA) Kit
- Recall Start Date
- 2010-07-05
- Recall Number
- 55630
- Hazard Classification
- Type II
- Model or Catalog #
- DSL 8700
- Lot or Serial #
- 992637, 100226A, 100319A, 100409A
- Reason for Recall
- Beckman Coulter has determined that the Estrone RIA Kit lots identified were inaccurately calibrated to their internal reference calibrator.
- Manufacturer
- Immuontech A.S.
- Recall Posting Date
- 2010-08-30
- Trade Name
- IGFBP-3 Immunoradiometric Assay (IRMA)
- Recall Start Date
- 2010-08-09
- Recall Number
- 56099
- Hazard Classification
- Type II
- Model or Catalog #
- DSL-6600
- Lot or Serial #
- 100115A, 100319A, 100409A, 100430A, 991225, 991694, 991938, 992171, 992372
- Reason for Recall
- Beckman Coulter has confirmed that IGFBP-3 measurements using IGFBP-3 IRMA Kit (REF DSL-6600) lots identified may result in IGFBP-3 levels that are not in agreement with simultaneously measured levels of IGF-1.
- Manufacturer
- INO Therapeutics, LLC DBA IKARIA
- Recall Posting Date
- 2010-08-16
- Trade Name
- INOMAX DS Drug Delivery-System
- Recall Start Date
- 2010-07-23
- Recall Number
- 55907
- Hazard Classification
- Type II
- Model or Catalog #
- DS 20050006
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- A component within the pressure switch which monitors for loss of drug supply may tear, causing Inomax to leak from the rear of the system. As a result, the Inomax cylinder empties more quickly than normal. An interruption or delay in the administration of Inomax therapy may cause worsening of systemic oxygenation indices (i.e. hypoxemia, especially manifested as decreased arterial oxygenation saturation), hypotension, and/or increase in pulmonary arterial pressure.
- Manufacturer
- Institut Straumann AG
- Recall Posting Date
- 2010-07-19
- Trade Name
- Impression Set for WN Solid Abutment
- Recall Start Date
- 2010-06-15
- Recall Number
- 55311
- Hazard Classification
- Type III
- Model or Catalog #
- 040.38
- Lot or Serial #
- Y4021
- Reason for Recall
- The product name on the label incorrectly identifies the product as "Impression Set for RN (regular neck) Solid Abutment" when it should read "Impression set for WN (wide neck) Solid Abutment".
- Manufacturer
- Instrumentation Laboratory Co.
- Recall Posting Date
- 2010-09-27
- Trade Name
- A) ACL Top 500 CTS
B) ACL Top, ACL Top 700 LAS, ACL Top CTS
- Recall Start Date
- 2010-06-01
- Recall Number
- 56433
- Hazard Classification
- Type II
- Model or Catalog #
- A) 0000280040
B) 0000280000, 0000280020, 0000280030
- Lot or Serial #
- A) 08010100 to 10061231
B) 03110101 to 1006111, 08120100 to 10020126, 06020101 to 10030659
- Reason for Recall
- Software V4.3.0 (with test parameters P-14.6) introduces an additional data check designed to minimize performance issues from factors such as patient condition and improper sample or reagent preparation by the user. Specifically, a new data CJECL, "multiple threshold check" was added to identify and fail abnormal (bimodal) clot curves for hemosil synthasil (APTT reagent, K060688) and hemosil recombiplastin 2g (PT reagent, K070005). The flag will recognize any APTT/PT curve that has a signal that increases in absorbance to a certain threshold level more than once.
- Manufacturer
- Integra Neurosciences Implants S.A.
- Recall Posting Date
- 2010-08-02
- Trade Name
- Neuro Balloon Catheter
- Recall Start Date
- 2010-07-07
- Recall Number
- 55602
- Hazard Classification
- Type II
- Model or Catalog #
- 7CBD10
- Lot or Serial #
- 0151870, 0158739, 0159938
- Reason for Recall
- Certain Integra Neuro Balloon Catheters have the potential to improperly inflate or deflate under certain conditions. It was determined that this condition could exist during pre-implant testing or during the procedure with some lots of Integra Neuro Balloon Catheter products.
- Manufacturer
- International Technidyne Corp.
- Recall Posting Date
- 2010-07-05
- Trade Name
- Hemochron Response Instrument, IDMS
- Recall Start Date
- 2010-06-14
- Recall Number
- 55307
- Hazard Classification
- Type III
- Model or Catalog #
- HRS.110
- Lot or Serial #
- 72900321, 72900376, 72900367
- Reason for Recall
- There is a tabulation error in the integrated data management system (IDMS) software, product code 464100, versions 7.1 through 7.3, used with the Hemochron Response Instrument, that affects the operator and the device performance summary reports for two electronic quality control (EQC) tests.
- Manufacturer
- International Technidyne Corp.
- Recall Posting Date
- 2010-08-16
- Trade Name
- Directcheck Whole Blood Control
- Recall Start Date
- 2010-07-19
- Recall Number
- 55897
- Hazard Classification
- Type III
- Model or Catalog #
- DCJPT-A
- Lot or Serial #
- G9DPA005
- Reason for Recall
- The acceptable performance range published in the Directcheck microcoagulation quality control (product code DCJPT-A, lot number G9DPA005 only) package insert is inaccurate. The posted range printed on the package insert reads "3.7 - 6.5 INR". The correct range is 4.1 - 8.7 INR.
- Manufacturer
- Inviro Medical, Inc.
- Recall Posting Date
- 2010-07-19
- Trade Name
- Invirosnap Safety Syringe
- Recall Start Date
- 2010-05-14
- Recall Number
- 55079
- Hazard Classification
- Type III
- Model or Catalog #
- 100002, 100018, 100019, 100017, 100038, 100047, 100034
- Lot or Serial #
- 2006-06, 2006-09, 061911, 2006-10
- Reason for Recall
- When the plunger is retracted after an injection, the plunger with needle can be pulled out of the back of the syringe.
- Manufacturer
- Inviro Medical, Inc.
- Recall Posting Date
- 2010-07-19
- Trade Name
- Invirosnap Safety Syringe
- Recall Start Date
- 2010-05-14
- Recall Number
- 55085
- Hazard Classification
- Type III
- Model or Catalog #
- 100019, 100038, 100032, 100034
- Lot or Serial #
- 063804, 061911, 2006-08
- Reason for Recall
- The needle length is out the specifications.
- Manufacturer
- Johnson & Johnson Vision Care, Inc.
- Recall Posting Date
- 2010-08-30
- Trade Name
- Acuvue Oasys Brand Contact Lenses for Astigmatism
- Recall Start Date
- 2010-08-02
- Recall Number
- 56160
- Hazard Classification
- Type III
- Model or Catalog #
- N/A
- Lot or Serial #
- B006X641
- Reason for Recall
- Some product may have been mislabeled on the secondary carton (yellow carton) and there is a discrepancy with the primary package (blister pack).
- Manufacturer
- Levo AG
- Recall Posting Date
- 2010-08-02
- Trade Name
- Levo C3 Wheelchair
- Recall Start Date
- 2010-06-25
- Recall Number
- 55674
- Hazard Classification
- Type II
- Model or Catalog #
- N/A
- Lot or Serial #
- 913021
- Reason for Recall
- A quality problem was discovered with the stand-up wheelchair model Levo C3. When using the stand-up mechanism after a short time a screw may break at the support or the stand-up actuator unit propping up the seat to a 45 degree angle position.
- Manufacturer
- Lifescan Inc.
- Recall Posting Date
- 2010-09-27
- Trade Name
- A) OneTouch DataLink Data Management System
B) OneTouch DataLink Management System
- Recall Start Date
- 2010-10-23
- Recall Number
- 56430
- Hazard Classification
- Type I
- Model or Catalog #
- A) 021-323, 021-969
B) 021-309, 021-970
- Lot or Serial #
- A) All lots
B) All lots
- Reason for Recall
- Possibility of data loss at healthcare facilities when downloading Surestep Flexx Meters to the OneTouch Datalink Data Management System under very specific circumstances (including having more than one meterlink instance installed & a meter that has not downloaded within 25 hrs).
- Manufacturer
- Liko AB
- Recall Posting Date
- 2010-07-19
- Trade Name
- Viking M Patient Lift
- Recall Start Date
- 2010-05-25
- Recall Number
- 54859
- Hazard Classification
- Type II
- Model or Catalog #
- 2040015
- Lot or Serial #
- 7500401 to 7503472
- Reason for Recall
- Failure of the Viking M Lift as it relates to the control box (CBL 20)and the battery (BAL 20). The product will stop functioning. Note that there is a mechanical means to safely lower the patient during transfer.
- Manufacturer
- Liko AB
- Recall Posting Date
- 2010-08-02
- Trade Name
- Scale Adapters
- Recall Start Date
- 2010-07-12
- Recall Number
- 55601
- Hazard Classification
- Type II
- Model or Catalog #
- 200, 350, 400
- Lot or Serial #
- N/A
- Reason for Recall
- If you twist the sling bar when using the scale and scale adapter at a right angle in the lifting direction, the adapter can get locked in this position when the lift has a flexlink. If patient is lifted with the equipment while adapter is in this position, adapter may be damaged or break, allowing a patient to fall.
- Manufacturer
- Linvatec Corporation D.B.A. Conmed Linvatec
- Recall Posting Date
- 2010-08-16
- Trade Name
- 87K Arthroscopy Tubing Set
- Recall Start Date
- 2010-07-06
- Recall Number
- 55704
- Hazard Classification
- Type II
- Model or Catalog #
- 87100
- Lot or Serial #
- Lot # date codes beginning with 0904151 to 1005201
- Reason for Recall
- A possibility the product may have a breach in the packaging seal that potentially could compromise the product sterility.
- Manufacturer
- Luxottica North America Distribution LLC
- Recall Posting Date
- 2010-08-02
- Trade Name
- Sport Goggles - Liberty Sport
- Recall Start Date
- 2010-07-02
- Recall Number
- 55657
- Hazard Classification
- Type II
- Model or Catalog #
- MAXMORPHEUS, REC SPEC MAXX
- Lot or Serial #
- N/A
- Reason for Recall
- Some Sport Goggles may have prescription lens which do not meet the lens thickness specifications established by the frame manufacturer. Potential for lens breakage or disengagement from frame in event of impact.
- Manufacturer
- Maquet Cardiopulmonary AG
- Recall Posting Date
- 2010-09-27
- Trade Name
- Straight Connector
- Recall Start Date
- 2010-08-30
- Recall Number
- 56358
- Hazard Classification
- Type II
- Model or Catalog #
- HG0286
- Lot or Serial #
- 70045443, 70050983
- Reason for Recall
- A leak has been found at the shank during priming before perfusion, a manufacturer issue was detected at the supplier level.
- Manufacturer
- Medical Device Technologies, Inc. DBA Angiotech
- Recall Posting Date
- 2010-08-02
- Trade Name
- Tru-Core II
- Recall Start Date
- 2010-07-02
- Recall Number
- 55562
- Hazard Classification
- Type II
- Model or Catalog #
- MD763114200X, MD763118100X, MD763118200X
- Lot or Serial #
- 92661WY0, 93491YG7, 00621ZQ1, 011720OQ, 93431YBI, 92661WY7
- Reason for Recall
- Pouch holes at the chevron edge on the polyethylene side of the pouch or along the chevron side of the seal can compromise the sterile barrier of the packaging.
- Manufacturer
- Medline Industries Inc
- Recall Posting Date
- 2010-09-27
- Trade Name
- A) Nygh Dental Pack
B) Nygh Minor and Major Pack
C) C-Section Pack
D) Nygh T & A Pack
E) Nygh Septoplasty Pack
F) Nygh Laparoscopy Pack
G) Grace Total Hip Pack
H) Nygh Vascular Pack
I) Cardiac Pack
- Recall Start Date
- 2010-08-23
- Recall Number
- 56350
- Hazard Classification
- Type II
- Model or Catalog #
- A) DYNJ23545
B) DYNJ23312, DYNJ23313
C) OBT242X
D) DYNJ28887
E) DYNJ23314
F) DYNJ23528
G) DYNJ23234
H) DYNJ23547
I) DYNJ31903X
- Lot or Serial #
- A) 07AD1212, 07AD4192, 07GD1673
B) 07AD1541, 07ED3906, 07AD2649, 07BD3551
C) 10FD0997, 10FD1415, 10GD0786, 10GD2489, 09CD2228, 09CD02261, 10DD0174, 09ED0001, 09ED0887, 09ED3576, 09GD2287, 09HD3244, 09JD3472, 08DD0257, 08ED1747, 08GD4104, 08HD1333, 08KD1091, 09KD3436
D) 07AD2162, 07BD0755, 07CD1423, 07DD1506, 07DD4242
E) 07AD2160, 07BD2175, 07BD2537, 07CD1090, 07DD3113
F) 07AD2290, 07BD0186, 07FD2924, 07GD3752
G) 10BD2818, 10AD0648, 09LD1851, 09HD1616, 09GD0515, 09FD3755, 09FD3241, 09FD0698, 08FD1984, 08FD1109, 08ED3801, 08DD1563, 08DD0368, 08CD2385, 08AD0583, 07LD0379, 07KD0907, 07KD0906, 07JD1514, 07ID1730, 07FD2226
H) 07AD4449, 07AD4545, 07BD1859, 07FD0509, 07GD3003
I) 09ED0655, 09ED3002, 09FD0041, 09FD3115, 09JD1513, 09KD1837, 09LD0272, 09LD2807, 10AD0647, 10AD3494
- Reason for Recall
- These custom procedure packs contain the Cardinal Medi-Vac Non-Conductive Suction Tubing that has been recalled by Cardinal Health. Medline Industries was alerted by Cardinal regarding a product recall for their Sterile Medi-Vac Non-Conductive Suction Tubing due to the possibility that packaging seals may be compromised.
- Manufacturer
- Medtronic Inc.
- Recall Posting Date
- 2010-07-19
- Trade Name
- Carelink - Monitor
- Recall Start Date
- 2010-06-10
- Recall Number
- 55207
- Hazard Classification
- Type II
- Model or Catalog #
- 2490C
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- During an update, a subset of Carelink monitors of a previous design received firmware intended for monitors of a newer design. This mismatch makes affected monitors non-functional & unable to transmit data.
- Manufacturer
- Medtronic Inc.
- Recall Posting Date
- 2010-08-02
- Trade Name
- Enrhythm Dual Chamber Rate Responsive Pacemaker
- Recall Start Date
- 2010-02-11
- Recall Number
- 53125
- Hazard Classification
- Type II
- Model or Catalog #
- P1501DR
- Lot or Serial #
- >1000 Numbers, Contact Manufacturer
- Reason for Recall
- 1) Battery voltage at device interrogation lower than what is tracked to provide data for elective replacement indicator notification. 2) Battery voltage could decrease sooner than expected.
- Manufacturer
- Medtronic Inc.
- Recall Posting Date
- 2010-08-02
- Trade Name
- HMS System - Instrument
- Recall Start Date
- 2010-03-08
- Recall Number
- 53511
- Hazard Classification
- Type III
- Model or Catalog #
- 305-14
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- Control test for heparin assay (HPT) cartridges may run longer than normal & in some cases time out after setting instrument parameters to "IU".
- Manufacturer
- Medtronic Inc.
- Recall Posting Date
- 2010-08-02
- Trade Name
- Sutureless Unipolar Myocardial Lead
- Recall Start Date
- 2010-02-25
- Recall Number
- 53385
- Hazard Classification
- Type II
- Model or Catalog #
- 5071
- Lot or Serial #
- >100 Numbers, Contact Manufacturer
- Reason for Recall
- Possible inadequate sterile packaging seals that could potentially affect product sterility.
- Manufacturer
- Medtronic Navigation Israel Ltd.
- Recall Posting Date
- 2010-08-16
- Trade Name
- Polestar N-20 - Magnetic Resonance Diagnostic Device
- Recall Start Date
- 2010-07-16
- Recall Number
- 55746
- Hazard Classification
- Type II
- Model or Catalog #
- Flexicoil - 9733063
- Lot or Serial #
- 7110519203, 8051519694, 9210040532
- Reason for Recall
- Electronics box on some flexicoils can overheat and may cause a skin burn when in direct contact with the patient.
- Manufacturer
- Medtronic Navigation, Inc.
- Recall Posting Date
- 2010-07-19
- Trade Name
- Stealthstation S7 - Stealth 3D
- Recall Start Date
- 2010-06-21
- Recall Number
- 55374
- Hazard Classification
- Type II
- Model or Catalog #
- 9734046
- Lot or Serial #
- S/N's 4500883854, 4500961922, 4500816943, 4500816944, 4501132005, 4500707425
- Reason for Recall
- Numerous electrical issues related to the power cord retractors.
- Manufacturer
- Medtronic Navigation, Inc.
- Recall Posting Date
- 2010-08-02
- Trade Name
- Orthopedic Dovetail Clamp
- Recall Start Date
- 2010-02-10
- Recall Number
- 53129
- Hazard Classification
- Type II
- Model or Catalog #
- 9730289
- Lot or Serial #
- 200304080750, 200308061080, 2003040807, 2003080610, 200308061080 & 5081710533
- Reason for Recall
- Current IFU does not indicate that the clamp is designed for exclusive use with Zimmer brand acetabular cup positioner/impactor. Otherwise the clamp can slip axially, resulting in movement of the reference frame.
- Manufacturer
- Medtronic Sofamor Danek USA, Inc.
- Recall Posting Date
- 2010-07-19
- Trade Name
- Nim-Eclipse System - Module, Surface Preamp
- Recall Start Date
- 2010-07-02
- Recall Number
- 55529
- Hazard Classification
- Type II
- Model or Catalog #
- 945SPM608
- Lot or Serial #
- 0923052D
- Reason for Recall
- Under worst case conditions, such as sweating under warming blankets and long term contact periods, the patient's skin can become discoloured at the electrode contact site.
- Manufacturer
- Metrex Research Corporation
- Recall Posting Date
- 2010-09-27
- Trade Name
- Curing Light Sleeves
- Recall Start Date
- 2010-08-09
- Recall Number
- 56429
- Hazard Classification
- Type III
- Model or Catalog #
- 4500, 4513
- Lot or Serial #
- 100622, 100733
- Reason for Recall
- Part no. 4500 labelled as 7mm/8mm curing light sleeves actually contained 13mm sleeves. Part no. 4513 labelled as 13mm curing light sleeves actually contained 7mm/8mm sleeves.
- Manufacturer
- Minntech Corporation
- Recall Posting Date
- 2010-08-02
- Trade Name
- Endoscope Disinfector
- Recall Start Date
- 2010-07-08
- Recall Number
- 55606
- Hazard Classification
- Type II
- Model or Catalog #
- DSD-110-HU0109
- Lot or Serial #
- N/A
- Reason for Recall
- Misprinted DSD hookup application guide that recommended the use of the DSD-110-HU0109 hookup rather than DSD-110-HU0144 hookup. May result in inadequate high-level disinfection of either the balloon filling channel or the elevator wire channel depending on the method of connection.
- Manufacturer
- Mizuho America, Inc.
- Recall Posting Date
- 2010-07-05
- Trade Name
- Mizuha Doppler Probe
- Recall Start Date
- 2010-03-22
- Recall Number
- 53996
- Hazard Classification
- Type III
- Model or Catalog #
- 07-150-10
- Lot or Serial #
- 9781-34600, 9782-34598, 9785-34634, 9780-34586
- Reason for Recall
- The affected lots contain probes that may produce no audio signal. The lack of audio signal will be apparent to the user once the probe is connected to the Doppler transceiver and there is not a reasonable probability that the use of, or exposure to, the product will cause serious adverse health consequences.
- Manufacturer
- Ohmeda Medical, A Division of Datex Ohmeda Inc.
- Recall Posting Date
- 2010-07-05
- Trade Name
- Giraffe In-Bed Scales
- Recall Start Date
- 2010-06-08
- Recall Number
- 55120
- Hazard Classification
- Type II
- Model or Catalog #
- N/A
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- During the weighing procedure, the scale is programmed to give only 6 seconds to place the baby on the mattress, if it takes more than 6 seconds to place the baby on the mattress, the scale will display an inaccurate weight.
- Manufacturer
- Oridion Medical 1987 Ltd.
- Recall Posting Date
- 2010-08-16
- Trade Name
- A) Filterline Set
B) Smart Capnoline Plus ADU/INT 25/100 UN
C) Smart Capnoline Plus O2 ADU/INT 25UN
D) Smart Capnoline H Plus O2 ADU/INT 25UN
E) Niv-Line Adult 25UN
- Recall Start Date
- 2010-05-26
- Recall Number
- 55493
- Hazard Classification
- Type II
- Model or Catalog #
- A) XS04620
B) 009818, 010209
C) 009822
D) 010433
E) 008174
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
C) >10 Numbers, Contact Manufacturer
D) >10 Numbers, Contact Manufacturer
E) >10 Numbers, Contact Manufacturer
- Reason for Recall
- As a result of a production issue, Microstream Filterlines may not be recognized by the monitor when connected to the Microstream Filterline input connector and co2 monitoring will not commence unless the filterline is replaced. No co2 readings will be displayed and no alarms will appear until co2 monitoring starts.
- Manufacturer
- Ortho-Clinical Diagnostics Inc.
- Recall Posting Date
- 2010-08-30
- Trade Name
- Vitros Chemistry Products MALB Reagent
- Recall Start Date
- 2010-08-10
- Recall Number
- 56101
- Hazard Classification
- Type II
- Model or Catalog #
- 6801740
- Lot or Serial #
- Generation 2, 3 & 4
- Reason for Recall
- Ortho Clinical Diagnostics has received several customer complaints of lower than expected albumin results for samples that were expected to generate results significantly above the measuring (reportable) range of 0.6 - 19.0 mg/dl. These samples may be negatively biassed and give values within the reportable range without an initial dilution.
- Manufacturer
- Ortho-Clinical Diagnostics Inc.
- Recall Posting Date
- 2010-09-27
- Trade Name
- A) Vitros 5600 System Software 1.4 & Below
B) Vitros 3600 System Software 1.4 & Below
- Recall Start Date
- 2010-07-28
- Recall Number
- 55916
- Hazard Classification
- Type II
- Model or Catalog #
- A) 680 2863, 680 2864
B) 680 2865, 680 2866
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) 36000264, 36000130, 36000116
- Reason for Recall
- Ortho-Clinical Diagnostics has identified two scenarios on the Vitros 5600 Integrated System (software version 1.4 and below) and Vitros 3600 Immunodiagnostic System (software version 1.4 and below) in which Microwell and/or Microtip reagent inventory may not be properly updated, resulting in misidentified reagent packs and incorrect results or 'no result' being reported by the system.
- Manufacturer
- Ortho-Clinical Diagnostics Inc.
- Recall Posting Date
- 2010-09-27
- Trade Name
- Ortho HTLV-I/II AB-Capture Elisa Test
- Recall Start Date
- 2010-09-09
- Recall Number
- 56514
- Hazard Classification
- Type III
- Model or Catalog #
- 934200
- Lot or Serial #
- HTV208
- Reason for Recall
- Ortho-Clinical Diagnostics has received customer complaints regarding intermittent invalid plates due to elevated optical densities (OD) for the negative calibrator when using Ortho HTLV-I/II AB-Capture Elisa Test System, lot HTV208. A negative calibrator OD that is outside the package insert specification would cause an invalid plate. If this occurs the entire plate is aborted and no results are reported.
- Manufacturer
- Pelton and Crane
- Recall Posting Date
- 2010-08-02
- Trade Name
- Dental Chair Magnetic Headrest Assembly
- Recall Start Date
- 2010-03-15
- Recall Number
- 54641
- Hazard Classification
- Type II
- Model or Catalog #
- N/A
- Lot or Serial #
- Option - 40R501-XX
- Reason for Recall
- Magnetic headrest may temporarily affect the function or programming of some implantable pacemakers or defibrillators if the device is programmed to respond to a magnet. People with such devices should avoid using magnetic headrests.
- Manufacturer
- Philips Medical Systems
- Recall Posting Date
- 2010-07-05
- Trade Name
- Heartstart FR2+ Semi-Automatic Defibrillator - Main Unit
- Recall Start Date
- 2010-04-23
- Recall Number
- 55119
- Hazard Classification
- Type II
- Model or Catalog #
- M3860A, M3861A
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- Certain Heartstart FR2+ Automated External Defibrillators (AEDS) may contain a component called a voltage detector that may have a higher than expected rate of failure. Failure of the voltage detector can cause the AED battery to drain more rapidly than normal or to render the AED unusable.
- Manufacturer
- Philips Medical Systems
- Recall Posting Date
- 2010-08-02
- Trade Name
- A) Philips Heartstart HS1 Onsite Defibrillator
B) Philips Heartstart HS1 Home Defibrillator
- Recall Start Date
- 2010-04-23
- Recall Number
- 55525
- Hazard Classification
- Type II
- Model or Catalog #
- A) M5066A
B) M5068A
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
- Reason for Recall
- There is a potential for users to misread the quick reference guide (QRG) and render the device ineffective.
- Manufacturer
- Philips Medical Systems
- Recall Posting Date
- 2010-08-30
- Trade Name
- Heartstart MRX Defibrillator / Monitor
- Recall Start Date
- 2010-07-12
- Recall Number
- 56084
- Hazard Classification
- Type II
- Model or Catalog #
- M3535A, M3536A
- Lot or Serial #
- >100 Numbers, Contact Manufacturer
- Reason for Recall
- Philips has become aware that users may not fully understand the advanced features related to primary ECG lead configuration settings and the automatic lead selection behaviour of the Heartstart MRX.
- Manufacturer
- Philips Medical Systems (Cleveland), Inc.
- Recall Posting Date
- 2010-07-05
- Trade Name
- 1.5T Edge Eclipse System - Main Unit
- Recall Start Date
- 2010-04-12
- Recall Number
- 55221
- Hazard Classification
- Type II
- Model or Catalog #
- OFF-ML-1
- Lot or Serial #
- N/A
- Reason for Recall
- There is a potential for misdiagnosis - slice grids associated with images acquired after couch movement will be shown in an incorrect position when viewed on images acquired before the couch was moved.
- Manufacturer
- Philips Medical Systems (Cleveland), Inc.
- Recall Posting Date
- 2010-09-27
- Trade Name
- Mx8000 IDT Multislice CT Imaging System
- Recall Start Date
- 2010-07-14
- Recall Number
- 56344
- Hazard Classification
- Type II
- Model or Catalog #
- OFF-EI-590
- Lot or Serial #
- 3056
- Reason for Recall
- The Mx8000 IDT Multislice CT Imaging System may become inoperable if the user does not adjust the site specific protocols manually after the software upgrade.
- Manufacturer
- Philips Medizin Systeme Boeblingen GMBH
- Recall Posting Date
- 2010-08-02
- Trade Name
- A) Intellivue MP70 Patient Monitor System - Main Unit
B) Intellivue MP60 Patient Monitor System - Main Unit
- Recall Start Date
- 2010-06-24
- Recall Number
- 55526
- Hazard Classification
- Type II
- Model or Catalog #
- A) M8007A
B) M8005A
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
- Reason for Recall
- Philips has recently received reports of monitors becoming loose and detaching from the variable height mount, with the potential of causing injury.
- Manufacturer
- Philips Medzin Systeme Boeblingen GMBH
- Recall Posting Date
- 2010-08-30
- Trade Name
- Anesthetic Gas Monitor Water Trap
- Recall Start Date
- 2010-07-15
- Recall Number
- 56013
- Hazard Classification
- Type II
- Model or Catalog #
- M1657B
- Lot or Serial #
- >100 Numbers, Contact Manufacturer
- Reason for Recall
- Philips has received reports that when changing the anaesthetic gas monitor water trap (M1657B), the manifold seals on the Philips gas analysers, models M1013A, M1019A and M1026B, may be unintentionally removed. A missing seal results in leakage and may cause incorrect monitor readings of inspiratory or expiratory gas levels.
- Manufacturer
- Prollenium Medical Technologies Inc.
- Recall Posting Date
- 2010-08-30
- Trade Name
- Dermal Roller SR
- Recall Start Date
- 2010-08-09
- Recall Number
- 56052
- Hazard Classification
- Type III
- Model or Catalog #
- DR10, DR5
- Lot or Serial #
- 10-1209, 05-1209
- Reason for Recall
- The Dermal SR Roller (transdermal delivery microneedle therapy) is being recalled because of lack of assurance of sterility.
- Manufacturer
- Rehabilitation Centre for Children
- Recall Posting Date
- 2010-08-02
- Trade Name
- Tilt Standing Frame
- Recall Start Date
- 2010-06-23
- Recall Number
- 55699
- Hazard Classification
- Type I
- Model or Catalog #
- N/A
- Lot or Serial #
- Contact Manufacturer
- Reason for Recall
- If an individual on the frame is moved to an upright position without the safety straps, knee blocks and tray in place, the individual may fall forward, tipping the tilt standing frame, posing a risk of serious injury.
- Manufacturer
- Remel Inc.
- Recall Posting Date
- 2010-07-19
- Trade Name
- TB Potassium Permanganate Counterstain
- Recall Start Date
- 2010-06-15
- Recall Number
- 55303
- Hazard Classification
- Type II
- Model or Catalog #
- RE40092, RE40192
- Lot or Serial #
- 152898, 152899
- Reason for Recall
- Product deterioration, may affect the ability of the counterstain to quench background fluorescent during performance of staining procedure.
- Manufacturer
- Roche Diagnostics GMBH
- Recall Posting Date
- 2010-08-16
- Trade Name
- IgG-2 (Tina Quant IgG Gen.2)
- Recall Start Date
- 2010-07-15
- Recall Number
- 55734
- Hazard Classification
- Type III
- Model or Catalog #
- 03507432190
- Lot or Serial #
- 63009201
- Reason for Recall
- Quality control release of the Cobas C501 IgG-2 (Tina Quant IgG Gen. 2) cassette showed a too low recovery of the precinorm PUC control target value while using the cerebrospinal fluid (CSF) application.
- Manufacturer
- Roche Diagnostics GMBH
- Recall Posting Date
- 2010-08-16
- Trade Name
- Anti-CCP for Elecsys and Cobas E
- Recall Start Date
- 2010-07-21
- Recall Number
- 55874
- Hazard Classification
- Type III
- Model or Catalog #
- 05031656190
- Lot or Serial #
- 157557
- Reason for Recall
- The manufacturer identified a potential decreased recovery of the precicontrol Anti-CCP in combination with the Cobas E-Elecsys Anti-CCP Reagent Kit.
- Manufacturer
- Roche Diagnostics GMBH
- Recall Posting Date
- 2010-08-30
- Trade Name
- Elecsys Vitamin D3 assay
- Recall Start Date
- 2010-08-03
- Recall Number
- 55964
- Hazard Classification
- Type III
- Model or Catalog #
- 03314847190
- Lot or Serial #
- 159526
- Reason for Recall
- Elecsys Vitamin D3 Monoclonal Assay shows deteriorating traceability to the reference method and a negative bias when compared to the previous polyclonal antibody based version.
- Manufacturer
- Roche Diagnostics GMBH
- Recall Posting Date
- 2010-08-30
- Trade Name
- Elecsys Free T4
- Recall Start Date
- 2010-08-06
- Recall Number
- 56041
- Hazard Classification
- Type III
- Model or Catalog #
- 11731297122
- Lot or Serial #
- 155674, 157677
- Reason for Recall
- A calibration factor close to or even below 0.8 can be observed with reagent kits stored on-board for more than four weeks. In such case, the rack pack calibration cannot be released by the system.
- Manufacturer
- Roche Diagnostics GMBH
- Recall Posting Date
- 2010-08-30
- Trade Name
- A) Coaguchek XS Monitor
B) Coaguchek XS Plus Monitor
- Recall Start Date
- 2010-07-07
- Recall Number
- 56053
- Hazard Classification
- Type II
- Model or Catalog #
- A) 3666794
B) 04800842190
- Lot or Serial #
- A) All lots
B) All lots
- Reason for Recall
- The recommended cleaning procedures for the Coaguchek XS and Coaguchek XS Plus systems may cause a meter malfunction which could lead to falsely elevated PT/INR results.
- Manufacturer
- Roche Diagnostics GMBH
- Recall Posting Date
- 2010-08-30
- Trade Name
- A) Tina-Quant D-Dimer Gen.2 for Cobas Integra and Cobas C-System
B) Tina-Quant D-Dimer Gen.2 for Roche/Hitachi 917 and Modular P Analyzers
C) Tina-Quant D-Dimer Gen.2 for Roche Cobas C111 Analyzers
- Recall Start Date
- 2010-08-12
- Recall Number
- 56170
- Hazard Classification
- Type I
- Model or Catalog #
- A) 04912551190
B) 04912497190
C) 05077753190
- Lot or Serial #
- A) 62837701
B) 62838001
C) 62841101
- Reason for Recall
- The impacted lots show unusual systematic negative bias of 15% when used together with serum based calibrators and controls. The false negative result could pose a patient at risk of not being timely diagnosed for thrombosis and embolism.
- Manufacturer
- Roche Diagnostics GMBH
- Recall Posting Date
- 2010-08-30
- Trade Name
- Precicontrol Universal
- Recall Start Date
- 2010-06-25
- Recall Number
- 55469
- Hazard Classification
- Type II
- Model or Catalog #
- 11731416190
- Lot or Serial #
- 15418301, 15418305
- Reason for Recall
- The package insert for the Precicontrol Universal has been modified to exclude the TG component from the storage stability of 3 days at 2-8c after reconstitution due to instability related to an aging effect.
- Manufacturer
- Roche Diagnostics GMBH
- Recall Posting Date
- 2010-09-27
- Trade Name
- MagNa Pure LC 2.0 Instrument
- Recall Start Date
- 2010-08-27
- Recall Number
- 56376
- Hazard Classification
- Type III
- Model or Catalog #
- 05197686001
- Lot or Serial #
- All lots
- Reason for Recall
- There is potential for wrong error handling by the MagNa Pure LC 2.0 Instrument software where the extraction protocol is performed only during a portion of the estimated run time and then exits out of the extraction process.
- Manufacturer
- Roche Diagnostics GMBH
- Recall Posting Date
- 2010-09-27
- Trade Name
- Cobas Integra Glucose HK Gen. 3
- Recall Start Date
- 2010-08-31
- Recall Number
- 56482
- Hazard Classification
- Type III
- Model or Catalog #
- N/A
- Lot or Serial #
- All lots
- Reason for Recall
- A change must be made to the glucose urine application of the Cobas C501 unit of Cobas 6000 analyzer due to precision problems and outliers when using aqueous urine control materials at a concentration level of 1.4 mmol/l.
- Manufacturer
- Roche Diagnostics Ltd.
- Recall Posting Date
- 2010-07-19
- Trade Name
- MagNa Pure Compact Instrument
- Recall Start Date
- 2010-06-07
- Recall Number
- 55083
- Hazard Classification
- Type III
- Model or Catalog #
- 03730964001, 03730972001, 03731146001, 04802993001
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- Manufacturer internal investigation revealed that the transfer of the nucleic acid contained in the eluates from the reagent cartridge to the elution tube was impacted with a frequency of 5%. Only a certain portion of the eluated material was transferred to the elution tubes.
- Manufacturer
- Roche Diagnostics Ltd.
- Recall Posting Date
- 2010-08-16
- Trade Name
- A) Amplilink Software Version 3.1.1
B) Amplilink Software Version 3.1.2
C) Amplilink Software Version 3.2
D) Amplilink Software Version 3.2.1
E) Amplilink Software Version 3.2.2
F) Amplilink Software Version 3.2.3
- Recall Start Date
- 2010-06-18
- Recall Number
- 55272
- Hazard Classification
- Type II
- Model or Catalog #
- A) 04807197001
B) 04855094001
C) 04862392001
D) 05093236001
E) 05201357001
F) 05666678001
- Lot or Serial #
- A) N/A
B) N/A
C) N/A
D) N/A
E) N/A
F) N/A
- Reason for Recall
- Out-of-range control failures flagged as failed did not invalidate patient samples in the batch for the customers using the Amplilink SW 3.1.X & 3.2X in combination with Cobas Amplicor assays.
- Manufacturer
- Roche Diagnostics Ltd.
- Recall Posting Date
- 2010-09-13
- Trade Name
- A) Amplilink Software Version 3.1.1
B) Amplilink Software Version 3.1.2
C) Amplilink Software Version 3.2
D) Amplilink Software Version 3.2.1
E) Amplilink Software Version 3.2.2
F) Amplilink Software Version 3.2.3
Updated recall # 55272 previously posted on 2010-08-16.
- Recall Start Date
- 2010-06-18
- Recall Number
- 55272
- Hazard Classification
- Type II
- Model or Catalog #
- A) 04807197001
B) 04855094001
C) 04862392001
D) 05093236001
E) 05201357001
F) 05666678001
- Lot or Serial #
- A) N/A
B) N/A
C) N/A
D) N/A
E) N/A
F) N/A
- Reason for Recall
- Out-of-range control failures flagged as failed did not invalidate patient samples in the batch for the customers using the Amplilink SW 3.1.X & 3.2X in combination with COBAS© AmpliPrep / COBAS© TaqMan©.
- Manufacturer
- Roche Diagnostics Ltd.
- Recall Posting Date
- 2010-09-27
- Trade Name
- Cobas IT 3000 Software
- Recall Start Date
- 2010-09-01
- Recall Number
- 56465
- Hazard Classification
- Type I
- Model or Catalog #
- 05161401001
- Lot or Serial #
- Version 2.03.09
- Reason for Recall
- Critical software issue when using automatic medical validation for Cobas IT 3000 version 2.03.09. A programming error was notice which causes the automatic medical validation to release results which should be held for review.
- Manufacturer
- Sequal Technologies Inc.
- Recall Posting Date
- 2010-09-13
- Trade Name
- Sequal Eclipse Oxygen System
- Recall Start Date
- 2009-12-07
- Recall Number
- 53321
- Hazard Classification
- Type I
- Model or Catalog #
- N/A
- Lot or Serial #
- 073002972, 063005184, 073006282, 073000585, 073007840, 073006062, 073002301, 073004651,073003155,073004456
- Reason for Recall
- Eclipse oxygen system units may experience premature wear on electrical connector between battery bridge printed circuit board and power manager printed circuit board that could lead to temporary power interruption when operating from power cartridge, causing unit to sound an alarm and shut down.
- Manufacturer
- Siemens AG
- Recall Posting Date
- 2010-08-02
- Trade Name
- Syngo Imaging
- Recall Start Date
- 2010-06-21
- Recall Number
- 55479
- Hazard Classification
- Type III
- Model or Catalog #
- 10 014 063
- Lot or Serial #
- 1094, 1200, 1262
- Reason for Recall
- Potential malfunction and hence hazard to patients when using Syngo Imaging version lines VB20, VB30, VB35 in combination with RIS (radiology information system) that is violating dicoms (digital imaging and communication in medicine) standard by creating non-unique study instance UIDS (unique identifiers).
- Manufacturer
- Siemens AG
- Recall Posting Date
- 2010-08-16
- Trade Name
- A) Artis Zee Ceiling
B) Artis Zee Biplane
- Recall Start Date
- 2010-07-15
- Recall Number
- 55780
- Hazard Classification
- Type III
- Model or Catalog #
- A) 10094137
B) 10094141
- Lot or Serial #
- A) 146288, 146314
B) 153275
- Reason for Recall
- A potential malfunction and a possible hazard to patients when using the Artis systems with motor controller failures caused by tolerances on the electronic parts.
- Manufacturer
- Siemens AG
- Recall Posting Date
- 2010-08-30
- Trade Name
- A) Mevatron Linear Accelerators
B) Artiste MV-Linear Accelerator-Main Unit
- Recall Start Date
- 2010-08-11
- Recall Number
- 56152
- Hazard Classification
- Type III
- Model or Catalog #
- A) 1940035
B) 8139789
- Lot or Serial #
- A) 2473, 2559, 2618, 3235, 3236, 3376, 2459, 2616, 3291, 3356, 3378, 3382, 3406, 3742, 2239, 2267, 2264
B) 5414, 5485
- Reason for Recall
- Siemens received a report on a system with 160MLC, where a misalignment of treatment fields was detected during quality assurance. This misalignment was caused by a defective bearing of the collimator.
- Manufacturer
- Siemens AG
- Recall Posting Date
- 2010-09-13
- Trade Name
- A) Axiom Artis DTA
B) Axiom Artis FA Alpha
C) Axiom Artis MP Diagnostic
- Recall Start Date
- 2010-08-18
- Recall Number
- 56246
- Hazard Classification
- Type III
- Model or Catalog #
- A) 7008605
B) 5904441
C) 5904466
- Lot or Serial #
- A) 32070
B) 25006, 25055, 25076, 25147
C) 20069, 20070, 20151, 20163, 20215
- Reason for Recall
- A potential danger by the image intensifier system with perivision options. During perivision the image system can sporadically crash. The customer has to repeat the perivision and to apply contrast media and x-ray to the patient.
- Manufacturer
- Siemens Healthcare Diagnostics Inc.
- Recall Posting Date
- 2010-07-05
- Trade Name
- Dimension Tacrolimus Flex Reagent Cartridge, DF107
- Recall Start Date
- 2010-06-16
- Recall Number
- 55304
- Hazard Classification
- Type III
- Model or Catalog #
- DF107
- Lot or Serial #
- DA0334
- Reason for Recall
- Siemens has confirmed customer complaints regarding low recovery of the Tacrolimus Flex Reagent Cartridge DF107.
- Manufacturer
- Siemens Healthcare Diagnostics Inc.
- Recall Posting Date
- 2010-08-02
- Trade Name
- Dimension Vista System-Enzymatic Creatinine Assay (ECREA)
- Recall Start Date
- 2010-06-25
- Recall Number
- 55520
- Hazard Classification
- Type II
- Model or Catalog #
- K1270
- Lot or Serial #
- 09350AC, 10011AF, 10131AC
- Reason for Recall
- Siemens Healthcare Diagnostics received customer complaints regarding the Dimension Vista Assay Enzymatic Creatinine Flex Reagent Cartridge. Siemens has confirmed an issue which may cause enzymatic creatinine (ECREA) results to be falsely depressed up to 0.4 mg/dl or falsely elevated up to 0.6 mg/dl across the assay range.
- Manufacturer
- Siemens Healthcare Diagnostics Inc.
- Recall Posting Date
- 2010-08-30
- Trade Name
- A) ADVIA Centaur Hepatitis B Core Total (HBCT) Assay
B) ADVIA Centaur XP System-HBC Total (HBCT) assay (donor screening for transplantation)
C) Advia Centaur CP System-HBC Total (HBCT) assay (donor screening for transplantation)
- Recall Start Date
- 2010-07-05
- Recall Number
- 55559
- Hazard Classification
- Type III
- Model or Catalog #
- A) 07566733
B) 07566733
C) 07566733
- Lot or Serial #
- A) 49471033, 50694033, 51071033, 51628033, 52260033, 52809033, 53674033
B) 50694033, 49471033, 51071033, 51628033, 52260033, 52809033, 53674033
C) 49471033, 50694033, 51071033, 51628033, 52260033, 52809033, 53674033
- Reason for Recall
- Siemens Healthcare Diagnostics confirms lower than expected specificity rates that may increase the frequency of false positive test results for this assay. Siemens believes that false positive results will lead to further investigation, and that the possibility of misinterpretation and incorrect patient management are low.
- Manufacturer
- Siemens Healthcare Diagnostics Inc.
- Recall Posting Date
- 2010-07-19
- Trade Name
- Enzymatic Creatinine (EZCR) Flex Reagent Cartridge, DF270B
- Recall Start Date
- 2010-06-24
- Recall Number
- 55437
- Hazard Classification
- Type II
- Model or Catalog #
- DF270B
- Lot or Serial #
- BD1008, CB1050
- Reason for Recall
- Siemens Healthcare Diagnostics received customer complaints regarding the Dimension Enzymatic Creatinine (EZCR) Flex Reagent Cartridge. Siemens recently confirmed an issue which may cause EZCR elevations across the assay range. This issue may or may not be detected by quality control, as the degree and onset of elevated results may vary and is likely dependent on multiple factors such as method mix or instrument maintenance.
- Manufacturer
- Siemens Healthcare Diagnostics Products Limited
- Recall Posting Date
- 2010-08-16
- Trade Name
- Immulite 2500 System-Human Chorionic Gonadotropin (HCG) Assay
- Recall Start Date
- 2010-06-29
- Recall Number
- 55581
- Hazard Classification
- Type III
- Model or Catalog #
- L5KCG2, L5KCG6
- Lot or Serial #
- 123, 151
- Reason for Recall
- Siemens Healthcare Diagnostics Ltd has determined that the dose at which a patient sample may hook for lots 123 and above is lower than indicated in the original instructions for use (no hook effect seen up to 600,000 miu/ml, instead of claimed 2,860,000 miu/ml).
- Manufacturer
- Siemens Medical Solutions USA, Inc.
- Recall Posting Date
- 2010-08-16
- Trade Name
- Syngo Dynamics
- Recall Start Date
- 2010-07-27
- Recall Number
- 55938
- Hazard Classification
- Type III
- Model or Catalog #
- 100 90 480
- Lot or Serial #
- 92241
- Reason for Recall
- For customers that are using the IHE displayable report profile, signing off (depending on your configuration) a report will result in an automatic upload of that report to your information system. However, if you create an addendum to a report that has been previously uploaded, the addendum must be uploaded manually using the #send to#.# feature in the Syngo Dynamics reporting system. If the addendum is not sent manually, the older report in your information system will not match the more recent report in your Syngo Dynamics system.
- Manufacturer
- Siemens Medical Solutions USA, Inc.
- Recall Posting Date
- 2010-09-27
- Trade Name
- A) Acuson Antares Ultrasound System
B) Acuson Antares System with 5.0 Upgrades
- Recall Start Date
- 2010-08-25
- Recall Number
- 56322
- Hazard Classification
- Type III
- Model or Catalog #
- A) 5936518
B) 10032746
- Lot or Serial #
- A) 112267, 10134
B) 113469, 114573
- Reason for Recall
- Have determined that there is an error in the estimation of mean pressure gradient in the cardiac measurement package available on Antares systems at product versions 4.0 and 5.0.
- Manufacturer
- Siemens Medical Solutions USA, Inc.
- Recall Posting Date
- 2010-09-27
- Trade Name
- Acuson S2000
- Recall Start Date
- 2010-08-25
- Recall Number
- 56323
- Hazard Classification
- Type III
- Model or Catalog #
- 10041461
- Lot or Serial #
- 200252, 200288, 200334, 200926, 200979, 200995, 202301
- Reason for Recall
- Have determined that there is an error in the estimation of mean pressure gradient in the cardiac measurement package available on Acuson S2000 systems.
- Manufacturer
- Siemens Medical Solutions USA, Inc.
Molecular Imaging Group
- Recall Posting Date
- 2010-08-16
- Trade Name
- A) E-Cam Profile Refurbished System
B) Collimators for E Cam
C) E Cam Signature Series
- Recall Start Date
- 2010-07-07
- Recall Number
- 55779
- Hazard Classification
- Type III
- Model or Catalog #
- A) 52 44 012
B) 43 55 108
C) 7333243
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) 1005, 3005, 3034, 1004
C) >10 Numbers, Contact Manufacturer
- Reason for Recall
- The possibility for the patient getting injured while holding on the E-Cam patient handling system (PHS) and the technologist puts the PHS to its home position. This could possibly pose a potential safety hazard.
- Manufacturer
- Smith & Nephew, Inc.
- Recall Posting Date
- 2010-07-05
- Trade Name
- Gigli Saw Blade, 12 Inch
- Recall Start Date
- 2010-06-15
- Recall Number
- 55302
- Hazard Classification
- Type II
- Model or Catalog #
- 210482
- Lot or Serial #
- 08KM00086
- Reason for Recall
- Smith & Nephew has found that the sterile pouches that contain the Gigli Saw Blade are not sealed. One entire end of the package is totally void of a seal.
- Manufacturer
- Smith & Nephew, Inc., Endoscopy Division
- Recall Posting Date
- 2010-08-30
- Trade Name
- Dytonics Bonecutter Electroblade
- Recall Start Date
- 2010-08-04
- Recall Number
- 56143
- Hazard Classification
- Type II
- Model or Catalog #
- 7205962
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- Smith & Nephew has received complaints in which a small number of patients have received skin burns associated with the usage of the Dyonics Bonecutter Electroblade.
- Manufacturer
- Smith & Nephew, Inc., Endoscopy Division
- Recall Posting Date
- 2010-09-27
- Trade Name
- BIORAPTOR 2.9mm Suture Anchor
- Recall Start Date
- 2010-08-30
- Recall Number
- 56446
- Hazard Classification
- Type III
- Model or Catalog #
- 72200774
- Lot or Serial #
- 50306820
- Reason for Recall
- Smith & Nephew has identified a single lot of the BIORAPTOR 2.9mm Suture Anchor as having been assembled with 2 suture braids rather than 1.
- Manufacturer
- Smiths Medical ASD, Inc.
- Recall Posting Date
- 2010-07-19
- Trade Name
- A) Medfusion Syringe Infusion Pump, 3010
B) Medfusion Syringe Infusion Pump, 3010A
- Recall Start Date
- 2010-06-18
- Recall Number
- 55335
- Hazard Classification
- Type II
- Model or Catalog #
- A) 3010
B) 3010A
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
- Reason for Recall
- Smiths Medical has found that pumps with obsolete software versions 2.0.2, 2.0.3, and 2.0.4, may continue to run past the set volume limit, if all of the following occur: 1) the volume over time (delivery method) mode is used, 2) the volume over time mode is re-accessed through the "recall last settings" function, and 3) the syringe is over-filled.
- Manufacturer
- Southmedic Inc.
- Recall Posting Date
- 2010-09-27
- Trade Name
- Stainless Safety Scalpel Sterile
- Recall Start Date
- 2010-09-02
- Recall Number
- 56459
- Hazard Classification
- Type II
- Model or Catalog #
- 73-0610, 73-0611, 73-0615
- Lot or Serial #
- 070610, 071510, 070910
- Reason for Recall
- Southmedic Inc. is conducting a voluntary recall of specific lot numbers of the Southmedic safety scalpel due to a potential of unsealed sterile pouches.
- Manufacturer
- Spacelabs Medical, Inc.
- Recall Posting Date
- 2010-09-13
- Trade Name
- A) Ultraview SL2600 Compact Monitor
B) Ultraview SL2200 Compact Monitor
C) Ultraview SL Compact Monitor
- Recall Start Date
- 2009-11-12
- Recall Number
- 56205
- Hazard Classification
- Type II
- Model or Catalog #
- A) 91370
B) 91367
C) 91369
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) 1367-000488, 1367-000447, 1367-000448, 1367-000449, 1367-000484
C) >10 Numbers, Contact Manufacturer
- Reason for Recall
- Affected Spacelabs' portable patient monitors may stop working when on AC power or shut down and do not power on again. The failure mode is caused by the capacitor identified as C329 malfunctioning and initiating a monitor shutdown.
- Manufacturer
- Sterngold Dental, LLC
- Recall Posting Date
- 2010-09-27
- Trade Name
- A) ERA Implant Master Prosthetic Kit
B) ERA Mini Dental Implant System,Complete
- Recall Start Date
- 2010-08-31
- Recall Number
- 56385
- Hazard Classification
- Type III
- Model or Catalog #
- A) 811914Z
B) 905119Z
- Lot or Serial #
- A) Z3303
B) Z3561
- Reason for Recall
- Products do not contain instructions for use. These instructions contain important safety information related to user cleaning and re-sterilization.
- Manufacturer
- Sterngold Dental, LLC
- Recall Posting Date
- 2010-09-27
- Trade Name
- Sterngold/Hader Torque Wrench Kit
- Recall Start Date
- 2010-08-31
- Recall Number
- 56387
- Hazard Classification
- Type III
- Model or Catalog #
- 905104Z
- Lot or Serial #
- Z3379
- Reason for Recall
- Product does not contain the correct instructions for use. These instructions contain important safety information related to user cleaning and re-sterilization.
- Manufacturer
- Stryker Endoscopy
- Recall Posting Date
- 2010-08-02
- Trade Name
- Stryker Disposable Scope Warmers
- Recall Start Date
- 2010-06-28
- Recall Number
- 55499
- Hazard Classification
- Type II
- Model or Catalog #
- 0502-360-000
- Lot or Serial #
- All lots
- Reason for Recall
- Stryker Endoscopy has become aware that there is potential for a breach in the sterile packaging of the Disposable Scope Warmers.
- Manufacturer
- Stryker Endoscopy
- Recall Posting Date
- 2010-08-02
- Trade Name
- Crossfire System - Console
- Recall Start Date
- 2010-07-12
- Recall Number
- 55664
- Hazard Classification
- Type II
- Model or Catalog #
- 475-000-000
- Lot or Serial #
- All lots
- Reason for Recall
- Stryker Endoscopy has become aware that there is potential for the continuous activation of the radio frequency powered shaver handpieces when using the iswitch wireless footswitch with the crossfire console resulting in prolongation of surgery.
- Manufacturer
- Stryker Instruments
- Recall Posting Date
- 2010-07-05
- Trade Name
- A) Commandf II Microelectric System
B) Core Drills
- Recall Start Date
- 2010-06-08
- Recall Number
- 55206
- Hazard Classification
- Type II
- Model or Catalog #
- A) 2296-301-000
B) 5400-300-000
- Lot or Serial #
- A) All lots
B) All lots
- Reason for Recall
- Stryker instruments has become aware that the inner diameter of the bur guards may be out of specification. It was discovered that the injection moulding tool was leaving additional material on the inner surface of the bur guard that is closest to the rotating nose of the impaction drill.
- Manufacturer
- Stryker Instruments
- Recall Posting Date
- 2010-08-30
- Trade Name
- Instructions for use (IFU) for Painpump2 Block-Aid
- Recall Start Date
- 2010-08-05
- Recall Number
- 56046
- Hazard Classification
- Type II
- Model or Catalog #
- 0575-100-000
- Lot or Serial #
- 09355012
- Reason for Recall
- Stryker Instruments has revised the instructions for use (IFU) for the Painpump2 Block-Aid.
- Manufacturer
- Stryker Medical
- Recall Posting Date
- 2010-07-19
- Trade Name
- A) Instruction Manuals for Epic II Critical Care Beds
B) Instruction Manuals for Zoom II Critical Care Beds
- Recall Start Date
- 2010-06-21
- Recall Number
- 55436
- Hazard Classification
- Type II
- Model or Catalog #
- A) 2030-000-000
B) 2040-000-000
- Lot or Serial #
- A) N/A
B) N/A
- Reason for Recall
- Stryker Medical has become aware that the Stryker Critical Care Beds were reportedly displaying symptoms of unsmooth and noisy fowler motion. Continuous use could potentially result in a situation in which the manual CPR drop on the bed may become inoperable at fowler angles of approximately 45 degrees or more.
- Manufacturer
- Stryker Medical
- Recall Posting Date
- 2010-08-02
- Trade Name
- M1 Ambulance Cot
- Recall Start Date
- 2010-06-10
- Recall Number
- 55697
- Hazard Classification
- Type II
- Model or Catalog #
- 6100-000-000
- Lot or Serial #
- N/A
- Reason for Recall
- Stryker Medical has become aware that the head end slide tube, bushings, and head end release rod have the potential to break during use, potentially resulting in the collapse of the head end of the cot.
- Manufacturer
- Stryker Medtech K.K.
- Recall Posting Date
- 2010-08-16
- Trade Name
- Stryker Sonopet Ultrasonic Aspirator Console
- Recall Start Date
- 2010-07-19
- Recall Number
- 55828
- Hazard Classification
- Type II
- Model or Catalog #
- 5450-850-000
- Lot or Serial #
- All lots
- Reason for Recall
- Stryker Instruments became aware that the Sonopet console did not meet the IEC 60601-1 standard (electromedical safety standard).
- Manufacturer
- Stryker Orthopeadics
- Recall Posting Date
- 2010-08-30
- Trade Name
- Scorpio T72 Insert Tibial Trials
- Recall Start Date
- 2010-07-27
- Recall Number
- 55981
- Hazard Classification
- Type II
- Model or Catalog #
- T-72
- Lot or Serial #
- >100 Numbers, Contact Manufacturer
- Reason for Recall
- Stryker Orthopaedics has become aware that there is the potential for the Scorpio T72 Insert Tibial Trials to be oversized by a maximum of .024"(0.6mm). It was determined that an oversized Tibial Insert Trial may sit on the anterior edge of the tibial baseplate if used for secondary trialing in a procedure. If there is an inability to seat the insert trial and impaction is used to seat the trial, fracture of the device may occur.
- Manufacturer
- Sybron Endo
- Recall Posting Date
- 2010-08-16
- Trade Name
- K3 Gutta Percha
- Recall Start Date
- 2010-07-14
- Recall Number
- 55901
- Hazard Classification
- Type III
- Model or Catalog #
- 825-0420
- Lot or Serial #
- 020210
- Reason for Recall
- Mislabeled packages of the K3 Gutta Percha .04/20 points (part number 825-0420) contained the K3 Gutta Percha .04/40.
- Manufacturer
- Synthes (Canada) Ltd.
- Recall Posting Date
- 2010-09-13
- Trade Name
- Insight Retractor System Frames
- Recall Start Date
- 2010-08-12
- Recall Number
- 56200
- Hazard Classification
- Type II
- Model or Catalog #
- 03.615.002, 03.615.100
- Lot or Serial #
- N/A
- Reason for Recall
- The holding pin on the retractor frame was reported as unscrewing and falling out. If this were to occur while the retractor is in use, the retraction will not be affected but the fallens screw has the potential to enter the surgical wound undetected. This may result in an un-retrieved device fragment.
- Manufacturer
- Synthes (Canada) Ltd.
- Recall Posting Date
- 2010-09-13
- Trade Name
- Implant Holder for Synfix-LR - System Technique Guide
- Recall Start Date
- 2010-08-13
- Recall Number
- 56201
- Hazard Classification
- Type II
- Model or Catalog #
- 03.802.039
- Lot or Serial #
- N/A
- Reason for Recall
- Synthes (Canada), Ltd is initiating a medical device labelling recall related to the Synthes Synfix-LR System Technique Guide, specifically regarding use of the Synfix implant holder. The surgical technique presented in the guide has been revised and emphasizes the importance of ensuring that the Synfix-LR implant holder is not cross threaded when attached to the Synfix-LR implant. The revision also describes the use of holding forceps.
- Manufacturer
- Sysmex Corporation
- Recall Posting Date
- 2010-09-27
- Trade Name
- A) XT-2000I Analyzer - Sysmex Hematology Analyzers
B) Sysmex XT-1800I Analayzer - Sysmex Hematology Analyzers
C) XT-4000I Hematology Analyzer
- Recall Start Date
- 2010-08-27
- Recall Number
- 56348
- Hazard Classification
- Type II
- Model or Catalog #
- A) 013-2531-8
B) 023-0531-6
C) 013-2553-0
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) 16975
C) 11165, 11232
- Reason for Recall
- A design error introduced in the field programmable gate array (FPGA) was discovered. This FPGA is a logic circuit on a chip embedded on the printed computer board (PCB) in the main unit of the XT-series analyzers. The error can present as an increase in the platelet (PLT) count of up to double the correct count and/or the generation of excessive PLT sampling errors. The increase in count and errors is caused by the circuit seeing a single platelet and creating two pulses that are included in the reported PLT count.
- Manufacturer
- Technidata S.A.S.
- Recall Posting Date
- 2010-07-05
- Trade Name
- TD-Synergy Software
- Recall Start Date
- 2009-12-30
- Recall Number
- 52751
- Hazard Classification
- Type III
- Model or Catalog #
- TD-SYN-01
- Lot or Serial #
- All lots
- Reason for Recall
- If some microbiology results have not been consolidated yet, and an event triggers replication from the production database to the management database, then the non-consolidated results are overwritten by blank results.
- Manufacturer
- Technidata S.A.S.
- Recall Posting Date
- 2010-08-30
- Trade Name
- TD-Synergy
- Recall Start Date
- 2010-07-02
- Recall Number
- 55962
- Hazard Classification
- Type II
- Model or Catalog #
- TD-SYN-01
- Lot or Serial #
- N/A
- Reason for Recall
- In the ERM session, when results are of coded text type, incorrect data could be displayed in the tool tip on the current result field.
- Manufacturer
- Technidata S.A.S.
- Recall Posting Date
- 2010-08-30
- Trade Name
- A) TDLims
B) TD-Synergy
- Recall Start Date
- 2010-07-02
- Recall Number
- 55963
- Hazard Classification
- Type II
- Model or Catalog #
- A) TD-LIS-01
B) TD-SYN-01
- Lot or Serial #
- A) N/A
B) N/A
- Reason for Recall
- Using F8 key in the ERM session to add a combined test can result in the modification of the value of a result that has already been entered for a test of the request. This could only occur if the combined test added by F8 includes at least one test with the "duplicate result" property set to ?yes?.
- Manufacturer
- Teleflex Medical
- Recall Posting Date
- 2010-08-02
- Trade Name
- Rusch Care Premium Drain Bag
- Recall Start Date
- 2010-07-07
- Recall Number
- 55666
- Hazard Classification
- Type II
- Model or Catalog #
- 390000
- Lot or Serial #
- 20100105
- Reason for Recall
- During an inspection wet product was found that was due to improper aeration during sterilization of the product and the sterile integrity could not be guaranteed.
- Manufacturer
- Teleflex Medical
- Recall Posting Date
- 2010-08-16
- Trade Name
- IMA / ENT Blade Electrode
- Recall Start Date
- 2010-07-30
- Recall Number
- 55952
- Hazard Classification
- Type II
- Model or Catalog #
- 809335
- Lot or Serial #
- All lots
- Reason for Recall
- There is a potential for peeling and melting of the insulation of the cautery tip of the IMA / ENT Blade Electrodes. This can result in the tip losing integrity leading to sparking or arcing of the current at the blade tip and result in a burn or electrical shock.
- Manufacturer
- Terumo Cardiovascular System Corporation
- Recall Posting Date
- 2010-07-05
- Trade Name
- Sarns Disposable Centrifugal Pump
- Recall Start Date
- 2010-06-02
- Recall Number
- 55167
- Hazard Classification
- Type II
- Model or Catalog #
- 164275
- Lot or Serial #
- N/A
- Reason for Recall
- Terumo Cardiovascular Systems Corp. received reports of loss of forward with the Sarns Disposable Centrifugal Pump. Terumo CVS investigation determined that in the cases reported that impeller in the disposable shaft separated from the shaft causing it to stop rotating.
- Manufacturer
- Terumo Cardiovascular System Corporation
- Recall Posting Date
- 2010-07-05
- Trade Name
- Sarns Cardioplegia Set with MP-4 Module
- Recall Start Date
- 2010-06-10
- Recall Number
- 55224
- Hazard Classification
- Type II
- Model or Catalog #
- 71032
- Lot or Serial #
- ME19
- Reason for Recall
- Terumo Cardiovascular Systems (Terumo CVS) has received four reports of an occluded vent port on the Sarns Cardioplegia Set with MP-4 module. In all reports the defect was evident during priming. Terumo CVS investigation found that for some units in certain lots of MP-4 modules, a damaged pin in the injection mold allowed the vent port to fill with excess plastic.
- Manufacturer
- Tomotherapy Incorporated
- Recall Posting Date
- 2010-07-19
- Trade Name
- TomoTherapy HI-ART System
- Recall Start Date
- 2010-06-23
- Recall Number
- 55439
- Hazard Classification
- Type II
- Model or Catalog #
- 101264
- Lot or Serial #
- 10037, 10169
- Reason for Recall
- The planning station's printed plan reports may indicate statistics which are not correct. The TomoTherapy Hi-Art treatment planning system presents DVH, isodose and dose statistics information on the user interface (or on-screen) which are accurate and unaffected by this issue. However, if erroneous dose statistics were printed and then provided to third parties to guide therapy in retreatment situations, improper therapy could result.
- Manufacturer
- Toshiba Medical Systems Corporation
- Recall Posting Date
- 2010-07-19
- Trade Name
- Kalare
- Recall Start Date
- 2010-06-07
- Recall Number
- 55449
- Hazard Classification
- Type II
- Model or Catalog #
- DUA-450F
- Lot or Serial #
- 99A0562042, 99A05X2054, 99A05X2055, 99A05Y2057, 99A0612074, 99A0622078, 99A0672105, A4612006
- Reason for Recall
- In diagnostic x-ray system Kalare tabletop lateral movement cannot be performed, or tabletop lateral movement is performed at extremely slow speed. As a result, positioning of the patient cannot be performed by tabletop movement.
- Manufacturer
- Toshiba Medical Systems Corporation
- Recall Posting Date
- 2010-08-16
- Trade Name
- Aquilion One System - Main Unit
- Recall Start Date
- 2010-07-22
- Recall Number
- 55835
- Hazard Classification
- Type II
- Model or Catalog #
- TSX-301A/2
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- In Aquilion One systems, communication between units in the console may be cut off due to failure of a component (hub).
- Manufacturer
- Toshiba Medical Systems Corporation
- Recall Posting Date
- 2010-09-13
- Trade Name
- Aquilion One
- Recall Start Date
- 2010-08-06
- Recall Number
- 56173
- Hazard Classification
- Type II
- Model or Catalog #
- TSX-301A/2
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- In Aquilion One systems, in some cases the pre-scan dose values displayed in the exam plan editing window is lower than what the displayed value should be when the scanner executes the exam plan.
- Manufacturer
- Toshiba Medical Systems Corporation
- Recall Posting Date
- 2010-09-13
- Trade Name
- Aplio Artida
- Recall Start Date
- 2010-08-05
- Recall Number
- 56227
- Hazard Classification
- Type II
- Model or Catalog #
- SSH-880CV
- Lot or Serial #
- W1B0862082, W1F0982235, WIF0982239, WIF0982240
- Reason for Recall
- In Artida Ultrasound Systems model SSH-880CV, there are two issues noted regarding (1) some measurement values are displayed incorrectly on the report page, and (2) temperature-related warnings are not consistently displayed.
- Manufacturer
- Trinity Biotech USA
- Recall Posting Date
- 2010-09-13
- Trade Name
- Captia VZV IgM Kit
- Recall Start Date
- 2010-07-17
- Recall Number
- 56303
- Hazard Classification
- Type II
- Model or Catalog #
- 2325650
- Lot or Serial #
- 122, 123, 124, 125
- Reason for Recall
- Disproportionate drop in absorbance of calibrator causes ISR to run out of range. Risk of a false positive.
- Manufacturer
- Trinity Biotech USA
- Recall Posting Date
- 2010-09-27
- Trade Name
- Captia EBV VCA IGM
- Recall Start Date
- 2010-07-28
- Recall Number
- 56527
- Hazard Classification
- Type II
- Model or Catalog #
- 2325760
- Lot or Serial #
- 084
- Reason for Recall
- Low positive control range for the low positive control (5760-4-003078) included in the Elisa Kit is incorrect. The correct range for the control is (0.91-2.3).
- Manufacturer
- Tyco Healthcare Liebel-Flarsheim Company
- Recall Posting Date
- 2010-07-05
- Trade Name
- Hydra Vision DR60/80 System, Generators
- Recall Start Date
- 2010-06-14
- Recall Number
- 55273
- Hazard Classification
- Type II
- Model or Catalog #
- 700539, 700540
- Lot or Serial #
- All lots
- Reason for Recall
- A smell of burning/smoke is emanating from the x-ray generators. This is caused by a failed printed circuit board electronic component and an internal electronic transformer. These failures will cause the generator to become inoperative.
- Manufacturer
- United States Surgical, A Division of Tyco Healthcare Group
- Recall Posting Date
- 2010-07-19
- Trade Name
- Auto Suture Protack Fixation Device
- Recall Start Date
- 2010-06-28
- Recall Number
- 55465
- Hazard Classification
- Type II
- Model or Catalog #
- 174006
- Lot or Serial #
- P0B0753
- Reason for Recall
- Certain Protack Fixation devices fire one helical tack and then become inoperable.
- Manufacturer
- Varian Medical Systems, Inc.
- Recall Posting Date
- 2010-08-16
- Trade Name
- A) Clinac Linear Accelerator
B) Trilogy Radiotherapy Delivery System
C) Clinac IX Radio Therapy Delivery System
D) Novalis TX Linear Accelerator
- Recall Start Date
- 2010-05-27
- Recall Number
- 55377
- Hazard Classification
- Type II
- Model or Catalog #
- A) 2100C, 21EX, 21EXD, 21EXS
B) IX, IXS
C) IX, IXS, TRILOGY
D) NOVALIS TX
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) H140836, H291166, H291172, H293787, H293883, H294134, H294403, H294551, H294750
C) >10 Numbers, Contact Manufacturer
D) H294674
- Reason for Recall
- The auto "goto" or auto "setup" functions in C-series version 7 software ignore couch angle exceeded motion zone unless a tolerance is defined by the user (gantry rotation exclusion now only applies by default if OBI is present and deployed. Customer facilities using both C-series version 6 & 7 may get confused because couch safety zone no longer exists by default in version 7.
- Manufacturer
- Varian Medical Systems, Inc.
- Recall Posting Date
- 2010-08-16
- Trade Name
- A) Varian Clinac Linear Accelerator
B) Trilogy Radiotherapy Delivery System
C) Varian Clinac IX Radiotherapy Delivery System
D) Novalis TX Linear Accelerator
- Recall Start Date
- 2010-06-23
- Recall Number
- 55612
- Hazard Classification
- Type II
- Model or Catalog #
- A) N/A
B) N/A
C) N/A
D) N/A
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
C) >10 Numbers, Contact Manufacturer
D) >10 Numbers, Contact Manufacturer
- Reason for Recall
- Anomalies have been identified with the Clinac Collimator and couches that can result in small patient or field positioning errors that may adversely affect high - dose treatments requiring small positioning tolerances, such as stereotactic radiosurgery (SRS) and stereotactic radiotherapy (SRT). Refer to 'C- series Clinacs when used for hypo-fractioned radiotherapy' dated 06/23/2010 for further information.
- Manufacturer
- Varian Medical Systems, Inc.
- Recall Posting Date
- 2010-09-13
- Trade Name
- Eclipse Treatment Planning (BrachyVision)
- Recall Start Date
- 2010-08-01
- Recall Number
- 56293
- Hazard Classification
- Type II
- Model or Catalog #
- 8.2, 8.5, 8.6, 8.9
- Lot or Serial #
- N/A
- Reason for Recall
- When incorrect syntax is used in the template file, a misinterpretation of the desired value can occur.
- Manufacturer
- Varian Medical Systems, Inc.
- Recall Posting Date
- 2010-07-05
- Trade Name
- A) Varis Vision or Aria (Version 6.5)
B) Eclipse
- Recall Start Date
- 2010-05-13
- Recall Number
- 55200
- Hazard Classification
- Type II
- Model or Catalog #
- A) Version 6.5
B) Version 6.5
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
- Reason for Recall
- An anomaly was discovered in which a second or subsequent user can access a patient record which is already open at another workstation, and, if they ignore the warning message that the record is being edited by another user, and proceed to edit the plan and save the record, a corrupted but usable patient plan record may result. Errors can include, but are not limited to, incorrect MU, incorrect beam modifiers, incorrect gantry angles, incorrect field sizes, etc.
- Manufacturer
- Vascular Solutions, Inc.
- Recall Posting Date
- 2010-09-27
- Trade Name
- Langston Dual Lumen Catheter
- Recall Start Date
- 2010-08-31
- Recall Number
- 56416
- Hazard Classification
- Type II
- Model or Catalog #
- 5540
- Lot or Serial #
- 548110, 548227, 548364, 548430, 548554, 548723
- Reason for Recall
- Some lots of the Langston Dual-Lumen Pigtail Catheter, model 5540 are missing a sterile barrier pouch seal.
- Manufacturer
- Vital Diagnostics, Inc.
- Recall Posting Date
- 2010-08-30
- Trade Name
- Excyte Sliders
- Recall Start Date
- 2010-04-14
- Recall Number
- 54156
- Hazard Classification
- Type III
- Model or Catalog #
- EX-10552
- Lot or Serial #
- All lots
- Reason for Recall
- There are Excyte ESR tubes in the field that do not fit properly in the Excyte sliders. The product will not be replaced. This product is discontinued.
- Manufacturer
- Wilson-Cook Medical, Inc.
- Recall Posting Date
- 2010-09-27
- Trade Name
- A) Six-Shooter Saeed Multi-Band Ligator
B) Ten-Shooter Saeed Multi-Band Ligator
- Recall Start Date
- 2010-09-07
- Recall Number
- 56462
- Hazard Classification
- Type II
- Model or Catalog #
- A) MBL-6-OV, MBL-6-OV-XS
B) MBL-10
- Lot or Serial #
- A) W2783447, W2758500, W2758501, W2772743, W2753989, W2772759, W2777465
B)W2758472, W2758474, W2764330, W2772073
- Reason for Recall
- Possibility of a section of the ligator barrel separating from the barrel's friction fit adapter that attaches to patient end of the endoscope, which could in turn result in a section of the ligator barrel detaching inside the patient's gastrointestinal tract.
- Manufacturer
- Zimmer Inc.
- Recall Posting Date
- 2010-08-02
- Trade Name
- Versys Hip System Femoral Head 12/14
- Recall Start Date
- 2010-07-09
- Recall Number
- 55632
- Hazard Classification
- Type II
- Model or Catalog #
- 00801802801, 00801803201
- Lot or Serial #
- 61424803, 61275855
- Reason for Recall
- This action is being conducted because there is a potential for a package to be labelled as a 28mm femoral head but contain a 32mm femoral head. In addition, there is a potential for the patient record label to be incorrect.
- Manufacturer
- Zimmer Inc.
- Recall Posting Date
- 2010-08-02
- Trade Name
- ITST Fixation Locking Bolt Extractor
- Recall Start Date
- 2010-07-09
- Recall Number
- 55641
- Hazard Classification
- Type II
- Model or Catalog #
- 00-2258-051-01
- Lot or Serial #
- 53951300
- Reason for Recall
- This action is being conducted because Zimmer has determined the instruments were manufactured using a grade of material that is nonconforming to specifications. As manufactured, the instruments are more brittle, which increases the potential for fracture.
- Manufacturer
- Zimmer Inc.
- Recall Posting Date
- 2010-09-13
- Trade Name
- Trabecular Metal Acetabular Porous Shell
- Recall Start Date
- 2010-08-20
- Recall Number
- 56268
- Hazard Classification
- Type II
- Model or Catalog #
- 00-6202-056-22
- Lot or Serial #
- 61091653
- Reason for Recall
- This action is being conducted because the threads on the device (Trabecular Metal Modular Acetabular 56mm Porous Shell with Cluster Holes) were not completely machined and therefore will not mate with the corresponding cup positioner.
- Manufacturer
- Zimmer Inc.
- Recall Posting Date
- 2010-09-13
- Trade Name
- Bigliani/Flatow Fukuda Retractors
- Recall Start Date
- 2010-08-23
- Recall Number
- 56271
- Hazard Classification
- Type III
- Model or Catalog #
- 00-4305-019-30, 00-4305-019-40
- Lot or Serial #
- 60795817, 60834923, 60916795, 61017041, 61038697, 60768814, 60806551, 60893241, 60946820, 60999665
- Reason for Recall
- This action is being conducted due to fractures occurring at the grooves around the perimeter of the blades. The Bigliani/Flatow complete shoulder solution Fukuda Retractors (small and large sizes) are used to gain exposure to the glenoid during surgery.
- Manufacturer
- Zimmer Trabecular Metal Technology, Inc.
- Recall Posting Date
- 2010-08-02
- Trade Name
- Trabecular Metal Femoral Cone Augment
- Recall Start Date
- 2010-07-09
- Recall Number
- 55634
- Hazard Classification
- Type II
- Model or Catalog #
- 00-5450-012-31
- Lot or Serial #
- 60683028
- Reason for Recall
- This action is being conducted because there is a potential for the packages to be mislabeled. Specifically, the left femoral cone augment component may be labelled as a right and the right femoral cone augment component may be labelled as a left.