Health Canada
Symbol of the Government of Canada

Institutional links

Drugs and Health Products

Medical Device Recall Listings - Sorted by Manufacturer

(From July 2010 - September 2010)

Manufacturer
Abbott Laboratories Diagnostic Division
Recall Posting Date
2010-08-30
Trade Name
A) Cell-Dyn 4000 Vent Needle and Accessory Kit
B) Cell-Dyn Sapphire Vent Head Assembly and Accessory Kit
Recall Start Date
2010-08-06
Recall Number
56047
Hazard Classification
Type III
Model or Catalog #
A) 01H01-01
B) 08H00-01
Lot or Serial #
A) N/A
B) N/A
Reason for Recall
During manufacturing, the aspiration probe was hitting the top portion of the vent head assembly and would not go all the way down. Upon further evaluation, it was found that the vent needle was not manufactured correctly by the supplier.
Manufacturer
Abbott Laboratories Diagnostic Division
Recall Posting Date
2010-09-27
Trade Name
A) Cell-Dyn 3700 Haematology System-SL Analyser
B) Cell-Dyn 3200SL System
C) Cell-Dyn 3700 Haematology System-CS Analyser
Recall Start Date
2010-08-31
Recall Number
56461
Hazard Classification
Type III
Model or Catalog #
A) 02H31-01
B) 4H60-01
C) 02H30-01
Lot or Serial #
A) All serial numbers
B) All serial numbers
C) All serial numbers
Reason for Recall
Cell-Dyn 3700 customers had reported occurrences of visible fire and smoke from the analyzer. In addition, a Cell-Dyn 3200 customer reported smoke from the analyzer. Returned power supply units were analysed and were identified to contain the incorrect fuse for the electrical configuration at the customer site.
Manufacturer
Abbott Point of Care Inc.
Recall Posting Date
2010-08-30
Trade Name
i-Stat Level 2 Control Value Assignment Sheet (VAS)
Recall Start Date
2010-08-02
Recall Number
55982
Hazard Classification
Type III
Model or Catalog #
06F13-01
Lot or Serial #
B09342, B09154
Reason for Recall
Abbott has determined that the i-Stat level 2 Control Value Assignment Sheets contain the incorrect range for PCO2, blood urea nitrogen (BUN) and glucose for the i-Stat cartridges. Abbott has made the decision to update the affected value assignment sheets to ensure the performance of their products.
Manufacturer
Accuray, Inc.
Recall Posting Date
2010-08-02
Trade Name
CyberKnife Robotic Radiosurgery System - Base CyberKnife
Recall Start Date
2010-07-06
Recall Number
55783
Hazard Classification
Type II
Model or Catalog #
028000, 030000, 032000
Lot or Serial #
C0231, C0202
Reason for Recall
When the cosmetic linac cover (the 'nose' cover) is reattached to the linac after removal, the rotating latches that secure the cover may not be properly engaged with the connection pins on the linac. If the latches are not fully engaged, the linac cover can become loose and may unexpectedly drop off the linac. If a patient treatment is initiated while the cover is loose, there is potential for it to contact the patient and lead to injury depending on the position and height of the linac at that time.
Manufacturer
Acumed, LLC
Recall Posting Date
2010-07-05
Trade Name
Trial Gage
Recall Start Date
2010-05-11
Recall Number
54806
Hazard Classification
Type III
Model or Catalog #
TR-TG01
Lot or Serial #
211365, 214346, 220365, 222678, 224183
Reason for Recall
These parts are being recalled as they were manufactured from a different grade of aluminum than required. The part therefore has an increased risk of breaking which could occur during a surgery.
Manufacturer
Advanced Neuromodulation Systems, Inc.
Recall Posting Date
2010-07-19
Trade Name
A) Genesis IPG Neurostimulator - Lamitrode (Surgical) Leads
B) Lamitrode S-4 Lead Kits
C) Lamitrode S-8 Lead Kits
D) Lamitrode C Series Leads
E) Lamitrode Neurostimulation Lead Kit
F) Lamitrode S-Series Neurostimulation Lead Kit
G) Lamitrode Tripole 8 Lead Kit
H) Lamitrode Tripole 8C Lead Kit
I) Lamitrode Tripole 16C Lead Kit
J) Lamitrode Exclaim Lead Kit
K) Lamitrode Tripole 16 Lead Kit
Recall Start Date
2010-06-18
Recall Number
55435
Hazard Classification
Type III
Model or Catalog #
A) 3240, 3244, 3280
B) 3243, 3246
C) 3283, 3286
D) 3245
E) 3254, 3255, 3262, 3263, 3264, 3265
F) 3266, 3267, 3268, 3269
G) 3208
H) 3210
I) 3214
J) 3224
K) 3219
Lot or Serial #
A) N/A
B) N/A
C) N/A
D) N/A
E) N/A
F) N/A
G) N/A
H) N/A
I) N/A
J) N/A
K) N/A
Reason for Recall
St. Jude Medical (SJM) has identified the potential for the rare occurrence of neurological deficit adverse events such as autonomic symptoms, pain (manifesting itself as chest or abdominal pain) or paralysis associated with implanting the paddle leads listed above.
Manufacturer
AGFA Healthcare N.V.
Recall Posting Date
2010-08-16
Trade Name
A) NX8000 workstations used in combination with 19'' Barco monitor and mounting stand
B) NX8200 workstations used in combination with 19'' Barco monitor and mounting stand
Recall Start Date
2010-06-30
Recall Number
55730
Hazard Classification
Type II
Model or Catalog #
A) E5LHV, E5LJX
B) X.0.8200
Lot or Serial #
A) N/A
B) N/A
Reason for Recall
Attempt to reposition the monitor by pulling or pushing the monitor housing will cause strain on the monitor. Depending on the extent of stress and after an extended period of time the monitor may fail at or near the mounting bracket pins.
Manufacturer
American Medical Systems
Recall Posting Date
2010-09-27
Trade Name
GreenLight HPS fibers, P/N 0010-2090
Recall Start Date
2010-01-29
Recall Number
53499
Hazard Classification
Type III
Model or Catalog #
0010-2090A, 0010-2090B, 0010-2090C,
0010-2090D
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
The laser fibres had intermittent packaging and labelling issues. Three scenarios: HPS 0010-2090 as PV2079, HPS0010-2090 as PV 0010-2090, and product insert may be missing.
Manufacturer
Amo Manufacturing USA, LLC
Recall Posting Date
2010-09-13
Trade Name
Wavescan Wavefront System and Offline Programming Module (OPM)
Recall Start Date
2010-08-04
Recall Number
56048
Hazard Classification
Type III
Model or Catalog #
HS-1
Lot or Serial #
N/A
Reason for Recall
An incorrect patient treatment table may be generated by a coding error within the Wavescan Wavefront system software. The field correction is in regard to the software upgrade of Wavescan Wavefront system and offline programming module (OPM), which is used to determine patient treatment.
Manufacturer
Arrow International, Inc.
Recall Posting Date
2010-08-16
Trade Name
Central Venous Catheterization Kit
Recall Start Date
2010-07-23
Recall Number
55893
Hazard Classification
Type II
Model or Catalog #
CA-22702
Lot or Serial #
RF0033779
Reason for Recall
Arrow has received complaints regarding the introducer needle being "flimsy" and being difficult to insert into the blood vessel. It was determined that some finished products lots had incorrect cannula with a blunt bevel.
Manufacturer
B. Braun Medical Inc.
Recall Posting Date
2010-08-02
Trade Name
AddEase Binary Connector
Recall Start Date
2010-07-06
Recall Number
55733
Hazard Classification
Type I
Model or Catalog #
N7990, N7993
Lot or Serial #
60984562, 61019650, 61031167, 61057867
Reason for Recall
B.Braun pab bags demonstrate coring when used with AddEase connectors and third party needles. The incidence rate appears to be higher with the use of AddEase connectors.
Manufacturer
B. Braun Medical Inc.
Recall Posting Date
2010-08-16
Trade Name
Addease Binary Connector
Recall Start Date
2010-07-22
Recall Number
55733
Hazard Classification
Type I
Model or Catalog #
N7990, N7993
Lot or Serial #
60984562, 61019650, 61031167, 61057867
Reason for Recall
B.Braun pab bags demonstrate coring when used with Addease connectors and third party needles. The incidence rate appears to be higher with the use of Addease connectors.
Manufacturer
Baxter Healthcare Corporation
Recall Posting Date
2010-09-27
Trade Name
A) Micro-Volume Extension Sets with Filter
B) Micro-Volume Extension Set
Recall Start Date
2010-09-07
Recall Number
56507
Hazard Classification
Type III
Model or Catalog #
A) 2N3350
B) 2N3347
Lot or Serial #
A) UR07I25092, UR08B04109, UR08B09058, UR09E12013, UR09F04083, UR192179X, UR242057, UR253450, UR261362
B) UE103267
Reason for Recall
Baxter has identified a labelling issue affecting all lots of extension sets 2N3347 and 2N3350, whereby the product configuration and label illustration have the slide clamp above the filter, but the written instructions tell the user to close the clamp below filter when set is not in use. The current clamp configuration could result in the unintended fluid administration of approximately 0.2ml.
Manufacturer
Beckman Coulter, Inc.
Recall Posting Date
2010-07-05
Trade Name
A) P3, P4 Plates for the PK7200 Automated Microplate System-Microplate Reader
B) P3, P4 Plates for the PK7200 Automated Microplate System-Analyzer Class 3
C) P3, P4 Plates for the PK7300 Automated Microplate System-Instrument Class 3
D) P3, P4 Plates for the PK7300 Automated Microplated System-Instrument
Recall Start Date
2010-04-26
Recall Number
55300
Hazard Classification
Type II
Model or Catalog #
A) PK7200
B) PK7200
C) PK7300
D) PK7300
Lot or Serial #
A) P3-91A,92A,93A,95D,96D, P4-93B,94B,99C
B) P3-91A,92A,93A,95D,96D, P4-93B,94B,99C
C) P3-91A,92A,93A,95D,96D, P4-93B,94B,99C
D) P3-91A,92A,93A,95D,96D, P4-93B,94B,99C
Reason for Recall
Beckman Coulter has confirmed the presence of deformed well terraces in the PK Microplate wells may cause fold over or slip down of positive reaction patterns in the PK7200 and PK7300 Automated Microplate System.
Manufacturer
Beckman Coulter, Inc.
Recall Posting Date
2010-07-19
Trade Name
Access Thyroglobulin Antibody II Reagent
Recall Start Date
2010-05-17
Recall Number
55500
Hazard Classification
Type II
Model or Catalog #
A32898
Lot or Serial #
916271, 917073, 918433
Reason for Recall
Beckman Coulter has confirmed that the three lots of Access Thyroglobulin Antibody II Reagent may produce significantly different results in a small subset of patient samples when compared to previously released reagent lots.
Manufacturer
Beckman Coulter, Inc.
Recall Posting Date
2010-08-02
Trade Name
A) Unicel DXI 600 Access Immunoassay System - Instrument Class 2
B) Unicel DXI 600 Access Immunoassay System - Instrument Class 3
C) Unicel DXI 800 Access Immunoassay - Analyzer-Class 3
D) Unicel DXI 800 Access Immunoassay System - Analyzer - Class II
Recall Start Date
2010-07-05
Recall Number
55603
Hazard Classification
Type II
Model or Catalog #
A) A30260
B) A30260
C) 973100
D) 973100
Lot or Serial #
A) All lots
B) All lots
C) All lots
D) All lots
Reason for Recall
Beckman Coulter has confirmed the following issue with the software version 4.3 being used in the Unicel DXI 600 and DXI 800 Access Immunoassay Systems. If a laboratory changed a default sample type setting on the tests screen while running Unicel DXI System software version 4.3, the new setting would not be saved to the system database. The changed sample type setting would reset to the previously saved setting when power was interrupted to the system, or if an operator rebooted the system.
Manufacturer
Beckman Coulter, Inc.
Recall Posting Date
2010-08-30
Trade Name
Sex Hormone Binding Globulin (SHBG) Reagent
Recall Start Date
2010-07-26
Recall Number
55983
Hazard Classification
Type II
Model or Catalog #
A48617
Lot or Serial #
009046, 010202, 829586, 909837, 914557, 915628, 917613, 919184
Reason for Recall
The instruction for use (IFU) for access Sex Hormone Binding Globulin (SHBG) does not specify the units of measure to be used when calculating the free androgen index (FAI %). Access SHBG (nmol/l) values and access testosterone (ng/ml) values are reported in different default units. The testosterone value (when reported in ng/ml) requires conversion to nmol/l before the FAI % can be calculated..
Manufacturer
Beckman Coulter, Inc.
Recall Posting Date
2010-08-30
Trade Name
Synchron System(s) Creatine Reagent
Recall Start Date
2010-08-03
Recall Number
55984
Hazard Classification
Type II
Model or Catalog #
472525
Lot or Serial #
M911428, M911519
Reason for Recall
Beckman Coulter Inc. has confirmed reports that black/brown particulates were observed in some bottles of the creatinine alkaline buffer in modular Creatinine Reagent Kit lots M911428 and M911519.
Manufacturer
Beckman Coulter, Inc.
Recall Posting Date
2010-08-30
Trade Name
Synchron Systems Glucose Reagent (GLU)
Recall Start Date
2010-08-09
Recall Number
56165
Hazard Classification
Type II
Model or Catalog #
442640
Lot or Serial #
All non-expired lots
Reason for Recall
Synchron Systems Glucose Reagent (GLU) when used in conjunction with Unicel DXC and Synchron LX instruments may give falsely high results for moderately hemolysed samples.
Manufacturer
Beckman Coulter, Inc.
Recall Posting Date
2010-09-13
Trade Name
Navios Flow Cytometer
Recall Start Date
2010-06-21
Recall Number
55631
Hazard Classification
Type III
Model or Catalog #
A52101, A52102, A52103
Lot or Serial #
AN29103-AS15099, AN06024-AS03008, AM51001-AS15102
Reason for Recall
Beckman Coulter has confirmed an issue with the Navios Flow Cytometer with version 1.0 software: red and violet laser related errors are not displayed in the event of a laser failure. The blue laser is not impacted by this issue.
Manufacturer
Beckman Coulter, Inc.
Recall Posting Date
2010-09-13
Trade Name
Synchron Systems Enzymatic CO2 (CO2E) Reagent
Recall Start Date
2010-07-12
Recall Number
55665
Hazard Classification
Type III
Model or Catalog #
A60290
Lot or Serial #
M907574
Reason for Recall
Blank rate high (SH) instrument flag may be generated when less than 30 tests left in the CO2E cartridge and recalibration attempts may fail or may not be maintained.
Manufacturer
Beckman Coulter, Inc.
Recall Posting Date
2010-09-13
Trade Name
A) Access System - Immunoassay Aanalyzers Class 2
B) Unicel DXC 600I Synchron Access Clinical Analyzer
C) Unicel DXI 600 Access Immunoassay System - Instrument Class 2
D) Unicel DXI 600 Access Immunoassay System - Instrument Class 3
E) Unicel DXC 880I Synchron Access Clinical Analyzer
F) Access Immunoassay Analyzers - Class 3
G) Access Immunoassay System - Analyzer Class 3
H) Access System Access 2 Analyzer - Class 2
I) Unicel DXI 800 Access Immunoassay System - Analyzer - Class III
J) Unicel DXI 800 Access Immunoassay System - Analyzer - Class II
K) Synchron LXI 725
L) Unicel DXC 660I Synchron Access Clinical Analyzer
M) Unicel DXC 680I Synchron Access Clinical Analyzer
N) Unicel DXC 860I Synchron Access Clinical Analyzer
Recall Start Date
2010-08-09
Recall Number
56259
Hazard Classification
Type II
Model or Catalog #
A) 81600
B) 4767
C) A30260
D) A30260
E) 4768
F) 81600
G) 81600N
H) 81600N
I) 973100
J) 973100
K) 476501
L) 4771
M) 4772
N) 4773
Lot or Serial #
A) All lots
B) All lots
C) All lots
D) All lots
E) All lots
F) All lots
G) All lots
H) All lots
I) All lots
J) All lots
K) All lots
L) All lots
M) All lots
N) All lots
Reason for Recall
If a reagent pack is not loaded onto Access and DXI instruments in accordance with the instrument operator's guide, it is possible for the instrument to pipette insufficient or incorrect reagents and generate erroneous quality control or patient results.
Manufacturer
Beckman Coulter, Inc.
Recall Posting Date
2010-09-13
Trade Name
SYNCHRON System(s) Immunoglobulin A (Ig-A) Reagent
Recall Start Date
2010-06-28
Recall Number
55480
Hazard Classification
Type II
Model or Catalog #
467920
Lot or Serial #
M812164, M902345
Reason for Recall
Ig-A Reagent lots M812164 and M902345 contain an unidentified cross reacting antibody that may yield falsely high Ig-A results in survey samples for immunoglobulins which are formulated with Ig-M monoclonal antibodies.
Manufacturer
Beckman Coulter, Inc.
Recall Posting Date
2010-09-27
Trade Name
A) AU5400 Clinical Chemistry Analyzer
B) AU2700 Clinical Chemistry Analyzer
C) AU400/AU400E Clinical Chemistry Analyzers
D) AU600/AU640/AU640E Clinical Chemistry Analyzers
E) AU680 Clinical Chemistry Analyzer
F) AU480 Clinical Chemistry Analyzer
Recall Start Date
2010-07-12
Recall Number
56411
Hazard Classification
Type I
Model or Catalog #
A) AU5400
B) AU2700
C) AU400, AU400E
D) AU600, AU640, AU640E
E) AU680
F) AU480
Lot or Serial #
A) All serial numbers
B) All serial numbers
C) All serial numbers
D) All serial numbers
E) All serial numbers
F) All serial numbers
Reason for Recall
Company has received reports of incorrect results generated by analyzers as a result of cuvette overflow. Current system software unable to complete real-time monitoring of reaction profiles to identify impact of cuvette overflow on results, yielding incorrect results without any flags or alarms to alert operator.
Manufacturer
Becton Dickinson and Company
Recall Posting Date
2010-09-13
Trade Name
BD Gaspak EZ Large Incubation Container
Recall Start Date
2010-08-17
Recall Number
56228
Hazard Classification
Type III
Model or Catalog #
260672
Lot or Serial #
7110573
Reason for Recall
Product may not maintain air tight environment due to a manufacturing defect. The defect is a missing clear plastic post on the inside of the container lid that is found in the middle position.
Manufacturer
Becton Dickinson and Company
Recall Posting Date
2010-09-27
Trade Name
BBL Staphyloslide Latex Test Kit
Recall Start Date
2010-07-19
Recall Number
56555
Hazard Classification
Type III
Model or Catalog #
240952, 240953
Lot or Serial #
16605410, 16640910, 16679310, 16778710, 16781210, 16781510,
16821684
Reason for Recall
Kits may exhibit false agglutination with non-staphylococcus aureus colonies as well as the negative control reagent. Quality control should be performed each day the kit is used. The kit should not be used if the reactions with the control organisms are incorrect.
Manufacturer
Becton Dickinson Diagnostic Instrument Systems
Recall Posting Date
2010-09-13
Trade Name
Bactec FX
Recall Start Date
2010-08-12
Recall Number
56174
Hazard Classification
Type III
Model or Catalog #
441385
Lot or Serial #
FT001 to FT0906, FT0922
Reason for Recall
Patient sample database on the Bactec FX are sometimes found to be unusable upon rebooting the instrument. The reboot proceeds normally until it tries to load the database file, at which point the instrument reboots and then repeats this cycle indefinitely or customers see a message on the display that says "please insert a bootable device". The system will continue to agitate and incubate the samples.
Manufacturer
Becton Dickinson Infusion Therapy Systems Inc.
Recall Posting Date
2010-08-30
Trade Name
A) BD Q-Syte Extension Set
B) MPS Acadia Extension Set With Rightbore
C) MPS Acadia Specialty Tri-Extension Set
D) Dual Port BD Nexiva Closed IV Catheter
Updated recall #51661 previously posted on 2009-12-07.
Recall Start Date
2009-11-02
Recall Number
51661
Hazard Classification
Type I
Model or Catalog #
A) 385102
B) 385150, 385151
C) 385164
D) 383530, 383531
Lot or Serial #
A) 8308330
B) A1967, A2399
C) A1951
D) 8238450, 8345951, 9013867
Reason for Recall
Potential risk of end-users inadvertently taking a BD Q-Syte device from BD Nexiva product or package, and placing it on a central venous catheter.
Manufacturer
Biomerieux SA
Recall Posting Date
2010-08-16
Trade Name
Vidas System Rub IGG II Assay
Recall Start Date
2010-07-21
Recall Number
55865
Hazard Classification
Type II
Model or Catalog #
30221
Lot or Serial #
837810801
Reason for Recall
One or the two pouch of lot #837810801 may contain out-of-specification cones that were affected by a manufacturing incident.
Manufacturer
Biomerieux, Inc.
Recall Posting Date
2010-08-30
Trade Name
A) Vitek 2 Compact 30 System - Test Cards
B) Vitek 2 Compact 60 System - Test Cards
C) Vitek 2 System - Test Cards
Recall Start Date
2010-08-02
Recall Number
55961
Hazard Classification
Type II
Model or Catalog #
A) 22225
B) 22225
C) 22225
Lot or Serial #
A) 345158910
B) 345158910
C) 345158910
Reason for Recall
There are small slits on the backsides of Vitek 2 AST-N085 lot 345158910 foil pouches near the seam separating the desiccant from the main body of the pouch.
Manufacturer
Biomerieux, Inc. Portland
Recall Posting Date
2010-08-16
Trade Name
O-F Dextrose Tubed Media
Recall Start Date
2010-06-08
Recall Number
55201
Hazard Classification
Type III
Model or Catalog #
T7141
Lot or Serial #
313295-1
Reason for Recall
O-F Dextrose labelling information on the tubes incorrectly described the product as O-F Staph.
Manufacturer
Biomet Microfixation, Inc.
Recall Posting Date
2010-09-13
Trade Name
Rees Insulated Forceps
Recall Start Date
2010-06-30
Recall Number
55527
Hazard Classification
Type II
Model or Catalog #
02-0461
Lot or Serial #
042308D08, 112108D08, 031309C09, 041009D09, 042310B10, 021910B10, 092707I07, 103607J07, 032607C07, 041607D07
Reason for Recall
Cracking or peeling off of the insulation of the forceps, causing burns to the patients.
Manufacturer
Biomet Sports Medicine Inc.
Recall Posting Date
2010-07-05
Trade Name
Artificial Ligament Fixation Device
Recall Start Date
2010-03-31
Recall Number
53910
Hazard Classification
Type II
Model or Catalog #
904755
Lot or Serial #
050400, 089710, 648470, 192250, 228610
Reason for Recall
The ziploop extended toggleloc device may contain an out of specification saddle length (too long), which could result in insufficient graft being introduced into the surgical tunnel.
Manufacturer
Bio-Rad
Recall Posting Date
2010-07-05
Trade Name
Strep B Select
Recall Start Date
2010-06-08
Recall Number
55233
Hazard Classification
Type II
Model or Catalog #
63750
Lot or Serial #
0D0039
Reason for Recall
The lot 0D0039 gives an abnormal rate of blue colonies that were not further confirmed as GBS (group B streptococcus). On this particular lot, some enterococci could produce blue colonies instead of the pink to violet colonies that are expected for these organisms.
Manufacturer
Bio-Rad Laboratories Deeside
Recall Posting Date
2010-08-30
Trade Name
IN2IT - A1C Test Cartridges
Recall Start Date
2010-08-10
Recall Number
56054
Hazard Classification
Type II
Model or Catalog #
2810001EX
Lot or Serial #
All lots
Reason for Recall
Bio-Rad notifies in2it users that when performing in2it testing at altitudes above 1 000 meters (3 280 feet) they may obtain high results. This problem can occur on all lots of A1C test cartridges.
Manufacturer
Boston Scientific Corporation
Recall Posting Date
2010-07-19
Trade Name
A) Synchro 2 Soft Guidewire
B) Synchro 2 Standard Guidewire
C) Synchro PVS 1300 Neuro Guidewire
Recall Start Date
2010-06-07
Recall Number
55223
Hazard Classification
Type II
Model or Catalog #
A) M00326010, M00326310
B) M00326410, M00326510
C) M00313010, M00313020, M00313310
Lot or Serial #
A) N/A
B) N/A
C) N/A
Reason for Recall
Product may exhibit a PTFE coating issue that could impact patient safety, more specifically, Boston Scientific Corporation has identified scrapes on the coating during production of the affected lots, which could lead to flaking.
Manufacturer
Boston Scientific Corporation
Recall Posting Date
2010-08-30
Trade Name
A) Sterling Monorail PTA Balloon
B) Sterling Over-The-Wire PTA Balloon
Recall Start Date
2010-05-11
Recall Number
54721
Hazard Classification
Type III
Model or Catalog #
A) H74939135152010, H74939135202010, H74939135203010, H74939135204010, H74939135252010, H74939135253010, H74939135254010, H74939135302010, H74939135303010, H74939135304010, H74939135352010, H74939135354010, H74939135402010, H74939135403010, H74939135404010
B) H74939134152010, H74939134202010, H74939134203010, H74939134204010, H74939134252010, H74939134253010, H74939134254010, H74939134302010, H74939134303010, H74939134304010, H74939134352010, H74939134353010, H74939134354010, H74939134402010, H74939134403010, H74939134404010
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
Reason for Recall
Directions for use for the affected lots do not include certain language previously requested.
Manufacturer
Boston Scientific Corporation
Recall Posting Date
2010-09-13
Trade Name
Ultraflex Precision Colonic Stent
Recall Start Date
2010-07-16
Recall Number
55796
Hazard Classification
Type II
Model or Catalog #
M00557360
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
Boston Scientific has identified an issue that may potentially result in the failure to deploy the stent due to the deployment suture breaking.
Manufacturer
Brainlab AG
Recall Posting Date
2010-09-13
Trade Name
A) Kolibri Navigation Station 2.0 with Mobile Camera Stand 2.0
B) Kolibri Mobile Camera Stand 2.0
Recall Start Date
2010-08-18
Recall Number
56243
Hazard Classification
Type II
Model or Catalog #
A) 19506C
B) 19801A
Lot or Serial #
A) 000000005042210001, 000000005042310001, 000000003934009001, 000000002185609001, 000000004809910001, 000000003509109001, 000000004367509001
B) N/A
Reason for Recall
For the Spectra camera interfaces assembled within the Kolibri Mobile Camera Stand, a long-term reliable fixation of the camera unit cannot be guaranteed due to a manufacturing error.
Manufacturer
Brentwood Medical Technology, Corp. DBA Midmark Diagnostics Group
Recall Posting Date
2010-07-19
Trade Name
IQmanager Diagnostic Workstation
Recall Start Date
2010-05-26
Recall Number
55107
Hazard Classification
Type III
Model or Catalog #
4-100-1200
Lot or Serial #
Versions 8.3, 8.3.1, 8.3.2
Reason for Recall
When the user access the QT dialogue box to manually accept or edit the QT interval, the software recalculates the QTC value using Hodges equation for all patients rather than for only patients 16 years and older.
Manufacturer
Cardiac Pacemakers Incorporated
Recall Posting Date
2010-09-13
Trade Name
A) Contak Renewal 3 RF CRT-D
B) Contak Renewal 3 RF HE CRT-D
C) Contak Renewal 4 CRT-D
D) Contak Renewal 4 HE CRT-D
Recall Start Date
2010-07-23
Recall Number
55829
Hazard Classification
Type III
Model or Catalog #
A) H215
B) H217
C) H190
D) H-197, H-199
Lot or Serial #
A) N/A
B) N/A
C) N/A
D) N/A
Reason for Recall
Application of a magnet (typically in a clinic hospital environment) may cause a magnetic reed switch located within the device to become permanently stuck in a closed position and prevent delivery of programmed tachy therapy.
Manufacturer
Cardinal Health
Recall Posting Date
2010-08-30
Trade Name
A) Sterile Craniotomy Pack
B) Sterile Ear Pack
Recall Start Date
2010-08-04
Recall Number
56051
Hazard Classification
Type II
Model or Catalog #
A) SNE30CRXXX
B) SEN30EAXXX, SEN30MEXXX
Lot or Serial #
A) 887069, 905131, 909364, 919774, 926904, 933512, 953095, 885710, 890855, 894055, 900358, 918868, 933456, 940554, 945788
B) 887885, 908312, 912899, 916104, 916587, 890561, 906447, 935731, 953954
Reason for Recall
The possibility of reduced adhesion between the patties and the blue radio-opaque x-ray stripe. This may result in the x-ray stripe becoming detached from the pattie during use and remaining in surgical site.
Manufacturer
Cardinal Health
Recall Posting Date
2010-09-27
Trade Name
Medi-Vac Non Conductive Suction Tubing
Recall Start Date
2010-08-18
Recall Number
56305
Hazard Classification
Type II
Model or Catalog #
PN510, PN512, PN56A, PN610, PN612, PN66A, PN710, PN712, PN76A
Lot or Serial #
All Lot Numbers Prior to Y10KXXXX (10K is August 2010)
Reason for Recall
The recall has been initiated due to the possibility that packaging seals on the referenced lot numbers may be compromised.
Manufacturer
CaridianBCT, Inc.
Recall Posting Date
2010-08-02
Trade Name
A) COBE Spectra White Blood Cell Sets
B) COBE Spectra Therapeutic Plasma Exchange Sets
Recall Start Date
2010-07-16
Recall Number
55784
Hazard Classification
Type II
Model or Catalog #
A) 70620
B) 70600
Lot or Serial #
A) 01S15239
B) 01S15238
Reason for Recall
Potential leaking at the return luer connection on two manufacturing lots of COBE Spectra White Blood Cell Disposable Tubing Sets.
Manufacturer
CaridianBCT, Inc.
Recall Posting Date
2010-08-30
Trade Name
COBE Spectra White Blood Cell Sets
Updated recall # 55784 previously posted 2010-08-02. / Mise … jour du retrait # 55784 afflich‚ le 2010-08-02.
Recall Start Date
2010-07-16
Recall Number
55784
Hazard Classification
Type II
Model or Catalog #
70620
Lot or Serial #
01S15239
Reason for Recall
Potential leaking at the return luer connection on two manufacturing lots of COBE Spectra White Blood Cell Disposable Tubing Sets.Note that COBE Spectra Therapeutic Plasma Exchange Sets has been removed from this recall.
Manufacturer
CaridianBCT, Inc.
Recall Posting Date
2010-09-27
Trade Name
Cobe Spectra Apheresis System wbc
Recall Start Date
2010-07-16
Recall Number
55784
Hazard Classification
Type II
Model or Catalog #
70620
Lot or Serial #
01S15239, 01S15238
Reason for Recall
Leak at the return luer connection on 2 MFR lots of Cobe Spectra white blood cell disposable tubing sets. Leak is observed as a very small fluid leak originating between the male Luer on the return line and the attached customer needle or as an observation of air in the line when blood warmer tubing is attached to the male Luer.
Manufacturer
Ciba Vision Corporation
Recall Posting Date
2010-09-13
Trade Name
Softperm
Recall Start Date
2010-07-03
Recall Number
55841
Hazard Classification
Type II
Model or Catalog #
Softperm
Lot or Serial #
>100 Numbers, Contact Manufacturer
Reason for Recall
Out-of-specification in a product stability study resulting from a change in the vial stopper.
Manufacturer
Codman & Shurtleff Inc.
Recall Posting Date
2010-08-16
Trade Name
Codman Surgical Patties, X-Ray Detectable
Recall Start Date
2010-07-22
Recall Number
55895
Hazard Classification
Type II
Model or Catalog #
80-1396, 80-1399
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
Codman Surgical Patties, which are labelled with a radio-opaque marker, and were distributed by Codman between May 2009 and May 2010, may have reduced adhesion between the patty and the blue radio-opaque x-ray stripe due to a recently discovered processing problem. This reduced adhesion may result in the x-ray stripe becoming detached from the patty during use and remaining in the surgical site inadvertently.
Manufacturer
Cook Urological Inc.
Recall Posting Date
2010-08-02
Trade Name
Modified Novy Cornual Cannualtion Set
Recall Start Date
2010-01-09
Recall Number
52817
Hazard Classification
Type II
Model or Catalog #
J-NCS-503570
Lot or Serial #
U1799467
Reason for Recall
The curved tip of the introducing catheter has the potential to separate during use. If the affected product has already been used successfully, there is not additional patient risk.
Manufacturer
Cook, Inc.
Recall Posting Date
2010-09-13
Trade Name
A) Ciaglia Blue Rhino Percutaneous
B) Ciaglia Blue Dolphin Percutaneous
Recall Start Date
2010-04-23
Recall Number
54482
Hazard Classification
Type I
Model or Catalog #
A) C-PTIS-100-HC-PERC6, C-PTIS-100-HC-PERC6-CA, C-PTIS-100-HC-PERC8, C-PTIS-100-HC-PERC8-CA
B) C-PTBS-2600-PERC6, C-PTBS-2800-PERC8, C-PTBSY-2600-PERC6, C-PTBSY-2800-PERC8
Lot or Serial #
A) G12403, G12566, G13165, G36037, G13166, G12523, G36038, G12565, G53178, G53169, G53170
B) G36116, G36118, G51971, G36117, G49507, G51972
Reason for Recall
Certain cuffed Shiley tracheostomy tubes, in which the cuff does not hold air as a result of leaks in the pilot balloon inflation assembly.
Manufacturer
Coopersurgical Inc.
Recall Posting Date
2010-08-16
Trade Name
A) INCA Infant Nasal Cannulae Assembly for CPAP
B) INCA Infant Nasal Cannulae Assembly for CPAP Replacement Set
Recall Start Date
2010-07-27
Recall Number
55932
Hazard Classification
Type III
Model or Catalog #
A) 44-2707, 44-2709, 44-2710, 44-2712, 44-2715
B) 44-0707, 44-0709, 44-0710, 44-0712, 44-0715
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
Reason for Recall
As a result of Coopersurgical's voluntary recall for the Infant Nasal Cannulae (CPAP)(INCA) assembly the classification of the device was reviewed by Health Canada. Subsequently, Coopersurgical is recalling the INCA since it was sold in Canada without a license.
Manufacturer
Depuy International Ltd.
Recall Posting Date
2010-09-27
Trade Name
A) Depuy ASR Cup System - Heads
B) ASR XL Acetabular Cup System - Heads
C) ASR XL Acetabular Cup - Sleeve Adaptors
D) Depuy ASR Acetabular Cup System - Shells
E) ASR 300 Spiked Acetabular Shells
Recall Start Date
2010-08-25
Recall Number
56326
Hazard Classification
Type II
Model or Catalog #
A) 9998-03-239, 9998-03-441, 9998-03-643, 9998-03-845, 9998-03-946, 9998-04-047, 9998-04-249, 9998-04-451, 9998-04-653, 9998-04-855, 9998-05-057, 9998-05-359, 9998-05-561, 9998-05-763
B) 9998-90-139, 9998-90-141, 9998-90-143, 9998-90-145, 9998-90-146, 9998-90-147, 9998-90-149, 9998-90-151, 9998-90-153, 9998-90-155, 9998-90-157, 9998-90-159, 9998-90-161, 9998-90-163
C) 9998-00-102, 9998-00-105, 9998-00-108, 9998-00-200, 9998-00-203, 9998-00-206, 9998-00-209, 9998-00-300, 9998-00-303, 9998-00-313
D) 9998-03-944, 9998-04-146, 9998-04-348, 9998-04-550, 9998-04-652, 9998-04-754, 9998-04-956, 9998-05-158, 9998-05-360, 9998-05-562, 9998-05-764, 9998-05-966, 9998-06-168, 9998-06-370
E) 999830744, 999830746, 999830748, 999830750, 999830752, 999830754, 999830756, 999830758, 999830760, 999830762, 999830764, 999830766, 999830768, 999830770
Lot or Serial #
A) >500 Numbers, Contact Manufacturer
B) >100 Numbers, Contact Manufacturer
C) >500 Numbers, Contact Manufacturer
D) >500 Numbers, Contact Manufacturer
E) >100 Numbers, Contact Manufacturer
Reason for Recall
Depuy issued a field safety notice in March 2010 (recall number 53553) after receiving data that demonstrated the ASR system had a higher than expected revision rate at three years when used with smaller head sizes. New data now shows a higher than expected revision rate at five years, across the entire size range. The manufacturer is therefore issuing a voluntary recall of all ASR products.
Manufacturer
Depuy Mitek
Recall Posting Date
2010-07-05
Trade Name
A) VAPR S90 Suction Electrode (Integrated)
B) VAPR S50 Knee Electrode (Integrated)
C) VAPR Premiere90 Electrode (Integrated)
D) VAPR Premiere50 Electrode (Integrated)
Recall Start Date
2010-06-09
Recall Number
55222
Hazard Classification
Type II
Model or Catalog #
A) 225370
B) 227355
C) 227204
D) 227504
Lot or Serial #
A) >100 Numbers, Contact Manufacturer
B) 805107, 904050, 904051, 904056, 904061
C) >10 Numbers, Contact Manufacturer
D) 906032
Reason for Recall
Some of the products within the recall list did not meet the required packaging specifications for pouch seal width dimensions, which could potentially lead to a breach in sterility. The recall applies to all non-expired inventory with the exception of inventory that is designated with round green stickers.
Manufacturer
Depuy Orthopaedics, Inc.
Recall Posting Date
2010-08-02
Trade Name
Summit Broach Handle
Recall Start Date
2010-07-14
Recall Number
55756
Hazard Classification
Type III
Model or Catalog #
2750-00-000
Lot or Serial #
A1109, A1209, A0210, A0310
Reason for Recall
Some of Depuy's straight broach handles are experiencing early failure due to a weakened handle locking mechanism (leaf spring).
Manufacturer
Diasorin Inc.
Recall Posting Date
2010-08-30
Trade Name
Liason Analyzer
Recall Start Date
2010-08-10
Recall Number
56139
Hazard Classification
Type II
Model or Catalog #
15970
Lot or Serial #
All Serial Numbers
Reason for Recall
During system initialization when pipettors are washing, if sample or reagent area flap are opened, the shaker rod may stop moving. This will affect the suspension of magnetic particles resulting in low RLU values for calibrators, controls & patient samples giving unexpected positive or negative results or incorrect dose.
Manufacturer
Draeger Medical AG & Co. KG
Recall Posting Date
2010-08-02
Trade Name
NBP Cuffs NEO Sizes 1 to 5
Recall Start Date
2010-06-29
Recall Number
55447
Hazard Classification
Type I
Model or Catalog #
MP00901, MP00902, MP00903, MP00904, MP00905
Lot or Serial #
Devices Delivered Between 12-01-2010 and 12-06-2010.
Reason for Recall
Cases where measurements with the affected neonatal blood pressure cuffs deviate from the actual values. This deviation can be up to 50%, meaning the measured and displayed value can be 50% higher than the actual one.
Manufacturer
Draeger Medical AG & Co. KG
Recall Posting Date
2010-09-13
Trade Name
Oxylog 3000
Recall Start Date
2010-08-16
Recall Number
56204
Hazard Classification
Type I
Model or Catalog #
2M86955
Lot or Serial #
SRZB-0059, SRZA-0128, SRYK-0134, SRYK-0133, SRYK-0132, SRYK-0131, SRYK-0069, SRYK-0068, SRYB-0074, SRYA-0195, SRYA-0014
Reason for Recall
Patients have been inadequately ventilated because the dead space of the breathing circuit was not completely considered by the user when setting the ventilation parameters. The information for use (IFU) for the Oxylog 3000 devices delivered from September 2008 contains additional warnings and cautions. A safety notice with the IFU amendment will be sent out to users who own affected Oxylog 3000. No patient injury has been reported in relation to that dead space issue.
Manufacturer
Draeger Medical AG & Co. KG
Recall Posting Date
2010-09-13
Trade Name
Fabius GS Anesthesia System
Recall Start Date
2010-08-25
Recall Number
56304
Hazard Classification
Type II
Model or Catalog #
8604700
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
The Fabius GS or Fabius Tiro delivered between June 2005 and June 2006 could have been equipped with a power cord of the manufacturer Electri-Cord Manufacturing Corporation (ECM). The US Food and Drug Administration (FDA) has associated a risk of sparks, fire, device failures or electrical shocks with these power cords. Although Draeger Medical has not received any reports of failures of power cords manufactured by ECM, Draeger Medical will replace the possible affected power cords.
Manufacturer
Draeger Medical Systems, Inc.
Recall Posting Date
2010-07-19
Trade Name
A) Gamma XXL
B) Delta Monitor
Recall Start Date
2010-05-24
Recall Number
55089
Hazard Classification
Type II
Model or Catalog #
A) MS18852
B) MS18597
Lot or Serial #
A) 5399463952, 5399468849, 5399469251, 5399494054, 5399498745, 5399532156, 5399532753, 5399533850, 5399662354, 5399664949, 6000446771, 6000796170
B) >10 Numbers, Contact Manufacturer
Reason for Recall
Investigation determined that the monitor's fixed keys may become inoperative or activate without user interaction due to a contamination present in the fixed key panel material. This is a concern because the worst case scenario of this malfunction could be that the monitor may discharge to a patient automatically.
Manufacturer
Elekta Business Area Software Systems ATA Computerized Medical Systems Inc.
Recall Posting Date
2010-07-05
Trade Name
Focal Software
Recall Start Date
2010-06-10
Recall Number
55234
Hazard Classification
Type II
Model or Catalog #
N/A
Lot or Serial #
Release 4.40.00 & Up
Reason for Recall
The couch position can be moved (X or Y direction) between scans of the same CT study. When images are imported, the software is not checking the Dicom image position (patient) tag & is not aligning the image sets. A shift between image sets will exist of the same amount that the CT couch was moved.
Manufacturer
Excelsior Medical Corporation
Recall Posting Date
2010-09-27
Trade Name
Prefilled Flush Syringe
Recall Start Date
2010-09-02
Recall Number
56438
Hazard Classification
Type II
Model or Catalog #
S-5
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
During a routine test for the syringe integrity, a possible integrity failure was discovered.
Manufacturer
Focus Medical, LLC
Recall Posting Date
2010-09-27
Trade Name
Naturalase 1064 Laser
Recall Start Date
2010-09-01
Recall Number
55902
Hazard Classification
Type II
Model or Catalog #
LT100
Lot or Serial #
N/A
Reason for Recall
The recall is being conducted because the licence to market and sell these systems in Canada had expired at the time the units were sold
Manufacturer
Fujirebio Diagnostics, Inc.
Recall Posting Date
2010-08-16
Trade Name
Mesomark Kits
Recall Start Date
2010-07-08
Recall Number
55955
Hazard Classification
Type II
Model or Catalog #
801-900
Lot or Serial #
57K04510, 57K04409 (Already Expired)
Reason for Recall
Certain lots of Mesomark Kits may generate inaccurate absorbance values due to the presence of a residue on the underside of some microtiter wells that can be detected visually.
Manufacturer
GE Healthcare
Recall Posting Date
2010-07-05
Trade Name
A) Prospeed Scanner System
B) SYTEC6000
Recall Start Date
2009-03-26
Recall Number
47109
Hazard Classification
Type II
Model or Catalog #
A) B7970YZ
B) P9183AH
Lot or Serial #
A) 298920YMO & 455302YMO
B) 121615YM9
Reason for Recall
The x-ray leakage from the diagnostic source assembly (tube+collimator) of the Zj/Qj families of computed tomography systems exceeded 21 CFR regulatory requirement and IEC standard.
Manufacturer
GE Healthcare IT
Recall Posting Date
2010-07-19
Trade Name
Centricity Ultra Laboratory Information System
Recall Start Date
2010-06-15
Recall Number
55072
Hazard Classification
Type II
Model or Catalog #
CORE
Lot or Serial #
All Software Versions
Reason for Recall
Under specific conditions, the laboratory system may report incorrect results such as truncating a value after a comma is encountered. For example, a result of 1,234 would be reported as 1.0.
Manufacturer
GE Healthcare Japan Corporation
Recall Posting Date
2010-08-16
Trade Name
HiSpeed DX/I LX/I CT System
Recall Start Date
2010-07-26
Recall Number
55695
Hazard Classification
Type III
Model or Catalog #
B7510ZZ
Lot or Serial #
00000532977YM6, 00000715131YM9
Reason for Recall
Computed tomography dose index (CTDI) values on dose display & in operator manual are incorrect for scan settings over 3mm slice thickness. The provided CTDI value may be more than 20% higher than actual CTDI value.
Manufacturer
GE Medical Systems Information Technologies, Inc.
Recall Posting Date
2010-09-27
Trade Name
A) Transport Pro V2 Patient Monitor
B) Transport Pro Patient Monitor (PDM Compatible)
Recall Start Date
2010-08-30
Recall Number
56100
Hazard Classification
Type II
Model or Catalog #
A) TRANS-AC-HXXX
B) TRANS-BC-HXXX
Lot or Serial #
A) AAD05360553GA, AAD06031134GA, AAD06131474GA, AAD06352093GA, AAD06352095GA, AAD05350491GA
B) >10 Numbers, Contact Manufacturer
Reason for Recall
Transport Pro stops communication with Carescape patient data modules after 414 days of continuous run time, resulting in the loss of the displayed parameter info. (Eg: waveforms, numerical data, alarms).
Manufacturer
GE Medical Systems Kretztechnik GMBH & CO OHG
Recall Posting Date
2010-07-05
Trade Name
A) Voluson E8 Ultrasound System Main Unit
B) Voluson E8 Expert Ultrasound System
Updated reall # 51227 previously posted on 2009/11/09
Recall Start Date
2009-10-21
Recall Number
51227
Hazard Classification
Type II
Model or Catalog #
A) H48651PT
B) H48661N
Lot or Serial #
A) S/N D00229
B) S/N D04842, D05589 & D50233
Reason for Recall
When using combination of 2D-compound resolution imaging (CRI) and a pulsed wave Doppler single gate (PW) with steered Doppler angle, the displayed velocity scale is incorrect, which could delay treatment.
Manufacturer
GE Medical Systems SCS
Recall Posting Date
2010-07-05
Trade Name
Innova 2000 Cardiovascular System
Recall Start Date
2010-06-09
Recall Number
54810
Hazard Classification
Type II
Model or Catalog #
S1872LC
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
Source to image distance (SID) failure could occur, which may result in loss of x-ray generation. An error message may or may not be displayed, depending on certain conditions.
Manufacturer
GE Medical Systems Ultrasound & Primary Care Diagnostics LLC
Recall Posting Date
2010-09-27
Trade Name
Logiq E9 Ultrasound System-Sector Transducer
Recall Start Date
2010-09-02
Recall Number
56175
Hazard Classification
Type III
Model or Catalog #
H4908SC
Lot or Serial #
94900496
Reason for Recall
Rating plate of S1-5 probe identifies name and model number of S4-10 probe. Correct S1-5 name appears on transducer handle and packaging.
Manufacturer
Gyrus Acmi, Inc.
Recall Posting Date
2010-08-30
Trade Name
Pleatman Sac Tissue Removal System
Recall Start Date
2010-08-04
Recall Number
56025
Hazard Classification
Type II
Model or Catalog #
004942-903
Lot or Serial #
186540FC
Reason for Recall
Trays used to package the Pleatman Sac product had cracks and/or dimples on them. Further investigation revealed that the dimples could form cracks when the tray is manipulated. This could lead to a breach in the sterile barrier which could compromise sterility of the product.
Manufacturer
Gyrus ACMI, Inc.
Recall Posting Date
2010-08-16
Trade Name
Dissector PlasmaKnife (DPK)
Recall Start Date
2010-07-15
Recall Number
55925
Hazard Classification
Type II
Model or Catalog #
7035-3005
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
A non conformance report (NCR) was initiated at Gyrus ACMI facility because a qualification run of fluid electrodes failed International Safe Transit Association (ISTA) testing due to a failure of the seal on the sterile packaging. A breach in the sterile barrier would compromise the sterility of the product.
Manufacturer
Haemonetics Corporation
Recall Posting Date
2010-07-05
Trade Name
Plasma Bowl, 0625B-00
Recall Start Date
2010-06-10
Recall Number
55256
Hazard Classification
Type II
Model or Catalog #
0625B-00
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
There is a defect in a very small number of plasma bowls in which a portion of the wall of the bowl may be too thin. There is a potential that a defective bowl may leak or fail within the centrifuge chamber during a plasma collection procedure. If a leak or failure were to occur, the PCS2 system is designed to detect the problem and safely terminate the collection procedure. However, it is possible that blood could escape from the device and spray into the surrounding environment and onto surrounding objects or people.
Manufacturer
Heine Optotechnik GMBH & Co. KG
Recall Posting Date
2010-08-16
Trade Name
XP Laryngoscope Blabe Mac 3
Recall Start Date
2010-07-08
Recall Number
55909
Hazard Classification
Type III
Model or Catalog #
F-000.22.763
Lot or Serial #
62117
Reason for Recall
There is a tolerance issue on a certain production batch from April 2010. The result is that the heel of the blade may be too wide to be inserted into the head of laryngoscope handles manufactured to the "green standard".
Manufacturer
Hill-Rom, Inc.
Recall Posting Date
2010-07-05
Trade Name
Hill-Rom 100 Low Bed
Recall Start Date
2010-05-03
Recall Number
54548
Hazard Classification
Type II
Model or Catalog #
P3930
Lot or Serial #
N/A
Reason for Recall
Potential for siderails used on the Hill-Rom 100 Low Bed to operate incorrectly (false latching, unintentional lowering) or become inoperate.
Manufacturer
Hill-Rom, Inc.
Recall Posting Date
2010-07-05
Trade Name
A) Hill-Rom 1000 Beds
B) Care Assist P1170
Recall Start Date
2009-07-13
Recall Number
54567
Hazard Classification
Type II
Model or Catalog #
A) P1600
B) P1170
Lot or Serial #
A) N/A
B) N/A
Reason for Recall
There is a potential for a shearing action between the siderail arm and the bracket used to secure the siderail to the bed.
Manufacturer
Hill-Rom, Inc.
Recall Posting Date
2010-07-19
Trade Name
A) Hill-Rom 1000 Beds
B) Careassist Hospital Bed
Recall Start Date
2010-06-10
Recall Number
55218
Hazard Classification
Type II
Model or Catalog #
A) P1160
B) P1170
Lot or Serial #
A) N/A
B) N/A
Reason for Recall
Failure of the plastic housing used to support the trapeze handle from the patient helper.
Manufacturer
Hologic, Inc.
Recall Posting Date
2010-09-13
Trade Name
A) Atec Breast Biopsy & Excision System Sapphire
B) Atec Breast Biopsy & Excision System Emerald
C) Atec Breast Biopsy & Excision System Pearl
Recall Start Date
2010-08-09
Recall Number
56247
Hazard Classification
Type II
Model or Catalog #
A) N/A
B) N/A
C) N/A
Lot or Serial #
A) 003016
B) 003016
C) 003016
Reason for Recall
The aperture of the needle may be a 20mm size, versus the 12mm size indicated on the label.
Manufacturer
Hospira Inc.
Recall Posting Date
2010-08-30
Trade Name
Gemstar Pump Set (110 Yellow)(AMB.INFU)
Recall Start Date
2010-08-10
Recall Number
56130
Hazard Classification
Type II
Model or Catalog #
13261, 13273
Lot or Serial #
740545H, 680295H, 770878H
Reason for Recall
Under delivery reports occurring during clinical use under low rate settings (less than 10ml/hr).
Manufacturer
Howmedica Osteonics Corporation
Recall Posting Date
2010-08-30
Trade Name
Osteonics Series 7000 Total Knee Standard Tibial Tray
Recall Start Date
2010-08-04
Recall Number
56012
Hazard Classification
Type II
Model or Catalog #
7115-0007
Lot or Serial #
MJK68N
Reason for Recall
Stryker Howmedica Osteonics has become aware that there was a cross mix between the Osteonics Series 7000 Total Knee Standard Tibial Tray and the Triathlon Total Knee System Primary Cemented Tibial Base-Plate. The packaging was labelled as a Series 7000 Total Knee Standard Tibial Tray, while the package actually contained a Triathlon Total Knee System Primary Cemented Tibial Base-Plate.
Manufacturer
Immucor Inc.
Recall Posting Date
2010-07-19
Trade Name
Anti-FY Blood Grouping Reagent
Recall Start Date
2010-06-16
Recall Number
55381
Hazard Classification
Type III
Model or Catalog #
0007594
Lot or Serial #
613007, 613007-1
Reason for Recall
Lots 613007 and 613007-1 demonstrate weaker reactivity than previous lots of Anti-FY.
Manufacturer
Immucor Inc.
Recall Posting Date
2010-08-16
Trade Name
Panoscreen I, II and III
Recall Start Date
2010-07-16
Recall Number
55873
Hazard Classification
Type I
Model or Catalog #
0002377
Lot or Serial #
24223
Reason for Recall
Panoscreen I, II and III lot 24223 does not include an S positive donor cell.
Manufacturer
Immucor Inc.
Recall Posting Date
2010-08-16
Trade Name
Galileo Blood Bank Analyzer
Recall Start Date
2010-07-28
Recall Number
55949
Hazard Classification
Type II
Model or Catalog #
0064999
Lot or Serial #
N/A
Reason for Recall
If an unrecognized rack is removed, and a sample is than removed or switched in the rack, followed by reinserting the rack into the lane, it is possible for the sample ID that originally occupied the position to be assigned to the newly loaded sample or the empty position. Under these conditions, it is possible that results will be associated with the wrong sample.
Manufacturer
Immucor Inc.
Recall Posting Date
2010-09-13
Trade Name
Galileo Echo Blood Bank Analyzer
Recall Start Date
2010-08-12
Recall Number
56238
Hazard Classification
Type II
Model or Catalog #
0087000
Lot or Serial #
N/A
Reason for Recall
When a sample rack is partially removed after testing is complete and another rack is inserted into an adjacent position, it is possible for the partially removed rack to be inadvertently loaded ahead of the second rack. In this instance, the second rack's sample ID's will be incorrectly assigned to the original rack.
Manufacturer
Immucor Inc.
Recall Posting Date
2010-09-27
Trade Name
Panocell-20
Recall Start Date
2010-08-30
Recall Number
56415
Hazard Classification
Type II
Model or Catalog #
0005020
Lot or Serial #
29304
Reason for Recall
Cell 5 (donor B3907) of lot 29304 is demonstrating weak positive reactions in direct antiglobulin testing (DAT). Tests performed with this cell may produce unexpected positive reactions at the anti-human globulin (AHG) phase of testing.
Manufacturer
Immunotech A.S.
Recall Posting Date
2010-08-02
Trade Name
Estrone Radioimmunoassay (RIA) Kit
Recall Start Date
2010-07-05
Recall Number
55630
Hazard Classification
Type II
Model or Catalog #
DSL 8700
Lot or Serial #
992637, 100226A, 100319A, 100409A
Reason for Recall
Beckman Coulter has determined that the Estrone RIA Kit lots identified were inaccurately calibrated to their internal reference calibrator.
Manufacturer
Immuontech A.S.
Recall Posting Date
2010-08-30
Trade Name
IGFBP-3 Immunoradiometric Assay (IRMA)
Recall Start Date
2010-08-09
Recall Number
56099
Hazard Classification
Type II
Model or Catalog #
DSL-6600
Lot or Serial #
100115A, 100319A, 100409A, 100430A, 991225, 991694, 991938, 992171, 992372
Reason for Recall
Beckman Coulter has confirmed that IGFBP-3 measurements using IGFBP-3 IRMA Kit (REF DSL-6600) lots identified may result in IGFBP-3 levels that are not in agreement with simultaneously measured levels of IGF-1.
Manufacturer
INO Therapeutics, LLC DBA IKARIA
Recall Posting Date
2010-08-16
Trade Name
INOMAX DS Drug Delivery-System
Recall Start Date
2010-07-23
Recall Number
55907
Hazard Classification
Type II
Model or Catalog #
DS 20050006
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
A component within the pressure switch which monitors for loss of drug supply may tear, causing Inomax to leak from the rear of the system. As a result, the Inomax cylinder empties more quickly than normal. An interruption or delay in the administration of Inomax therapy may cause worsening of systemic oxygenation indices (i.e. hypoxemia, especially manifested as decreased arterial oxygenation saturation), hypotension, and/or increase in pulmonary arterial pressure.
Manufacturer
Institut Straumann AG
Recall Posting Date
2010-07-19
Trade Name
Impression Set for WN Solid Abutment
Recall Start Date
2010-06-15
Recall Number
55311
Hazard Classification
Type III
Model or Catalog #
040.38
Lot or Serial #
Y4021
Reason for Recall
The product name on the label incorrectly identifies the product as "Impression Set for RN (regular neck) Solid Abutment" when it should read "Impression set for WN (wide neck) Solid Abutment".
Manufacturer
Instrumentation Laboratory Co.
Recall Posting Date
2010-09-27
Trade Name
A) ACL Top 500 CTS
B) ACL Top, ACL Top 700 LAS, ACL Top CTS
Recall Start Date
2010-06-01
Recall Number
56433
Hazard Classification
Type II
Model or Catalog #
A) 0000280040
B) 0000280000, 0000280020, 0000280030
Lot or Serial #
A) 08010100 to 10061231
B) 03110101 to 1006111, 08120100 to 10020126, 06020101 to 10030659
Reason for Recall
Software V4.3.0 (with test parameters P-14.6) introduces an additional data check designed to minimize performance issues from factors such as patient condition and improper sample or reagent preparation by the user. Specifically, a new data CJECL, "multiple threshold check" was added to identify and fail abnormal (bimodal) clot curves for hemosil synthasil (APTT reagent, K060688) and hemosil recombiplastin 2g (PT reagent, K070005). The flag will recognize any APTT/PT curve that has a signal that increases in absorbance to a certain threshold level more than once.
Manufacturer
Integra Neurosciences Implants S.A.
Recall Posting Date
2010-08-02
Trade Name
Neuro Balloon Catheter
Recall Start Date
2010-07-07
Recall Number
55602
Hazard Classification
Type II
Model or Catalog #
7CBD10
Lot or Serial #
0151870, 0158739, 0159938
Reason for Recall
Certain Integra Neuro Balloon Catheters have the potential to improperly inflate or deflate under certain conditions. It was determined that this condition could exist during pre-implant testing or during the procedure with some lots of Integra Neuro Balloon Catheter products.
Manufacturer
International Technidyne Corp.
Recall Posting Date
2010-07-05
Trade Name
Hemochron Response Instrument, IDMS
Recall Start Date
2010-06-14
Recall Number
55307
Hazard Classification
Type III
Model or Catalog #
HRS.110
Lot or Serial #
72900321, 72900376, 72900367
Reason for Recall
There is a tabulation error in the integrated data management system (IDMS) software, product code 464100, versions 7.1 through 7.3, used with the Hemochron Response Instrument, that affects the operator and the device performance summary reports for two electronic quality control (EQC) tests.
Manufacturer
International Technidyne Corp.
Recall Posting Date
2010-08-16
Trade Name
Directcheck Whole Blood Control
Recall Start Date
2010-07-19
Recall Number
55897
Hazard Classification
Type III
Model or Catalog #
DCJPT-A
Lot or Serial #
G9DPA005
Reason for Recall
The acceptable performance range published in the Directcheck microcoagulation quality control (product code DCJPT-A, lot number G9DPA005 only) package insert is inaccurate. The posted range printed on the package insert reads "3.7 - 6.5 INR". The correct range is 4.1 - 8.7 INR.
Manufacturer
Inviro Medical, Inc.
Recall Posting Date
2010-07-19
Trade Name
Invirosnap Safety Syringe
Recall Start Date
2010-05-14
Recall Number
55079
Hazard Classification
Type III
Model or Catalog #
100002, 100018, 100019, 100017, 100038, 100047, 100034
Lot or Serial #
2006-06, 2006-09, 061911, 2006-10
Reason for Recall
When the plunger is retracted after an injection, the plunger with needle can be pulled out of the back of the syringe.
Manufacturer
Inviro Medical, Inc.
Recall Posting Date
2010-07-19
Trade Name
Invirosnap Safety Syringe
Recall Start Date
2010-05-14
Recall Number
55085
Hazard Classification
Type III
Model or Catalog #
100019, 100038, 100032, 100034
Lot or Serial #
063804, 061911, 2006-08
Reason for Recall
The needle length is out the specifications.
Manufacturer
Johnson & Johnson Vision Care, Inc.
Recall Posting Date
2010-08-30
Trade Name
Acuvue Oasys Brand Contact Lenses for Astigmatism
Recall Start Date
2010-08-02
Recall Number
56160
Hazard Classification
Type III
Model or Catalog #
N/A
Lot or Serial #
B006X641
Reason for Recall
Some product may have been mislabeled on the secondary carton (yellow carton) and there is a discrepancy with the primary package (blister pack).
Manufacturer
Levo AG
Recall Posting Date
2010-08-02
Trade Name
Levo C3 Wheelchair
Recall Start Date
2010-06-25
Recall Number
55674
Hazard Classification
Type II
Model or Catalog #
N/A
Lot or Serial #
913021
Reason for Recall
A quality problem was discovered with the stand-up wheelchair model Levo C3. When using the stand-up mechanism after a short time a screw may break at the support or the stand-up actuator unit propping up the seat to a 45 degree angle position.
Manufacturer
Lifescan Inc.
Recall Posting Date
2010-09-27
Trade Name
A) OneTouch DataLink Data Management System
B) OneTouch DataLink Management System
Recall Start Date
2010-10-23
Recall Number
56430
Hazard Classification
Type I
Model or Catalog #
A) 021-323, 021-969
B) 021-309, 021-970
Lot or Serial #
A) All lots
B) All lots
Reason for Recall
Possibility of data loss at healthcare facilities when downloading Surestep Flexx Meters to the OneTouch Datalink Data Management System under very specific circumstances (including having more than one meterlink instance installed & a meter that has not downloaded within 25 hrs).
Manufacturer
Liko AB
Recall Posting Date
2010-07-19
Trade Name
Viking M Patient Lift
Recall Start Date
2010-05-25
Recall Number
54859
Hazard Classification
Type II
Model or Catalog #
2040015
Lot or Serial #
7500401 to 7503472
Reason for Recall
Failure of the Viking M Lift as it relates to the control box (CBL 20)and the battery (BAL 20). The product will stop functioning. Note that there is a mechanical means to safely lower the patient during transfer.
Manufacturer
Liko AB
Recall Posting Date
2010-08-02
Trade Name
Scale Adapters
Recall Start Date
2010-07-12
Recall Number
55601
Hazard Classification
Type II
Model or Catalog #
200, 350, 400
Lot or Serial #
N/A
Reason for Recall
If you twist the sling bar when using the scale and scale adapter at a right angle in the lifting direction, the adapter can get locked in this position when the lift has a flexlink. If patient is lifted with the equipment while adapter is in this position, adapter may be damaged or break, allowing a patient to fall.
Manufacturer
Linvatec Corporation D.B.A. Conmed Linvatec
Recall Posting Date
2010-08-16
Trade Name
87K Arthroscopy Tubing Set
Recall Start Date
2010-07-06
Recall Number
55704
Hazard Classification
Type II
Model or Catalog #
87100
Lot or Serial #
Lot # date codes beginning with 0904151 to 1005201
Reason for Recall
A possibility the product may have a breach in the packaging seal that potentially could compromise the product sterility.
Manufacturer
Luxottica North America Distribution LLC
Recall Posting Date
2010-08-02
Trade Name
Sport Goggles - Liberty Sport
Recall Start Date
2010-07-02
Recall Number
55657
Hazard Classification
Type II
Model or Catalog #
MAXMORPHEUS, REC SPEC MAXX
Lot or Serial #
N/A
Reason for Recall
Some Sport Goggles may have prescription lens which do not meet the lens thickness specifications established by the frame manufacturer. Potential for lens breakage or disengagement from frame in event of impact.
Manufacturer
Maquet Cardiopulmonary AG
Recall Posting Date
2010-09-27
Trade Name
Straight Connector
Recall Start Date
2010-08-30
Recall Number
56358
Hazard Classification
Type II
Model or Catalog #
HG0286
Lot or Serial #
70045443, 70050983
Reason for Recall
A leak has been found at the shank during priming before perfusion, a manufacturer issue was detected at the supplier level.
Manufacturer
Medical Device Technologies, Inc. DBA Angiotech
Recall Posting Date
2010-08-02
Trade Name
Tru-Core II
Recall Start Date
2010-07-02
Recall Number
55562
Hazard Classification
Type II
Model or Catalog #
MD763114200X, MD763118100X, MD763118200X
Lot or Serial #
92661WY0, 93491YG7, 00621ZQ1, 011720OQ, 93431YBI, 92661WY7
Reason for Recall
Pouch holes at the chevron edge on the polyethylene side of the pouch or along the chevron side of the seal can compromise the sterile barrier of the packaging.
Manufacturer
Medline Industries Inc
Recall Posting Date
2010-09-27
Trade Name
A) Nygh Dental Pack
B) Nygh Minor and Major Pack
C) C-Section Pack
D) Nygh T & A Pack
E) Nygh Septoplasty Pack
F) Nygh Laparoscopy Pack
G) Grace Total Hip Pack
H) Nygh Vascular Pack
I) Cardiac Pack
Recall Start Date
2010-08-23
Recall Number
56350
Hazard Classification
Type II
Model or Catalog #
A) DYNJ23545
B) DYNJ23312, DYNJ23313
C) OBT242X
D) DYNJ28887
E) DYNJ23314
F) DYNJ23528
G) DYNJ23234
H) DYNJ23547
I) DYNJ31903X
Lot or Serial #
A) 07AD1212, 07AD4192, 07GD1673
B) 07AD1541, 07ED3906, 07AD2649, 07BD3551
C) 10FD0997, 10FD1415, 10GD0786, 10GD2489, 09CD2228, 09CD02261, 10DD0174, 09ED0001, 09ED0887, 09ED3576, 09GD2287, 09HD3244, 09JD3472, 08DD0257, 08ED1747, 08GD4104, 08HD1333, 08KD1091, 09KD3436
D) 07AD2162, 07BD0755, 07CD1423, 07DD1506, 07DD4242
E) 07AD2160, 07BD2175, 07BD2537, 07CD1090, 07DD3113
F) 07AD2290, 07BD0186, 07FD2924, 07GD3752
G) 10BD2818, 10AD0648, 09LD1851, 09HD1616, 09GD0515, 09FD3755, 09FD3241, 09FD0698, 08FD1984, 08FD1109, 08ED3801, 08DD1563, 08DD0368, 08CD2385, 08AD0583, 07LD0379, 07KD0907, 07KD0906, 07JD1514, 07ID1730, 07FD2226
H) 07AD4449, 07AD4545, 07BD1859, 07FD0509, 07GD3003
I) 09ED0655, 09ED3002, 09FD0041, 09FD3115, 09JD1513, 09KD1837, 09LD0272, 09LD2807, 10AD0647, 10AD3494
Reason for Recall
These custom procedure packs contain the Cardinal Medi-Vac Non-Conductive Suction Tubing that has been recalled by Cardinal Health. Medline Industries was alerted by Cardinal regarding a product recall for their Sterile Medi-Vac Non-Conductive Suction Tubing due to the possibility that packaging seals may be compromised.
Manufacturer
Medtronic Inc.
Recall Posting Date
2010-07-19
Trade Name
Carelink - Monitor
Recall Start Date
2010-06-10
Recall Number
55207
Hazard Classification
Type II
Model or Catalog #
2490C
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
During an update, a subset of Carelink monitors of a previous design received firmware intended for monitors of a newer design. This mismatch makes affected monitors non-functional & unable to transmit data.
Manufacturer
Medtronic Inc.
Recall Posting Date
2010-08-02
Trade Name
Enrhythm Dual Chamber Rate Responsive Pacemaker
Recall Start Date
2010-02-11
Recall Number
53125
Hazard Classification
Type II
Model or Catalog #
P1501DR
Lot or Serial #
>1000 Numbers, Contact Manufacturer
Reason for Recall
1) Battery voltage at device interrogation lower than what is tracked to provide data for elective replacement indicator notification. 2) Battery voltage could decrease sooner than expected.
Manufacturer
Medtronic Inc.
Recall Posting Date
2010-08-02
Trade Name
HMS System - Instrument
Recall Start Date
2010-03-08
Recall Number
53511
Hazard Classification
Type III
Model or Catalog #
305-14
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
Control test for heparin assay (HPT) cartridges may run longer than normal & in some cases time out after setting instrument parameters to "IU".
Manufacturer
Medtronic Inc.
Recall Posting Date
2010-08-02
Trade Name
Sutureless Unipolar Myocardial Lead
Recall Start Date
2010-02-25
Recall Number
53385
Hazard Classification
Type II
Model or Catalog #
5071
Lot or Serial #
>100 Numbers, Contact Manufacturer
Reason for Recall
Possible inadequate sterile packaging seals that could potentially affect product sterility.
Manufacturer
Medtronic Navigation Israel Ltd.
Recall Posting Date
2010-08-16
Trade Name
Polestar N-20 - Magnetic Resonance Diagnostic Device
Recall Start Date
2010-07-16
Recall Number
55746
Hazard Classification
Type II
Model or Catalog #
Flexicoil - 9733063
Lot or Serial #
7110519203, 8051519694, 9210040532
Reason for Recall
Electronics box on some flexicoils can overheat and may cause a skin burn when in direct contact with the patient.
Manufacturer
Medtronic Navigation, Inc.
Recall Posting Date
2010-07-19
Trade Name
Stealthstation S7 - Stealth 3D
Recall Start Date
2010-06-21
Recall Number
55374
Hazard Classification
Type II
Model or Catalog #
9734046
Lot or Serial #
S/N's 4500883854, 4500961922, 4500816943, 4500816944, 4501132005, 4500707425
Reason for Recall
Numerous electrical issues related to the power cord retractors.
Manufacturer
Medtronic Navigation, Inc.
Recall Posting Date
2010-08-02
Trade Name
Orthopedic Dovetail Clamp
Recall Start Date
2010-02-10
Recall Number
53129
Hazard Classification
Type II
Model or Catalog #
9730289
Lot or Serial #
200304080750, 200308061080, 2003040807, 2003080610, 200308061080 & 5081710533
Reason for Recall
Current IFU does not indicate that the clamp is designed for exclusive use with Zimmer brand acetabular cup positioner/impactor. Otherwise the clamp can slip axially, resulting in movement of the reference frame.
Manufacturer
Medtronic Sofamor Danek USA, Inc.
Recall Posting Date
2010-07-19
Trade Name
Nim-Eclipse System - Module, Surface Preamp
Recall Start Date
2010-07-02
Recall Number
55529
Hazard Classification
Type II
Model or Catalog #
945SPM608
Lot or Serial #
0923052D
Reason for Recall
Under worst case conditions, such as sweating under warming blankets and long term contact periods, the patient's skin can become discoloured at the electrode contact site.
Manufacturer
Metrex Research Corporation
Recall Posting Date
2010-09-27
Trade Name
Curing Light Sleeves
Recall Start Date
2010-08-09
Recall Number
56429
Hazard Classification
Type III
Model or Catalog #
4500, 4513
Lot or Serial #
100622, 100733
Reason for Recall
Part no. 4500 labelled as 7mm/8mm curing light sleeves actually contained 13mm sleeves. Part no. 4513 labelled as 13mm curing light sleeves actually contained 7mm/8mm sleeves.
Manufacturer
Minntech Corporation
Recall Posting Date
2010-08-02
Trade Name
Endoscope Disinfector
Recall Start Date
2010-07-08
Recall Number
55606
Hazard Classification
Type II
Model or Catalog #
DSD-110-HU0109
Lot or Serial #
N/A
Reason for Recall
Misprinted DSD hookup application guide that recommended the use of the DSD-110-HU0109 hookup rather than DSD-110-HU0144 hookup. May result in inadequate high-level disinfection of either the balloon filling channel or the elevator wire channel depending on the method of connection.
Manufacturer
Mizuho America, Inc.
Recall Posting Date
2010-07-05
Trade Name
Mizuha Doppler Probe
Recall Start Date
2010-03-22
Recall Number
53996
Hazard Classification
Type III
Model or Catalog #
07-150-10
Lot or Serial #
9781-34600, 9782-34598, 9785-34634, 9780-34586
Reason for Recall
The affected lots contain probes that may produce no audio signal. The lack of audio signal will be apparent to the user once the probe is connected to the Doppler transceiver and there is not a reasonable probability that the use of, or exposure to, the product will cause serious adverse health consequences.
Manufacturer
Ohmeda Medical, A Division of Datex Ohmeda Inc.
Recall Posting Date
2010-07-05
Trade Name
Giraffe In-Bed Scales
Recall Start Date
2010-06-08
Recall Number
55120
Hazard Classification
Type II
Model or Catalog #
N/A
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
During the weighing procedure, the scale is programmed to give only 6 seconds to place the baby on the mattress, if it takes more than 6 seconds to place the baby on the mattress, the scale will display an inaccurate weight.
Manufacturer
Oridion Medical 1987 Ltd.
Recall Posting Date
2010-08-16
Trade Name
A) Filterline Set
B) Smart Capnoline Plus ADU/INT 25/100 UN
C) Smart Capnoline Plus O2 ADU/INT 25UN
D) Smart Capnoline H Plus O2 ADU/INT 25UN
E) Niv-Line Adult 25UN
Recall Start Date
2010-05-26
Recall Number
55493
Hazard Classification
Type II
Model or Catalog #
A) XS04620
B) 009818, 010209
C) 009822
D) 010433
E) 008174
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
C) >10 Numbers, Contact Manufacturer
D) >10 Numbers, Contact Manufacturer
E) >10 Numbers, Contact Manufacturer
Reason for Recall
As a result of a production issue, Microstream Filterlines may not be recognized by the monitor when connected to the Microstream Filterline input connector and co2 monitoring will not commence unless the filterline is replaced. No co2 readings will be displayed and no alarms will appear until co2 monitoring starts.
Manufacturer
Ortho-Clinical Diagnostics Inc.
Recall Posting Date
2010-08-30
Trade Name
Vitros Chemistry Products MALB Reagent
Recall Start Date
2010-08-10
Recall Number
56101
Hazard Classification
Type II
Model or Catalog #
6801740
Lot or Serial #
Generation 2, 3 & 4
Reason for Recall
Ortho Clinical Diagnostics has received several customer complaints of lower than expected albumin results for samples that were expected to generate results significantly above the measuring (reportable) range of 0.6 - 19.0 mg/dl. These samples may be negatively biassed and give values within the reportable range without an initial dilution.
Manufacturer
Ortho-Clinical Diagnostics Inc.
Recall Posting Date
2010-09-27
Trade Name
A) Vitros 5600 System Software 1.4 & Below
B) Vitros 3600 System Software 1.4 & Below
Recall Start Date
2010-07-28
Recall Number
55916
Hazard Classification
Type II
Model or Catalog #
A) 680 2863, 680 2864
B) 680 2865, 680 2866
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) 36000264, 36000130, 36000116
Reason for Recall
Ortho-Clinical Diagnostics has identified two scenarios on the Vitros 5600 Integrated System (software version 1.4 and below) and Vitros 3600 Immunodiagnostic System (software version 1.4 and below) in which Microwell and/or Microtip reagent inventory may not be properly updated, resulting in misidentified reagent packs and incorrect results or 'no result' being reported by the system.
Manufacturer
Ortho-Clinical Diagnostics Inc.
Recall Posting Date
2010-09-27
Trade Name
Ortho HTLV-I/II AB-Capture Elisa Test
Recall Start Date
2010-09-09
Recall Number
56514
Hazard Classification
Type III
Model or Catalog #
934200
Lot or Serial #
HTV208
Reason for Recall
Ortho-Clinical Diagnostics has received customer complaints regarding intermittent invalid plates due to elevated optical densities (OD) for the negative calibrator when using Ortho HTLV-I/II AB-Capture Elisa Test System, lot HTV208. A negative calibrator OD that is outside the package insert specification would cause an invalid plate. If this occurs the entire plate is aborted and no results are reported.
Manufacturer
Pelton and Crane
Recall Posting Date
2010-08-02
Trade Name
Dental Chair Magnetic Headrest Assembly
Recall Start Date
2010-03-15
Recall Number
54641
Hazard Classification
Type II
Model or Catalog #
N/A
Lot or Serial #
Option - 40R501-XX
Reason for Recall
Magnetic headrest may temporarily affect the function or programming of some implantable pacemakers or defibrillators if the device is programmed to respond to a magnet. People with such devices should avoid using magnetic headrests.
Manufacturer
Philips Medical Systems
Recall Posting Date
2010-07-05
Trade Name
Heartstart FR2+ Semi-Automatic Defibrillator - Main Unit
Recall Start Date
2010-04-23
Recall Number
55119
Hazard Classification
Type II
Model or Catalog #
M3860A, M3861A
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
Certain Heartstart FR2+ Automated External Defibrillators (AEDS) may contain a component called a voltage detector that may have a higher than expected rate of failure. Failure of the voltage detector can cause the AED battery to drain more rapidly than normal or to render the AED unusable.
Manufacturer
Philips Medical Systems
Recall Posting Date
2010-08-02
Trade Name
A) Philips Heartstart HS1 Onsite Defibrillator
B) Philips Heartstart HS1 Home Defibrillator
Recall Start Date
2010-04-23
Recall Number
55525
Hazard Classification
Type II
Model or Catalog #
A) M5066A
B) M5068A
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
Reason for Recall
There is a potential for users to misread the quick reference guide (QRG) and render the device ineffective.
Manufacturer
Philips Medical Systems
Recall Posting Date
2010-08-30
Trade Name
Heartstart MRX Defibrillator / Monitor
Recall Start Date
2010-07-12
Recall Number
56084
Hazard Classification
Type II
Model or Catalog #
M3535A, M3536A
Lot or Serial #
>100 Numbers, Contact Manufacturer
Reason for Recall
Philips has become aware that users may not fully understand the advanced features related to primary ECG lead configuration settings and the automatic lead selection behaviour of the Heartstart MRX.
Manufacturer
Philips Medical Systems (Cleveland), Inc.
Recall Posting Date
2010-07-05
Trade Name
1.5T Edge Eclipse System - Main Unit
Recall Start Date
2010-04-12
Recall Number
55221
Hazard Classification
Type II
Model or Catalog #
OFF-ML-1
Lot or Serial #
N/A
Reason for Recall
There is a potential for misdiagnosis - slice grids associated with images acquired after couch movement will be shown in an incorrect position when viewed on images acquired before the couch was moved.
Manufacturer
Philips Medical Systems (Cleveland), Inc.
Recall Posting Date
2010-09-27
Trade Name
Mx8000 IDT Multislice CT Imaging System
Recall Start Date
2010-07-14
Recall Number
56344
Hazard Classification
Type II
Model or Catalog #
OFF-EI-590
Lot or Serial #
3056
Reason for Recall
The Mx8000 IDT Multislice CT Imaging System may become inoperable if the user does not adjust the site specific protocols manually after the software upgrade.
Manufacturer
Philips Medizin Systeme Boeblingen GMBH
Recall Posting Date
2010-08-02
Trade Name
A) Intellivue MP70 Patient Monitor System - Main Unit
B) Intellivue MP60 Patient Monitor System - Main Unit
Recall Start Date
2010-06-24
Recall Number
55526
Hazard Classification
Type II
Model or Catalog #
A) M8007A
B) M8005A
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
Reason for Recall
Philips has recently received reports of monitors becoming loose and detaching from the variable height mount, with the potential of causing injury.
Manufacturer
Philips Medzin Systeme Boeblingen GMBH
Recall Posting Date
2010-08-30
Trade Name
Anesthetic Gas Monitor Water Trap
Recall Start Date
2010-07-15
Recall Number
56013
Hazard Classification
Type II
Model or Catalog #
M1657B
Lot or Serial #
>100 Numbers, Contact Manufacturer
Reason for Recall
Philips has received reports that when changing the anaesthetic gas monitor water trap (M1657B), the manifold seals on the Philips gas analysers, models M1013A, M1019A and M1026B, may be unintentionally removed. A missing seal results in leakage and may cause incorrect monitor readings of inspiratory or expiratory gas levels.
Manufacturer
Prollenium Medical Technologies Inc.
Recall Posting Date
2010-08-30
Trade Name
Dermal Roller SR
Recall Start Date
2010-08-09
Recall Number
56052
Hazard Classification
Type III
Model or Catalog #
DR10, DR5
Lot or Serial #
10-1209, 05-1209
Reason for Recall
The Dermal SR Roller (transdermal delivery microneedle therapy) is being recalled because of lack of assurance of sterility.
Manufacturer
Rehabilitation Centre for Children
Recall Posting Date
2010-08-02
Trade Name
Tilt Standing Frame
Recall Start Date
2010-06-23
Recall Number
55699
Hazard Classification
Type I
Model or Catalog #
N/A
Lot or Serial #
Contact Manufacturer
Reason for Recall
If an individual on the frame is moved to an upright position without the safety straps, knee blocks and tray in place, the individual may fall forward, tipping the tilt standing frame, posing a risk of serious injury.
Manufacturer
Remel Inc.
Recall Posting Date
2010-07-19
Trade Name
TB Potassium Permanganate Counterstain
Recall Start Date
2010-06-15
Recall Number
55303
Hazard Classification
Type II
Model or Catalog #
RE40092, RE40192
Lot or Serial #
152898, 152899
Reason for Recall
Product deterioration, may affect the ability of the counterstain to quench background fluorescent during performance of staining procedure.
Manufacturer
Roche Diagnostics GMBH
Recall Posting Date
2010-08-16
Trade Name
IgG-2 (Tina Quant IgG Gen.2)
Recall Start Date
2010-07-15
Recall Number
55734
Hazard Classification
Type III
Model or Catalog #
03507432190
Lot or Serial #
63009201
Reason for Recall
Quality control release of the Cobas C501 IgG-2 (Tina Quant IgG Gen. 2) cassette showed a too low recovery of the precinorm PUC control target value while using the cerebrospinal fluid (CSF) application.
Manufacturer
Roche Diagnostics GMBH
Recall Posting Date
2010-08-16
Trade Name
Anti-CCP for Elecsys and Cobas E
Recall Start Date
2010-07-21
Recall Number
55874
Hazard Classification
Type III
Model or Catalog #
05031656190
Lot or Serial #
157557
Reason for Recall
The manufacturer identified a potential decreased recovery of the precicontrol Anti-CCP in combination with the Cobas E-Elecsys Anti-CCP Reagent Kit.
Manufacturer
Roche Diagnostics GMBH
Recall Posting Date
2010-08-30
Trade Name
Elecsys Vitamin D3 assay
Recall Start Date
2010-08-03
Recall Number
55964
Hazard Classification
Type III
Model or Catalog #
03314847190
Lot or Serial #
159526
Reason for Recall
Elecsys Vitamin D3 Monoclonal Assay shows deteriorating traceability to the reference method and a negative bias when compared to the previous polyclonal antibody based version.
Manufacturer
Roche Diagnostics GMBH
Recall Posting Date
2010-08-30
Trade Name
Elecsys Free T4
Recall Start Date
2010-08-06
Recall Number
56041
Hazard Classification
Type III
Model or Catalog #
11731297122
Lot or Serial #
155674, 157677
Reason for Recall
A calibration factor close to or even below 0.8 can be observed with reagent kits stored on-board for more than four weeks. In such case, the rack pack calibration cannot be released by the system.
Manufacturer
Roche Diagnostics GMBH
Recall Posting Date
2010-08-30
Trade Name
A) Coaguchek XS Monitor
B) Coaguchek XS Plus Monitor
Recall Start Date
2010-07-07
Recall Number
56053
Hazard Classification
Type II
Model or Catalog #
A) 3666794
B) 04800842190
Lot or Serial #
A) All lots
B) All lots
Reason for Recall
The recommended cleaning procedures for the Coaguchek XS and Coaguchek XS Plus systems may cause a meter malfunction which could lead to falsely elevated PT/INR results.
Manufacturer
Roche Diagnostics GMBH
Recall Posting Date
2010-08-30
Trade Name
A) Tina-Quant D-Dimer Gen.2 for Cobas Integra and Cobas C-System
B) Tina-Quant D-Dimer Gen.2 for Roche/Hitachi 917 and Modular P Analyzers
C) Tina-Quant D-Dimer Gen.2 for Roche Cobas C111 Analyzers
Recall Start Date
2010-08-12
Recall Number
56170
Hazard Classification
Type I
Model or Catalog #
A) 04912551190
B) 04912497190
C) 05077753190
Lot or Serial #
A) 62837701
B) 62838001
C) 62841101
Reason for Recall
The impacted lots show unusual systematic negative bias of 15% when used together with serum based calibrators and controls. The false negative result could pose a patient at risk of not being timely diagnosed for thrombosis and embolism.
Manufacturer
Roche Diagnostics GMBH
Recall Posting Date
2010-08-30
Trade Name
Precicontrol Universal
Recall Start Date
2010-06-25
Recall Number
55469
Hazard Classification
Type II
Model or Catalog #
11731416190
Lot or Serial #
15418301, 15418305
Reason for Recall
The package insert for the Precicontrol Universal has been modified to exclude the TG component from the storage stability of 3 days at 2-8c after reconstitution due to instability related to an aging effect.
Manufacturer
Roche Diagnostics GMBH
Recall Posting Date
2010-09-27
Trade Name
MagNa Pure LC 2.0 Instrument
Recall Start Date
2010-08-27
Recall Number
56376
Hazard Classification
Type III
Model or Catalog #
05197686001
Lot or Serial #
All lots
Reason for Recall
There is potential for wrong error handling by the MagNa Pure LC 2.0 Instrument software where the extraction protocol is performed only during a portion of the estimated run time and then exits out of the extraction process.
Manufacturer
Roche Diagnostics GMBH
Recall Posting Date
2010-09-27
Trade Name
Cobas Integra Glucose HK Gen. 3
Recall Start Date
2010-08-31
Recall Number
56482
Hazard Classification
Type III
Model or Catalog #
N/A
Lot or Serial #
All lots
Reason for Recall
A change must be made to the glucose urine application of the Cobas C501 unit of Cobas 6000 analyzer due to precision problems and outliers when using aqueous urine control materials at a concentration level of 1.4 mmol/l.
Manufacturer
Roche Diagnostics Ltd.
Recall Posting Date
2010-07-19
Trade Name
MagNa Pure Compact Instrument
Recall Start Date
2010-06-07
Recall Number
55083
Hazard Classification
Type III
Model or Catalog #
03730964001, 03730972001, 03731146001, 04802993001
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
Manufacturer internal investigation revealed that the transfer of the nucleic acid contained in the eluates from the reagent cartridge to the elution tube was impacted with a frequency of 5%. Only a certain portion of the eluated material was transferred to the elution tubes.
Manufacturer
Roche Diagnostics Ltd.
Recall Posting Date
2010-08-16
Trade Name
A) Amplilink Software Version 3.1.1
B) Amplilink Software Version 3.1.2
C) Amplilink Software Version 3.2
D) Amplilink Software Version 3.2.1
E) Amplilink Software Version 3.2.2
F) Amplilink Software Version 3.2.3
Recall Start Date
2010-06-18
Recall Number
55272
Hazard Classification
Type II
Model or Catalog #
A) 04807197001
B) 04855094001
C) 04862392001
D) 05093236001
E) 05201357001
F) 05666678001
Lot or Serial #
A) N/A
B) N/A
C) N/A
D) N/A
E) N/A
F) N/A
Reason for Recall
Out-of-range control failures flagged as failed did not invalidate patient samples in the batch for the customers using the Amplilink SW 3.1.X & 3.2X in combination with Cobas Amplicor assays.
Manufacturer
Roche Diagnostics Ltd.
Recall Posting Date
2010-09-13
Trade Name
A) Amplilink Software Version 3.1.1
B) Amplilink Software Version 3.1.2
C) Amplilink Software Version 3.2
D) Amplilink Software Version 3.2.1
E) Amplilink Software Version 3.2.2
F) Amplilink Software Version 3.2.3
Updated recall # 55272 previously posted on 2010-08-16.
Recall Start Date
2010-06-18
Recall Number
55272
Hazard Classification
Type II
Model or Catalog #
A) 04807197001
B) 04855094001
C) 04862392001
D) 05093236001
E) 05201357001
F) 05666678001
Lot or Serial #
A) N/A
B) N/A
C) N/A
D) N/A
E) N/A
F) N/A
Reason for Recall
Out-of-range control failures flagged as failed did not invalidate patient samples in the batch for the customers using the Amplilink SW 3.1.X & 3.2X in combination with COBAS© AmpliPrep / COBAS© TaqMan©.
Manufacturer
Roche Diagnostics Ltd.
Recall Posting Date
2010-09-27
Trade Name
Cobas IT 3000 Software
Recall Start Date
2010-09-01
Recall Number
56465
Hazard Classification
Type I
Model or Catalog #
05161401001
Lot or Serial #
Version 2.03.09
Reason for Recall
Critical software issue when using automatic medical validation for Cobas IT 3000 version 2.03.09. A programming error was notice which causes the automatic medical validation to release results which should be held for review.
Manufacturer
Sequal Technologies Inc.
Recall Posting Date
2010-09-13
Trade Name
Sequal Eclipse Oxygen System
Recall Start Date
2009-12-07
Recall Number
53321
Hazard Classification
Type I
Model or Catalog #
N/A
Lot or Serial #
073002972, 063005184, 073006282, 073000585, 073007840, 073006062, 073002301, 073004651,073003155,073004456
Reason for Recall
Eclipse oxygen system units may experience premature wear on electrical connector between battery bridge printed circuit board and power manager printed circuit board that could lead to temporary power interruption when operating from power cartridge, causing unit to sound an alarm and shut down.
Manufacturer
Siemens AG
Recall Posting Date
2010-08-02
Trade Name
Syngo Imaging
Recall Start Date
2010-06-21
Recall Number
55479
Hazard Classification
Type III
Model or Catalog #
10 014 063
Lot or Serial #
1094, 1200, 1262
Reason for Recall
Potential malfunction and hence hazard to patients when using Syngo Imaging version lines VB20, VB30, VB35 in combination with RIS (radiology information system) that is violating dicoms (digital imaging and communication in medicine) standard by creating non-unique study instance UIDS (unique identifiers).
Manufacturer
Siemens AG
Recall Posting Date
2010-08-16
Trade Name
A) Artis Zee Ceiling
B) Artis Zee Biplane
Recall Start Date
2010-07-15
Recall Number
55780
Hazard Classification
Type III
Model or Catalog #
A) 10094137
B) 10094141
Lot or Serial #
A) 146288, 146314
B) 153275
Reason for Recall
A potential malfunction and a possible hazard to patients when using the Artis systems with motor controller failures caused by tolerances on the electronic parts.
Manufacturer
Siemens AG
Recall Posting Date
2010-08-30
Trade Name
A) Mevatron Linear Accelerators
B) Artiste MV-Linear Accelerator-Main Unit
Recall Start Date
2010-08-11
Recall Number
56152
Hazard Classification
Type III
Model or Catalog #
A) 1940035
B) 8139789
Lot or Serial #
A) 2473, 2559, 2618, 3235, 3236, 3376, 2459, 2616, 3291, 3356, 3378, 3382, 3406, 3742, 2239, 2267, 2264
B) 5414, 5485
Reason for Recall
Siemens received a report on a system with 160MLC, where a misalignment of treatment fields was detected during quality assurance. This misalignment was caused by a defective bearing of the collimator.
Manufacturer
Siemens AG
Recall Posting Date
2010-09-13
Trade Name
A) Axiom Artis DTA
B) Axiom Artis FA Alpha
C) Axiom Artis MP Diagnostic
Recall Start Date
2010-08-18
Recall Number
56246
Hazard Classification
Type III
Model or Catalog #
A) 7008605
B) 5904441
C) 5904466
Lot or Serial #
A) 32070
B) 25006, 25055, 25076, 25147
C) 20069, 20070, 20151, 20163, 20215
Reason for Recall
A potential danger by the image intensifier system with perivision options. During perivision the image system can sporadically crash. The customer has to repeat the perivision and to apply contrast media and x-ray to the patient.
Manufacturer
Siemens Healthcare Diagnostics Inc.
Recall Posting Date
2010-07-05
Trade Name
Dimension Tacrolimus Flex Reagent Cartridge, DF107
Recall Start Date
2010-06-16
Recall Number
55304
Hazard Classification
Type III
Model or Catalog #
DF107
Lot or Serial #
DA0334
Reason for Recall
Siemens has confirmed customer complaints regarding low recovery of the Tacrolimus Flex Reagent Cartridge DF107.
Manufacturer
Siemens Healthcare Diagnostics Inc.
Recall Posting Date
2010-08-02
Trade Name
Dimension Vista System-Enzymatic Creatinine Assay (ECREA)
Recall Start Date
2010-06-25
Recall Number
55520
Hazard Classification
Type II
Model or Catalog #
K1270
Lot or Serial #
09350AC, 10011AF, 10131AC
Reason for Recall
Siemens Healthcare Diagnostics received customer complaints regarding the Dimension Vista Assay Enzymatic Creatinine Flex Reagent Cartridge. Siemens has confirmed an issue which may cause enzymatic creatinine (ECREA) results to be falsely depressed up to 0.4 mg/dl or falsely elevated up to 0.6 mg/dl across the assay range.
Manufacturer
Siemens Healthcare Diagnostics Inc.
Recall Posting Date
2010-08-30
Trade Name
A) ADVIA Centaur Hepatitis B Core Total (HBCT) Assay
B) ADVIA Centaur XP System-HBC Total (HBCT) assay (donor screening for transplantation)
C) Advia Centaur CP System-HBC Total (HBCT) assay (donor screening for transplantation)
Recall Start Date
2010-07-05
Recall Number
55559
Hazard Classification
Type III
Model or Catalog #
A) 07566733
B) 07566733
C) 07566733
Lot or Serial #
A) 49471033, 50694033, 51071033, 51628033, 52260033, 52809033, 53674033
B) 50694033, 49471033, 51071033, 51628033, 52260033, 52809033, 53674033
C) 49471033, 50694033, 51071033, 51628033, 52260033, 52809033, 53674033
Reason for Recall
Siemens Healthcare Diagnostics confirms lower than expected specificity rates that may increase the frequency of false positive test results for this assay. Siemens believes that false positive results will lead to further investigation, and that the possibility of misinterpretation and incorrect patient management are low.
Manufacturer
Siemens Healthcare Diagnostics Inc.
Recall Posting Date
2010-07-19
Trade Name
Enzymatic Creatinine (EZCR) Flex Reagent Cartridge, DF270B
Recall Start Date
2010-06-24
Recall Number
55437
Hazard Classification
Type II
Model or Catalog #
DF270B
Lot or Serial #
BD1008, CB1050
Reason for Recall
Siemens Healthcare Diagnostics received customer complaints regarding the Dimension Enzymatic Creatinine (EZCR) Flex Reagent Cartridge. Siemens recently confirmed an issue which may cause EZCR elevations across the assay range. This issue may or may not be detected by quality control, as the degree and onset of elevated results may vary and is likely dependent on multiple factors such as method mix or instrument maintenance.
Manufacturer
Siemens Healthcare Diagnostics Products Limited
Recall Posting Date
2010-08-16
Trade Name
Immulite 2500 System-Human Chorionic Gonadotropin (HCG) Assay
Recall Start Date
2010-06-29
Recall Number
55581
Hazard Classification
Type III
Model or Catalog #
L5KCG2, L5KCG6
Lot or Serial #
123, 151
Reason for Recall
Siemens Healthcare Diagnostics Ltd has determined that the dose at which a patient sample may hook for lots 123 and above is lower than indicated in the original instructions for use (no hook effect seen up to 600,000 miu/ml, instead of claimed 2,860,000 miu/ml).
Manufacturer
Siemens Medical Solutions USA, Inc.
Recall Posting Date
2010-08-16
Trade Name
Syngo Dynamics
Recall Start Date
2010-07-27
Recall Number
55938
Hazard Classification
Type III
Model or Catalog #
100 90 480
Lot or Serial #
92241
Reason for Recall
For customers that are using the IHE displayable report profile, signing off (depending on your configuration) a report will result in an automatic upload of that report to your information system. However, if you create an addendum to a report that has been previously uploaded, the addendum must be uploaded manually using the #send to#.# feature in the Syngo Dynamics reporting system. If the addendum is not sent manually, the older report in your information system will not match the more recent report in your Syngo Dynamics system.
Manufacturer
Siemens Medical Solutions USA, Inc.
Recall Posting Date
2010-09-27
Trade Name
A) Acuson Antares Ultrasound System
B) Acuson Antares System with 5.0 Upgrades
Recall Start Date
2010-08-25
Recall Number
56322
Hazard Classification
Type III
Model or Catalog #
A) 5936518
B) 10032746
Lot or Serial #
A) 112267, 10134
B) 113469, 114573
Reason for Recall
Have determined that there is an error in the estimation of mean pressure gradient in the cardiac measurement package available on Antares systems at product versions 4.0 and 5.0.
Manufacturer
Siemens Medical Solutions USA, Inc.
Recall Posting Date
2010-09-27
Trade Name
Acuson S2000
Recall Start Date
2010-08-25
Recall Number
56323
Hazard Classification
Type III
Model or Catalog #
10041461
Lot or Serial #
200252, 200288, 200334, 200926, 200979, 200995, 202301
Reason for Recall
Have determined that there is an error in the estimation of mean pressure gradient in the cardiac measurement package available on Acuson S2000 systems.
Manufacturer
Siemens Medical Solutions USA, Inc.
Molecular Imaging Group
Recall Posting Date
2010-08-16
Trade Name
A) E-Cam Profile Refurbished System
B) Collimators for E Cam
C) E Cam Signature Series
Recall Start Date
2010-07-07
Recall Number
55779
Hazard Classification
Type III
Model or Catalog #
A) 52 44 012
B) 43 55 108
C) 7333243
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) 1005, 3005, 3034, 1004
C) >10 Numbers, Contact Manufacturer
Reason for Recall
The possibility for the patient getting injured while holding on the E-Cam patient handling system (PHS) and the technologist puts the PHS to its home position. This could possibly pose a potential safety hazard.
Manufacturer
Smith & Nephew, Inc.
Recall Posting Date
2010-07-05
Trade Name
Gigli Saw Blade, 12 Inch
Recall Start Date
2010-06-15
Recall Number
55302
Hazard Classification
Type II
Model or Catalog #
210482
Lot or Serial #
08KM00086
Reason for Recall
Smith & Nephew has found that the sterile pouches that contain the Gigli Saw Blade are not sealed. One entire end of the package is totally void of a seal.
Manufacturer
Smith & Nephew, Inc., Endoscopy Division
Recall Posting Date
2010-08-30
Trade Name
Dytonics Bonecutter Electroblade
Recall Start Date
2010-08-04
Recall Number
56143
Hazard Classification
Type II
Model or Catalog #
7205962
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
Smith & Nephew has received complaints in which a small number of patients have received skin burns associated with the usage of the Dyonics Bonecutter Electroblade.
Manufacturer
Smith & Nephew, Inc., Endoscopy Division
Recall Posting Date
2010-09-27
Trade Name
BIORAPTOR 2.9mm Suture Anchor
Recall Start Date
2010-08-30
Recall Number
56446
Hazard Classification
Type III
Model or Catalog #
72200774
Lot or Serial #
50306820
Reason for Recall
Smith & Nephew has identified a single lot of the BIORAPTOR 2.9mm Suture Anchor as having been assembled with 2 suture braids rather than 1.
Manufacturer
Smiths Medical ASD, Inc.
Recall Posting Date
2010-07-19
Trade Name
A) Medfusion Syringe Infusion Pump, 3010
B) Medfusion Syringe Infusion Pump, 3010A
Recall Start Date
2010-06-18
Recall Number
55335
Hazard Classification
Type II
Model or Catalog #
A) 3010
B) 3010A
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
Reason for Recall
Smiths Medical has found that pumps with obsolete software versions 2.0.2, 2.0.3, and 2.0.4, may continue to run past the set volume limit, if all of the following occur: 1) the volume over time (delivery method) mode is used, 2) the volume over time mode is re-accessed through the "recall last settings" function, and 3) the syringe is over-filled.
Manufacturer
Southmedic Inc.
Recall Posting Date
2010-09-27
Trade Name
Stainless Safety Scalpel Sterile
Recall Start Date
2010-09-02
Recall Number
56459
Hazard Classification
Type II
Model or Catalog #
73-0610, 73-0611, 73-0615
Lot or Serial #
070610, 071510, 070910
Reason for Recall
Southmedic Inc. is conducting a voluntary recall of specific lot numbers of the Southmedic safety scalpel due to a potential of unsealed sterile pouches.
Manufacturer
Spacelabs Medical, Inc.
Recall Posting Date
2010-09-13
Trade Name
A) Ultraview SL2600 Compact Monitor
B) Ultraview SL2200 Compact Monitor
C) Ultraview SL Compact Monitor
Recall Start Date
2009-11-12
Recall Number
56205
Hazard Classification
Type II
Model or Catalog #
A) 91370
B) 91367
C) 91369
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) 1367-000488, 1367-000447, 1367-000448, 1367-000449, 1367-000484
C) >10 Numbers, Contact Manufacturer
Reason for Recall
Affected Spacelabs' portable patient monitors may stop working when on AC power or shut down and do not power on again. The failure mode is caused by the capacitor identified as C329 malfunctioning and initiating a monitor shutdown.
Manufacturer
Sterngold Dental, LLC
Recall Posting Date
2010-09-27
Trade Name
A) ERA Implant Master Prosthetic Kit
B) ERA Mini Dental Implant System,Complete
Recall Start Date
2010-08-31
Recall Number
56385
Hazard Classification
Type III
Model or Catalog #
A) 811914Z
B) 905119Z
Lot or Serial #
A) Z3303
B) Z3561
Reason for Recall
Products do not contain instructions for use. These instructions contain important safety information related to user cleaning and re-sterilization.
Manufacturer
Sterngold Dental, LLC
Recall Posting Date
2010-09-27
Trade Name
Sterngold/Hader Torque Wrench Kit
Recall Start Date
2010-08-31
Recall Number
56387
Hazard Classification
Type III
Model or Catalog #
905104Z
Lot or Serial #
Z3379
Reason for Recall
Product does not contain the correct instructions for use. These instructions contain important safety information related to user cleaning and re-sterilization.
Manufacturer
Stryker Endoscopy
Recall Posting Date
2010-08-02
Trade Name
Stryker Disposable Scope Warmers
Recall Start Date
2010-06-28
Recall Number
55499
Hazard Classification
Type II
Model or Catalog #
0502-360-000
Lot or Serial #
All lots
Reason for Recall
Stryker Endoscopy has become aware that there is potential for a breach in the sterile packaging of the Disposable Scope Warmers.
Manufacturer
Stryker Endoscopy
Recall Posting Date
2010-08-02
Trade Name
Crossfire System - Console
Recall Start Date
2010-07-12
Recall Number
55664
Hazard Classification
Type II
Model or Catalog #
475-000-000
Lot or Serial #
All lots
Reason for Recall
Stryker Endoscopy has become aware that there is potential for the continuous activation of the radio frequency powered shaver handpieces when using the iswitch wireless footswitch with the crossfire console resulting in prolongation of surgery.
Manufacturer
Stryker Instruments
Recall Posting Date
2010-07-05
Trade Name
A) Commandf II Microelectric System
B) Core Drills
Recall Start Date
2010-06-08
Recall Number
55206
Hazard Classification
Type II
Model or Catalog #
A) 2296-301-000
B) 5400-300-000
Lot or Serial #
A) All lots
B) All lots
Reason for Recall
Stryker instruments has become aware that the inner diameter of the bur guards may be out of specification. It was discovered that the injection moulding tool was leaving additional material on the inner surface of the bur guard that is closest to the rotating nose of the impaction drill.
Manufacturer
Stryker Instruments
Recall Posting Date
2010-08-30
Trade Name
Instructions for use (IFU) for Painpump2 Block-Aid
Recall Start Date
2010-08-05
Recall Number
56046
Hazard Classification
Type II
Model or Catalog #
0575-100-000
Lot or Serial #
09355012
Reason for Recall
Stryker Instruments has revised the instructions for use (IFU) for the Painpump2 Block-Aid.
Manufacturer
Stryker Medical
Recall Posting Date
2010-07-19
Trade Name
A) Instruction Manuals for Epic II Critical Care Beds
B) Instruction Manuals for Zoom II Critical Care Beds
Recall Start Date
2010-06-21
Recall Number
55436
Hazard Classification
Type II
Model or Catalog #
A) 2030-000-000
B) 2040-000-000
Lot or Serial #
A) N/A
B) N/A
Reason for Recall
Stryker Medical has become aware that the Stryker Critical Care Beds were reportedly displaying symptoms of unsmooth and noisy fowler motion. Continuous use could potentially result in a situation in which the manual CPR drop on the bed may become inoperable at fowler angles of approximately 45 degrees or more.
Manufacturer
Stryker Medical
Recall Posting Date
2010-08-02
Trade Name
M1 Ambulance Cot
Recall Start Date
2010-06-10
Recall Number
55697
Hazard Classification
Type II
Model or Catalog #
6100-000-000
Lot or Serial #
N/A
Reason for Recall
Stryker Medical has become aware that the head end slide tube, bushings, and head end release rod have the potential to break during use, potentially resulting in the collapse of the head end of the cot.
Manufacturer
Stryker Medtech K.K.
Recall Posting Date
2010-08-16
Trade Name
Stryker Sonopet Ultrasonic Aspirator Console
Recall Start Date
2010-07-19
Recall Number
55828
Hazard Classification
Type II
Model or Catalog #
5450-850-000
Lot or Serial #
All lots
Reason for Recall
Stryker Instruments became aware that the Sonopet console did not meet the IEC 60601-1 standard (electromedical safety standard).
Manufacturer
Stryker Orthopeadics
Recall Posting Date
2010-08-30
Trade Name
Scorpio T72 Insert Tibial Trials
Recall Start Date
2010-07-27
Recall Number
55981
Hazard Classification
Type II
Model or Catalog #
T-72
Lot or Serial #
>100 Numbers, Contact Manufacturer
Reason for Recall
Stryker Orthopaedics has become aware that there is the potential for the Scorpio T72 Insert Tibial Trials to be oversized by a maximum of .024"(0.6mm). It was determined that an oversized Tibial Insert Trial may sit on the anterior edge of the tibial baseplate if used for secondary trialing in a procedure. If there is an inability to seat the insert trial and impaction is used to seat the trial, fracture of the device may occur.
Manufacturer
Sybron Endo
Recall Posting Date
2010-08-16
Trade Name
K3 Gutta Percha
Recall Start Date
2010-07-14
Recall Number
55901
Hazard Classification
Type III
Model or Catalog #
825-0420
Lot or Serial #
020210
Reason for Recall
Mislabeled packages of the K3 Gutta Percha .04/20 points (part number 825-0420) contained the K3 Gutta Percha .04/40.
Manufacturer
Synthes (Canada) Ltd.
Recall Posting Date
2010-09-13
Trade Name
Insight Retractor System Frames
Recall Start Date
2010-08-12
Recall Number
56200
Hazard Classification
Type II
Model or Catalog #
03.615.002, 03.615.100
Lot or Serial #
N/A
Reason for Recall
The holding pin on the retractor frame was reported as unscrewing and falling out. If this were to occur while the retractor is in use, the retraction will not be affected but the fallens screw has the potential to enter the surgical wound undetected. This may result in an un-retrieved device fragment.
Manufacturer
Synthes (Canada) Ltd.
Recall Posting Date
2010-09-13
Trade Name
Implant Holder for Synfix-LR - System Technique Guide
Recall Start Date
2010-08-13
Recall Number
56201
Hazard Classification
Type II
Model or Catalog #
03.802.039
Lot or Serial #
N/A
Reason for Recall
Synthes (Canada), Ltd is initiating a medical device labelling recall related to the Synthes Synfix-LR System Technique Guide, specifically regarding use of the Synfix implant holder. The surgical technique presented in the guide has been revised and emphasizes the importance of ensuring that the Synfix-LR implant holder is not cross threaded when attached to the Synfix-LR implant. The revision also describes the use of holding forceps.
Manufacturer
Sysmex Corporation
Recall Posting Date
2010-09-27
Trade Name
A) XT-2000I Analyzer - Sysmex Hematology Analyzers
B) Sysmex XT-1800I Analayzer - Sysmex Hematology Analyzers
C) XT-4000I Hematology Analyzer
Recall Start Date
2010-08-27
Recall Number
56348
Hazard Classification
Type II
Model or Catalog #
A) 013-2531-8
B) 023-0531-6
C) 013-2553-0
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) 16975
C) 11165, 11232
Reason for Recall
A design error introduced in the field programmable gate array (FPGA) was discovered. This FPGA is a logic circuit on a chip embedded on the printed computer board (PCB) in the main unit of the XT-series analyzers. The error can present as an increase in the platelet (PLT) count of up to double the correct count and/or the generation of excessive PLT sampling errors. The increase in count and errors is caused by the circuit seeing a single platelet and creating two pulses that are included in the reported PLT count.
Manufacturer
Technidata S.A.S.
Recall Posting Date
2010-07-05
Trade Name
TD-Synergy Software
Recall Start Date
2009-12-30
Recall Number
52751
Hazard Classification
Type III
Model or Catalog #
TD-SYN-01
Lot or Serial #
All lots
Reason for Recall
If some microbiology results have not been consolidated yet, and an event triggers replication from the production database to the management database, then the non-consolidated results are overwritten by blank results.
Manufacturer
Technidata S.A.S.
Recall Posting Date
2010-08-30
Trade Name
TD-Synergy
Recall Start Date
2010-07-02
Recall Number
55962
Hazard Classification
Type II
Model or Catalog #
TD-SYN-01
Lot or Serial #
N/A
Reason for Recall
In the ERM session, when results are of coded text type, incorrect data could be displayed in the tool tip on the current result field.
Manufacturer
Technidata S.A.S.
Recall Posting Date
2010-08-30
Trade Name
A) TDLims
B) TD-Synergy
Recall Start Date
2010-07-02
Recall Number
55963
Hazard Classification
Type II
Model or Catalog #
A) TD-LIS-01
B) TD-SYN-01
Lot or Serial #
A) N/A
B) N/A
Reason for Recall
Using F8 key in the ERM session to add a combined test can result in the modification of the value of a result that has already been entered for a test of the request. This could only occur if the combined test added by F8 includes at least one test with the "duplicate result" property set to ?yes?.
Manufacturer
Teleflex Medical
Recall Posting Date
2010-08-02
Trade Name
Rusch Care Premium Drain Bag
Recall Start Date
2010-07-07
Recall Number
55666
Hazard Classification
Type II
Model or Catalog #
390000
Lot or Serial #
20100105
Reason for Recall
During an inspection wet product was found that was due to improper aeration during sterilization of the product and the sterile integrity could not be guaranteed.
Manufacturer
Teleflex Medical
Recall Posting Date
2010-08-16
Trade Name
IMA / ENT Blade Electrode
Recall Start Date
2010-07-30
Recall Number
55952
Hazard Classification
Type II
Model or Catalog #
809335
Lot or Serial #
All lots
Reason for Recall
There is a potential for peeling and melting of the insulation of the cautery tip of the IMA / ENT Blade Electrodes. This can result in the tip losing integrity leading to sparking or arcing of the current at the blade tip and result in a burn or electrical shock.
Manufacturer
Terumo Cardiovascular System Corporation
Recall Posting Date
2010-07-05
Trade Name
Sarns Disposable Centrifugal Pump
Recall Start Date
2010-06-02
Recall Number
55167
Hazard Classification
Type II
Model or Catalog #
164275
Lot or Serial #
N/A
Reason for Recall
Terumo Cardiovascular Systems Corp. received reports of loss of forward with the Sarns Disposable Centrifugal Pump. Terumo CVS investigation determined that in the cases reported that impeller in the disposable shaft separated from the shaft causing it to stop rotating.
Manufacturer
Terumo Cardiovascular System Corporation
Recall Posting Date
2010-07-05
Trade Name
Sarns Cardioplegia Set with MP-4 Module
Recall Start Date
2010-06-10
Recall Number
55224
Hazard Classification
Type II
Model or Catalog #
71032
Lot or Serial #
ME19
Reason for Recall
Terumo Cardiovascular Systems (Terumo CVS) has received four reports of an occluded vent port on the Sarns Cardioplegia Set with MP-4 module. In all reports the defect was evident during priming. Terumo CVS investigation found that for some units in certain lots of MP-4 modules, a damaged pin in the injection mold allowed the vent port to fill with excess plastic.
Manufacturer
Tomotherapy Incorporated
Recall Posting Date
2010-07-19
Trade Name
TomoTherapy HI-ART System
Recall Start Date
2010-06-23
Recall Number
55439
Hazard Classification
Type II
Model or Catalog #
101264
Lot or Serial #
10037, 10169
Reason for Recall
The planning station's printed plan reports may indicate statistics which are not correct. The TomoTherapy Hi-Art treatment planning system presents DVH, isodose and dose statistics information on the user interface (or on-screen) which are accurate and unaffected by this issue. However, if erroneous dose statistics were printed and then provided to third parties to guide therapy in retreatment situations, improper therapy could result.
Manufacturer
Toshiba Medical Systems Corporation
Recall Posting Date
2010-07-19
Trade Name
Kalare
Recall Start Date
2010-06-07
Recall Number
55449
Hazard Classification
Type II
Model or Catalog #
DUA-450F
Lot or Serial #
99A0562042, 99A05X2054, 99A05X2055, 99A05Y2057, 99A0612074, 99A0622078, 99A0672105, A4612006
Reason for Recall
In diagnostic x-ray system Kalare tabletop lateral movement cannot be performed, or tabletop lateral movement is performed at extremely slow speed. As a result, positioning of the patient cannot be performed by tabletop movement.
Manufacturer
Toshiba Medical Systems Corporation
Recall Posting Date
2010-08-16
Trade Name
Aquilion One System - Main Unit
Recall Start Date
2010-07-22
Recall Number
55835
Hazard Classification
Type II
Model or Catalog #
TSX-301A/2
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
In Aquilion One systems, communication between units in the console may be cut off due to failure of a component (hub).
Manufacturer
Toshiba Medical Systems Corporation
Recall Posting Date
2010-09-13
Trade Name
Aquilion One
Recall Start Date
2010-08-06
Recall Number
56173
Hazard Classification
Type II
Model or Catalog #
TSX-301A/2
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
In Aquilion One systems, in some cases the pre-scan dose values displayed in the exam plan editing window is lower than what the displayed value should be when the scanner executes the exam plan.
Manufacturer
Toshiba Medical Systems Corporation
Recall Posting Date
2010-09-13
Trade Name
Aplio Artida
Recall Start Date
2010-08-05
Recall Number
56227
Hazard Classification
Type II
Model or Catalog #
SSH-880CV
Lot or Serial #
W1B0862082, W1F0982235, WIF0982239, WIF0982240
Reason for Recall
In Artida Ultrasound Systems model SSH-880CV, there are two issues noted regarding (1) some measurement values are displayed incorrectly on the report page, and (2) temperature-related warnings are not consistently displayed.
Manufacturer
Trinity Biotech USA
Recall Posting Date
2010-09-13
Trade Name
Captia VZV IgM Kit
Recall Start Date
2010-07-17
Recall Number
56303
Hazard Classification
Type II
Model or Catalog #
2325650
Lot or Serial #
122, 123, 124, 125
Reason for Recall
Disproportionate drop in absorbance of calibrator causes ISR to run out of range. Risk of a false positive.
Manufacturer
Trinity Biotech USA
Recall Posting Date
2010-09-27
Trade Name
Captia EBV VCA IGM
Recall Start Date
2010-07-28
Recall Number
56527
Hazard Classification
Type II
Model or Catalog #
2325760
Lot or Serial #
084
Reason for Recall
Low positive control range for the low positive control (5760-4-003078) included in the Elisa Kit is incorrect. The correct range for the control is (0.91-2.3).
Manufacturer
Tyco Healthcare Liebel-Flarsheim Company
Recall Posting Date
2010-07-05
Trade Name
Hydra Vision DR60/80 System, Generators
Recall Start Date
2010-06-14
Recall Number
55273
Hazard Classification
Type II
Model or Catalog #
700539, 700540
Lot or Serial #
All lots
Reason for Recall
A smell of burning/smoke is emanating from the x-ray generators. This is caused by a failed printed circuit board electronic component and an internal electronic transformer. These failures will cause the generator to become inoperative.
Manufacturer
United States Surgical, A Division of Tyco Healthcare Group
Recall Posting Date
2010-07-19
Trade Name
Auto Suture Protack Fixation Device
Recall Start Date
2010-06-28
Recall Number
55465
Hazard Classification
Type II
Model or Catalog #
174006
Lot or Serial #
P0B0753
Reason for Recall
Certain Protack Fixation devices fire one helical tack and then become inoperable.
Manufacturer
Varian Medical Systems, Inc.
Recall Posting Date
2010-08-16
Trade Name
A) Clinac Linear Accelerator
B) Trilogy Radiotherapy Delivery System
C) Clinac IX Radio Therapy Delivery System
D) Novalis TX Linear Accelerator
Recall Start Date
2010-05-27
Recall Number
55377
Hazard Classification
Type II
Model or Catalog #
A) 2100C, 21EX, 21EXD, 21EXS
B) IX, IXS
C) IX, IXS, TRILOGY
D) NOVALIS TX
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) H140836, H291166, H291172, H293787, H293883, H294134, H294403, H294551, H294750
C) >10 Numbers, Contact Manufacturer
D) H294674
Reason for Recall
The auto "goto" or auto "setup" functions in C-series version 7 software ignore couch angle exceeded motion zone unless a tolerance is defined by the user (gantry rotation exclusion now only applies by default if OBI is present and deployed. Customer facilities using both C-series version 6 & 7 may get confused because couch safety zone no longer exists by default in version 7.
Manufacturer
Varian Medical Systems, Inc.
Recall Posting Date
2010-08-16
Trade Name
A) Varian Clinac Linear Accelerator
B) Trilogy Radiotherapy Delivery System
C) Varian Clinac IX Radiotherapy Delivery System
D) Novalis TX Linear Accelerator
Recall Start Date
2010-06-23
Recall Number
55612
Hazard Classification
Type II
Model or Catalog #
A) N/A
B) N/A
C) N/A
D) N/A
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
C) >10 Numbers, Contact Manufacturer
D) >10 Numbers, Contact Manufacturer
Reason for Recall
Anomalies have been identified with the Clinac Collimator and couches that can result in small patient or field positioning errors that may adversely affect high - dose treatments requiring small positioning tolerances, such as stereotactic radiosurgery (SRS) and stereotactic radiotherapy (SRT). Refer to 'C- series Clinacs when used for hypo-fractioned radiotherapy' dated 06/23/2010 for further information.
Manufacturer
Varian Medical Systems, Inc.
Recall Posting Date
2010-09-13
Trade Name
Eclipse Treatment Planning (BrachyVision)
Recall Start Date
2010-08-01
Recall Number
56293
Hazard Classification
Type II
Model or Catalog #
8.2, 8.5, 8.6, 8.9
Lot or Serial #
N/A
Reason for Recall
When incorrect syntax is used in the template file, a misinterpretation of the desired value can occur.
Manufacturer
Varian Medical Systems, Inc.
Recall Posting Date
2010-07-05
Trade Name
A) Varis Vision or Aria (Version 6.5)
B) Eclipse
Recall Start Date
2010-05-13
Recall Number
55200
Hazard Classification
Type II
Model or Catalog #
A) Version 6.5
B) Version 6.5
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
Reason for Recall
An anomaly was discovered in which a second or subsequent user can access a patient record which is already open at another workstation, and, if they ignore the warning message that the record is being edited by another user, and proceed to edit the plan and save the record, a corrupted but usable patient plan record may result. Errors can include, but are not limited to, incorrect MU, incorrect beam modifiers, incorrect gantry angles, incorrect field sizes, etc.
Manufacturer
Vascular Solutions, Inc.
Recall Posting Date
2010-09-27
Trade Name
Langston Dual Lumen Catheter
Recall Start Date
2010-08-31
Recall Number
56416
Hazard Classification
Type II
Model or Catalog #
5540
Lot or Serial #
548110, 548227, 548364, 548430, 548554, 548723
Reason for Recall
Some lots of the Langston Dual-Lumen Pigtail Catheter, model 5540 are missing a sterile barrier pouch seal.
Manufacturer
Vital Diagnostics, Inc.
Recall Posting Date
2010-08-30
Trade Name
Excyte Sliders
Recall Start Date
2010-04-14
Recall Number
54156
Hazard Classification
Type III
Model or Catalog #
EX-10552
Lot or Serial #
All lots
Reason for Recall
There are Excyte ESR tubes in the field that do not fit properly in the Excyte sliders. The product will not be replaced. This product is discontinued.
Manufacturer
Wilson-Cook Medical, Inc.
Recall Posting Date
2010-09-27
Trade Name
A) Six-Shooter Saeed Multi-Band Ligator
B) Ten-Shooter Saeed Multi-Band Ligator
Recall Start Date
2010-09-07
Recall Number
56462
Hazard Classification
Type II
Model or Catalog #
A) MBL-6-OV, MBL-6-OV-XS
B) MBL-10
Lot or Serial #
A) W2783447, W2758500, W2758501, W2772743, W2753989, W2772759, W2777465
B)W2758472, W2758474, W2764330, W2772073
Reason for Recall
Possibility of a section of the ligator barrel separating from the barrel's friction fit adapter that attaches to patient end of the endoscope, which could in turn result in a section of the ligator barrel detaching inside the patient's gastrointestinal tract.
Manufacturer
Zimmer Inc.
Recall Posting Date
2010-08-02
Trade Name
Versys Hip System Femoral Head 12/14
Recall Start Date
2010-07-09
Recall Number
55632
Hazard Classification
Type II
Model or Catalog #
00801802801, 00801803201
Lot or Serial #
61424803, 61275855
Reason for Recall
This action is being conducted because there is a potential for a package to be labelled as a 28mm femoral head but contain a 32mm femoral head. In addition, there is a potential for the patient record label to be incorrect.
Manufacturer
Zimmer Inc.
Recall Posting Date
2010-08-02
Trade Name
ITST Fixation Locking Bolt Extractor
Recall Start Date
2010-07-09
Recall Number
55641
Hazard Classification
Type II
Model or Catalog #
00-2258-051-01
Lot or Serial #
53951300
Reason for Recall
This action is being conducted because Zimmer has determined the instruments were manufactured using a grade of material that is nonconforming to specifications. As manufactured, the instruments are more brittle, which increases the potential for fracture.
Manufacturer
Zimmer Inc.
Recall Posting Date
2010-09-13
Trade Name
Trabecular Metal Acetabular Porous Shell
Recall Start Date
2010-08-20
Recall Number
56268
Hazard Classification
Type II
Model or Catalog #
00-6202-056-22
Lot or Serial #
61091653
Reason for Recall
This action is being conducted because the threads on the device (Trabecular Metal Modular Acetabular 56mm Porous Shell with Cluster Holes) were not completely machined and therefore will not mate with the corresponding cup positioner.
Manufacturer
Zimmer Inc.
Recall Posting Date
2010-09-13
Trade Name
Bigliani/Flatow Fukuda Retractors
Recall Start Date
2010-08-23
Recall Number
56271
Hazard Classification
Type III
Model or Catalog #
00-4305-019-30, 00-4305-019-40
Lot or Serial #
60795817, 60834923, 60916795, 61017041, 61038697, 60768814, 60806551, 60893241, 60946820, 60999665
Reason for Recall
This action is being conducted due to fractures occurring at the grooves around the perimeter of the blades. The Bigliani/Flatow complete shoulder solution Fukuda Retractors (small and large sizes) are used to gain exposure to the glenoid during surgery.
Manufacturer
Zimmer Trabecular Metal Technology, Inc.
Recall Posting Date
2010-08-02
Trade Name
Trabecular Metal Femoral Cone Augment
Recall Start Date
2010-07-09
Recall Number
55634
Hazard Classification
Type II
Model or Catalog #
00-5450-012-31
Lot or Serial #
60683028
Reason for Recall
This action is being conducted because there is a potential for the packages to be mislabeled. Specifically, the left femoral cone augment component may be labelled as a right and the right femoral cone augment component may be labelled as a left.