Health Canada
www.hc-sc.gc.ca
Home
> Drugs & Health Products
> Compliance & Enforcement
> Recalls
> Recalls
Institutional links
-
Back to
-
Explore...
-
Proactive Disclosure
Medical Device Recall Listings - Sorted by Trade Name
(From October 2009 - December 2009)
- Trade Name
- 070 Neuron Delivery Catheter
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Penumbra Inc.
- Recall Start Date
- 2009/09/17
- Recall Number
- 51330
- Hazard Classification
- Type II
- Model or Catalog #
- N/A
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- Design modifications were made to the Neuron Delivery Catheter in March 2009 to improve the catheter kink resistance. Penumbra initiated voluntary field removal of the original catheter design from the marketplace.
- Trade Name
- A Locus High Resolution SSP Unitray
- Recall Posting Date
- 2009/11/23
- Manufacturer
- Invitrogen Corporation
- Recall Start Date
- 2009/07/24
- Recall Number
- 51366
- Hazard Classification
- Type III
- Model or Catalog #
- 47290-10, 47291-10
- Lot or Serial #
- 1145687
- Reason for Recall
- Lane 23 in the 'A' high resolution SSP Unitray, lot and batch number identified, will produce a false positive band at the correct base pair size in the presence of A*0301 samples. This will affect typing results on samples that are Aa*0301 homozygote, since they could be erroneously reported as A*0301/A*1101.
- Trade Name
- A) Access Immunoassay System - Troponin
B) Unicel DXI600 Access Immunoassay System
C) Unicel DXI800 Access Immunoassay System
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Beckman Coulter Inc.
- Recall Start Date
- 2009/10/23
- Recall Number
- 51585
- Hazard Classification
- Type III
- Model or Catalog #
- A) A78803
B) A78803
C) A78803
- Lot or Serial #
- A) 915417
B) 915417
C) 915417
- Reason for Recall
- The reagents in wells three (RLC) and four (RLD) of the reagent pack were inadvertently switched during manufacturing. Lot number 915417 contains reagent RLD in well three and RLC in well four.
- Trade Name
- A) Accura Breast Localisation Needles
B) Hawkins II Breast Localisation Needles
C) Hawkins I Breast Localization Needles
D) Homer Mammalok Breast Localisation Needles
E) Manan Co-Axial Introducer Needles
F) Manan Introducer Sheath Needles
G) Manan Guidewire Introducer Needles
H) Manan Seldinger Needles
I) Manan Modified Potts/Cournand Needles
J) Manan Galactography Trays
K) Skater-Centesis Catheter
L) Chiba Style Manual
M) Franseen Style
N) Green Style
O) Spinal Style
P) Westcott Style
Q) Hawkins III Breast Localization Needles
R) Pro-Mag Disposable Biopsy Needles
S) Maxi-Cell Biopsy Needles
T) Ultracore Disposable Biopsy Needles
U) Lorad Needle Guide
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Medical Device Technologies, Inc. DBA Angiotech
- Recall Start Date
- 2009/09/01
- Recall Number
- 50935
- Hazard Classification
- Type II
- Model or Catalog #
- A) N/A
B) N/A
C) N/A
D) N/A
E) N/A
F) N/A
G) N/A
H) N/A
I) N/A
J) N/A
K) N/A
L) N/A
M) N/A
N) N/A
O) N/A
P) N/A
Q) N/A
R) N/A
S) N/A
T) N/A
U) N/A
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
C) >10 Numbers, Contact Manufacturer
D) >10 Numbers, Contact Manufacturer
E) >10 Numbers, Contact Manufacturer
F) >10 Numbers, Contact Manufacturer
G) >10 Numbers, Contact Manufacturer
H) >10 Numbers, Contact Manufacturer
I) >10 Numbers, Contact Manufacturer
J) >10 Numbers, Contact Manufacturer
K) >10 Numbers, Contact Manufacturer
L) >10 Numbers, Contact Manufacturer
M) >10 Numbers, Contact Manufacturer
N) >10 Numbers, Contact Manufacturer
O) >10 Numbers, Contact Manufacturer
P) >10 Numbers, Contact Manufacturer
Q) >10 Numbers, Contact Manufacturer
R) >10 Numbers, Contact Manufacturer
S) >10 Numbers, Contact Manufacturer
T) >10 Numbers, Contact Manufacturer
U) >10 Numbers, Contact Manufacturer
- Reason for Recall
- The complaints were reported as "run off" (non-contiguous) seals. A run off seal is a seal in which the Angiotech applied seal is mis-aligned.
- Trade Name
- A) Acuson Antares Software Version 4.0
B) Acuson Antares Main Unit/5.0 Upgrades
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Siemens Medical Solutions USA, Inc.
- Recall Start Date
- 2009/10/08
- Recall Number
- 51249
- Hazard Classification
- Type III
- Model or Catalog #
- A) 05936518
B) 10032746
- Lot or Serial #
- A) 112267, 10134
B) 114573, 113469
- Reason for Recall
- Potential software problem with Acuson and Sonoline Antares systems product version 5.0 with cardiac option installed below software version 200.0.079.
- Trade Name
- A) Antistreptolysin O Assay
B) Antistrepolysin O Assay
C) IGM Flex Assay
D) ASL Flex Reagent
- Recall Posting Date
- 2009/10/26
- Manufacturer
- Siemens Healthcare Diagnostics Products GMBH
- Recall Start Date
- 2009/09/24
- Recall Number
- 50979
- Hazard Classification
- Type III
- Model or Catalog #
- A) K7058
B) K7022
C) 10445904
D) 10445888
- Lot or Serial #
- A) 07310MA, 08029MA, 08252MA, 08336MA
B) 07311MA, 08043MA, 09077MA, 07177MA
C) 07310MA, 08029MA, 08252MA, 08336MA
D) 07177MA, 07311MA, 08043MA, 09077MA
- Reason for Recall
- Under certain conditions the software falsely elevated the results particularly if the preceding serum sample has a higher volume and concentration of beta2-microglobulin, anti-strepolysin o, or immunoglobulin M analytes.
- Trade Name
- A) Artiste MV-Syngo RT Therapist Assist
B) Artiste MV-Syngo RT Therapist
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Siemens Medical Solutions USA, Inc. Oncology Care Systems
- Recall Start Date
- 2009/11/05
- Recall Number
- 51605
- Hazard Classification
- Type III
- Model or Catalog #
- A) 08162807
B) 08162815
- Lot or Serial #
- A) 10024
B) 10040
- Reason for Recall
- Two or more treatment plans with different ISOcenter positions are stored together in one course in Lantis, e.g. main treatment plan and boost plan. If the cone beam is acquired using calculation of table shift coordinates will be based on ISOcenter position. This will lead to an incorrect table shift.
- Trade Name
- A) Artiste Syngo RT Therapist Assist
B) Artiste Syngo RT Therapist
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Siemens Medical Solutions USA, Inc.
Oncology Care Systems
- Recall Start Date
- 2009/09/23
- Recall Number
- 51246
- Hazard Classification
- Type III
- Model or Catalog #
- A) 08162807
B) 08162815
- Lot or Serial #
- A) 10024
B) 10040
- Reason for Recall
- Fractions delivered on the Artiste might not be recorded in Lantis and subsequently patients might be mistreated.
- Trade Name
- A) AW Volumeshare
B) AW Volumeshare2
- Recall Posting Date
- 2009/10/12
- Manufacturer
- GE Medical Systems SCS
- Recall Start Date
- 2009/09/15
- Recall Number
- 49927
- Hazard Classification
- Type II
- Model or Catalog #
- A) S30321ZK
B) 5183859-3, 5183859-4, 5183859-5
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) S/N'S A8177079, A9062007, 9093001, A7274009, M8281004, 8354008
- Reason for Recall
- Measurement values provided by Seno Advantage Workstations 1x. 2.0 & 2.1 may be incorrect when applied to magnification images acquired on non-GE digital mammography systems, which could lead to overestimate of size of breast lesion.
- Trade Name
- A) BD Q-Syte Extension Set
B) MPS Acadia Extension Set With Rightbore
C) MPS Acadia Specialty Tri-Extension Set
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Becton Dickinson Infusion Therapy Systems Inc.
- Recall Start Date
- 2009/11/02
- Recall Number
- 51661
- Hazard Classification
- Type I
- Model or Catalog #
- A) 385102
B) 385150, 385151
C) 385164
- Lot or Serial #
- A) 8308330
B) A1967, A2399
C) A1951
- Reason for Recall
- Product may function improperly. This may result in the risk of air bubbles leaking into the infusion system through the BD Q-Syte Luer access split septum. This may lead to complications should the air bubbles not be detected.
- Trade Name
- A) Calcar Planer Surgical Protocols
B) Accolade Surgical Protocol
- Recall Posting Date
- 2009/10/12
- Manufacturer
- Howmedica Osteonics Corporation
- Recall Start Date
- 2009/09/14
- Recall Number
- 50823
- Hazard Classification
- Type II
- Model or Catalog #
- A) LSP42, LSP46, LSP47M, LSP48
B) LASST
- Lot or Serial #
- A) N/A
B) N/A
- Reason for Recall
- Stryker has identified a need to provide specific instructions for initiating power to the Calcar Planer prior to engagement with bone in the cutting edge, Omnifit HFX, Secur-Fit Max/Secur-Fit Plus Max, and Omnifit Eon Surgical Protocols, as well as to provide specific instructions for rasping dense bone and/or narrow intra medullary canals for the Accolade Surgical Protocol.
- Trade Name
- A) C-Series Clinac, Trilogy & Novalis Tx
B) Trilogy Radiotherapy Delivery System
C) Novalis TX Linear Accelerator
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Varian Medical Systems, Inc.
- Recall Start Date
- 2009/10/08
- Recall Number
- 51773
- Hazard Classification
- Type II
- Model or Catalog #
- A) H18, H27, H29
B) H14, H29
C) H14, H29
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
C) >10 Numbers, Contact Manufacturer
- Reason for Recall
- SRS treatments occurred with conical collimator accessory not inserted, or with conical collimator accessory correctly mounted, but with a primary collimator field size setting exceeding outside diameter of that conical collimator. With wrong field size, unintended radiation is given to patient outside the outer edge of cone.
- Trade Name
- A) Definium 8000 Digital Radiographic System
B) Proteus XR/A General Purpose Radiographic
C) Revolution XR/D Digital Radiographic System
- Recall Posting Date
- 2009/10/12
- Manufacturer
- GE Healthcare
- Recall Start Date
- 2009/09/21
- Recall Number
- 47780
- Hazard Classification
- Type II
- Model or Catalog #
- A) 2259988-2, 5131070
B) 2259973-3, 2259976, 2259988, S3918KD/KE
C) 2351505, 5131070, S3918LD
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) >100 Numbers, Contact Manufacturer
C) >10 Numbers, Contact Manufacturer
- Reason for Recall
- Table locks may not actuate, causing tabletop to unexpectedly float in either the lateral axes or both the longitudinal and lateral axes, which could result in patient or operator injury.
- Trade Name
- A) Imager II Angiographic Catheters
B) Imager II Catheters BERN
C) Imager II Catheters SIM
D) Imager II Catheters Contra2
E) Imager II Catheters C
F) Imager II Catheters RC
G) Imager II Catheters RDC
H) Imager II Catheters Pigtail
I) Imager II Catheters Contralateral
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Boston Scientific Corporation
- Recall Start Date
- 2009/11/03
- Recall Number
- 51535
- Hazard Classification
- Type II
- Model or Catalog #
- A) 31-400, 31-402, 31-510, 31-516, 31-520, 31-602
B) 31-407, 31-408, 31-409, 31-410, 31-607, 31-609
C) 31-430, 31-631
D) 31-458, 31-459, 31-658
E) 31-466, 31-467, 31-468, 31-469, 31-669
F) 31-485, 31-486, 31-685
G) 31-488, 31-489, 31-491
H) 31-509, 31-513, 31-519
I) 31-525, 31-528, 31-531
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
C) >10 Numbers, Contact Manufacturer
D) >10 Numbers, Contact Manufacturer
E) >10 Numbers, Contact Manufacturer
F) >10 Numbers, Contact Manufacturer
G) >10 Numbers, Contact Manufacturer
H) >10 Numbers, Contact Manufacturer
I) >10 Numbers, Contact Manufacturer
- Reason for Recall
- The sterile barrier in the packaging (pouch) that contains the catheter may be breached. The breach of sterility could lead to a contamination of the device with subsequent transfer of infectious agents to the patient.
- Trade Name
- A) Immulite System-Reactive Protein (CRP) Assay
B) Immulite 2500 System-HIH Sensit CRP Assay
C) Immulite 1000 System-Cardiac HSCRP Assay
D) Immulite System-Cardiac HSCRP Assay
E) Immulite 2000 System-Cardiac HSCRP Assay
F) Immulite 2500 System-HSCRP Assay
G) Immulite 1000 System-High Sens CRP Assay
H) Immulite 3000 Systems-High Sens CRP Assay
I) Immulite System-High Sensitivity CRP Assay
J) Immulite 1000 System-C-Reactive Protein
- Recall Posting Date
- 2009/10/26
- Manufacturer
- Siemens Healthcare Diagnostics Products Limited
- Recall Start Date
- 2009/09/11
- Recall Number
- 50797
- Hazard Classification
- Type II
- Model or Catalog #
- A) LCRCM
B) LCRCM
C) LCRCM
D) LCRCM
E) L2KCRP2
F) LCRCM
G) LCRCM
H) L2KCRP2
I) LCRCM
J) LCRCM
- Lot or Serial #
- A) 022L, 025L
B) 022L, 025L
C) 022L, 025L
D) 022L, 025L
E) 243
F) 022L, 025L
G) 022L, 025L
H) 243
I) 022L, 025L
J) 022L, 025L
- Reason for Recall
- Siemens Healthcare Diagnostics received customer complaints regarding instabilities of high sensitivity C-reactive protein adjusters and controls giving an average positive bias of 20% compared with results from previous lots.
- Trade Name
- A) Intravenous Infusion Accessory IVMED20
B) Intravenous Infusion Accessory IVMED60
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Medxl Inc.
- Recall Start Date
- 2009/11/06
- Recall Number
- 51696
- Hazard Classification
- Type II
- Model or Catalog #
- A) N/A
B) N/A
- Lot or Serial #
- A) N/A
B) N/A
- Reason for Recall
- A warning letter was sent to the distributor requesting that sales be suspended until a licence has been obtained.
- Trade Name
- A) Intro-Flex Introducer Sets
B) Percutaneous Sheath Introducer Kit
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Edwards Lifesciences LLC
- Recall Start Date
- 2009/05/21
- Recall Number
- 48857
- Hazard Classification
- Type II
- Model or Catalog #
- A) 1350BF85
B) I652BF85H, S5507BF85
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) 58511631, 58549300, 58606253, 58648812, 58496258, 58535112, 58604814, 58644423
- Reason for Recall
- Blue connector proximal to Tuohy-Borst hemostasis valve may fatigue, crack or separate, resulting in patency failure & potential blood loss where Intro-Flex Introducer secures a venous catheter.
- Trade Name
- A) Navigation Software Knee Zimmer
B) Brainlab Knee Essential Software
C) Brainlab Knee Unlimited Software
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Brainlab AG
- Recall Start Date
- 2009/09/11
- Recall Number
- 50954
- Hazard Classification
- Type III
- Model or Catalog #
- A) 22086-03D, 22086-03E
B) 21100-01(2.0), 21100-01C(2.1)
C) 70001A, 70002A(2.1)
- Lot or Serial #
- A) Version 2.1.1 & All Previous
B) Version 2.1.1 & All Previous
C) Version 2.1.1 & All Previous
- Reason for Recall
- Incorrect Zimmer implant data in Brainlab Software Knee leads to incorrect information for Zimmer Innex implants.
- Trade Name
- A) Nexgen Rotating Hinge Knee - Femoral
B) Nexgen Rotating Hinge Knee - Tibial
C) Nexgen Rotating Hinge Knee,Tibial Plate
D) Nexgen Rotating Hinge Knee - Femoral, B
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Zimmer Inc.
- Recall Start Date
- 2009/11/06
- Recall Number
- 51622
- Hazard Classification
- Type III
- Model or Catalog #
- A) 00-5880-013-01, 00-5880-014-01, 00-5880-014-02, 00-5880-015-01, 00-5880-015-02, 00-5880-016-01, 00-5880-016-02
B) 00-5880-002-00, 00-5880-003-00, 00-5880-004-00, 00-5880-005-00, 00-5880-006-00
C) 00-5880-001-02
D) 00-5880-012-01
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
C) 60433397, 60461749
D) 60186249 H5
- Reason for Recall
- This action is being conducted because these components are labelled with a 10 year expiry period when in fact these components expire 5 years after manufacture. Expiration dating refers to 'use by' dates. There is no risk to patients who receive these implants prior to January 2010, when devices from the affected lots begin to expire.
- Trade Name
- A) Online TDM Gentamicin for Roche/Hitachi
B) Online TDM Gentamicin (GENT2) for Cobas
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Roche Diagnostics GMBH
- Recall Start Date
- 2009/10/07
- Recall Number
- 51194
- Hazard Classification
- Type II
- Model or Catalog #
- A) 03800504190
B) 04490843190
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
- Reason for Recall
- The online TDM Gentamicin assays may generate falsely elevated control and patient results when used with the current lot of Preciset TDM 1 Calibrator on the Roche/Hitachi analyzers.
- Trade Name
- A) PDS II Absorbable Sutures
B) Coated Vicryl Absorbable Sutures
- Recall Posting Date
- 2009/11/23
- Manufacturer
- Ethicon Inc.
- Recall Start Date
- 2009/10/26
- Recall Number
- 51428
- Hazard Classification
- Type II
- Model or Catalog #
- A) Z316H
B) J318H
- Lot or Serial #
- A) ACM008
B) ACM473
- Reason for Recall
- Specific lots of PDS II (polydioxanone) Absorbable Sutures are being recalled because some of these products may have a small void in the foil outer package. The void in the foil may lead to suture degradation and the potential for a compromise in the sterility of the product.
- Trade Name
- A) Powerhart G3 Automated External Defibrillator
B) Powerhart AED G3 Pro
C) Powerhart AED G3 Plus
D) Powerhart G3 Plus Automatic
E) Powerhart - AED G3 (Biphasic)
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Cardiac Science Corporation
- Recall Start Date
- 2009/10/08
- Recall Number
- 51818
- Hazard Classification
- Type II
- Model or Catalog #
- A) 9300A-501PPC
B) 9300P-601
C) 9390E-501
D) 9390A-501
E) 9300E-501
- Lot or Serial #
- A) 4230260ppc
B) 4226121
C) 4237595
D) 4183220
E) 4218441, 4153329, 4153683
- Reason for Recall
- Review of test records indicated that the seven impacted Powerhart G3 AEDS manufactured by Cardiac Science Corporation (CSC) may not have received adequate testing prior to being shipped from the factory.
- Trade Name
- A) PowerProMax Battery Modular Handpiece
B) PowerProMax Battery 2-Trigger/Full F HP
C) PowerProMax Oscilator Handpiece
D) PowerProMax Reciprocator Saw Handpiece
E) MPower Battery Two Trigger
F) MPower Battery Sternum Saw
G) MPower Battery Handpieces
H) PowerPro Battery Trigger Handpiece
Updated recall # 50403 previously posted on
2009-09-28.
- Recall Posting Date
- 2009/10/26
- Manufacturer
- Linvatec Corporation D.B.A. Conmed Linvatec
- Recall Start Date
- 2009/08/14
- Recall Number
- 50403
- Hazard Classification
- Type II
- Model or Catalog #
- A) PRO5100M
B) PRO5200M
C) PRO5300M
D) PRO5400M
E) PRO6202
F) PRO6450
G) PRO6200, PRO6300, PRO6400
H) PRO5200, PRO5100, PRO5300, PRO5400
- Lot or Serial #
- A) BBB97037, BBB97730, BBB97731, BBB97734, BBB98124, BBB98125, BBB99622, BBB99925, BBC07409, BBC07410
B) >10 Numbers, Contact Manufacturer
C) >10 Numbers, Contact Manufacturer
D) >10 Numbers, Contact Manufacturer
E) >10 Numbers, Contact Manufacturer
F) BBC98784, BBC98787, BBD08568
G) >100 Numbers, Contact Manufacturer
H) >10 Numbers, Contact Manufacturer
- Reason for Recall
- Update: The manufacturer now requests return of the affected devices. The manufacturer has determined there is a remote possibility that certain handpieces may self activate.
- Trade Name
- A) Renegade Fiber Braided Microcatheter
B) Renegade Hi-Flow Microtainer
- Recall Posting Date
- 2009/10/12
- Manufacturer
- Boston Scientific Cork Ltd.
- Recall Start Date
- 2009/08/26
- Recall Number
- 50811
- Hazard Classification
- Type II
- Model or Catalog #
- A) 18-251, 18-252, 18-253
B) 18-285, 18-286, 18-288, 18-289, 18-290, 18-291, 18-298, 18-299, 18-300, 18-302, 18-303
- Lot or Serial #
- A) >100 Numbers, Contact Manufacturer
B) >100 Numbers, Contact Manufacturer
- Reason for Recall
- Sterile barrier in the packaging of affected product may be compromised. Pouch seal (top seal) may be breached.
- Trade Name
- A) Screwplant Implants
B) Screwplant Healing Collars
C) Legacy Healing Collars
D) Screwindirect Implants
E) Screwplus Implants
F) Screwdirect Implants
G) Replant Implants
H) Replus Implants
Updated recall # 50598 previously posted on 2009-09-14.
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Implant Direct LLC
- Recall Start Date
- 2009/08/18
- Recall Number
- 50598
- Hazard Classification
- Type II
- Model or Catalog #
- A) N/A
B) N/A
C) N/A
D) N/A
E) N/A
F) N/A
G) N/A
H) N/A
- Lot or Serial #
- A) > 100 Numbers, Contact Manufacturer
B) > 10 Numbers, Contact Manufacturer
C) > 10 Numbers, Contact Manufacturer
D) > 10 Numbers, Contact Manufacturer
E) > 10 Numbers, Contact Manufacturer
F) > 10 Numbers, Contact Manufacturer
G) > 10 Numbers, Contact Manufacturer
H) > 10 Numbers, Contact Manufacturer
- Reason for Recall
- Cover caps exhibited visible cracks post sterilization. Testing indicates the integrity of the packaging was maintained by an inner o-ring seal. The date range of distribution for the affected products is 2007/04/26 to 2007/08/01.
- Trade Name
- A) SDC HD Capture Recorder
B) SDC HD Capture Device
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Stryker Endoscopy
- Recall Start Date
- 2009/10/30
- Recall Number
- 51583
- Hazard Classification
- Type II
- Model or Catalog #
- A) 240-050-888
B) 240-050-890
- Lot or Serial #
- A) All Lot Numbers
B) All Lot Numbers
- Reason for Recall
- Stryker Endoscopy was aware that the SDC HD picture/video capture device would not be able to assign the correct time/date stamp for saved files after December 31st 2009. On January 1st 2010, the system date for all SDC HD units would re-start from October 1st 2001 and begin incrementing from that point onwards. This event would not affect how pictures or videos were captured using SDC HD during surgery, however the system date would be incorrect and users would not be able to correct this issue without a software upgrade.
- Trade Name
- A) Signa 1.5T Excite HD - Gradients
B) 3.0T Signa HDX MR System - Main Unit
C) 1.5T Signa HDX MR System - Hardware
D) Discovery MR750 MR System - Main Unit
E) 1.5T Signa Infinity Twinspeed Excite
F) Signa 1.5T Infinity Excite MR System
G) Signa Excite 3.0T Electronics
H) Signa HDXT 1.5T MR System - Main Unit
- Recall Posting Date
- 2009/12/07
- Manufacturer
- GE Medical Systems LLC
- Recall Start Date
- 2009/11/05
- Recall Number
- 50777
- Hazard Classification
- Type II
- Model or Catalog #
- A) M3000LF
B) M3335JK
C) M3335PB
D) M7000BF
E) M3000TD
F) 2294302
G) M3333TA
H) SIGNA HDXT 1.5T MR SYSTEM
- Lot or Serial #
- A) HDMR10368, 7547YR3, 278606MR9, 266072MR8, 274168MR4, 270572MR1
B) 7555YR6
C) >10 Numbers, Contact Manufacturer
D) 291192MR3, 290992MR7
E) >10 Numbers, Contact Manufacturer
F) 243409MR0, 239046MR6, 253630MR8, 239043MR3, 256216MR3, 246610MR0, 258415MR9, 247728MR9, 242181MR6
G) 6690YR2, 266690MR7
H) 30BD4960
- Reason for Recall
- Use of CADstream with "pure" software for Dynamic MRI imaging may require a modification to the study preference settings to present colour overlays equivalent to non-pure studies.
- Trade Name
- A) TCD Drainage Catheters
B) SKATER Single Step Drainage Set
- Recall Posting Date
- 2009/11/09
- Manufacturer
- PBN Medical Denmark A/S
- Recall Start Date
- 2009/09/22
- Recall Number
- 51091
- Hazard Classification
- Type II
- Model or Catalog #
- A) 716084020
B) 756007025, 756008025, 756010025, 756012025, 756506025, 756507025, 756508020, 756512020
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
- Reason for Recall
- Packaging integrity may be compromised as a result of transportation and distribution of the products.
- Trade Name
- A) Tegaderm Hydrocolloid Dressing
B) Medipore Soft Cloth Surgical Tape
- Recall Posting Date
- 2009/12/21
- Manufacturer
- 3M Health Care
- Recall Start Date
- 2009/11/23
- Recall Number
- 51873
- Hazard Classification
- Type III
- Model or Catalog #
- A) 90002
B) 2961
- Lot or Serial #
- A) 2012-06 AA
B) 2012-10 XT
- Reason for Recall
- The master carton and the individual dressing packages are labelled correctly, however the boxes that contain the individual dressing packages were labelled for Tegaderm Hydrocolloid thin dressing, catalogue # 90022.
- Trade Name
- A) Therakair Visio
B) InfoV.A.C. Therapy Unit
C) First Step All in One
D) ActiV.A.C. Therapy Unit
E) First Step Select - Pump
F) Kinair IV
G) V.A.C. ATS Therapy Unit
H) Proficare Pump Unit
I) Therapulse ATP
- Recall Posting Date
- 2009/12/07
- Manufacturer
- KCI USA, Inc.
- Recall Start Date
- 2009/11/09
- Recall Number
- 51634
- Hazard Classification
- Type II
- Model or Catalog #
- A) M6257763
B) M8260027
C) 227500
D) 340013
E) 215200
F) 216050
G) M8259994
H) M8259988
I) 227080.S, 227800
- Lot or Serial #
- A) N/A
B) N/A
C) N/A
D) N/A
E) N/A
F) N/A
G) N/A
H) N/A
I) N/A
- Reason for Recall
- Power cords may crack and fail leading to potential risks of electrical shock, delay in setup and therapy, interruption of therapy and fires which may also occur in an oxygen-rich environment.
- Trade Name
- A) Transport Stretcher
B) M-Series Stretcher
C) Trauma Stretcher
D) Eye Surgery Stretcher
- Recall Posting Date
- 2009/10/12
- Manufacturer
- Stryker Medical
- Recall Start Date
- 2009/08/31
- Recall Number
- 50938
- Hazard Classification
- Type II
- Model or Catalog #
- A) 0737-000-000, 0738-000-000
B) 1007-000-000 (SM104)
C) 1027-000-000
D) 1079-000-000
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) >100 Numbers, Contact Manufacturer
C) >10 Numbers, Contact Manufacturer
D) 0811100540, 0812101547, 0812102798, 0901034284, 0901034285
- Reason for Recall
- The stretchers have the potential for the bushing in the braking system to fracture, which could result in the brake/steer pedal seizing. If the pedal seizes, the braking system may become inoperable and/or steering around corners may become more difficult.
- Trade Name
- A) Triporteur Star 3
B) Quadriporteur Star 4
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Orthofab Inc.
- Recall Start Date
- 2009/09/22
- Recall Number
- 50934
- Hazard Classification
- Type II
- Model or Catalog #
- A) 180-32040-22
B) 180-32000-07, 180-32000-22
- Lot or Serial #
- A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
- Reason for Recall
- Lack of surface treatment on one of the gears of the transmission gearbox.
- Trade Name
- A) Unicel DXC600I Synch.Acc.Clin. Analyzer
B) Unicel DXC880I Synch.Acc.Clin. Analyzer
C) Unicel DXC660I Synch.Acc.Clin. Analyzer
D) Unicel DXC680I Synch.Acc.Clin. Analyzer
E) Unicel DXC860I Synch.Acc.Clin. Analyzer
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Beckman Coulter, Inc.
- Recall Start Date
- 2009/10/22
- Recall Number
- 51584
- Hazard Classification
- Type III
- Model or Catalog #
- A) 4767
B) 4768
C) 4771
D) 4772
E) 4773
- Lot or Serial #
- A) Cup Lot Numbers Starting With "09" Or Later
B) Cup Lot Numbers Starting With "09" Or Later
C) Cup Lot Numbers Starting With "09" Or Later
D) Cup Lot Numbers Starting With "09" Or Later
E) Cup Lot Numbers Starting With "09" Or Later
- Reason for Recall
- The access 1.0 ml insert cup, used on Unicel DXC Synchron Clinical Systems, might be the cause of some sample wheel motion errors when run using 13x100 sample tubes.
- Trade Name
- A) Unicel DXI600 Access Immunoassay System Class 2
B) Unicel DXI600 Access Immunoassay System Class 3
C) Unicel DXC880I Access Clinical Analyzer
D) Unicel DXI800 Access Analyzer Class 3
E) Unicel DXI800 Access System Analyzer Class 2
F) Unicel DXC660I Access Clinical Analyzer
G) Unicel DXC680I Access Clinical Analyzer
H) Unicel DXC860I Access Clinical Analyzer
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Beckman Coulter Inc.
- Recall Start Date
- 2009/10/26
- Recall Number
- 51422
- Hazard Classification
- Type II
- Model or Catalog #
- A) A30260, A71461
B) A71461
C) 4768
D) 973100, A71457
E) 973100, A71457
F) 4771
G) 4772
H) 4773
- Lot or Serial #
- A) N/A
B) N/A
C) N/A
D) N/A
E) N/A
F) N/A
G) N/A
H) N/A
- Reason for Recall
- A software issue was identified in system software 4.2 and 4.2.1, which may result in the incorrect handling of a wash arm dispense plate motion failure. Instead of generating a red 'warning' event and stopping sample processing, the system will post a yellow 'caution' event and continue to operate with the dispense plate in an elevated position until the instrument is reinitialized. Potentially affected results are not flagged.
- Trade Name
- A) VNS Therapy Demipulse Programmable Generator
B) VNS Therapy Demipulse Duo Generator
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Cyberonics Inc.
- Recall Start Date
- 2009/12/03
- Recall Number
- 52025
- Hazard Classification
- Type II
- Model or Catalog #
- A) N/A
B) N/A
- Lot or Serial #
- A) All Lot Numbers
B) All Lot Numbers
- Reason for Recall
- Enhanced labelling to prevent end of service (EOS) warning messages & reduced battery life resulting from electromagnetic induction (EMI), electrostatic discharge (ESD), or electro cautery exposure during surgery.
- Trade Name
- A) Voluson E8 Ultrasound System Main Unit
B) Voluson E8 Expert Ultrasound System
- Recall Posting Date
- 2009/11/09
- Manufacturer
- GE Medical Systems Kretztechnik GMBH & CO OHG
- Recall Start Date
- 2009/10/21
- Recall Number
- 51227
- Hazard Classification
- Type III
- Model or Catalog #
- A) H48651PT
B) H48661N
- Lot or Serial #
- A) S/N D00229
B) S/N D04842, D05589 & D50233
- Reason for Recall
- When using combination of 2D-compound resolution imaging (CRI) and a pulsed wave Doppler single gate (PW) with steered Doppler angle, the displayed velocity scale is incorrect, which could delay treatment.
- Trade Name
- Acuson S2000
- Recall Posting Date
- 2009/10/12
- Manufacturer
- Siemens Medical Solutions USA, Inc.
- Recall Start Date
- 2009/09/08
- Recall Number
- 50793
- Hazard Classification
- Type II
- Model or Catalog #
- 10041461
- Lot or Serial #
- 200252, 200288, 200334, 200926, 200979, 200995
- Reason for Recall
- There is an intermittent series of events in which a print or store event failure may cause the following to occur: A) The last image of a patient study can be stored with the next new patient study performed. B) If the first patient's last image had measurement results they will appear on the next patient's image. C) When a distance ratio or area ratio measurement is performed with L Testis Mass 2 measurement label, it appears in the report as R Testis Mass 2 in English only. D) When more than 4 measurements of the same label are performed they will not appear on-screen or on the printed report.
- Trade Name
- A-Dec Decade Plus 1221 Dental Chair
- Recall Posting Date
- 2009/11/23
- Manufacturer
- A-Dec, Inc.
- Recall Start Date
- 2009/04/06
- Recall Number
- 51399
- Hazard Classification
- Type III
- Model or Catalog #
- 1221
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- When the chair back was returned up to the 'patient entry / exit' position, a pinch existed between the moving chair back support and the stationary underside of the armrest.
- Trade Name
- AdvanSync Orthodontic Device
- Recall Posting Date
- 2009/10/26
- Manufacturer
- Ormco Corporation
- Recall Start Date
- 2009/10/05
- Recall Number
- 51125
- Hazard Classification
- Type III
- Model or Catalog #
- 600-4000
- Lot or Serial #
- 080901333, 080902356, 070981416, 060971299, 060976084, 080902357, 080901335, 070981417, 060971304, 060976079
- Reason for Recall
- Five complaints involving braze joint failures of AdvanSync Telescopic Rod Assemblies resulting in a recall.
- Trade Name
- Advantage, 2-Button Turbo Handpiece
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Linvatec Corporation D.B.A. Conmed Linvatec
- Recall Start Date
- 2009/11/09
- Recall Number
- 51507
- Hazard Classification
- Type II
- Model or Catalog #
- D9924
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- The manufacturer has determined there is a remote possibility that the device may self activate.
- Trade Name
- Advent 4.5mmd Platform Tapered Abutment
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Zimmer Dental Inc.
- Recall Start Date
- 2009/11/06
- Recall Number
- 51588
- Hazard Classification
- Type II
- Model or Catalog #
- AVACT
- Lot or Serial #
- 61233652, 61238956, 61245049, 61252577, 61256798, 61263053, 61314067, 61341224
- Reason for Recall
- Zimmer Dental has received field complaints reporting that during application of torque to the abutment, the upper portion of the abutment had fractured and separated from the lower shaft. The investigation has determined that the product is not manufactured to specification. The pilot broach for threading the internal threads of the implant is too deep and extends into the shaft of the screw.
- Trade Name
- Amplatz PTFE-Coated Guidewire
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Gyrus Acmi, Inc.
- Recall Start Date
- 2009/08/25
- Recall Number
- 50973
- Hazard Classification
- Type III
- Model or Catalog #
- 13BX
- Lot or Serial #
- 96158DB
- Reason for Recall
- Amplatz PTFE-Coated Guidewires, part number 13BX, sold as lot 96158DB, may have been mis-packaged with product from lot number 96159DB. In this case, the outer box is labelled with the errant 96158DB lot number while the inner pouches are correctly labelled as lot 96159DB.
- Trade Name
- Amplicor STD Specimen & Transport Kit
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Roche Molecular Systems, Inc.
- Recall Start Date
- 2009/11/27
- Recall Number
- 51858
- Hazard Classification
- Type II
- Model or Catalog #
- 20753920122
- Lot or Serial #
- K13333, K13334, K15491, K15492, 2036
- Reason for Recall
- Potential for the male swab to present a defectiveness where the cotton head of the swab may dislodge and/or unravel with minimal force. This could result in having cotton fibres remaining within the urethra during specimen collection of male patients.
- Trade Name
- Apexpro Telemetry System - Clinical Info.
- Recall Posting Date
- 2009/11/23
- Manufacturer
- GE Medical Systems Information Technologies
- Recall Start Date
- 2009/10/26
- Recall Number
- 48250
- Hazard Classification
- Type II
- Model or Catalog #
- 418779-010, APRSRVR16=XXX, CICV3APRO=XXX, CICV3APRO=XXX/XXX , CICV4APRO=XXX
- Lot or Serial #
- >100 Numbers, Contact Manufacturer
- Reason for Recall
- Operator's manual does not warn against the unintended use of Apexpro Telemetry to monitor SPO2 without concurrent ECG monitoring and does not identify correct alarms and messages during loss of SPO2 data condition.
- Trade Name
- ASP Automatic Endoscope Reprocessor
- Recall Posting Date
- 2009/11/23
- Manufacturer
- Advanced Sterilization Products,
A Division Of Ethicon
- Recall Start Date
- 2009/09/28
- Recall Number
- 51459
- Hazard Classification
- Type III
- Model or Catalog #
- 20300, 20301
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- Advanced Sterilization Products (ASP) is sending a letter to all Automated Endoscope Reprocessor (AER) customers to reinforce the importance of properly setting the disinfectant cycle time when using the ASP AER and to remind users to observe the illuminated temperature light during the cycle. ASP has received user reports about improper times set for each disinfectant cycle.
- Trade Name
- Auto D Dimer Kit CRS 126B
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Trinity Biotech PLC
- Recall Start Date
- 2009/08/19
- Recall Number
- 51650
- Hazard Classification
- Type II
- Model or Catalog #
- CRS126-B
- Lot or Serial #
- T065095, T082003, T119007
- Reason for Recall
- The response of the two highest points of the kit's calibration curve are the same and consequently very high D dimer values may be underestimated.
- Trade Name
- Bard Magnum Core Tissue Biopsy Needle
- Recall Posting Date
- 2009/11/23
- Manufacturer
- Bard Peripheral Vascular, Inc.
- Recall Start Date
- 2009/10/16
- Recall Number
- 51391
- Hazard Classification
- Type II
- Model or Catalog #
- MN1610, MN1613, MN1616, MN1620
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- Some of the needles may exhibit an exposed sample chamber due to extended stylet length from a loose hub connection, resulting in an inability to obtain the biopsy sample.
- Trade Name
- BCI 9015 4-Channel Analog Output Cable
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Smiths Medical PM, Inc.
- Recall Start Date
- 2009/10/19
- Recall Number
- 51331
- Hazard Classification
- Type III
- Model or Catalog #
- 9015
- Lot or Serial #
- W72407
- Reason for Recall
- When interfaced via A 9015 cable to an external data collection device, A 9004 monitor configured with CCAP 1.13 software will not communicate reliably through either channel 3 or 4. This issue only occurs when the monitor's analog output is set to external.
- Trade Name
- BD 30ml Luer-Lok Syringes
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Becton Dickinson Canada Inc.
- Recall Start Date
- 2009/11/26
- Recall Number
- 51921
- Hazard Classification
- Type II
- Model or Catalog #
- 309650
- Lot or Serial #
- 9175046
- Reason for Recall
- BD has determined that a small number of unit packages may have open seals which can adversely impact product package integrity/product sterility.
- Trade Name
- BHR Streamline Cup Introducers
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Smith & Nephew, Inc.
- Recall Start Date
- 2009/11/11
- Recall Number
- 51665
- Hazard Classification
- Type II
- Model or Catalog #
- 90128285
- Lot or Serial #
- S811270
- Reason for Recall
- The identified lot number may mechanically seize up, preventing release of the cable tightening mechanism, during surgery or before use. As a result, this could prolong surgery time. Also, if implanted, this could cause a possibility of malpositioning of the cup due to use of the device in a mechanically seized condition.
- Trade Name
- Bioflex 2200- 665-1850 Vasculitis Kit
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Bio-Rad Laboratories Diagnostics Group
- Recall Start Date
- 2009/11/06
- Recall Number
- 51677
- Hazard Classification
- Type II
- Model or Catalog #
- 6651850
- Lot or Serial #
- 950759, 959261, 961363
- Reason for Recall
- Bioplex 2200 vasculitis reagent kits could produce false results when analyzing diluted samples.
- Trade Name
- Bryan Cervical Disk System
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Medtronic Sofamor Danek USA, Inc.
- Recall Start Date
- 2009/10/09
- Recall Number
- 51184
- Hazard Classification
- Type II
- Model or Catalog #
- 6470114, 6470115, 6470116, 6470117,
6470118
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- There is a slight possibility of outer pouch failure during shipping, which could result in the introduction of an inner pouch having non-sterile outer surfaces into the sterile field. The innermost packaging has been shown to remain intact. Proper dispensing technique must be followed, placing the product, rather than the inner pouch, into the sterile field.
- Trade Name
- Buretrol IV Solution Adm.Sets Clearlink
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Baxter Healthcare Corporation
- Recall Start Date
- 2009/11/05
- Recall Number
- 51663
- Hazard Classification
- Type II
- Model or Catalog #
- 2C8864
- Lot or Serial #
- UR09E21238
- Reason for Recall
- Separation of tubing from Distal Clearlink Y-site could occur, which may result in blood loss, infection, interruption of or delay in the start of intravenous therapy.
- Trade Name
- Carto 3 EP Mapping System
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Biosense Webster (Israel) Ltd.
- Recall Start Date
- 2009/11/25
- Recall Number
- 51915
- Hazard Classification
- Type II
- Model or Catalog #
- FG540000
- Lot or Serial #
- 3257, 2246, 4203, 4166, 4184
- Reason for Recall
- Biosense Webster, Inc. has recently become aware of a potential hazard that may occur during the operation of the bloom DTU-215b cardiac stimulator (not a Johnson & Johnson device) in conjunction with RF generators and EP equipment, including the Carto 3 system. If RF energy and pacing stimuli are delivered simultaneously, across the same electrode pair, a DC voltage may be generated. This can cause micro-bubbles in the blood within the heart chamber.
- Trade Name
- CBC II Blood Conservation Kit
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Stryker Instruments
- Recall Start Date
- 2009/11/13
- Recall Number
- 51764
- Hazard Classification
- Type II
- Model or Catalog #
- 225-414-000
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- There is a potential that the package pouch in the CBC II Blood Conservation Kits may have a pin hole, a weakened seal, or an opened seal that could result in the product not being sterile.
- Trade Name
- Centricity PACS Workstation - Software
- Recall Posting Date
- 2009/12/07
- Manufacturer
- GE Healthcare IT
- Recall Start Date
- 2009/10/30
- Recall Number
- 51228
- Hazard Classification
- Type II
- Model or Catalog #
- K2020YS, K2030JN
- Lot or Serial #
- >100 Numbers, Contact Manufacturer
- Reason for Recall
- If Dicom node config. is set to accept, user may unintentionally add new images to an approved exam, under certain conditions images will be deleted from PACS or from dictated, transcribed or completed status exam.
- Trade Name
- Colleague Volumetric Infusion Pumps
Updated recall # 44735 previously posted on 2008-11-24.
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Baxter Healthcare Corporation
- Recall Start Date
- 2008/11/24
- Recall Number
- 44735
- Hazard Classification
- Type III
- Model or Catalog #
- 2M9163
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- Customers were erroneously provided with version 0719H0166 of the operator's manual instead of the correct version 0719I0166.
- Trade Name
- Complement Control Cells
- Recall Posting Date
- 2009/11/23
- Manufacturer
- Immucor Inc.
- Recall Start Date
- 2009/10/26
- Recall Number
- 51502
- Hazard Classification
- Type III
- Model or Catalog #
- N/A
- Lot or Serial #
- 32616
- Reason for Recall
- There is evidence of potential microbial contamination in some vials of complement control cells (lot # 32616), based on gram stain results.
- Trade Name
- Connection Bracket 1180.36AO
- Recall Posting Date
- 2009/10/26
- Manufacturer
- Maquet GMBH & Co. KG
- Recall Start Date
- 2009/09/30
- Recall Number
- 51053
- Hazard Classification
- Type II
- Model or Catalog #
- 1180.360AO
- Lot or Serial #
- N/A
- Reason for Recall
- The Connection Bracket show too much mechanical play at the operating room table interface for some procedures. The cause of this play, is that the hole in the lock washer, which is responsible for locking the accessory to the operating room table interface, is too large.
- Trade Name
- Contour SE Microspheres
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Boston Scientific Corporation
- Recall Start Date
- 2009/08/28
- Recall Number
- 50810
- Hazard Classification
- Type II
- Model or Catalog #
- M001761101, M001761122, M001761141, M001761162, M001761201, M001761222, M001761241, M001761262, M001761301, M001761322, M001761341, M001761362, M001761401, M001761422, M001761441, M001761462, M001761501, M001761522, M001761541, M001761562
- Lot or Serial #
- >100 Numbers, Contact Manufacturer
- Reason for Recall
- All affected product is packaged within a vial. Boston Scientific has determined that, for affected products, the sterile barrier in the packaging (pouch) that contains the vial may be breached.
- Trade Name
- CyberKnife Robotic Radiosurgery System
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Accuray, Inc.
- Recall Start Date
- 2009/06/22
- Recall Number
- 50252
- Hazard Classification
- Type II
- Model or Catalog #
- 28000
- Lot or Serial #
- C202
- Reason for Recall
- A healthcare facility reported their end-to-end targeting accuracy was out of specification. The position of one of the robotic manipulator axes differed from the position determined by the robotic controller. Error 103 (karel e-stop detected) and error 342 (deviation in absolute position value DSE - RDC) were displayed.
- Trade Name
- CyberKnife Robotic Radiosurgery System
Updated recall # 50252 previously posted on 2009-08-31.
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Accuray, Inc.
- Recall Start Date
- 2009/06/22
- Recall Number
- 50252
- Hazard Classification
- Type II
- Model or Catalog #
- 28000
- Lot or Serial #
- C202
- Reason for Recall
- A healthcare facility reported their end-to-end targeting accuracy was out of specification. The position of one of the robotic manipulator axes differed from the position determined by the robotic controller. Error 103 (karel e-stop detected) and error 342 (Deviation in absolute position value DSE - RDC) were displayed. Recalls 50114 and 50252 are identical. For future reference, refer to recall 50252.
- Trade Name
- Da Vinci S Surgical System, Model IS2000
- Recall Posting Date
- 2009/10/26
- Manufacturer
- Intuitive Surgical Inc.
- Recall Start Date
- 2009/07/01
- Recall Number
- 51026
- Hazard Classification
- Type III
- Model or Catalog #
- IS2000
- Lot or Serial #
- N/A
- Reason for Recall
- Intuitive Surgical Inc. observed an electronics subsystem failure that resulted in partial intermittent closure of the instrument grip (IS2000 system). A software coding error in P7 caused the error for the MTM grip.
- Trade Name
- Dade Hepzyme (B4240-10)
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Siemens Healthcare Diagnostics Products GMBH
- Recall Start Date
- 2009/12/02
- Recall Number
- 51994
- Hazard Classification
- Type III
- Model or Catalog #
- B4240-10
- Lot or Serial #
- 528752
- Reason for Recall
- Customer complaints regarding incorrect high APTT values that may be observed on heparin and heparin-free samples treated with lot 528752 of Dade Hepzyme when using lot 528752 of Dade Hepzyme with non-Siemens APTT reagents.
- Trade Name
- Diagnostic Ultrasound Transducers / Probe
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Toshiba Medical Systems Corporation
- Recall Start Date
- 2009/10/30
- Recall Number
- 51870
- Hazard Classification
- Type II
- Model or Catalog #
- PVT-681MV
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- In ultrasound systems using the Ultrasound Convex Transducer PVT-681MV, the volume image acquired in 'plane C' is reversed horizontally.
- Trade Name
- Dimension Vista System-Chem Calibrators
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Siemens Healthcare Diagnostics Inc.
- Recall Start Date
- 2009/10/19
- Recall Number
- 51373
- Hazard Classification
- Type III
- Model or Catalog #
- KC130
- Lot or Serial #
- 9ED058
- Reason for Recall
- Customer complaints regarding Dimension Vista Amon method. Siemens has confirmed a negative bias of up to 35% for level 1 quality control when calibrated with Chem 3 Calibrator lot 9ED058 vs. 9BD064.
- Trade Name
- Dimension Vista System-Urea Nitrogen Assay
- Recall Posting Date
- 2009/10/26
- Manufacturer
- Siemens Healthcare Diagnostics Inc.
- Recall Start Date
- 2009/09/28
- Recall Number
- 51016
- Hazard Classification
- Type III
- Model or Catalog #
- K1021
- Lot or Serial #
- 09187AA
- Reason for Recall
- Received customer complaints regarding inaccurate results with the Bun Flex Reagent cartridge lots 09154AE and 09187AA.The test results may be falsely depressed or elevated.
- Trade Name
- Dimension V-Lyte Integrated Multisensor
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Siemens Healthcare Diagnostics Inc.
- Recall Start Date
- 2009/10/29
- Recall Number
- 51421
- Hazard Classification
- Type III
- Model or Catalog #
- K800, K820, K825, K835
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- Received customer complaints regarding Dimension Vista V-Lyte. Siemens confirms that urine sodium results become elevated as urine specimens are run on a sensor.
- Trade Name
- Dupaque X-Ray Detectable Sponges
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Derma Sciences, Inc.
- Recall Start Date
- 2009/10/27
- Recall Number
- 51464
- Hazard Classification
- Type II
- Model or Catalog #
- 94675-5
- Lot or Serial #
- 4408
- Reason for Recall
- Received two complaints from a customer indicating that the sponge separated at the end during surgery and was left behind.
- Trade Name
- Duralite Pyramid Receiver Prox. Adaptor
- Recall Posting Date
- 2009/12/07
- Manufacturer
- The Ohio Willow Wood Company
- Recall Start Date
- 2009/11/09
- Recall Number
- 51740
- Hazard Classification
- Type II
- Model or Catalog #
- N/A
- Lot or Serial #
- All Lot Numbers
- Reason for Recall
- There is a potential for the adapter to crack or break while in use by active amputees.
- Trade Name
- Durom Acetabular Cup
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Zimmer GMBH
- Recall Start Date
- 2009/11/15
- Recall Number
- 51631
- Hazard Classification
- Type II
- Model or Catalog #
- 01.00214.044, 01.00214.046, 01.00214.048, 01.00214.050, 01.00214.052, 01.00214.054, 01.00214.056, 01.00214.058, 01.00214.060, 01.00214.062, 01.00214.064, 01.00214.066
- Lot or Serial #
- >1000 Numbers, Contact Manufacturer
- Reason for Recall
- There have been reports of revisions of Zimmer's Durom Acetabular Cup and the Metasul LDH (large diameter head) femoral head in certain European markets. Based on Zimmer's investigation, the most probable cause is using a surgical technique which differs from that prescribed in the surgical technique for the Durom Acetabular Cup. This field safety notice updates surgical technique instructions.
- Trade Name
- Endo Femoral Aimers
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Smith & Nephew, Inc., Endoscopy Division
- Recall Start Date
- 2009/09/22
- Recall Number
- 50942
- Hazard Classification
- Type I
- Model or Catalog #
- 72201714, 72201716
- Lot or Serial #
- For product code 72201714: (L) 50278012
For product code 72201716: (L)502780141 and (L)50278014
- Reason for Recall
- Specific product codes and lot numbers of the Endo Femoral Aimers may have been improperly marked indicating the wrong size and part number. The actual 4mm Aimer is marked as 6mm and the actual 6mm Aimer is marked as 4mm.
- Trade Name
- Endoclamp Aortic Catheter
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Edwards Lifesciences
- Recall Start Date
- 2009/10/01
- Recall Number
- 51118
- Hazard Classification
- Type II
- Model or Catalog #
- EC1001
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- Possibility of the balloon rupturing during use.
- Trade Name
- Engen Automated Sample Handling System
- Recall Posting Date
- 2009/10/26
- Manufacturer
- Thermo Fisher Scientific OY
- Recall Start Date
- 2009/10/06
- Recall Number
- 51117
- Hazard Classification
- Type I
- Model or Catalog #
- ENGEN
- Lot or Serial #
- JCT20Q3J
- Reason for Recall
- Ortho Clinical Diagnostics is aware of one confirmed report of misreported patient results with the Engen Laboratory Automation System. When used with TCAutomation software versions 2.8.2 or below, incorrect sample IDs may be electronically written to the radio frequency ID tags on the sample carriers.
- Trade Name
- Equalizer Balloon Catheter
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Boston Scientific Cork Ltd.
- Recall Start Date
- 2009/09/30
- Recall Number
- 51119
- Hazard Classification
- Type II
- Model or Catalog #
- M001171060, M001171080, M001171100, M001171120
- Lot or Serial #
- >100 Numbers, Contact Manufacturer
- Reason for Recall
- Sterile barrier in the packaging of the affected product may be compromised. It has been reported that the outer pouch seal may be breached.
- Trade Name
- Excor System-Connecting Set For Cannula
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Berlin Heart GMBH
- Recall Start Date
- 2009/10/02
- Recall Number
- 51703
- Hazard Classification
- Type I
- Model or Catalog #
- A06-009, A09-012, A12-016
- Lot or Serial #
- 526337
- Reason for Recall
- It was observed that a too short cut of the adapter cannula combined together with certain conditions of stress may have a effect of leakage of the cannula.
- Trade Name
- Express SD Premounted Stent System
- Recall Posting Date
- 2009/10/12
- Manufacturer
- Boston Scientific Corporation
- Recall Start Date
- 2009/04/14
- Recall Number
- 48011
- Hazard Classification
- Type III
- Model or Catalog #
- H74937911515900, H74937911614900, H74937911715900, H74937912519900, H74937912618150, H74937912618900
- Lot or Serial #
- 12045170, 12070363, 12073173, 12054585, 12054587, 12054580, 12082319, 12082341, 12045167, 12085875, 12089749, 12089752, 12129028
- Reason for Recall
- Packaging for the applicable devices may not include a written copy of the product directions for use.
- Trade Name
- EZManager Max Diabetes Management Software
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Animas Corporation
- Recall Start Date
- 2009/11/16
- Recall Number
- 51762
- Hazard Classification
- Type III
- Model or Catalog #
- 100-467-23, 100-467-50
- Lot or Serial #
- 2937776 for EZManager Max CD, French
2936068, 2938247, 2962490 for EZManager Max CD, English
- Reason for Recall
- Compact discs received for EZManager Max Diabetes Management Software packages were identified as being blank when used in a computer.
- Trade Name
- Galileo Echo Blood Bank Analyzer
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Immucor Inc.
- Recall Start Date
- 2009/11/06
- Recall Number
- 51706
- Hazard Classification
- Type I
- Model or Catalog #
- 87000
- Lot or Serial #
- N/A
- Reason for Recall
- Two lot numbers of reagents or strips are allowed to be tested while performing QC testing on the echo. The result is that the final lot tested will pass QC even though it did not undergo complete QC testing.
- Trade Name
- Galileo Echo Blood Bank Analyzer
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Immucor Inc.
- Recall Start Date
- 2009/11/06
- Recall Number
- 51708
- Hazard Classification
- Type I
- Model or Catalog #
- 87000
- Lot or Serial #
- N/A
- Reason for Recall
- Under some circumstances, a manually entered sample ID could be assigned to the wrong sample.
- Trade Name
- Galileo Echo Blood Bank Analyzer
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Immucor Inc.
- Recall Start Date
- 2009/11/25
- Recall Number
- 51956
- Hazard Classification
- Type II
- Model or Catalog #
- 87000
- Lot or Serial #
- N/A
- Reason for Recall
- The instrument may generate a negative well interpretation for capture-R ready-screen or capture-R ready-ID assays and subsequent visual interpretation of those reactions are weak positive or questionable (equivocal).
- Trade Name
- Gammacoat Plasma Renin Activity 123I RIA Kit
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Diasorin Inc./S.R.L.
- Recall Start Date
- 2009/10/19
- Recall Number
- 51278
- Hazard Classification
- Type II
- Model or Catalog #
- N/A
- Lot or Serial #
- 123574, 123612, 123620
- Reason for Recall
- The Renin Activity Control provided with the kit has an incorrect range printed on the label. Printed range on the label is 5.5 - 10.3 ng/ml/hr. The correct range is 4.0 -7.6 ng/ml/hr.
- Trade Name
- Getinge Steam Sterilizer
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Getinge Sourcing LLC.
- Recall Start Date
- 2009/09/01
- Recall Number
- 50976
- Hazard Classification
- Type II
- Model or Catalog #
- 422HC, 433HC, 522HC, 533HC, 633HC, 733HC, 833HC
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- RTD temperature sensor component was failing prematurely. Temperature control of the sterilization process is maintained through the use of the sensor used in the sterilizer jacket and chamber.
- Trade Name
- Giraffe & Panda T-Piece Resuscitation System
- Recall Posting Date
- 2009/10/26
- Manufacturer
- Ohmeda Medical,
A Division Of Datex Ohmeda Inc.
- Recall Start Date
- 2009/09/22
- Recall Number
- 51025
- Hazard Classification
- Type II
- Model or Catalog #
- M1091316, M1091335, M1091365
- Lot or Serial #
- N/A
- Reason for Recall
- 1. Moulding flash during manufacturing of T-piece at supplier caused out of specification aperture. 2. Severity in risk analysis control document (RAC) scored low leading to less strict supplier inspection and process control requirements.
- Trade Name
- Goldmann Applanation Tonometer Prism
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Haag-Streit, AG
- Recall Start Date
- 2009/10/13
- Recall Number
- 51262
- Hazard Classification
- Type III
- Model or Catalog #
- 1000855
- Lot or Serial #
- >100 Numbers, Contact Manufacturer
- Reason for Recall
- In Europe, Hagg-Streit received 3 reports concerning leaking Tonometer Prisms. A known reason of failing Tonometer Prisms is the use of disinfectants not approved and not released by Haag-Streit, mainly cool wipes. This might lead to irritation of cornea and in rare cases to reversible corneal erosions.
- Trade Name
- GT Obturators
- Recall Posting Date
- 2009/10/12
- Manufacturer
- Dentsply Tulsa Dental Specialties Trading As Dentsply Endodontics
- Recall Start Date
- 2009/09/25
- Recall Number
- 50931
- Hazard Classification
- Type III
- Model or Catalog #
- GTOB20042
- Lot or Serial #
- 12709123
- Reason for Recall
- The product, GT Obturators, GTOB20042, lot 012709123 was found to contain mixed instruments, size 20, taper .04 (20/.04) and size 20, taper .08 (20/.08) obturators.
- Trade Name
- Guider Softip Guiding Catheter
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Medi-Tech Boston Scientific Corporation
- Recall Start Date
- 2009/10/07
- Recall Number
- 51219
- Hazard Classification
- Type II
- Model or Catalog #
- 10042, 10043, 10044, 10046, 10047, 10048, 10050, 10051, 10052, 10062, 10063, 10064, 10142, 10143, 10144, 10146, 10147, 10148, 10150, 10151, 10152, 10162, 10163, 10164
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- It is believed that vast swings in environmental conditions can impact catheter products over time. Improper storage may cause polymer degradation in portions of the catheter. Use of degraded catheters can expose patients to potential adverse events including a delay in procedure time due to catheter exchange, vessel trauma and stroke. Boston Scientific Corp. is issuing a safety alert reminding users to follow directions for use.
- Trade Name
- Gynecare Morcellex Tissue Morcellator
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Ethicon Inc. Also Trading As Ethicon Women's
Health And Urology
- Recall Start Date
- 2009/10/09
- Recall Number
- 51327
- Hazard Classification
- Type II
- Model or Catalog #
- MX0100
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- Ethicon is conducting a voluntary product recall of all lots of Gynecare Morcellex Tissue Morcellator due to the potential for the duckbill seal to become dislodged, and a confirmed event in which the seal was retained in a patient.
- Trade Name
- Heartmate XVE And II LVAS Battery Clip
- Recall Posting Date
- 2009/11/23
- Manufacturer
- Thoratec Corporation
- Recall Start Date
- 2009/10/14
- Recall Number
- 51438
- Hazard Classification
- Type II
- Model or Catalog #
- N/A
- Lot or Serial #
- >100 Numbers, Contact Manufacturer
- Reason for Recall
- The threaded connector of the battery clip housing can become loose from the battery clip housing thereby creating a potential to interrupt pump function.
- Trade Name
- Heater Wire Adaptors
- Recall Posting Date
- 2009/10/12
- Manufacturer
- Fisher And Paykel Healthcare Ltd.
- Recall Start Date
- 2009/07/23
- Recall Number
- 50850
- Hazard Classification
- Type II
- Model or Catalog #
- 900MR800
- Lot or Serial #
- All Lot Numbers
- Reason for Recall
- Manufacturer is replacing these adaptors with new, more advanced product, which include surge protection hardware that will reduce the risk of breathing circuit heater wire malfunction.
- Trade Name
- Heater Wire Adaptors
Updated recall # 50850 previously posted on 2009-10-12.
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Fisher And Paykel Healthcare Ltd.
- Recall Start Date
- 2009/07/23
- Recall Number
- 50850
- Hazard Classification
- N/A
- Model or Catalog #
- N/A
- Lot or Serial #
- N/A
- Reason for Recall
- After further review it has been determined that this action by the manufacturer is not considered a recall under the definition of recall, but is considered a product withdrawal by the manufacturer.
- Trade Name
- Hintegra Talar Cutting Guides
- Recall Posting Date
- 2009/12/21
- Manufacturer
- New Deal SA
- Recall Start Date
- 2009/10/20
- Recall Number
- 51435
- Hazard Classification
- Type II
- Model or Catalog #
- Left Cutting Guide, Right Cutting Guide
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- The width of the defective Talar Cutting Guides is too large, thus the medio-lateral cuts are too large compared to the Talar trial implant dimensions.
- Trade Name
- Histopaque 1077-1
- Recall Posting Date
- 2009/10/12
- Manufacturer
- Sigma Diagnostics
- Recall Start Date
- 2009/09/21
- Recall Number
- 50914
- Hazard Classification
- Type III
- Model or Catalog #
- 1077-1
- Lot or Serial #
- 038K6170
- Reason for Recall
- Bacterial contamination (Pseudomonas mendocina) in the affected device.The device is a cell separation media and in the reported instance was used to isolate stem cells from bone marrow specimens. The isolated cells were placed into cell culture for further testing and contamination was evident after 14 days of culture. The laboratory confirmed the source of contamination was the device.
- Trade Name
- Hypergel 5 Gram
- Recall Posting Date
- 2009/10/12
- Manufacturer
- Molnlycke Health Care AB
- Recall Start Date
- 2009/09/22
- Recall Number
- 50961
- Hazard Classification
- Type III
- Model or Catalog #
- 360500
- Lot or Serial #
- 08D22-156-32913, 08D22-157
- Reason for Recall
- Hypergel 5 gram, product code 360500, lot numbers 08d22-156-32913 and 08d22-157, have been distributed with a lot number deficiency wherein the lot number listed on the packaging does not match the lot number on the individual piece.
- Trade Name
- i-CAT 3D Dental Imaging System
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Imaging Sciences International, Inc.
- Recall Start Date
- 2009/10/30
- Recall Number
- 51709
- Hazard Classification
- Type II
- Model or Catalog #
- 9140-0000-0000
- Lot or Serial #
- ICU070488, ICU070518, ICU070519
- Reason for Recall
- Fault in the brake system may result in gantry-overhead dropping quickly when the overhead is lowered. i-CAT classic containing electrical brake.
- Trade Name
- i-CAT 3D Dental Imaging System
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Imaging Sciences International, Inc.
- Recall Start Date
- 2009/10/30
- Recall Number
- 51944
- Hazard Classification
- Type II
- Model or Catalog #
- 9140-0000-0000
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- i-CAT classic units containing mechanical brake or containing dual brake system- fault in brake system may result in the gantry-overhead carriage moving below its limits.
- Trade Name
- Ilumavision Software Ver. 2.1.0
- Recall Posting Date
- 2009/10/12
- Manufacturer
- Imtec Imaging, LLC
- Recall Start Date
- 2009/08/03
- Recall Number
- 50774
- Hazard Classification
- Type II
- Model or Catalog #
- N/A
- Lot or Serial #
- Version 2.1.0
- Reason for Recall
- 1. Incorrect imaging resulting from 3rd party digital image (Dicom) incompatibility. 2. Potential incorrect patient identification on report. Both may result in inaccurate or inappropriate treatment.
- Trade Name
- Imager II Urology Torque Catheters
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Boston Scientific Corporation
- Recall Start Date
- 2009/11/03
- Recall Number
- 51536
- Hazard Classification
- Type II
- Model or Catalog #
- M0064003031
- Lot or Serial #
- 57271, 57616, 57269, 57613
- Reason for Recall
- The sterile barrier in the packaging (pouch) that contains the catheter may be breached. The breach of sterility could lead to a contamination of the device with subsequent transfer of infectious agents to the patient.
- Trade Name
- Impression Coping Closed Tray
- Recall Posting Date
- 2009/10/26
- Manufacturer
- Nobel Biocare AB
- Recall Start Date
- 2009/05/21
- Recall Number
- 48621
- Hazard Classification
- Type II
- Model or Catalog #
- N/A
- Lot or Serial #
- 687218
- Reason for Recall
- Nobel Biocare received several complaints from customers. After investigation it was determined that one batch of Impression Coping Closed Tray Nobelactive INT RP Q6x9mm was mis-packaged.
- Trade Name
- Innova 2121IQ Bi-Plane Cardio Imaging
- Recall Posting Date
- 2009/11/23
- Manufacturer
- GE Medical Systems SCS
- Recall Start Date
- 2009/10/30
- Recall Number
- 51374
- Hazard Classification
- Type II
- Model or Catalog #
- S18821AM
- Lot or Serial #
- S/N 553918BU4
- Reason for Recall
- Potential limit switch failure could occur due to collision between lateral plane of Innova positioner and other equipment, which may break x-ray and positioning switches and may not allow exam completion.
- Trade Name
- Insight Software Revision H,Bioview REV5.5.1
- Recall Posting Date
- 2009/11/23
- Manufacturer
- Sandhill Scientific Inc.
- Recall Start Date
- 2009/02/23
- Recall Number
- 47672
- Hazard Classification
- Type III
- Model or Catalog #
- S98-1925-2H
- Lot or Serial #
- All lots are affected
- Reason for Recall
- The Revision "H" software calculated esophageal amplitudes incorrectly when the user generated a report. There were no issues with the acquisition of the data. Users could still acquire data, however there was a requirement to install the revision 'I' software upgrade before reports were printed to ensure data was reported (printed) correctly.
- Trade Name
- Intertan Hex Screw
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Smith & Nephew, Inc.
- Recall Start Date
- 2009/10/15
- Recall Number
- 51245
- Hazard Classification
- Type III
- Model or Catalog #
- 71642220
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- This recall was initiated as the product description included the word 'captured' and this was not a 'captured' screw. This could cause intra operative confusion due to the mislabeling. The problem originated in 2007 and all internal product was corrected, however the product in the field was not corrected.
- Trade Name
- IW900 Series Infant Warmer
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Fisher And Paykel Healthcare Ltd.
- Recall Start Date
- 2009/09/07
- Recall Number
- 51309
- Hazard Classification
- Type III
- Model or Catalog #
- IW910, IW920, IW930, IW932, IW934, IW950,
IW980, WITH SUFFIX JEU
- Lot or Serial #
- All Infant Warmers in Canada.
- Reason for Recall
- Notifying users in Canada of updating cleaning instructions for the IW900 Series Infant Radiant Warmers. The inclusion of the recommended and non-recommended proprietary cleaning products supplements the current operating instructions and technical manual held by Canadian customers which already cautions against using solutions containing specifically named chemicals.
- Trade Name
- IW900 Series Infant Warmer
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Fisher And Paykel Healthcare Ltd.
- Recall Start Date
- 2009/09/07
- Recall Number
- 51383
- Hazard Classification
- Type III
- Model or Catalog #
- Series With Suffix > IW910, IW920, IW930, IW932, IW934, IW950, IW980
- Lot or Serial #
- All Infant Warmers in Canada.
- Reason for Recall
- Notifying users in Canada of an updated test procedure for the power-fail alarm of the IW900 Series Infant Radiant Warmer. The service manual already includes a requirement for an annual check of the power-fail alarm. The test procedure has been revised to check that the power-fail alarm operates for at least 12 minutes. The revised test procedure accounts for any unexpected degradation over time of the supercapacitor that operates the power-fail alarm.
- Trade Name
- IW900 Series Infant Warmer
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Fisher And Paykel Healthcare Ltd.
- Recall Start Date
- 2009/09/07
- Recall Number
- 51309
- Hazard Classification
- Type III
- Model or Catalog #
- IW910, IW920, IW930, IW932, IW934, IW950, IW980, WITH SUFFIX JEU
- Lot or Serial #
- All Infant Warmers in Canada.
- Reason for Recall
- Notifying users in Canada of updating cleaning instructions for the IW900 series infant radiant warmers. The inclusion of the recommended and non-recommended proprietary cleaning products supplements the current operating instructions and technical manual held by Canadian customers which already cautions against using solutions containing specifically named chemicals.
- Trade Name
- Koordinat M/MP
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Siemens AG
- Recall Start Date
- 2009/08/13
- Recall Number
- 50747
- Hazard Classification
- Type III
- Model or Catalog #
- 05294003, 06239003
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- If the swivel locking mechanism is not completely engaged into position, during patient transfer, the table may unintentionally swivel, could result in uncontrolled patient movement during transfer, patient falling to the floor.
- Trade Name
- Kryptonite Bone Cement
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Doctors Research Group Inc.
- Recall Start Date
- 2009/09/17
- Recall Number
- 51492
- Hazard Classification
- Type III
- Model or Catalog #
- KRYP-X-03, KRYP-X-10, KRYP-Z-03, KRYP-Z-10
- Lot or Serial #
- 05072009, 05272009, 06292009, 03102009, 07142008, 10232008, 10272008, 10312008, 10282009
- Reason for Recall
- As a result of increased viscosity of component "A" in some Kryptonite Bone Cement Kits, the end user may observe difficulty in discharging component "A" from the syringe that it is packaged in.
- Trade Name
- Kyphx HV-R Bone Cement
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Medtronic Spine LLC
- Recall Start Date
- 2009/11/09
- Recall Number
- 51643
- Hazard Classification
- Type II
- Model or Catalog #
- C01A, C01B
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- Medtronic Spine LLC is recalling specific lots of Kyphx HV-R bone cement following complaints regarding extended time to reach the proper viscosity (doughy state) and inconsistency or lack of homogeneity during cement preparation, as well as when waiting to deliver the bone cement with the delivery device (e.g. bone filler device).
- Trade Name
- LabPro V3.01 Panel Update-01
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Siemens Healthcare Diagnostics Inc.
- Recall Start Date
- 2009/11/30
- Recall Number
- 51937
- Hazard Classification
- Type III
- Model or Catalog #
- 9800-3771
- Lot or Serial #
- Software Version 3.01
- Reason for Recall
- Siemens Healthcare Diagnostics received customer complaints regarding QC isolates when tested on the walkaway instrument. Following the installation of LabPro 3.01 panel update -01 and utilization of the panel type POS combo 33 (PC33), an issue has been confirmed for QC isolates when tested on the walkaway instrument. This issue only applies to those that have installed the LP V3.01 panel update-01 (9800-3768 compact disk). QC issue affects combo type panels.
- Trade Name
- Leksell gamma Knife Perfexion
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Elekta Instruments AB
- Recall Start Date
- 2009/08/18
- Recall Number
- 51437
- Hazard Classification
- Type II
- Model or Catalog #
- 715000
- Lot or Serial #
- Serial Number 6005
- Reason for Recall
- When incorrectly used the frame adapter may become damaged which in worst case could make it possible to lock the adapter so it does not properly fixate the frame.
- Trade Name
- Leksell Stereotactic System - Fixation Post
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Elekta Instruments AB
- Recall Start Date
- 2009/10/02
- Recall Number
- 51522
- Hazard Classification
- Type II
- Model or Catalog #
- 1003935, 1003937
- Lot or Serial #
- N/A
- Reason for Recall
- Excessive heating of the tip of the fixation screws with the standard non-insulated posts may cause minor burning of the contact point of patients during certain higher energy MRI usage.
- Trade Name
- Lexicon K-File
- Recall Posting Date
- 2009/10/26
- Manufacturer
- Dentsply Tulsa Dental Specialties Trading As Dentsply Endodontics
- Recall Start Date
- 2009/09/29
- Recall Number
- 51017
- Hazard Classification
- Type III
- Model or Catalog #
- KF01025
- Lot or Serial #
- 811000105
- Reason for Recall
- The product is labelled as 25 mm in length but the files are actually 31 mm in length.
- Trade Name
- Manual Pulmonary Resuscitator (MPR)
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Unomedical Inc.
- Recall Start Date
- 2009/10/08
- Recall Number
- 51151
- Hazard Classification
- Type I
- Model or Catalog #
- 724-E
- Lot or Serial #
- 03-25
- Reason for Recall
- On some units of the Manual Pulmonary Resuscitator (MPR), the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this were to occur, the MPR may loose its ability to generate the positive pressure necessary to function properly.
- Trade Name
- Maquet Hanaulux Model Blue 130
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Maquet S.A.
- Recall Start Date
- 2009/09/22
- Recall Number
- 50980
- Hazard Classification
- Type III
- Model or Catalog #
- 56077943
- Lot or Serial #
- 00001717, 00001719
- Reason for Recall
- There is potential for a crack on the front pivot of the Acrobat 2000 spring arm installed with a Blue 130 series surgical light that could cause the surgical light to swing freely. Defect rate is 0.35%.
- Trade Name
- MC5057 Universal Cable
Updated recall # 50401 previously posted on
2009-09-14.
- Recall Posting Date
- 2009/10/26
- Manufacturer
- Linvatec Corporation D.B.A. Conmed Linvatec
- Recall Start Date
- 2009/08/14
- Recall Number
- 50401
- Hazard Classification
- Type II
- Model or Catalog #
- 537702
- Lot or Serial #
- N/A
- Reason for Recall
- Update: The manufacturer now requests return of the affected devices. A sufficiently worn/damaged Universal Cable MC5057, used with Conmed Linvatec Electric Handpieces may cause the handpiece to self activate.
- Trade Name
- Medfusion Syringe Infusion Pumps
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Smiths Medical MD Inc.
- Recall Start Date
- 2009/11/16
- Recall Number
- 51668
- Hazard Classification
- Type II
- Model or Catalog #
- 3010A, 3500-306, 3500-415, 3500-500
- Lot or Serial #
- >100 Numbers, Contact Manufacturer
- Reason for Recall
- Smiths Medical has become aware of an increased trend in reports of "motor not running" ("MNR") and "motor rate error" ("MRE") alarm message events with certain Medfusion Syringe Infusion Pumps, models 3010a and 3500 ("pump"). These are high priority alarm messages which can occur for a variety of reasons as part of the usual self-test functionality of these pumps. The pump stops delivery when an MNR/MRE alarm message event occurs.
- Trade Name
- Medical Grade Flat Panel Display
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Spacelabs Medical, Inc.
- Recall Start Date
- 2009/10/30
- Recall Number
- 51662
- Hazard Classification
- Type II
- Model or Catalog #
- 94260-19
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- A potential safety hazard associated with the use of Spacelabs Medical 19" Flat Panel Displays, model 94260-19. There is a possibility that the internal power distribution circuit board will fail prematurely, causing the display to go blank. Audio alarms will continue to function normally.
- Trade Name
- Midas Touch Needle Electrode
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Olsen Medical
- Recall Start Date
- 2009/09/24
- Recall Number
- 51204
- Hazard Classification
- Type II
- Model or Catalog #
- 30-1014
- Lot or Serial #
- 011668, 011970
- Reason for Recall
- The recall was initiated due to recent tests that indicated some Midas Touch Needle Electrodes with modified tip might have an insulation breach and could generate a risk of superficial burn to the patient and/or end-user.
- Trade Name
- Model 7600 - Oximeter Pod
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Nonin Medical Inc.
- Recall Start Date
- 2009/10/23
- Recall Number
- 51731
- Hazard Classification
- Type II
- Model or Catalog #
- Model 7600
- Lot or Serial #
- 500849356, 500849358, 500849360, 500549361, 500823174, 500857064, 500874926
- Reason for Recall
- A malfunction with one of the components that may result in the overheating of the Oximeter pod has the potential to cause a burn if the pod is in direct contact with the patient.
- Trade Name
- MRSASelect Referance 63747 Coffret 20 Plaque
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Bio-Rad
- Recall Start Date
- 2009/10/23
- Recall Number
- 51525
- Hazard Classification
- Type II
- Model or Catalog #
- 63747
- Lot or Serial #
- 9H2414, 9H2417, 9H2425, 9J2429, 9J2432, 9J2435, 9J2437
- Reason for Recall
- Decrease of medium selectivity for methicillin-sensitive Saphyloccus aureus strains before mediums expiration date.
- Trade Name
- neoBLUE mini LED Phototherapy System
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Natus Medical Inc.
- Recall Start Date
- 2009/10/01
- Recall Number
- 51237
- Hazard Classification
- Type II
- Model or Catalog #
- 10101
- Lot or Serial #
- 60001-62656 serial number
- Reason for Recall
- There has been a trend observed in failure of the pole mount clamp shipped with a neoBlue mini LED Phototherapy Systems. When tightened with excessive force, some of these clamps have shown stress fractures or broken. The pole mount clamp has been redesigned by the manufacturer and customers of record are being advised to inspect their systems and to order a replacement clamp if necessary.
- Trade Name
- Neptune 2 Ultra Rover
- Recall Posting Date
- 2009/10/26
- Manufacturer
- Stryker Instruments
- Recall Start Date
- 2009/10/06
- Recall Number
- 51120
- Hazard Classification
- Type II
- Model or Catalog #
- 0702-001-000
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- The company has become aware that there is the potential that over time, a buildup of surgical waste on the receptacle bearing surfaces may increase the force required to rotate the suction manifold. If the force becomes too high the manifold cannot be rotated - therefore it cannot be inserted or removed. Once this occurs there will be no vacuum applied to the suction manifold so the primary function of the affected rover canister is lost.
- Trade Name
- OneTouch SureStep Test Strips
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Lifescan Inc.
- Recall Start Date
- 2009/09/24
- Recall Number
- 51926
- Hazard Classification
- Type III
- Model or Catalog #
- Part # 010-665
- Lot or Serial #
- 2749789
- Reason for Recall
- Expired OneTouch SureStep Test Strips (lot#274789) were shipped to four Canadian wholesalers.
- Trade Name
- Oxidase Touch Sticks
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Oxoid Limited
- Recall Start Date
- 2009/07/07
- Recall Number
- 51737
- Hazard Classification
- Type II
- Model or Catalog #
- BR0064A
- Lot or Serial #
- 702079
- Reason for Recall
- May produce weak positive or false negative results.
- Trade Name
- Periarticular Locking Plate System
- Recall Posting Date
- 2009/11/23
- Manufacturer
- Zimmer Inc.
- Recall Start Date
- 2009/10/20
- Recall Number
- 51408
- Hazard Classification
- Type III
- Model or Catalog #
- 00-2357-009-06
- Lot or Serial #
- 60594607
- Reason for Recall
- This action is being conducted because the part number on the locking plate (Zimmer Periarticular Locking Plate System distal lateral tibial plate) is incorrect. The part was etched as '2357-09-07', when it should have been etched '2357-09-06'.
- Trade Name
- Power Cord Component Of Sesmoid Plasty
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Zimmer CAS
- Recall Start Date
- 2009/11/05
- Recall Number
- 51683
- Hazard Classification
- Type II
- Model or Catalog #
- 500.923 SUB COMP 200.301
- Lot or Serial #
- SP-122, SP-197
- Reason for Recall
- 1. Female receptacle end: the terminal contact are not sufficiently recessed and protected from access. 2. The cable portion is not certified North America. 3. The male plug that plug to the power supply mains is missing a label.
- Trade Name
- pRK(Rabbit Kidney)
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Diagnostic Hybrids, Inc.
- Recall Start Date
- 2009/01/21
- Recall Number
- 51217
- Hazard Classification
- Type III
- Model or Catalog #
- 48-0600
- Lot or Serial #
- 480109
- Reason for Recall
- The Rabbit Kidney Culture(pRK), which consisted of a monolayer culture of rabbit kidney cells used to grow viruses from a patient specimen in order to detect or identify the virus, appeared to also contain another type of cells.
- Trade Name
- Pronto-Short Extraction Catheter
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Vascular Solutions, Inc.
- Recall Start Date
- 2009/11/20
- Recall Number
- 51819
- Hazard Classification
- Type II
- Model or Catalog #
- N/A
- Lot or Serial #
- Lot 544202, Exp. 11-2009, Lot 545693, Exp. 05-2010, Lot 546614, Exp. 09-2010
- Reason for Recall
- The expiration date of the sterile pouched guidewire was not considered when expiration dating was calculated for the finished device. Cannot assure sterility or function of the guidewire beyond its expiration date.
- Trade Name
- Quantiferon-TB Gold In-Tube
- Recall Posting Date
- 2009/10/12
- Manufacturer
- Cellestis Ltd.
- Recall Start Date
- 2009/09/01
- Recall Number
- 50725
- Hazard Classification
- Type III
- Model or Catalog #
- N/A
- Lot or Serial #
- A090504Z
- Reason for Recall
- It was observed that there were a higher than expected number of false positive results with lot A090504Z and the presence of higher levels of endotoxins was found.
- Trade Name
- Rapid Eric Electronic Compendium
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Remel Inc.
- Recall Start Date
- 2009/10/30
- Recall Number
- 51669
- Hazard Classification
- Type I
- Model or Catalog #
- RE8323600
- Lot or Serial #
- 1096277
- Reason for Recall
- Error limited to the Rapid Yeast Plus Database regarding differentiation of C. albicans from C. dubliniensis potentially leading to false ID of isolate.
- Trade Name
- Recon Plate 8 X 9mm, 3.5mm
- Recall Posting Date
- 2009/11/23
- Manufacturer
- Smith & Nephew, Inc.
- Recall Start Date
- 2009/10/14
- Recall Number
- 51369
- Hazard Classification
- Type III
- Model or Catalog #
- 7114-0108
- Lot or Serial #
- L06MT04447, L08ET19936
- Reason for Recall
- Smith & Nephew was made aware that all parts were missing the laser etching as identification information.
- Trade Name
- Reflex Hybrid Screw Extractor Inner Shaft
- Recall Posting Date
- 2009/11/23
- Manufacturer
- Stryker Spine SAS
- Recall Start Date
- 2009/10/13
- Recall Number
- 51370
- Hazard Classification
- Type II
- Model or Catalog #
- 48511905B, 48511905D
- Lot or Serial #
- All lots
- Reason for Recall
- The Reflex Hybrid Screw Extractor Inner Shaft was reported to fail at the threaded tip and fracture in the head of the screw. There had been reports of the threaded tip of the inner shaft being left in the screw due to the surgeon being unable to remove the screw. There had also been reports of damage to the screw and/or the plate while attempting to remove the screw.
- Trade Name
- Removable Magnet Hires 90K
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Advanced Bionics, LLC
- Recall Start Date
- 2009/08/25
- Recall Number
- 51652
- Hazard Classification
- Type II
- Model or Catalog #
- N/A
- Lot or Serial #
- N/A
- Reason for Recall
- Advanced Bionics (AB) is notifying all surgeons, clinicians and patients that the removable magnet component of the HiRes 90k cochlear implant (model no. CI-1400-01 and CI-1400-02H) is susceptible to corrosion. The affected units should be monitored and replaced if necessary.
- Trade Name
- Remstar M Series Heated Humidifier
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Respironics Inc.
- Recall Start Date
- 2009/11/20
- Recall Number
- 50202
- Hazard Classification
- Type II
- Model or Catalog #
- 1049109, 1049110, 1049124, 10499110, 1051157, 1051158, R1049109, R1049110, U1049109, U1049110
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- Some humidifiers may fail to operate due to potential malfunction of the internal connection between the heated plate and the printed circuit board which results in a 'blinking blue light' on the humidifier control knob.
- Trade Name
- Renegade Fiber Braided Microcatheter
- Recall Posting Date
- 2009/10/12
- Manufacturer
- Boston Scientific Cork Ltd.
- Recall Start Date
- 2009/08/26
- Recall Number
- 50941
- Hazard Classification
- Type II
- Model or Catalog #
- 18-257, 18-259
- Lot or Serial #
- >100 Numbers, Contact Manufacturer
- Reason for Recall
- Sterile barrier in the packaging of affected product may be compromised. Pouch seal (top seal) may be breached.
- Trade Name
- RET-Search II
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Sysmex Reagents America
- Recall Start Date
- 2009/11/03
- Recall Number
- 51623
- Hazard Classification
- Type II
- Model or Catalog #
- RED-700A
- Lot or Serial #
- ZY9021
- Reason for Recall
- Several boxes of RET-Search II (Red-700A) manufactured on July 15, 2009 had an incorrect bar code label applied to the product box. The bar code label identified the product as stromatolyser-NR with the same lot number and expiration date. The box contains a reagent bottle and dye pouch, both of which were correctly labelled, and the contents of the bottle and dye pouch are of the proper chemical composition for RET-Search II and passed all in-process and final inspection test criteria.
- Trade Name
- Revlite
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Hoya Conbio
- Recall Start Date
- 2009/03/31
- Recall Number
- 49248
- Hazard Classification
- Type III
- Model or Catalog #
- N/A
- Lot or Serial #
- RL-1094
- Reason for Recall
- Advisory letter was sent to customers of Revlite System with Revision E1 software, declaring that if PTP keypad is pressed while the wavelength window is green, the system may not complete change to new wavelength as requested.
- Trade Name
- Revolution Centrifugal Blood Pump
- Recall Posting Date
- 2009/10/26
- Manufacturer
- Sorin Group Italia S.R.L.
- Recall Start Date
- 2009/09/23
- Recall Number
- 51015
- Hazard Classification
- Type II
- Model or Catalog #
- 50300700
- Lot or Serial #
- 0905220164, 0906240184
- Reason for Recall
- Sorin Group Italia has recently identified a weakness in the plastic packaging tray used for the Revolution Centrifugal Blood Pump that might lead the packaging tray to crack. Cracks in the packaging tray can compromise the product sterility.
- Trade Name
- Sarns Disposable Centrifugal Pump
- Recall Posting Date
- 2009/10/26
- Manufacturer
- Terumo Cardiovascular System Corporation
- Recall Start Date
- 2009/07/06
- Recall Number
- 51040
- Hazard Classification
- Type II
- Model or Catalog #
- 164275
- Lot or Serial #
- KN10, KP08, LA12, LC16, LE15
- Reason for Recall
- Terumo Cardiovascular Systems has received three reports of tubing disconnecting from the pump inlet of the Sarns Disposable Centrifugal Pump during cardiopulmonary bypass.
- Trade Name
- Sarns Disposable Centrifugal Pump
- Recall Posting Date
- 2009/10/26
- Manufacturer
- Termo Cardiovascular System Corporation
- Recall Start Date
- 2009/07/14
- Recall Number
- 51040
- Hazard Classification
- Type II
- Model or Catalog #
- 164275
- Lot or Serial #
- KN10, KP08, LA12, LC16, LE15
- Reason for Recall
- Tubing disconnecting from the pump inlet of the disposable centrifugal pump during cardiopulmonary bypass.
- Trade Name
- Sequencer For Mosaiq
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Impac Medical Systems Inc.
- Recall Start Date
- 2009/10/16
- Recall Number
- 51821
- Hazard Classification
- Type II
- Model or Catalog #
- N/A
- Lot or Serial #
- Mosaiq 1.40 (All Builds), Mosaiq 1.41 (All Builds), Mosaiq 1.50
(All Builds), 1.60 (Up To, Including T2SP4), 2.00 (All Builds)
- Reason for Recall
- Consolidated field sequencing (CFS) feature not working properly as X2 jaw moved when only the X1 jaw should have been moving.
- Trade Name
- Signa HDE 1.5T MRI System - Magnet
- Recall Posting Date
- 2009/12/07
- Manufacturer
- GE Healthcare Japan Corporation
- Recall Start Date
- 2009/11/05
- Recall Number
- 51766
- Hazard Classification
- Type II
- Model or Catalog #
- M3335SE
- Lot or Serial #
- HDMR2675, 9948YR1,10833YR2, 11827YR3, 8704YR9, 10685YR6, 7964YR0, 10023YR0
- Reason for Recall
- Use of CADstream with "pure" software for Dynamic MRI imaging may require a modification to the study preference settings to present colour overlays equivalent to non-pure studies.
- Trade Name
- Siregraph CF
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Siemens AG
- Recall Start Date
- 2009/10/08
- Recall Number
- 51376
- Hazard Classification
- Type III
- Model or Catalog #
- 04 98 717
- Lot or Serial #
- 1987, 3089
- Reason for Recall
- The potential hazard of fingers or hands getting caught exists when the patient holds on to the tomographic rod between the column/tube assembly and the spotfilm device.
- Trade Name
- Speedlink 5.5 Transverse Connector
- Recall Posting Date
- 2009/10/12
- Manufacturer
- Zimmer Spine Austin, Inc.
- Recall Start Date
- 2009/09/22
- Recall Number
- 50964
- Hazard Classification
- Type III
- Model or Catalog #
- 1302-55M
- Lot or Serial #
- 55QJ
- Reason for Recall
- This action is being conducted because a small quantity of titanium alloy was discovered to have a discrepancy from the alloying process at another manufacturer. Alloy segregation can affect material strength. It has been suggested that the metal used to manufacture parts from lot 55QJ may have a similar problem, though testing has not shown this.
- Trade Name
- Staphytect Plus/Dryspot, Test Kits
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Remel Europe Ltd.
- Recall Start Date
- 2009/07/09
- Recall Number
- 50862
- Hazard Classification
- Type II
- Model or Catalog #
- DR0100M, DR0850M
- Lot or Serial #
- J577910, J055610, J539110, J610610
- Reason for Recall
- A potential for false positive reactions with some non staphylococcus aureus strains.
- Trade Name
- Steinmann Pin
- Recall Posting Date
- 2009/11/23
- Manufacturer
- Zimmer Inc.
- Recall Start Date
- 2009/10/23
- Recall Number
- 51411
- Hazard Classification
- Type III
- Model or Catalog #
- 18700259
- Lot or Serial #
- 61211489
- Reason for Recall
- This action is being conducted because packages from the identified lot should contain style 5 (trocar-style three point) Steinmann pins. However, they actually contain style 6 (spade-type diamond point).
- Trade Name
- Steris Harmony LL500/700 Surgical Light
- Recall Posting Date
- 2009/10/12
- Manufacturer
- Steris Corporation
- Recall Start Date
- 2009/09/14
- Recall Number
- 50864
- Hazard Classification
- Type III
- Model or Catalog #
- YG18568
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- A welded seam at the joint of the spring-loaded arm of the surgical light may brake and the light head may become disengaged and fall.
- Trade Name
- Streamlab Analytical Workcell
- Recall Posting Date
- 2009/10/12
- Manufacturer
- Siemens Healthcare Diagnostics Products GMBH
- Recall Start Date
- 2009/09/09
- Recall Number
- 50779
- Hazard Classification
- Type III
- Model or Catalog #
- Workcell
- Lot or Serial #
- N/A
- Reason for Recall
- Customer complaints regarding Dimension Vista Systems connected to a Streamlab Analytical Workcell where it has been observed on rare occasions an intermittent low result.
- Trade Name
- Synex II System - Central Body
- Recall Posting Date
- 2009/10/12
- Manufacturer
- Synthes (Canada) Ltd.
- Recall Start Date
- 2009/09/23
- Recall Number
- 50981
- Hazard Classification
- Type II
- Model or Catalog #
- 04.808.001, 04.808.002, 04.808.003, 04.808.004, 04.808.005, 04.808.006, 04.808.007, 04.808.008, 04.808.009, 04.808.010, 04.808.011
- Lot or Serial #
- All Lot Numbers
- Reason for Recall
- Synthes (Canada), Ltd. was notified by their supplier, Synthes (USA), of six complaints reported outside the US associated with Synex II Implant (Central Body) which fully or partially collapsed in the original implant position after a period of six to fifteen months post operatively.
- Trade Name
- Synex II System - Central Body
Updated recall # 50981 previously posted on 2009-10-12.
- Recall Posting Date
- 2009/11/23
- Manufacturer
- Synthes (Canada) Ltd.
- Recall Start Date
- 2009/09/23
- Recall Number
- 50981
- Hazard Classification
- Type I
- Model or Catalog #
- 04.808.001, 04.808.002, 04.808.003, 04.808.004, 04.808.005, 04.808.006, 04.808.007, 04.808.008, 04.808.009, 04.808.010, 04.808.011
- Lot or Serial #
- All lots
- Reason for Recall
- Synthes (Canada), Ltd. was notified by their supplier, Synthes (USA), of six complaints reported outside the US associated with Synex II Implant (Central Body) which fully or partially collapsed in the original implant position after a period of six to fifteen months post operatively.
- Trade Name
- Syngo Dynamics C.S.A.N.-5.X,6.X,7.0
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Siemens Medical Solutions USA, Inc.
- Recall Start Date
- 2009/04/30
- Recall Number
- 48348
- Hazard Classification
- Type III
- Model or Catalog #
- 100 90 480
- Lot or Serial #
- 92241, 95243, 93477, 95071, 95248
- Reason for Recall
- To inform customers about a potential issue, when receiving the multiple PSV measurements from some ultrasound systems, the absolute value is not being used.
- Trade Name
- Syngo Imaging XS
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Siemens AG
- Recall Start Date
- 2009/10/13
- Recall Number
- 51220
- Hazard Classification
- Type III
- Model or Catalog #
- 104 96 279
- Lot or Serial #
- 1225, 1532, 1533
- Reason for Recall
- A potential of a single image of a series is displayed "distorted" after loading to a viewing application.
- Trade Name
- Syngo Imaging XS
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Siemens AG
- Recall Start Date
- 2009/10/13
- Recall Number
- 51221
- Hazard Classification
- Type III
- Model or Catalog #
- 075 020 29
- Lot or Serial #
- 14630
- Reason for Recall
- Potential loss of single image data of prior examinations when archived to network-attached storage (NAS) or hierarchical storage management (HSM).
- Trade Name
- Systeme PhD EIA/IFA (cat# 426-0246)
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Bio-Rad Laboratories Diagnostics Group
- Recall Start Date
- 2009/10/13
- Recall Number
- 51252
- Hazard Classification
- Type III
- Model or Catalog #
- 4226-0228, 426-0246
- Lot or Serial #
- N/A
- Reason for Recall
- Results passed on integrating lab systems may be erroneous when using competitive tests, if they are located outside of the standard range.
- Trade Name
- TECNIS 1-Piece Posterior Chamber IOL
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Advanced Medical Optics, Inc.
- Recall Start Date
- 2009/11/10
- Recall Number
- 51829
- Hazard Classification
- Type II
- Model or Catalog #
- ZCB00
- Lot or Serial #
- N/A
- Reason for Recall
- The TECNIS 1-Piece Intraocular Lenses (IOLS) are being recalled because the haptics of some TECNIS 1-Piece IOLS (model ZCB00) have adhered to the lens optic after insertion into the eye. This situation can lead to additional manipulation to separate the haptics from the optic during surgery which could lead to an injury.
- Trade Name
- Tg Gen.2 (Thyroglobulin) Assay
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Roche Diagnostics GMBH
- Recall Start Date
- 2009/11/12
- Recall Number
- 51712
- Hazard Classification
- Type II
- Model or Catalog #
- 5118921190
- Lot or Serial #
- 155166
- Reason for Recall
- The current lot #155166 of TG Gen.2 (Thyroglobulin) Assay shows a premature ageing effect which leads to a positive bias of approximately 1ng/ml TG when used in combination with all current lots # 151605, 152737, 155488 and 157297 of elecsys thyroglobulin calibrator. The manufacturer has determined that customers should not use it for this purpose anymore.
- Trade Name
- Therascreen K-Ras Mutation Kit
- Recall Posting Date
- 2009/11/23
- Manufacturer
- DXS Ltd.
- Recall Start Date
- 2009/09/23
- Recall Number
- 50937
- Hazard Classification
- Type III
- Model or Catalog #
- KR-41
- Lot or Serial #
- DK146-01, DK149-01
- Reason for Recall
- The mixed standard component of the Therascreen Mutation Kit may sporadically generate out-of-range high results when tested with the control assay.
- Trade Name
- TheraScreen K-Ras Mutation Kit
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Roche Diagnostics GMBH
- Recall Start Date
- 2009/09/23
- Recall Number
- 50937
- Hazard Classification
- Type III
- Model or Catalog #
- 5366259190
- Lot or Serial #
- DK146-01, DK149-01
- Reason for Recall
- The mixed standard component of the Therascreen Mutation may sporadically generate out-of-range high results when tested with the control assay.
- Trade Name
- Titan Anchor Accessory Kits
- Recall Posting Date
- 2009/11/23
- Manufacturer
- Medtronic Inc.
- Recall Start Date
- 2009/10/27
- Recall Number
- 51434
- Hazard Classification
- Type II
- Model or Catalog #
- 3550-39
- Lot or Serial #
- >100 Numbers, Contact Manufacturer
- Reason for Recall
- Medtronic has initiated a field action to notify managing and implanting physicians of Medtronic per cutaneous leads for spinal cord stimulation of the potential for separation of the titanium insert from the silicone portion of the Titan Anchor, model 3550-39. If this occurs it may result in lead migration and compromised therapy for the patient.
- Trade Name
- Triage Tox Drug Screen PN 94400
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Biosite Incorporated
- Recall Start Date
- 2009/05/28
- Recall Number
- 50030
- Hazard Classification
- Type III
- Model or Catalog #
- 94400
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- Biosite, Inc. initiates notice of correction for Triage Tox Drug Screen Kit, PN 94400. Boxes shipped after August 6, 2008 are misprinted on the right side of the box flap. The misprint incorrectly states "use EDTA whole blood or plasma sample only". The kit is intended for use with urine samples not blood or plasma.
- Trade Name
- Trinest Microplate Incubator And Shaker
- Recall Posting Date
- 2009/11/23
- Manufacturer
- Wallac OY
- Recall Start Date
- 2009/10/21
- Recall Number
- 51375
- Hazard Classification
- Type II
- Model or Catalog #
- 1296-0050
- Lot or Serial #
- 12960024, 12960028
- Reason for Recall
- The protective coating around certain electrical wires inside the Trinest Incubator/Shaker may be worn out due to contact with a round aluminum part rotating during the continuous shaking movement, which would result in a warning message and the instrument would shut down.
- Trade Name
- Triniclot APPT S, Cat. No. T1202
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Trinity Biotech PLC
- Recall Start Date
- 2009/04/23
- Recall Number
- 49895
- Hazard Classification
- Type II
- Model or Catalog #
- T1202
- Lot or Serial #
- S067009
- Reason for Recall
- Precision and open vial stability issues with this lot may result in increased or decreased APTT times across the range of the assay. The lot is not stable for the full 30 day period when opened and stored.
- Trade Name
- Triniclot APTT HS, Product Code T1204
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Trinity Biotech PLC
- Recall Start Date
- 2009/05/24
- Recall Number
- 49896
- Hazard Classification
- Type II
- Model or Catalog #
- T1204
- Lot or Serial #
- R324001
- Reason for Recall
- Precision issues with this lot may result in increased APTT times across the range of the assay.
- Trade Name
- Triniclot PT Excel, T1106
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Trinity Biotech PLC
- Recall Start Date
- 2009/08/06
- Recall Number
- 51651
- Hazard Classification
- Type III
- Model or Catalog #
- T1106
- Lot or Serial #
- S172057
- Reason for Recall
- Quality control materials not recovering within predefined ranges. Prothrombin times are prolonged in lot S172057 by about 1.2 seconds.
- Trade Name
- Twist Drill 1.1x50mm 3.5mm stop
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Walter Lorenz Surgical, Inc. Also trading as: Biomet Microfixation
- Recall Start Date
- 2009/11/18
- Recall Number
- 51857
- Hazard Classification
- Type III
- Model or Catalog #
- 01-7141
- Lot or Serial #
- 537407, 536306, 535933
- Reason for Recall
- Three manufacturing lots of Twist Drill 1.1 x 50mm, 3.5mm stop with J notch, part number 01-7141 were packaged and distributed without the instructions for use.
- Trade Name
- Uric Acid Assay
- Recall Posting Date
- 2009/10/12
- Manufacturer
- Randox Laboratories Ltd.
- Recall Start Date
- 2009/09/15
- Recall Number
- 50863
- Hazard Classification
- Type III
- Model or Catalog #
- UA233
- Lot or Serial #
- 154107
- Reason for Recall
- UA233, batch 154107, was incorrectly labelled with an expiry date of August 28, 2011. The correct expiry date is November 29, 2010.
- Trade Name
- Varis 4D Integrated Treatment Console
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Varian Medical Systems, Inc.
- Recall Start Date
- 2009/10/16
- Recall Number
- 51774
- Hazard Classification
- Type II
- Model or Catalog #
- 8.0.12, 8.0.23 SP2 , 8.1 SP2, 8.1.1.3, 8.1.1.4,
8.1.2, 8.1.3, 8.1.4, 8.3, 8.6, 8.6.1, 8.6.2,
8.8.0, H51
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- Malfunction when adding CBCT image. Present software has ability to acquire gantry angle when 'treatment approval enforced' and/or 'allow only couch parameter (vert, long, lat) editing & acquiring for approved plans' selected in treatment administration. Therapists can acquire gantry angle into plan, even without system rights to acquire gantry angle. Patient treated with wrong gantry angles one day.
- Trade Name
- Vertical Surgical Table
- Recall Posting Date
- 2009/12/07
- Manufacturer
- Stryker Communications
- Recall Start Date
- 2009/11/06
- Recall Number
- 51664
- Hazard Classification
- Type II
- Model or Catalog #
- 0788-100-000, 0788-100-001, 0788-100-0015
- Lot or Serial #
- All Lot Numbers
- Reason for Recall
- The table may experience involuntary movement as a result of fluid intrusion. This hazard can arise especially when the table is in the trendelenburg or anti-trendelenburg position allowing fluids, such as saline or water, to enter the table column.
- Trade Name
- Vitrax II (Sodium Hyaluronate)
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Advanced Medical Optics, Inc.
- Recall Start Date
- 2009/11/06
- Recall Number
- 51828
- Hazard Classification
- Type II
- Model or Catalog #
- VT465
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- The Vitrax II product is being recalled because during routine stability testing, it was discovered that a sample tested at the three month time point had a ph value of 7.70 which exceeded the ph specification range of 6.8 to 7.5 resulting in a lab investigation. The investigation revealed that some units from different lots post ETO sterilization have a ph greater than the upper ph limit of 7.5. The highest value was found to be 8.29.
- Trade Name
- Von Willebrand Reagent
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Siemens Healthcare Diagnostics Products GMBH
- Recall Start Date
- 2009/10/13
- Recall Number
- 51218
- Hazard Classification
- Type III
- Model or Catalog #
- OUBD23
- Lot or Serial #
- 537675, 537682
- Reason for Recall
- An incorrect sensitivity limit has been assigned which is reagent specific and could cause results to be erroneously elevated between 17% and 40%.
- Trade Name
- Wallach LL 100 Cryosurgery Device
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Cooper Surgical Inc. Also Trading As Wallach Surgical Devices
- Recall Start Date
- 2009/09/19
- Recall Number
- 51212
- Hazard Classification
- Type II
- Model or Catalog #
- REF-900000-900002, REF-900034, REF-900509-5
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- Under certain circumstances a restriction of the flow of carbon dioxide gas within the handle may result in a less than optimal temperature delivery to the treatment site.
- Trade Name
- Washer Disinfectors
- Recall Posting Date
- 2009/12/21
- Manufacturer
- Getinge Disinfection AB
- Recall Start Date
- 2009/04/23
- Recall Number
- 51707
- Hazard Classification
- Type II
- Model or Catalog #
- 9027, 9100, 9120, 9125, 9128
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- The ground wire of the supply transformer exhibited signs of high resistance in a complaint and may give high eddy currents.
- Trade Name
- XIO Radiation Treatment Planning System
- Recall Posting Date
- 2009/11/09
- Manufacturer
- Computerized Medical Systems, Inc. (DBA CMS, Inc.) The Elekta Group
- Recall Start Date
- 2009/10/13
- Recall Number
- 51241
- Hazard Classification
- Type II
- Model or Catalog #
- N/A
- Lot or Serial #
- Release 4.3.0 & Higher
- Reason for Recall
- If an auto port is created for a beam on focal, the plan sent to XIO and the beam edited, it is possible the graphic beam representation and dose display do not agree with the edited field on the text page.
- Trade Name
- XTEN, HLX 3004, HLX 3005
- Recall Posting Date
- 2009/10/26
- Manufacturer
- Maquet S.A.
- Recall Start Date
- 2009/09/24
- Recall Number
- 50977
- Hazard Classification
- Type III
- Model or Catalog #
- ARD 567801093, ARD 567801094, ARD 567910901, ARD 567910910
- Lot or Serial #
- >10 Numbers, Contact Manufacturer
- Reason for Recall
- Excessive strain may cause impairment of the interlocking between the arm joint and surgical light. Inappropriate handling may cause a crack on the front pivot of the Acrobat 2000 spring arm.