Health Canada
Symbol of the Government of Canada

Institutional links

Drugs and Health Products

Medical Device Recall Listings - Sorted by Trade Name

(From October 2009 - December 2009)

Trade Name
070 Neuron Delivery Catheter
Recall Posting Date
2009/11/09
Manufacturer
Penumbra Inc.
Recall Start Date
2009/09/17
Recall Number
51330
Hazard Classification
Type II
Model or Catalog #
N/A
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
Design modifications were made to the Neuron Delivery Catheter in March 2009 to improve the catheter kink resistance. Penumbra initiated voluntary field removal of the original catheter design from the marketplace.
Trade Name
A Locus High Resolution SSP Unitray
Recall Posting Date
2009/11/23
Manufacturer
Invitrogen Corporation
Recall Start Date
2009/07/24
Recall Number
51366
Hazard Classification
Type III
Model or Catalog #
47290-10, 47291-10
Lot or Serial #
1145687
Reason for Recall
Lane 23 in the 'A' high resolution SSP Unitray, lot and batch number identified, will produce a false positive band at the correct base pair size in the presence of A*0301 samples. This will affect typing results on samples that are Aa*0301 homozygote, since they could be erroneously reported as A*0301/A*1101.
Trade Name
A) Access Immunoassay System - Troponin
B) Unicel DXI600 Access Immunoassay System
C) Unicel DXI800 Access Immunoassay System
Recall Posting Date
2009/12/07
Manufacturer
Beckman Coulter Inc.
Recall Start Date
2009/10/23
Recall Number
51585
Hazard Classification
Type III
Model or Catalog #
A) A78803
B) A78803
C) A78803
Lot or Serial #
A) 915417
B) 915417
C) 915417
Reason for Recall
The reagents in wells three (RLC) and four (RLD) of the reagent pack were inadvertently switched during manufacturing. Lot number 915417 contains reagent RLD in well three and RLC in well four.
Trade Name
A) Accura Breast Localisation Needles
B) Hawkins II Breast Localisation Needles
C) Hawkins I Breast Localization Needles
D) Homer Mammalok Breast Localisation Needles
E) Manan Co-Axial Introducer Needles
F) Manan Introducer Sheath Needles
G) Manan Guidewire Introducer Needles
H) Manan Seldinger Needles
I) Manan Modified Potts/Cournand Needles
J) Manan Galactography Trays
K) Skater-Centesis Catheter
L) Chiba Style Manual
M) Franseen Style
N) Green Style
O) Spinal Style
P) Westcott Style
Q) Hawkins III Breast Localization Needles
R) Pro-Mag Disposable Biopsy Needles
S) Maxi-Cell Biopsy Needles
T) Ultracore Disposable Biopsy Needles
U) Lorad Needle Guide
Recall Posting Date
2009/12/07
Manufacturer
Medical Device Technologies, Inc. DBA Angiotech
Recall Start Date
2009/09/01
Recall Number
50935
Hazard Classification
Type II
Model or Catalog #
A) N/A
B) N/A
C) N/A
D) N/A
E) N/A
F) N/A
G) N/A
H) N/A
I) N/A
J) N/A
K) N/A
L) N/A
M) N/A
N) N/A
O) N/A
P) N/A
Q) N/A
R) N/A
S) N/A
T) N/A
U) N/A
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
C) >10 Numbers, Contact Manufacturer
D) >10 Numbers, Contact Manufacturer
E) >10 Numbers, Contact Manufacturer
F) >10 Numbers, Contact Manufacturer
G) >10 Numbers, Contact Manufacturer
H) >10 Numbers, Contact Manufacturer
I) >10 Numbers, Contact Manufacturer
J) >10 Numbers, Contact Manufacturer
K) >10 Numbers, Contact Manufacturer
L) >10 Numbers, Contact Manufacturer
M) >10 Numbers, Contact Manufacturer
N) >10 Numbers, Contact Manufacturer
O) >10 Numbers, Contact Manufacturer
P) >10 Numbers, Contact Manufacturer
Q) >10 Numbers, Contact Manufacturer
R) >10 Numbers, Contact Manufacturer
S) >10 Numbers, Contact Manufacturer
T) >10 Numbers, Contact Manufacturer
U) >10 Numbers, Contact Manufacturer
Reason for Recall
The complaints were reported as "run off" (non-contiguous) seals. A run off seal is a seal in which the Angiotech applied seal is mis-aligned.
Trade Name
A) Acuson Antares Software Version 4.0
B) Acuson Antares Main Unit/5.0 Upgrades
Recall Posting Date
2009/11/09
Manufacturer
Siemens Medical Solutions USA, Inc.
Recall Start Date
2009/10/08
Recall Number
51249
Hazard Classification
Type III
Model or Catalog #
A) 05936518
B) 10032746
Lot or Serial #
A) 112267, 10134
B) 114573, 113469
Reason for Recall
Potential software problem with Acuson and Sonoline Antares systems product version 5.0 with cardiac option installed below software version 200.0.079.
Trade Name
A) Antistreptolysin O Assay
B) Antistrepolysin O Assay
C) IGM Flex Assay
D) ASL Flex Reagent
Recall Posting Date
2009/10/26
Manufacturer
Siemens Healthcare Diagnostics Products GMBH
Recall Start Date
2009/09/24
Recall Number
50979
Hazard Classification
Type III
Model or Catalog #
A) K7058
B) K7022
C) 10445904
D) 10445888
Lot or Serial #
A) 07310MA, 08029MA, 08252MA, 08336MA
B) 07311MA, 08043MA, 09077MA, 07177MA
C) 07310MA, 08029MA, 08252MA, 08336MA
D) 07177MA, 07311MA, 08043MA, 09077MA
Reason for Recall
Under certain conditions the software falsely elevated the results particularly if the preceding serum sample has a higher volume and concentration of beta2-microglobulin, anti-strepolysin o, or immunoglobulin M analytes.
Trade Name
A) Artiste MV-Syngo RT Therapist Assist
B) Artiste MV-Syngo RT Therapist
Recall Posting Date
2009/12/07
Manufacturer
Siemens Medical Solutions USA, Inc. Oncology Care Systems
Recall Start Date
2009/11/05
Recall Number
51605
Hazard Classification
Type III
Model or Catalog #
A) 08162807
B) 08162815
Lot or Serial #
A) 10024
B) 10040
Reason for Recall
Two or more treatment plans with different ISOcenter positions are stored together in one course in Lantis, e.g. main treatment plan and boost plan. If the cone beam is acquired using calculation of table shift coordinates will be based on ISOcenter position. This will lead to an incorrect table shift.
Trade Name
A) Artiste Syngo RT Therapist Assist
B) Artiste Syngo RT Therapist
Recall Posting Date
2009/11/09
Manufacturer
Siemens Medical Solutions USA, Inc.
Oncology Care Systems
Recall Start Date
2009/09/23
Recall Number
51246
Hazard Classification
Type III
Model or Catalog #
A) 08162807
B) 08162815
Lot or Serial #
A) 10024
B) 10040
Reason for Recall
Fractions delivered on the Artiste might not be recorded in Lantis and subsequently patients might be mistreated.
Trade Name
A) AW Volumeshare
B) AW Volumeshare2
Recall Posting Date
2009/10/12
Manufacturer
GE Medical Systems SCS
Recall Start Date
2009/09/15
Recall Number
49927
Hazard Classification
Type II
Model or Catalog #
A) S30321ZK
B) 5183859-3, 5183859-4, 5183859-5
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) S/N'S A8177079, A9062007, 9093001, A7274009, M8281004, 8354008
Reason for Recall
Measurement values provided by Seno Advantage Workstations 1x. 2.0 & 2.1 may be incorrect when applied to magnification images acquired on non-GE digital mammography systems, which could lead to overestimate of size of breast lesion.
Trade Name
A) BD Q-Syte Extension Set
B) MPS Acadia Extension Set With Rightbore
C) MPS Acadia Specialty Tri-Extension Set
Recall Posting Date
2009/12/07
Manufacturer
Becton Dickinson Infusion Therapy Systems Inc.
Recall Start Date
2009/11/02
Recall Number
51661
Hazard Classification
Type I
Model or Catalog #
A) 385102
B) 385150, 385151
C) 385164
Lot or Serial #
A) 8308330
B) A1967, A2399
C) A1951
Reason for Recall
Product may function improperly. This may result in the risk of air bubbles leaking into the infusion system through the BD Q-Syte Luer access split septum. This may lead to complications should the air bubbles not be detected.
Trade Name
A) Calcar Planer Surgical Protocols
B) Accolade Surgical Protocol
Recall Posting Date
2009/10/12
Manufacturer
Howmedica Osteonics Corporation
Recall Start Date
2009/09/14
Recall Number
50823
Hazard Classification
Type II
Model or Catalog #
A) LSP42, LSP46, LSP47M, LSP48
B) LASST
Lot or Serial #
A) N/A
B) N/A
Reason for Recall
Stryker has identified a need to provide specific instructions for initiating power to the Calcar Planer prior to engagement with bone in the cutting edge, Omnifit HFX, Secur-Fit Max/Secur-Fit Plus Max, and Omnifit Eon Surgical Protocols, as well as to provide specific instructions for rasping dense bone and/or narrow intra medullary canals for the Accolade Surgical Protocol.
Trade Name
A) C-Series Clinac, Trilogy & Novalis Tx
B) Trilogy Radiotherapy Delivery System
C) Novalis TX Linear Accelerator
Recall Posting Date
2009/12/07
Manufacturer
Varian Medical Systems, Inc.
Recall Start Date
2009/10/08
Recall Number
51773
Hazard Classification
Type II
Model or Catalog #
A) H18, H27, H29
B) H14, H29
C) H14, H29
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
C) >10 Numbers, Contact Manufacturer
Reason for Recall
SRS treatments occurred with conical collimator accessory not inserted, or with conical collimator accessory correctly mounted, but with a primary collimator field size setting exceeding outside diameter of that conical collimator. With wrong field size, unintended radiation is given to patient outside the outer edge of cone.
Trade Name
A) Definium 8000 Digital Radiographic System
B) Proteus XR/A General Purpose Radiographic
C) Revolution XR/D Digital Radiographic System
Recall Posting Date
2009/10/12
Manufacturer
GE Healthcare
Recall Start Date
2009/09/21
Recall Number
47780
Hazard Classification
Type II
Model or Catalog #
A) 2259988-2, 5131070
B) 2259973-3, 2259976, 2259988, S3918KD/KE
C) 2351505, 5131070, S3918LD
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) >100 Numbers, Contact Manufacturer
C) >10 Numbers, Contact Manufacturer
Reason for Recall
Table locks may not actuate, causing tabletop to unexpectedly float in either the lateral axes or both the longitudinal and lateral axes, which could result in patient or operator injury.
Trade Name
A) Imager II Angiographic Catheters
B) Imager II Catheters BERN
C) Imager II Catheters SIM
D) Imager II Catheters Contra2
E) Imager II Catheters C
F) Imager II Catheters RC
G) Imager II Catheters RDC
H) Imager II Catheters Pigtail
I) Imager II Catheters Contralateral
Recall Posting Date
2009/12/21
Manufacturer
Boston Scientific Corporation
Recall Start Date
2009/11/03
Recall Number
51535
Hazard Classification
Type II
Model or Catalog #
A) 31-400, 31-402, 31-510, 31-516, 31-520, 31-602
B) 31-407, 31-408, 31-409, 31-410, 31-607, 31-609
C) 31-430, 31-631
D) 31-458, 31-459, 31-658
E) 31-466, 31-467, 31-468, 31-469, 31-669
F) 31-485, 31-486, 31-685
G) 31-488, 31-489, 31-491
H) 31-509, 31-513, 31-519
I) 31-525, 31-528, 31-531
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
C) >10 Numbers, Contact Manufacturer
D) >10 Numbers, Contact Manufacturer
E) >10 Numbers, Contact Manufacturer
F) >10 Numbers, Contact Manufacturer
G) >10 Numbers, Contact Manufacturer
H) >10 Numbers, Contact Manufacturer
I) >10 Numbers, Contact Manufacturer
Reason for Recall
The sterile barrier in the packaging (pouch) that contains the catheter may be breached. The breach of sterility could lead to a contamination of the device with subsequent transfer of infectious agents to the patient.
Trade Name
A) Immulite System-Reactive Protein (CRP) Assay
B) Immulite 2500 System-HIH Sensit CRP Assay
C) Immulite 1000 System-Cardiac HSCRP Assay
D) Immulite System-Cardiac HSCRP Assay
E) Immulite 2000 System-Cardiac HSCRP Assay
F) Immulite 2500 System-HSCRP Assay
G) Immulite 1000 System-High Sens CRP Assay
H) Immulite 3000 Systems-High Sens CRP Assay
I) Immulite System-High Sensitivity CRP Assay
J) Immulite 1000 System-C-Reactive Protein
Recall Posting Date
2009/10/26
Manufacturer
Siemens Healthcare Diagnostics Products Limited
Recall Start Date
2009/09/11
Recall Number
50797
Hazard Classification
Type II
Model or Catalog #
A) LCRCM
B) LCRCM
C) LCRCM
D) LCRCM
E) L2KCRP2
F) LCRCM
G) LCRCM
H) L2KCRP2
I) LCRCM
J) LCRCM
Lot or Serial #
A) 022L, 025L
B) 022L, 025L
C) 022L, 025L
D) 022L, 025L
E) 243
F) 022L, 025L
G) 022L, 025L
H) 243
I) 022L, 025L
J) 022L, 025L
Reason for Recall
Siemens Healthcare Diagnostics received customer complaints regarding instabilities of high sensitivity C-reactive protein adjusters and controls giving an average positive bias of 20% compared with results from previous lots.
Trade Name
A) Intravenous Infusion Accessory IVMED20
B) Intravenous Infusion Accessory IVMED60
Recall Posting Date
2009/12/07
Manufacturer
Medxl Inc.
Recall Start Date
2009/11/06
Recall Number
51696
Hazard Classification
Type II
Model or Catalog #
A) N/A
B) N/A
Lot or Serial #
A) N/A
B) N/A
Reason for Recall
A warning letter was sent to the distributor requesting that sales be suspended until a licence has been obtained.
Trade Name
A) Intro-Flex Introducer Sets
B) Percutaneous Sheath Introducer Kit
Recall Posting Date
2009/12/21
Manufacturer
Edwards Lifesciences LLC
Recall Start Date
2009/05/21
Recall Number
48857
Hazard Classification
Type II
Model or Catalog #
A) 1350BF85
B) I652BF85H, S5507BF85
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) 58511631, 58549300, 58606253, 58648812, 58496258, 58535112, 58604814, 58644423
Reason for Recall
Blue connector proximal to Tuohy-Borst hemostasis valve may fatigue, crack or separate, resulting in patency failure & potential blood loss where Intro-Flex Introducer secures a venous catheter.
Trade Name
A) Navigation Software Knee Zimmer
B) Brainlab Knee Essential Software
C) Brainlab Knee Unlimited Software
Recall Posting Date
2009/12/21
Manufacturer
Brainlab AG
Recall Start Date
2009/09/11
Recall Number
50954
Hazard Classification
Type III
Model or Catalog #
A) 22086-03D, 22086-03E
B) 21100-01(2.0), 21100-01C(2.1)
C) 70001A, 70002A(2.1)
Lot or Serial #
A) Version 2.1.1 & All Previous
B) Version 2.1.1 & All Previous
C) Version 2.1.1 & All Previous
Reason for Recall
Incorrect Zimmer implant data in Brainlab Software Knee leads to incorrect information for Zimmer Innex implants.
Trade Name
A) Nexgen Rotating Hinge Knee - Femoral
B) Nexgen Rotating Hinge Knee - Tibial
C) Nexgen Rotating Hinge Knee,Tibial Plate
D) Nexgen Rotating Hinge Knee - Femoral, B
Recall Posting Date
2009/12/07
Manufacturer
Zimmer Inc.
Recall Start Date
2009/11/06
Recall Number
51622
Hazard Classification
Type III
Model or Catalog #
A) 00-5880-013-01, 00-5880-014-01, 00-5880-014-02, 00-5880-015-01, 00-5880-015-02, 00-5880-016-01, 00-5880-016-02
B) 00-5880-002-00, 00-5880-003-00, 00-5880-004-00, 00-5880-005-00, 00-5880-006-00
C) 00-5880-001-02
D) 00-5880-012-01
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
C) 60433397, 60461749
D) 60186249 H5
Reason for Recall
This action is being conducted because these components are labelled with a 10 year expiry period when in fact these components expire 5 years after manufacture. Expiration dating refers to 'use by' dates. There is no risk to patients who receive these implants prior to January 2010, when devices from the affected lots begin to expire.
Trade Name
A) Online TDM Gentamicin for Roche/Hitachi
B) Online TDM Gentamicin (GENT2) for Cobas
Recall Posting Date
2009/11/09
Manufacturer
Roche Diagnostics GMBH
Recall Start Date
2009/10/07
Recall Number
51194
Hazard Classification
Type II
Model or Catalog #
A) 03800504190
B) 04490843190
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
Reason for Recall
The online TDM Gentamicin assays may generate falsely elevated control and patient results when used with the current lot of Preciset TDM 1 Calibrator on the Roche/Hitachi analyzers.
Trade Name
A) PDS II Absorbable Sutures
B) Coated Vicryl Absorbable Sutures
Recall Posting Date
2009/11/23
Manufacturer
Ethicon Inc.
Recall Start Date
2009/10/26
Recall Number
51428
Hazard Classification
Type II
Model or Catalog #
A) Z316H
B) J318H
Lot or Serial #
A) ACM008
B) ACM473
Reason for Recall
Specific lots of PDS II (polydioxanone) Absorbable Sutures are being recalled because some of these products may have a small void in the foil outer package. The void in the foil may lead to suture degradation and the potential for a compromise in the sterility of the product.
Trade Name
A) Powerhart G3 Automated External Defibrillator
B) Powerhart AED G3 Pro
C) Powerhart AED G3 Plus
D) Powerhart G3 Plus Automatic
E) Powerhart - AED G3 (Biphasic)
Recall Posting Date
2009/12/21
Manufacturer
Cardiac Science Corporation
Recall Start Date
2009/10/08
Recall Number
51818
Hazard Classification
Type II
Model or Catalog #
A) 9300A-501PPC
B) 9300P-601
C) 9390E-501
D) 9390A-501
E) 9300E-501
Lot or Serial #
A) 4230260ppc
B) 4226121
C) 4237595
D) 4183220
E) 4218441, 4153329, 4153683
Reason for Recall
Review of test records indicated that the seven impacted Powerhart G3 AEDS manufactured by Cardiac Science Corporation (CSC) may not have received adequate testing prior to being shipped from the factory.
Trade Name
A) PowerProMax Battery Modular Handpiece
B) PowerProMax Battery 2-Trigger/Full F HP
C) PowerProMax Oscilator Handpiece
D) PowerProMax Reciprocator Saw Handpiece
E) MPower Battery Two Trigger
F) MPower Battery Sternum Saw
G) MPower Battery Handpieces
H) PowerPro Battery Trigger Handpiece
Updated recall # 50403 previously posted on
2009-09-28.
Recall Posting Date
2009/10/26
Manufacturer
Linvatec Corporation D.B.A. Conmed Linvatec
Recall Start Date
2009/08/14
Recall Number
50403
Hazard Classification
Type II
Model or Catalog #
A) PRO5100M
B) PRO5200M
C) PRO5300M
D) PRO5400M
E) PRO6202
F) PRO6450
G) PRO6200, PRO6300, PRO6400
H) PRO5200, PRO5100, PRO5300, PRO5400
Lot or Serial #
A) BBB97037, BBB97730, BBB97731, BBB97734, BBB98124, BBB98125, BBB99622, BBB99925, BBC07409, BBC07410
B) >10 Numbers, Contact Manufacturer
C) >10 Numbers, Contact Manufacturer
D) >10 Numbers, Contact Manufacturer
E) >10 Numbers, Contact Manufacturer
F) BBC98784, BBC98787, BBD08568
G) >100 Numbers, Contact Manufacturer
H) >10 Numbers, Contact Manufacturer
Reason for Recall
Update: The manufacturer now requests return of the affected devices. The manufacturer has determined there is a remote possibility that certain handpieces may self activate.
Trade Name
A) Renegade Fiber Braided Microcatheter
B) Renegade Hi-Flow Microtainer
Recall Posting Date
2009/10/12
Manufacturer
Boston Scientific Cork Ltd.
Recall Start Date
2009/08/26
Recall Number
50811
Hazard Classification
Type II
Model or Catalog #
A) 18-251, 18-252, 18-253
B) 18-285, 18-286, 18-288, 18-289, 18-290, 18-291, 18-298, 18-299, 18-300, 18-302, 18-303
Lot or Serial #
A) >100 Numbers, Contact Manufacturer
B) >100 Numbers, Contact Manufacturer
Reason for Recall
Sterile barrier in the packaging of affected product may be compromised. Pouch seal (top seal) may be breached.
Trade Name
A) Screwplant Implants
B) Screwplant Healing Collars
C) Legacy Healing Collars
D) Screwindirect Implants
E) Screwplus Implants
F) Screwdirect Implants
G) Replant Implants
H) Replus Implants
Updated recall # 50598 previously posted on 2009-09-14.
Recall Posting Date
2009/11/09
Manufacturer
Implant Direct LLC
Recall Start Date
2009/08/18
Recall Number
50598
Hazard Classification
Type II
Model or Catalog #
A) N/A
B) N/A
C) N/A
D) N/A
E) N/A
F) N/A
G) N/A
H) N/A
Lot or Serial #
A) > 100 Numbers, Contact Manufacturer
B) > 10 Numbers, Contact Manufacturer
C) > 10 Numbers, Contact Manufacturer
D) > 10 Numbers, Contact Manufacturer
E) > 10 Numbers, Contact Manufacturer
F) > 10 Numbers, Contact Manufacturer
G) > 10 Numbers, Contact Manufacturer
H) > 10 Numbers, Contact Manufacturer
Reason for Recall
Cover caps exhibited visible cracks post sterilization. Testing indicates the integrity of the packaging was maintained by an inner o-ring seal. The date range of distribution for the affected products is 2007/04/26 to 2007/08/01.
Trade Name
A) SDC HD Capture Recorder
B) SDC HD Capture Device
Recall Posting Date
2009/12/07
Manufacturer
Stryker Endoscopy
Recall Start Date
2009/10/30
Recall Number
51583
Hazard Classification
Type II
Model or Catalog #
A) 240-050-888
B) 240-050-890
Lot or Serial #
A) All Lot Numbers
B) All Lot Numbers
Reason for Recall
Stryker Endoscopy was aware that the SDC HD picture/video capture device would not be able to assign the correct time/date stamp for saved files after December 31st 2009. On January 1st 2010, the system date for all SDC HD units would re-start from October 1st 2001 and begin incrementing from that point onwards. This event would not affect how pictures or videos were captured using SDC HD during surgery, however the system date would be incorrect and users would not be able to correct this issue without a software upgrade.
Trade Name
A) Signa 1.5T Excite HD - Gradients
B) 3.0T Signa HDX MR System - Main Unit
C) 1.5T Signa HDX MR System - Hardware
D) Discovery MR750 MR System - Main Unit
E) 1.5T Signa Infinity Twinspeed Excite
F) Signa 1.5T Infinity Excite MR System
G) Signa Excite 3.0T Electronics
H) Signa HDXT 1.5T MR System - Main Unit
Recall Posting Date
2009/12/07
Manufacturer
GE Medical Systems LLC
Recall Start Date
2009/11/05
Recall Number
50777
Hazard Classification
Type II
Model or Catalog #
A) M3000LF
B) M3335JK
C) M3335PB
D) M7000BF
E) M3000TD
F) 2294302
G) M3333TA
H) SIGNA HDXT 1.5T MR SYSTEM
Lot or Serial #
A) HDMR10368, 7547YR3, 278606MR9, 266072MR8, 274168MR4, 270572MR1
B) 7555YR6
C) >10 Numbers, Contact Manufacturer
D) 291192MR3, 290992MR7
E) >10 Numbers, Contact Manufacturer
F) 243409MR0, 239046MR6, 253630MR8, 239043MR3, 256216MR3, 246610MR0, 258415MR9, 247728MR9, 242181MR6
G) 6690YR2, 266690MR7
H) 30BD4960
Reason for Recall
Use of CADstream with "pure" software for Dynamic MRI imaging may require a modification to the study preference settings to present colour overlays equivalent to non-pure studies.
Trade Name
A) TCD Drainage Catheters
B) SKATER Single Step Drainage Set
Recall Posting Date
2009/11/09
Manufacturer
PBN Medical Denmark A/S
Recall Start Date
2009/09/22
Recall Number
51091
Hazard Classification
Type II
Model or Catalog #
A) 716084020
B) 756007025, 756008025, 756010025, 756012025, 756506025, 756507025, 756508020, 756512020
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
Reason for Recall
Packaging integrity may be compromised as a result of transportation and distribution of the products.
Trade Name
A) Tegaderm Hydrocolloid Dressing
B) Medipore Soft Cloth Surgical Tape
Recall Posting Date
2009/12/21
Manufacturer
3M Health Care
Recall Start Date
2009/11/23
Recall Number
51873
Hazard Classification
Type III
Model or Catalog #
A) 90002
B) 2961
Lot or Serial #
A) 2012-06 AA
B) 2012-10 XT
Reason for Recall
The master carton and the individual dressing packages are labelled correctly, however the boxes that contain the individual dressing packages were labelled for Tegaderm Hydrocolloid thin dressing, catalogue # 90022.
Trade Name
A) Therakair Visio
B) InfoV.A.C. Therapy Unit
C) First Step All in One
D) ActiV.A.C. Therapy Unit
E) First Step Select - Pump
F) Kinair IV
G) V.A.C. ATS Therapy Unit
H) Proficare Pump Unit
I) Therapulse ATP
Recall Posting Date
2009/12/07
Manufacturer
KCI USA, Inc.
Recall Start Date
2009/11/09
Recall Number
51634
Hazard Classification
Type II
Model or Catalog #
A) M6257763
B) M8260027
C) 227500
D) 340013
E) 215200
F) 216050
G) M8259994
H) M8259988
I) 227080.S, 227800
Lot or Serial #
A) N/A
B) N/A
C) N/A
D) N/A
E) N/A
F) N/A
G) N/A
H) N/A
I) N/A
Reason for Recall
Power cords may crack and fail leading to potential risks of electrical shock, delay in setup and therapy, interruption of therapy and fires which may also occur in an oxygen-rich environment.
Trade Name
A) Transport Stretcher
B) M-Series Stretcher
C) Trauma Stretcher
D) Eye Surgery Stretcher
Recall Posting Date
2009/10/12
Manufacturer
Stryker Medical
Recall Start Date
2009/08/31
Recall Number
50938
Hazard Classification
Type II
Model or Catalog #
A) 0737-000-000, 0738-000-000
B) 1007-000-000 (SM104)
C) 1027-000-000
D) 1079-000-000
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) >100 Numbers, Contact Manufacturer
C) >10 Numbers, Contact Manufacturer
D) 0811100540, 0812101547, 0812102798, 0901034284, 0901034285
Reason for Recall
The stretchers have the potential for the bushing in the braking system to fracture, which could result in the brake/steer pedal seizing. If the pedal seizes, the braking system may become inoperable and/or steering around corners may become more difficult.
Trade Name
A) Triporteur Star 3
B) Quadriporteur Star 4
Recall Posting Date
2009/11/09
Manufacturer
Orthofab Inc.
Recall Start Date
2009/09/22
Recall Number
50934
Hazard Classification
Type II
Model or Catalog #
A) 180-32040-22
B) 180-32000-07, 180-32000-22
Lot or Serial #
A) >10 Numbers, Contact Manufacturer
B) >10 Numbers, Contact Manufacturer
Reason for Recall
Lack of surface treatment on one of the gears of the transmission gearbox.
Trade Name
A) Unicel DXC600I Synch.Acc.Clin. Analyzer
B) Unicel DXC880I Synch.Acc.Clin. Analyzer
C) Unicel DXC660I Synch.Acc.Clin. Analyzer
D) Unicel DXC680I Synch.Acc.Clin. Analyzer
E) Unicel DXC860I Synch.Acc.Clin. Analyzer
Recall Posting Date
2009/12/07
Manufacturer
Beckman Coulter, Inc.
Recall Start Date
2009/10/22
Recall Number
51584
Hazard Classification
Type III
Model or Catalog #
A) 4767
B) 4768
C) 4771
D) 4772
E) 4773
Lot or Serial #
A) Cup Lot Numbers Starting With "09" Or Later
B) Cup Lot Numbers Starting With "09" Or Later
C) Cup Lot Numbers Starting With "09" Or Later
D) Cup Lot Numbers Starting With "09" Or Later
E) Cup Lot Numbers Starting With "09" Or Later
Reason for Recall
The access 1.0 ml insert cup, used on Unicel DXC Synchron Clinical Systems, might be the cause of some sample wheel motion errors when run using 13x100 sample tubes.
Trade Name
A) Unicel DXI600 Access Immunoassay System Class 2
B) Unicel DXI600 Access Immunoassay System Class 3
C) Unicel DXC880I Access Clinical Analyzer
D) Unicel DXI800 Access Analyzer Class 3
E) Unicel DXI800 Access System Analyzer Class 2
F) Unicel DXC660I Access Clinical Analyzer
G) Unicel DXC680I Access Clinical Analyzer
H) Unicel DXC860I Access Clinical Analyzer
Recall Posting Date
2009/12/07
Manufacturer
Beckman Coulter Inc.
Recall Start Date
2009/10/26
Recall Number
51422
Hazard Classification
Type II
Model or Catalog #
A) A30260, A71461
B) A71461
C) 4768
D) 973100, A71457
E) 973100, A71457
F) 4771
G) 4772
H) 4773
Lot or Serial #
A) N/A
B) N/A
C) N/A
D) N/A
E) N/A
F) N/A
G) N/A
H) N/A
Reason for Recall
A software issue was identified in system software 4.2 and 4.2.1, which may result in the incorrect handling of a wash arm dispense plate motion failure. Instead of generating a red 'warning' event and stopping sample processing, the system will post a yellow 'caution' event and continue to operate with the dispense plate in an elevated position until the instrument is reinitialized. Potentially affected results are not flagged.
Trade Name
A) VNS Therapy Demipulse Programmable Generator
B) VNS Therapy Demipulse Duo Generator
Recall Posting Date
2009/12/21
Manufacturer
Cyberonics Inc.
Recall Start Date
2009/12/03
Recall Number
52025
Hazard Classification
Type II
Model or Catalog #
A) N/A
B) N/A
Lot or Serial #
A) All Lot Numbers
B) All Lot Numbers
Reason for Recall
Enhanced labelling to prevent end of service (EOS) warning messages & reduced battery life resulting from electromagnetic induction (EMI), electrostatic discharge (ESD), or electro cautery exposure during surgery.
Trade Name
A) Voluson E8 Ultrasound System Main Unit
B) Voluson E8 Expert Ultrasound System
Recall Posting Date
2009/11/09
Manufacturer
GE Medical Systems Kretztechnik GMBH & CO OHG
Recall Start Date
2009/10/21
Recall Number
51227
Hazard Classification
Type III
Model or Catalog #
A) H48651PT
B) H48661N
Lot or Serial #
A) S/N D00229
B) S/N D04842, D05589 & D50233
Reason for Recall
When using combination of 2D-compound resolution imaging (CRI) and a pulsed wave Doppler single gate (PW) with steered Doppler angle, the displayed velocity scale is incorrect, which could delay treatment.
Trade Name
Acuson S2000
Recall Posting Date
2009/10/12
Manufacturer
Siemens Medical Solutions USA, Inc.
Recall Start Date
2009/09/08
Recall Number
50793
Hazard Classification
Type II
Model or Catalog #
10041461
Lot or Serial #
200252, 200288, 200334, 200926, 200979, 200995
Reason for Recall
There is an intermittent series of events in which a print or store event failure may cause the following to occur: A) The last image of a patient study can be stored with the next new patient study performed. B) If the first patient's last image had measurement results they will appear on the next patient's image. C) When a distance ratio or area ratio measurement is performed with L Testis Mass 2 measurement label, it appears in the report as R Testis Mass 2 in English only. D) When more than 4 measurements of the same label are performed they will not appear on-screen or on the printed report.
Trade Name
A-Dec Decade Plus 1221 Dental Chair
Recall Posting Date
2009/11/23
Manufacturer
A-Dec, Inc.
Recall Start Date
2009/04/06
Recall Number
51399
Hazard Classification
Type III
Model or Catalog #
1221
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
When the chair back was returned up to the 'patient entry / exit' position, a pinch existed between the moving chair back support and the stationary underside of the armrest.
Trade Name
AdvanSync Orthodontic Device
Recall Posting Date
2009/10/26
Manufacturer
Ormco Corporation
Recall Start Date
2009/10/05
Recall Number
51125
Hazard Classification
Type III
Model or Catalog #
600-4000
Lot or Serial #
080901333, 080902356, 070981416, 060971299, 060976084, 080902357, 080901335, 070981417, 060971304, 060976079
Reason for Recall
Five complaints involving braze joint failures of AdvanSync Telescopic Rod Assemblies resulting in a recall.
Trade Name
Advantage, 2-Button Turbo Handpiece
Recall Posting Date
2009/12/07
Manufacturer
Linvatec Corporation D.B.A. Conmed Linvatec
Recall Start Date
2009/11/09
Recall Number
51507
Hazard Classification
Type II
Model or Catalog #
D9924
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
The manufacturer has determined there is a remote possibility that the device may self activate.
Trade Name
Advent 4.5mmd Platform Tapered Abutment
Recall Posting Date
2009/12/07
Manufacturer
Zimmer Dental Inc.
Recall Start Date
2009/11/06
Recall Number
51588
Hazard Classification
Type II
Model or Catalog #
AVACT
Lot or Serial #
61233652, 61238956, 61245049, 61252577, 61256798, 61263053, 61314067, 61341224
Reason for Recall
Zimmer Dental has received field complaints reporting that during application of torque to the abutment, the upper portion of the abutment had fractured and separated from the lower shaft. The investigation has determined that the product is not manufactured to specification. The pilot broach for threading the internal threads of the implant is too deep and extends into the shaft of the screw.
Trade Name
Amplatz PTFE-Coated Guidewire
Recall Posting Date
2009/11/09
Manufacturer
Gyrus Acmi, Inc.
Recall Start Date
2009/08/25
Recall Number
50973
Hazard Classification
Type III
Model or Catalog #
13BX
Lot or Serial #
96158DB
Reason for Recall
Amplatz PTFE-Coated Guidewires, part number 13BX, sold as lot 96158DB, may have been mis-packaged with product from lot number 96159DB. In this case, the outer box is labelled with the errant 96158DB lot number while the inner pouches are correctly labelled as lot 96159DB.
Trade Name
Amplicor STD Specimen & Transport Kit
Recall Posting Date
2009/12/21
Manufacturer
Roche Molecular Systems, Inc.
Recall Start Date
2009/11/27
Recall Number
51858
Hazard Classification
Type II
Model or Catalog #
20753920122
Lot or Serial #
K13333, K13334, K15491, K15492, 2036
Reason for Recall
Potential for the male swab to present a defectiveness where the cotton head of the swab may dislodge and/or unravel with minimal force. This could result in having cotton fibres remaining within the urethra during specimen collection of male patients.
Trade Name
Apexpro Telemetry System - Clinical Info.
Recall Posting Date
2009/11/23
Manufacturer
GE Medical Systems Information Technologies
Recall Start Date
2009/10/26
Recall Number
48250
Hazard Classification
Type II
Model or Catalog #
418779-010, APRSRVR16=XXX, CICV3APRO=XXX, CICV3APRO=XXX/XXX , CICV4APRO=XXX
Lot or Serial #
>100 Numbers, Contact Manufacturer
Reason for Recall
Operator's manual does not warn against the unintended use of Apexpro Telemetry to monitor SPO2 without concurrent ECG monitoring and does not identify correct alarms and messages during loss of SPO2 data condition.
Trade Name
ASP Automatic Endoscope Reprocessor
Recall Posting Date
2009/11/23
Manufacturer
Advanced Sterilization Products,
A Division Of Ethicon
Recall Start Date
2009/09/28
Recall Number
51459
Hazard Classification
Type III
Model or Catalog #
20300, 20301
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
Advanced Sterilization Products (ASP) is sending a letter to all Automated Endoscope Reprocessor (AER) customers to reinforce the importance of properly setting the disinfectant cycle time when using the ASP AER and to remind users to observe the illuminated temperature light during the cycle. ASP has received user reports about improper times set for each disinfectant cycle.
Trade Name
Auto D Dimer Kit CRS 126B
Recall Posting Date
2009/12/07
Manufacturer
Trinity Biotech PLC
Recall Start Date
2009/08/19
Recall Number
51650
Hazard Classification
Type II
Model or Catalog #
CRS126-B
Lot or Serial #
T065095, T082003, T119007
Reason for Recall
The response of the two highest points of the kit's calibration curve are the same and consequently very high D dimer values may be underestimated.
Trade Name
Bard Magnum Core Tissue Biopsy Needle
Recall Posting Date
2009/11/23
Manufacturer
Bard Peripheral Vascular, Inc.
Recall Start Date
2009/10/16
Recall Number
51391
Hazard Classification
Type II
Model or Catalog #
MN1610, MN1613, MN1616, MN1620
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
Some of the needles may exhibit an exposed sample chamber due to extended stylet length from a loose hub connection, resulting in an inability to obtain the biopsy sample.
Trade Name
BCI 9015 4-Channel Analog Output Cable
Recall Posting Date
2009/11/09
Manufacturer
Smiths Medical PM, Inc.
Recall Start Date
2009/10/19
Recall Number
51331
Hazard Classification
Type III
Model or Catalog #
9015
Lot or Serial #
W72407
Reason for Recall
When interfaced via A 9015 cable to an external data collection device, A 9004 monitor configured with CCAP 1.13 software will not communicate reliably through either channel 3 or 4. This issue only occurs when the monitor's analog output is set to external.
Trade Name
BD 30ml Luer-Lok Syringes
Recall Posting Date
2009/12/21
Manufacturer
Becton Dickinson Canada Inc.
Recall Start Date
2009/11/26
Recall Number
51921
Hazard Classification
Type II
Model or Catalog #
309650
Lot or Serial #
9175046
Reason for Recall
BD has determined that a small number of unit packages may have open seals which can adversely impact product package integrity/product sterility.
Trade Name
BHR Streamline Cup Introducers
Recall Posting Date
2009/12/07
Manufacturer
Smith & Nephew, Inc.
Recall Start Date
2009/11/11
Recall Number
51665
Hazard Classification
Type II
Model or Catalog #
90128285
Lot or Serial #
S811270
Reason for Recall
The identified lot number may mechanically seize up, preventing release of the cable tightening mechanism, during surgery or before use. As a result, this could prolong surgery time. Also, if implanted, this could cause a possibility of malpositioning of the cup due to use of the device in a mechanically seized condition.
Trade Name
Bioflex 2200- 665-1850 Vasculitis Kit
Recall Posting Date
2009/12/07
Manufacturer
Bio-Rad Laboratories Diagnostics Group
Recall Start Date
2009/11/06
Recall Number
51677
Hazard Classification
Type II
Model or Catalog #
6651850
Lot or Serial #
950759, 959261, 961363
Reason for Recall
Bioplex 2200 vasculitis reagent kits could produce false results when analyzing diluted samples.
Trade Name
Bryan Cervical Disk System
Recall Posting Date
2009/11/09
Manufacturer
Medtronic Sofamor Danek USA, Inc.
Recall Start Date
2009/10/09
Recall Number
51184
Hazard Classification
Type II
Model or Catalog #
6470114, 6470115, 6470116, 6470117,
6470118
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
There is a slight possibility of outer pouch failure during shipping, which could result in the introduction of an inner pouch having non-sterile outer surfaces into the sterile field. The innermost packaging has been shown to remain intact. Proper dispensing technique must be followed, placing the product, rather than the inner pouch, into the sterile field.
Trade Name
Buretrol IV Solution Adm.Sets Clearlink
Recall Posting Date
2009/12/07
Manufacturer
Baxter Healthcare Corporation
Recall Start Date
2009/11/05
Recall Number
51663
Hazard Classification
Type II
Model or Catalog #
2C8864
Lot or Serial #
UR09E21238
Reason for Recall
Separation of tubing from Distal Clearlink Y-site could occur, which may result in blood loss, infection, interruption of or delay in the start of intravenous therapy.
Trade Name
Carto 3 EP Mapping System
Recall Posting Date
2009/12/21
Manufacturer
Biosense Webster (Israel) Ltd.
Recall Start Date
2009/11/25
Recall Number
51915
Hazard Classification
Type II
Model or Catalog #
FG540000
Lot or Serial #
3257, 2246, 4203, 4166, 4184
Reason for Recall
Biosense Webster, Inc. has recently become aware of a potential hazard that may occur during the operation of the bloom DTU-215b cardiac stimulator (not a Johnson & Johnson device) in conjunction with RF generators and EP equipment, including the Carto 3 system. If RF energy and pacing stimuli are delivered simultaneously, across the same electrode pair, a DC voltage may be generated. This can cause micro-bubbles in the blood within the heart chamber.
Trade Name
CBC II Blood Conservation Kit
Recall Posting Date
2009/12/07
Manufacturer
Stryker Instruments
Recall Start Date
2009/11/13
Recall Number
51764
Hazard Classification
Type II
Model or Catalog #
225-414-000
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
There is a potential that the package pouch in the CBC II Blood Conservation Kits may have a pin hole, a weakened seal, or an opened seal that could result in the product not being sterile.
Trade Name
Centricity PACS Workstation - Software
Recall Posting Date
2009/12/07
Manufacturer
GE Healthcare IT
Recall Start Date
2009/10/30
Recall Number
51228
Hazard Classification
Type II
Model or Catalog #
K2020YS, K2030JN
Lot or Serial #
>100 Numbers, Contact Manufacturer
Reason for Recall
If Dicom node config. is set to accept, user may unintentionally add new images to an approved exam, under certain conditions images will be deleted from PACS or from dictated, transcribed or completed status exam.
Trade Name
Colleague Volumetric Infusion Pumps
Updated recall # 44735 previously posted on 2008-11-24.
Recall Posting Date
2009/12/07
Manufacturer
Baxter Healthcare Corporation
Recall Start Date
2008/11/24
Recall Number
44735
Hazard Classification
Type III
Model or Catalog #
2M9163
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
Customers were erroneously provided with version 0719H0166 of the operator's manual instead of the correct version 0719I0166.
Trade Name
Complement Control Cells
Recall Posting Date
2009/11/23
Manufacturer
Immucor Inc.
Recall Start Date
2009/10/26
Recall Number
51502
Hazard Classification
Type III
Model or Catalog #
N/A
Lot or Serial #
32616
Reason for Recall
There is evidence of potential microbial contamination in some vials of complement control cells (lot # 32616), based on gram stain results.
Trade Name
Connection Bracket 1180.36AO
Recall Posting Date
2009/10/26
Manufacturer
Maquet GMBH & Co. KG
Recall Start Date
2009/09/30
Recall Number
51053
Hazard Classification
Type II
Model or Catalog #
1180.360AO
Lot or Serial #
N/A
Reason for Recall
The Connection Bracket show too much mechanical play at the operating room table interface for some procedures. The cause of this play, is that the hole in the lock washer, which is responsible for locking the accessory to the operating room table interface, is too large.
Trade Name
Contour SE Microspheres
Recall Posting Date
2009/11/09
Manufacturer
Boston Scientific Corporation
Recall Start Date
2009/08/28
Recall Number
50810
Hazard Classification
Type II
Model or Catalog #
M001761101, M001761122, M001761141, M001761162, M001761201, M001761222, M001761241, M001761262, M001761301, M001761322, M001761341, M001761362, M001761401, M001761422, M001761441, M001761462, M001761501, M001761522, M001761541, M001761562
Lot or Serial #
>100 Numbers, Contact Manufacturer
Reason for Recall
All affected product is packaged within a vial. Boston Scientific has determined that, for affected products, the sterile barrier in the packaging (pouch) that contains the vial may be breached.
Trade Name
CyberKnife Robotic Radiosurgery System
Recall Posting Date
2009/12/21
Manufacturer
Accuray, Inc.
Recall Start Date
2009/06/22
Recall Number
50252
Hazard Classification
Type II
Model or Catalog #
28000
Lot or Serial #
C202
Reason for Recall
A healthcare facility reported their end-to-end targeting accuracy was out of specification. The position of one of the robotic manipulator axes differed from the position determined by the robotic controller. Error 103 (karel e-stop detected) and error 342 (deviation in absolute position value DSE - RDC) were displayed.
Trade Name
CyberKnife Robotic Radiosurgery System
Updated recall # 50252 previously posted on 2009-08-31.
Recall Posting Date
2009/12/07
Manufacturer
Accuray, Inc.
Recall Start Date
2009/06/22
Recall Number
50252
Hazard Classification
Type II
Model or Catalog #
28000
Lot or Serial #
C202
Reason for Recall
A healthcare facility reported their end-to-end targeting accuracy was out of specification. The position of one of the robotic manipulator axes differed from the position determined by the robotic controller. Error 103 (karel e-stop detected) and error 342 (Deviation in absolute position value DSE - RDC) were displayed. Recalls 50114 and 50252 are identical. For future reference, refer to recall 50252.
Trade Name
Da Vinci S Surgical System, Model IS2000
Recall Posting Date
2009/10/26
Manufacturer
Intuitive Surgical Inc.
Recall Start Date
2009/07/01
Recall Number
51026
Hazard Classification
Type III
Model or Catalog #
IS2000
Lot or Serial #
N/A
Reason for Recall
Intuitive Surgical Inc. observed an electronics subsystem failure that resulted in partial intermittent closure of the instrument grip (IS2000 system). A software coding error in P7 caused the error for the MTM grip.
Trade Name
Dade Hepzyme (B4240-10)
Recall Posting Date
2009/12/21
Manufacturer
Siemens Healthcare Diagnostics Products GMBH
Recall Start Date
2009/12/02
Recall Number
51994
Hazard Classification
Type III
Model or Catalog #
B4240-10
Lot or Serial #
528752
Reason for Recall
Customer complaints regarding incorrect high APTT values that may be observed on heparin and heparin-free samples treated with lot 528752 of Dade Hepzyme when using lot 528752 of Dade Hepzyme with non-Siemens APTT reagents.
Trade Name
Diagnostic Ultrasound Transducers / Probe
Recall Posting Date
2009/12/21
Manufacturer
Toshiba Medical Systems Corporation
Recall Start Date
2009/10/30
Recall Number
51870
Hazard Classification
Type II
Model or Catalog #
PVT-681MV
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
In ultrasound systems using the Ultrasound Convex Transducer PVT-681MV, the volume image acquired in 'plane C' is reversed horizontally.
Trade Name
Dimension Vista System-Chem Calibrators
Recall Posting Date
2009/12/21
Manufacturer
Siemens Healthcare Diagnostics Inc.
Recall Start Date
2009/10/19
Recall Number
51373
Hazard Classification
Type III
Model or Catalog #
KC130
Lot or Serial #
9ED058
Reason for Recall
Customer complaints regarding Dimension Vista Amon method. Siemens has confirmed a negative bias of up to 35% for level 1 quality control when calibrated with Chem 3 Calibrator lot 9ED058 vs. 9BD064.
Trade Name
Dimension Vista System-Urea Nitrogen Assay
Recall Posting Date
2009/10/26
Manufacturer
Siemens Healthcare Diagnostics Inc.
Recall Start Date
2009/09/28
Recall Number
51016
Hazard Classification
Type III
Model or Catalog #
K1021
Lot or Serial #
09187AA
Reason for Recall
Received customer complaints regarding inaccurate results with the Bun Flex Reagent cartridge lots 09154AE and 09187AA.The test results may be falsely depressed or elevated.
Trade Name
Dimension V-Lyte Integrated Multisensor
Recall Posting Date
2009/12/07
Manufacturer
Siemens Healthcare Diagnostics Inc.
Recall Start Date
2009/10/29
Recall Number
51421
Hazard Classification
Type III
Model or Catalog #
K800, K820, K825, K835
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
Received customer complaints regarding Dimension Vista V-Lyte. Siemens confirms that urine sodium results become elevated as urine specimens are run on a sensor.
Trade Name
Dupaque X-Ray Detectable Sponges
Recall Posting Date
2009/12/07
Manufacturer
Derma Sciences, Inc.
Recall Start Date
2009/10/27
Recall Number
51464
Hazard Classification
Type II
Model or Catalog #
94675-5
Lot or Serial #
4408
Reason for Recall
Received two complaints from a customer indicating that the sponge separated at the end during surgery and was left behind.
Trade Name
Duralite Pyramid Receiver Prox. Adaptor
Recall Posting Date
2009/12/07
Manufacturer
The Ohio Willow Wood Company
Recall Start Date
2009/11/09
Recall Number
51740
Hazard Classification
Type II
Model or Catalog #
N/A
Lot or Serial #
All Lot Numbers
Reason for Recall
There is a potential for the adapter to crack or break while in use by active amputees.
Trade Name
Durom Acetabular Cup
Recall Posting Date
2009/12/07
Manufacturer
Zimmer GMBH
Recall Start Date
2009/11/15
Recall Number
51631
Hazard Classification
Type II
Model or Catalog #
01.00214.044, 01.00214.046, 01.00214.048, 01.00214.050, 01.00214.052, 01.00214.054, 01.00214.056, 01.00214.058, 01.00214.060, 01.00214.062, 01.00214.064, 01.00214.066
Lot or Serial #
>1000 Numbers, Contact Manufacturer
Reason for Recall
There have been reports of revisions of Zimmer's Durom Acetabular Cup and the Metasul LDH (large diameter head) femoral head in certain European markets. Based on Zimmer's investigation, the most probable cause is using a surgical technique which differs from that prescribed in the surgical technique for the Durom Acetabular Cup. This field safety notice updates surgical technique instructions.
Trade Name
Endo Femoral Aimers
Recall Posting Date
2009/11/09
Manufacturer
Smith & Nephew, Inc., Endoscopy Division
Recall Start Date
2009/09/22
Recall Number
50942
Hazard Classification
Type I
Model or Catalog #
72201714, 72201716
Lot or Serial #
For product code 72201714: (L) 50278012
For product code 72201716: (L)502780141 and (L)50278014
Reason for Recall
Specific product codes and lot numbers of the Endo Femoral Aimers may have been improperly marked indicating the wrong size and part number. The actual 4mm Aimer is marked as 6mm and the actual 6mm Aimer is marked as 4mm.
Trade Name
Endoclamp Aortic Catheter
Recall Posting Date
2009/11/09
Manufacturer
Edwards Lifesciences
Recall Start Date
2009/10/01
Recall Number
51118
Hazard Classification
Type II
Model or Catalog #
EC1001
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
Possibility of the balloon rupturing during use.
Trade Name
Engen Automated Sample Handling System
Recall Posting Date
2009/10/26
Manufacturer
Thermo Fisher Scientific OY
Recall Start Date
2009/10/06
Recall Number
51117
Hazard Classification
Type I
Model or Catalog #
ENGEN
Lot or Serial #
JCT20Q3J
Reason for Recall
Ortho Clinical Diagnostics is aware of one confirmed report of misreported patient results with the Engen Laboratory Automation System. When used with TCAutomation software versions 2.8.2 or below, incorrect sample ID’s may be electronically written to the radio frequency ID tags on the sample carriers.
Trade Name
Equalizer Balloon Catheter
Recall Posting Date
2009/11/09
Manufacturer
Boston Scientific Cork Ltd.
Recall Start Date
2009/09/30
Recall Number
51119
Hazard Classification
Type II
Model or Catalog #
M001171060, M001171080, M001171100, M001171120
Lot or Serial #
>100 Numbers, Contact Manufacturer
Reason for Recall
Sterile barrier in the packaging of the affected product may be compromised. It has been reported that the outer pouch seal may be breached.
Trade Name
Excor System-Connecting Set For Cannula
Recall Posting Date
2009/12/21
Manufacturer
Berlin Heart GMBH
Recall Start Date
2009/10/02
Recall Number
51703
Hazard Classification
Type I
Model or Catalog #
A06-009, A09-012, A12-016
Lot or Serial #
526337
Reason for Recall
It was observed that a too short cut of the adapter cannula combined together with certain conditions of stress may have a effect of leakage of the cannula.
Trade Name
Express SD Premounted Stent System
Recall Posting Date
2009/10/12
Manufacturer
Boston Scientific Corporation
Recall Start Date
2009/04/14
Recall Number
48011
Hazard Classification
Type III
Model or Catalog #
H74937911515900, H74937911614900, H74937911715900, H74937912519900, H74937912618150, H74937912618900
Lot or Serial #
12045170, 12070363, 12073173, 12054585, 12054587, 12054580, 12082319, 12082341, 12045167, 12085875, 12089749, 12089752, 12129028
Reason for Recall
Packaging for the applicable devices may not include a written copy of the product directions for use.
Trade Name
EZManager Max Diabetes Management Software
Recall Posting Date
2009/12/21
Manufacturer
Animas Corporation
Recall Start Date
2009/11/16
Recall Number
51762
Hazard Classification
Type III
Model or Catalog #
100-467-23, 100-467-50
Lot or Serial #
2937776 for EZManager Max CD, French
2936068, 2938247, 2962490 for EZManager Max CD, English
Reason for Recall
Compact discs received for EZManager Max Diabetes Management Software packages were identified as being blank when used in a computer.
Trade Name
Galileo Echo Blood Bank Analyzer
Recall Posting Date
2009/12/07
Manufacturer
Immucor Inc.
Recall Start Date
2009/11/06
Recall Number
51706
Hazard Classification
Type I
Model or Catalog #
87000
Lot or Serial #
N/A
Reason for Recall
Two lot numbers of reagents or strips are allowed to be tested while performing QC testing on the echo. The result is that the final lot tested will pass QC even though it did not undergo complete QC testing.
Trade Name
Galileo Echo Blood Bank Analyzer
Recall Posting Date
2009/12/07
Manufacturer
Immucor Inc.
Recall Start Date
2009/11/06
Recall Number
51708
Hazard Classification
Type I
Model or Catalog #
87000
Lot or Serial #
N/A
Reason for Recall
Under some circumstances, a manually entered sample ID could be assigned to the wrong sample.
Trade Name
Galileo Echo Blood Bank Analyzer
Recall Posting Date
2009/12/21
Manufacturer
Immucor Inc.
Recall Start Date
2009/11/25
Recall Number
51956
Hazard Classification
Type II
Model or Catalog #
87000
Lot or Serial #
N/A
Reason for Recall
The instrument may generate a negative well interpretation for capture-R ready-screen or capture-R ready-ID assays and subsequent visual interpretation of those reactions are weak positive or questionable (equivocal).
Trade Name
Gammacoat Plasma Renin Activity 123I RIA Kit
Recall Posting Date
2009/11/09
Manufacturer
Diasorin Inc./S.R.L.
Recall Start Date
2009/10/19
Recall Number
51278
Hazard Classification
Type II
Model or Catalog #
N/A
Lot or Serial #
123574, 123612, 123620
Reason for Recall
The Renin Activity Control provided with the kit has an incorrect range printed on the label. Printed range on the label is 5.5 - 10.3 ng/ml/hr. The correct range is 4.0 -7.6 ng/ml/hr.
Trade Name
Getinge Steam Sterilizer
Recall Posting Date
2009/12/07
Manufacturer
Getinge Sourcing LLC.
Recall Start Date
2009/09/01
Recall Number
50976
Hazard Classification
Type II
Model or Catalog #
422HC, 433HC, 522HC, 533HC, 633HC, 733HC, 833HC
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
RTD temperature sensor component was failing prematurely. Temperature control of the sterilization process is maintained through the use of the sensor used in the sterilizer jacket and chamber.
Trade Name
Giraffe & Panda T-Piece Resuscitation System
Recall Posting Date
2009/10/26
Manufacturer
Ohmeda Medical,
A Division Of Datex Ohmeda Inc.
Recall Start Date
2009/09/22
Recall Number
51025
Hazard Classification
Type II
Model or Catalog #
M1091316, M1091335, M1091365
Lot or Serial #
N/A
Reason for Recall
1. Moulding flash during manufacturing of T-piece at supplier caused out of specification aperture. 2. Severity in risk analysis control document (RAC) scored low leading to less strict supplier inspection and process control requirements.
Trade Name
Goldmann Applanation Tonometer Prism
Recall Posting Date
2009/11/09
Manufacturer
Haag-Streit, AG
Recall Start Date
2009/10/13
Recall Number
51262
Hazard Classification
Type III
Model or Catalog #
1000855
Lot or Serial #
>100 Numbers, Contact Manufacturer
Reason for Recall
In Europe, Hagg-Streit received 3 reports concerning leaking Tonometer Prisms. A known reason of failing Tonometer Prisms is the use of disinfectants not approved and not released by Haag-Streit, mainly cool wipes. This might lead to irritation of cornea and in rare cases to reversible corneal erosions.
Trade Name
GT Obturators
Recall Posting Date
2009/10/12
Manufacturer
Dentsply Tulsa Dental Specialties Trading As Dentsply Endodontics
Recall Start Date
2009/09/25
Recall Number
50931
Hazard Classification
Type III
Model or Catalog #
GTOB20042
Lot or Serial #
12709123
Reason for Recall
The product, GT Obturators, GTOB20042, lot 012709123 was found to contain mixed instruments, size 20, taper .04 (20/.04) and size 20, taper .08 (20/.08) obturators.
Trade Name
Guider Softip Guiding Catheter
Recall Posting Date
2009/11/09
Manufacturer
Medi-Tech Boston Scientific Corporation
Recall Start Date
2009/10/07
Recall Number
51219
Hazard Classification
Type II
Model or Catalog #
10042, 10043, 10044, 10046, 10047, 10048, 10050, 10051, 10052, 10062, 10063, 10064, 10142, 10143, 10144, 10146, 10147, 10148, 10150, 10151, 10152, 10162, 10163, 10164
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
It is believed that vast swings in environmental conditions can impact catheter products over time. Improper storage may cause polymer degradation in portions of the catheter. Use of degraded catheters can expose patients to potential adverse events including a delay in procedure time due to catheter exchange, vessel trauma and stroke. Boston Scientific Corp. is issuing a safety alert reminding users to follow directions for use.
Trade Name
Gynecare Morcellex Tissue Morcellator
Recall Posting Date
2009/11/09
Manufacturer
Ethicon Inc. Also Trading As Ethicon Women's
Health And Urology
Recall Start Date
2009/10/09
Recall Number
51327
Hazard Classification
Type II
Model or Catalog #
MX0100
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
Ethicon is conducting a voluntary product recall of all lots of Gynecare Morcellex Tissue Morcellator due to the potential for the duckbill seal to become dislodged, and a confirmed event in which the seal was retained in a patient.
Trade Name
Heartmate XVE And II LVAS Battery Clip
Recall Posting Date
2009/11/23
Manufacturer
Thoratec Corporation
Recall Start Date
2009/10/14
Recall Number
51438
Hazard Classification
Type II
Model or Catalog #
N/A
Lot or Serial #
>100 Numbers, Contact Manufacturer
Reason for Recall
The threaded connector of the battery clip housing can become loose from the battery clip housing thereby creating a potential to interrupt pump function.
Trade Name
Heater Wire Adaptors
Recall Posting Date
2009/10/12
Manufacturer
Fisher And Paykel Healthcare Ltd.
Recall Start Date
2009/07/23
Recall Number
50850
Hazard Classification
Type II
Model or Catalog #
900MR800
Lot or Serial #
All Lot Numbers
Reason for Recall
Manufacturer is replacing these adaptors with new, more advanced product, which include surge protection hardware that will reduce the risk of breathing circuit heater wire malfunction.
Trade Name
Heater Wire Adaptors
Updated recall # 50850 previously posted on 2009-10-12.
Recall Posting Date
2009/11/09
Manufacturer
Fisher And Paykel Healthcare Ltd.
Recall Start Date
2009/07/23
Recall Number
50850
Hazard Classification
N/A
Model or Catalog #
N/A
Lot or Serial #
N/A
Reason for Recall
After further review it has been determined that this action by the manufacturer is not considered a recall under the definition of recall, but is considered a product withdrawal by the manufacturer.
Trade Name
Hintegra Talar Cutting Guides
Recall Posting Date
2009/12/21
Manufacturer
New Deal SA
Recall Start Date
2009/10/20
Recall Number
51435
Hazard Classification
Type II
Model or Catalog #
Left Cutting Guide, Right Cutting Guide
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
The width of the defective Talar Cutting Guides is too large, thus the medio-lateral cuts are too large compared to the Talar trial implant dimensions.
Trade Name
Histopaque 1077-1
Recall Posting Date
2009/10/12
Manufacturer
Sigma Diagnostics
Recall Start Date
2009/09/21
Recall Number
50914
Hazard Classification
Type III
Model or Catalog #
1077-1
Lot or Serial #
038K6170
Reason for Recall
Bacterial contamination (Pseudomonas mendocina) in the affected device.The device is a cell separation media and in the reported instance was used to isolate stem cells from bone marrow specimens. The isolated cells were placed into cell culture for further testing and contamination was evident after 14 days of culture. The laboratory confirmed the source of contamination was the device.
Trade Name
Hypergel 5 Gram
Recall Posting Date
2009/10/12
Manufacturer
Molnlycke Health Care AB
Recall Start Date
2009/09/22
Recall Number
50961
Hazard Classification
Type III
Model or Catalog #
360500
Lot or Serial #
08D22-156-32913, 08D22-157
Reason for Recall
Hypergel 5 gram, product code 360500, lot numbers 08d22-156-32913 and 08d22-157, have been distributed with a lot number deficiency wherein the lot number listed on the packaging does not match the lot number on the individual piece.
Trade Name
i-CAT 3D Dental Imaging System
Recall Posting Date
2009/12/21
Manufacturer
Imaging Sciences International, Inc.
Recall Start Date
2009/10/30
Recall Number
51709
Hazard Classification
Type II
Model or Catalog #
9140-0000-0000
Lot or Serial #
ICU070488, ICU070518, ICU070519
Reason for Recall
Fault in the brake system may result in gantry-overhead dropping quickly when the overhead is lowered. i-CAT classic containing electrical brake.
Trade Name
i-CAT 3D Dental Imaging System
Recall Posting Date
2009/12/21
Manufacturer
Imaging Sciences International, Inc.
Recall Start Date
2009/10/30
Recall Number
51944
Hazard Classification
Type II
Model or Catalog #
9140-0000-0000
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
i-CAT classic units containing mechanical brake or containing dual brake system- fault in brake system may result in the gantry-overhead carriage moving below its limits.
Trade Name
Ilumavision Software Ver. 2.1.0
Recall Posting Date
2009/10/12
Manufacturer
Imtec Imaging, LLC
Recall Start Date
2009/08/03
Recall Number
50774
Hazard Classification
Type II
Model or Catalog #
N/A
Lot or Serial #
Version 2.1.0
Reason for Recall
1. Incorrect imaging resulting from 3rd party digital image (Dicom) incompatibility. 2. Potential incorrect patient identification on report. Both may result in inaccurate or inappropriate treatment.
Trade Name
Imager II Urology Torque Catheters
Recall Posting Date
2009/12/21
Manufacturer
Boston Scientific Corporation
Recall Start Date
2009/11/03
Recall Number
51536
Hazard Classification
Type II
Model or Catalog #
M0064003031
Lot or Serial #
57271, 57616, 57269, 57613
Reason for Recall
The sterile barrier in the packaging (pouch) that contains the catheter may be breached. The breach of sterility could lead to a contamination of the device with subsequent transfer of infectious agents to the patient.
Trade Name
Impression Coping Closed Tray
Recall Posting Date
2009/10/26
Manufacturer
Nobel Biocare AB
Recall Start Date
2009/05/21
Recall Number
48621
Hazard Classification
Type II
Model or Catalog #
N/A
Lot or Serial #
687218
Reason for Recall
Nobel Biocare received several complaints from customers. After investigation it was determined that one batch of Impression Coping Closed Tray Nobelactive INT RP Q6x9mm was mis-packaged.
Trade Name
Innova 2121IQ Bi-Plane Cardio Imaging
Recall Posting Date
2009/11/23
Manufacturer
GE Medical Systems SCS
Recall Start Date
2009/10/30
Recall Number
51374
Hazard Classification
Type II
Model or Catalog #
S18821AM
Lot or Serial #
S/N 553918BU4
Reason for Recall
Potential limit switch failure could occur due to collision between lateral plane of Innova positioner and other equipment, which may break x-ray and positioning switches and may not allow exam completion.
Trade Name
Insight Software Revision H,Bioview REV5.5.1
Recall Posting Date
2009/11/23
Manufacturer
Sandhill Scientific Inc.
Recall Start Date
2009/02/23
Recall Number
47672
Hazard Classification
Type III
Model or Catalog #
S98-1925-2H
Lot or Serial #
All lots are affected
Reason for Recall
The Revision "H" software calculated esophageal amplitudes incorrectly when the user generated a report. There were no issues with the acquisition of the data. Users could still acquire data, however there was a requirement to install the revision 'I' software upgrade before reports were printed to ensure data was reported (printed) correctly.
Trade Name
Intertan Hex Screw
Recall Posting Date
2009/11/09
Manufacturer
Smith & Nephew, Inc.
Recall Start Date
2009/10/15
Recall Number
51245
Hazard Classification
Type III
Model or Catalog #
71642220
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
This recall was initiated as the product description included the word 'captured' and this was not a 'captured' screw. This could cause intra operative confusion due to the mislabeling. The problem originated in 2007 and all internal product was corrected, however the product in the field was not corrected.
Trade Name
IW900 Series Infant Warmer
Recall Posting Date
2009/11/09
Manufacturer
Fisher And Paykel Healthcare Ltd.
Recall Start Date
2009/09/07
Recall Number
51309
Hazard Classification
Type III
Model or Catalog #
IW910, IW920, IW930, IW932, IW934, IW950,
IW980, WITH SUFFIX JEU
Lot or Serial #
All Infant Warmers in Canada.
Reason for Recall
Notifying users in Canada of updating cleaning instructions for the IW900 Series Infant Radiant Warmers. The inclusion of the recommended and non-recommended proprietary cleaning products supplements the current operating instructions and technical manual held by Canadian customers which already cautions against using solutions containing specifically named chemicals.
Trade Name
IW900 Series Infant Warmer
Recall Posting Date
2009/11/09
Manufacturer
Fisher And Paykel Healthcare Ltd.
Recall Start Date
2009/09/07
Recall Number
51383
Hazard Classification
Type III
Model or Catalog #
Series With Suffix > IW910, IW920, IW930, IW932, IW934, IW950, IW980
Lot or Serial #
All Infant Warmers in Canada.
Reason for Recall
Notifying users in Canada of an updated test procedure for the power-fail alarm of the IW900 Series Infant Radiant Warmer. The service manual already includes a requirement for an annual check of the power-fail alarm. The test procedure has been revised to check that the power-fail alarm operates for at least 12 minutes. The revised test procedure accounts for any unexpected degradation over time of the supercapacitor that operates the power-fail alarm.
Trade Name
IW900 Series Infant Warmer
Recall Posting Date
2009/12/21
Manufacturer
Fisher And Paykel Healthcare Ltd.
Recall Start Date
2009/09/07
Recall Number
51309
Hazard Classification
Type III
Model or Catalog #
IW910, IW920, IW930, IW932, IW934, IW950, IW980, WITH SUFFIX JEU
Lot or Serial #
All Infant Warmers in Canada.
Reason for Recall
Notifying users in Canada of updating cleaning instructions for the IW900 series infant radiant warmers. The inclusion of the recommended and non-recommended proprietary cleaning products supplements the current operating instructions and technical manual held by Canadian customers which already cautions against using solutions containing specifically named chemicals.
Trade Name
Koordinat M/MP
Recall Posting Date
2009/12/21
Manufacturer
Siemens AG
Recall Start Date
2009/08/13
Recall Number
50747
Hazard Classification
Type III
Model or Catalog #
05294003, 06239003
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
If the swivel locking mechanism is not completely engaged into position, during patient transfer, the table may unintentionally swivel, could result in uncontrolled patient movement during transfer, patient falling to the floor.
Trade Name
Kryptonite Bone Cement
Recall Posting Date
2009/12/07
Manufacturer
Doctors Research Group Inc.
Recall Start Date
2009/09/17
Recall Number
51492
Hazard Classification
Type III
Model or Catalog #
KRYP-X-03, KRYP-X-10, KRYP-Z-03, KRYP-Z-10
Lot or Serial #
05072009, 05272009, 06292009, 03102009, 07142008, 10232008, 10272008, 10312008, 10282009
Reason for Recall
As a result of increased viscosity of component "A" in some Kryptonite Bone Cement Kits, the end user may observe difficulty in discharging component "A" from the syringe that it is packaged in.
Trade Name
Kyphx HV-R Bone Cement
Recall Posting Date
2009/12/07
Manufacturer
Medtronic Spine LLC
Recall Start Date
2009/11/09
Recall Number
51643
Hazard Classification
Type II
Model or Catalog #
C01A, C01B
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
Medtronic Spine LLC is recalling specific lots of Kyphx HV-R bone cement following complaints regarding extended time to reach the proper viscosity (doughy state) and inconsistency or lack of homogeneity during cement preparation, as well as when waiting to deliver the bone cement with the delivery device (e.g. bone filler device).
Trade Name
LabPro V3.01 Panel Update-01
Recall Posting Date
2009/12/21
Manufacturer
Siemens Healthcare Diagnostics Inc.
Recall Start Date
2009/11/30
Recall Number
51937
Hazard Classification
Type III
Model or Catalog #
9800-3771
Lot or Serial #
Software Version 3.01
Reason for Recall
Siemens Healthcare Diagnostics received customer complaints regarding QC isolates when tested on the walkaway instrument. Following the installation of LabPro 3.01 panel update -01 and utilization of the panel type POS combo 33 (PC33), an issue has been confirmed for QC isolates when tested on the walkaway instrument. This issue only applies to those that have installed the LP V3.01 panel update-01 (9800-3768 compact disk). QC issue affects combo type panels.
Trade Name
Leksell gamma Knife Perfexion
Recall Posting Date
2009/12/07
Manufacturer
Elekta Instruments AB
Recall Start Date
2009/08/18
Recall Number
51437
Hazard Classification
Type II
Model or Catalog #
715000
Lot or Serial #
Serial Number 6005
Reason for Recall
When incorrectly used the frame adapter may become damaged which in worst case could make it possible to lock the adapter so it does not properly fixate the frame.
Trade Name
Leksell Stereotactic System - Fixation Post
Recall Posting Date
2009/12/07
Manufacturer
Elekta Instruments AB
Recall Start Date
2009/10/02
Recall Number
51522
Hazard Classification
Type II
Model or Catalog #
1003935, 1003937
Lot or Serial #
N/A
Reason for Recall
Excessive heating of the tip of the fixation screws with the standard non-insulated posts may cause minor burning of the contact point of patients during certain higher energy MRI usage.
Trade Name
Lexicon K-File
Recall Posting Date
2009/10/26
Manufacturer
Dentsply Tulsa Dental Specialties Trading As Dentsply Endodontics
Recall Start Date
2009/09/29
Recall Number
51017
Hazard Classification
Type III
Model or Catalog #
KF01025
Lot or Serial #
811000105
Reason for Recall
The product is labelled as 25 mm in length but the files are actually 31 mm in length.
Trade Name
Manual Pulmonary Resuscitator (MPR)
Recall Posting Date
2009/11/09
Manufacturer
Unomedical Inc.
Recall Start Date
2009/10/08
Recall Number
51151
Hazard Classification
Type I
Model or Catalog #
724-E
Lot or Serial #
03-25
Reason for Recall
On some units of the Manual Pulmonary Resuscitator (MPR), the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this were to occur, the MPR may loose its ability to generate the positive pressure necessary to function properly.
Trade Name
Maquet Hanaulux Model Blue 130
Recall Posting Date
2009/11/09
Manufacturer
Maquet S.A.
Recall Start Date
2009/09/22
Recall Number
50980
Hazard Classification
Type III
Model or Catalog #
56077943
Lot or Serial #
00001717, 00001719
Reason for Recall
There is potential for a crack on the front pivot of the Acrobat 2000 spring arm installed with a Blue 130 series surgical light that could cause the surgical light to swing freely. Defect rate is 0.35%.
Trade Name
MC5057 Universal Cable
Updated recall # 50401 previously posted on
2009-09-14.
Recall Posting Date
2009/10/26
Manufacturer
Linvatec Corporation D.B.A. Conmed Linvatec
Recall Start Date
2009/08/14
Recall Number
50401
Hazard Classification
Type II
Model or Catalog #
537702
Lot or Serial #
N/A
Reason for Recall
Update: The manufacturer now requests return of the affected devices. A sufficiently worn/damaged Universal Cable MC5057, used with Conmed Linvatec Electric Handpieces may cause the handpiece to self activate.
Trade Name
Medfusion Syringe Infusion Pumps
Recall Posting Date
2009/12/07
Manufacturer
Smiths Medical MD Inc.
Recall Start Date
2009/11/16
Recall Number
51668
Hazard Classification
Type II
Model or Catalog #
3010A, 3500-306, 3500-415, 3500-500
Lot or Serial #
>100 Numbers, Contact Manufacturer
Reason for Recall
Smiths Medical has become aware of an increased trend in reports of "motor not running" ("MNR") and "motor rate error" ("MRE") alarm message events with certain Medfusion Syringe Infusion Pumps, models 3010a and 3500 ("pump"). These are high priority alarm messages which can occur for a variety of reasons as part of the usual self-test functionality of these pumps. The pump stops delivery when an MNR/MRE alarm message event occurs.
Trade Name
Medical Grade Flat Panel Display
Recall Posting Date
2009/12/07
Manufacturer
Spacelabs Medical, Inc.
Recall Start Date
2009/10/30
Recall Number
51662
Hazard Classification
Type II
Model or Catalog #
94260-19
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
A potential safety hazard associated with the use of Spacelabs Medical 19" Flat Panel Displays, model 94260-19. There is a possibility that the internal power distribution circuit board will fail prematurely, causing the display to go blank. Audio alarms will continue to function normally.
Trade Name
Midas Touch Needle Electrode
Recall Posting Date
2009/11/09
Manufacturer
Olsen Medical
Recall Start Date
2009/09/24
Recall Number
51204
Hazard Classification
Type II
Model or Catalog #
30-1014
Lot or Serial #
011668, 011970
Reason for Recall
The recall was initiated due to recent tests that indicated some Midas Touch Needle Electrodes with modified tip might have an insulation breach and could generate a risk of superficial burn to the patient and/or end-user.
Trade Name
Model 7600 - Oximeter Pod
Recall Posting Date
2009/12/07
Manufacturer
Nonin Medical Inc.
Recall Start Date
2009/10/23
Recall Number
51731
Hazard Classification
Type II
Model or Catalog #
Model 7600
Lot or Serial #
500849356, 500849358, 500849360, 500549361, 500823174, 500857064, 500874926
Reason for Recall
A malfunction with one of the components that may result in the overheating of the Oximeter pod has the potential to cause a burn if the pod is in direct contact with the patient.
Trade Name
MRSASelect Referance 63747 Coffret 20 Plaque
Recall Posting Date
2009/12/21
Manufacturer
Bio-Rad
Recall Start Date
2009/10/23
Recall Number
51525
Hazard Classification
Type II
Model or Catalog #
63747
Lot or Serial #
9H2414, 9H2417, 9H2425, 9J2429, 9J2432, 9J2435, 9J2437
Reason for Recall
Decrease of medium selectivity for methicillin-sensitive Saphyloccus aureus strains before medium’s expiration date.
Trade Name
neoBLUE mini LED Phototherapy System
Recall Posting Date
2009/12/07
Manufacturer
Natus Medical Inc.
Recall Start Date
2009/10/01
Recall Number
51237
Hazard Classification
Type II
Model or Catalog #
10101
Lot or Serial #
60001-62656 serial number
Reason for Recall
There has been a trend observed in failure of the pole mount clamp shipped with a neoBlue mini LED Phototherapy Systems. When tightened with excessive force, some of these clamps have shown stress fractures or broken. The pole mount clamp has been redesigned by the manufacturer and customers of record are being advised to inspect their systems and to order a replacement clamp if necessary.
Trade Name
Neptune 2 Ultra Rover
Recall Posting Date
2009/10/26
Manufacturer
Stryker Instruments
Recall Start Date
2009/10/06
Recall Number
51120
Hazard Classification
Type II
Model or Catalog #
0702-001-000
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
The company has become aware that there is the potential that over time, a buildup of surgical waste on the receptacle bearing surfaces may increase the force required to rotate the suction manifold. If the force becomes too high the manifold cannot be rotated - therefore it cannot be inserted or removed. Once this occurs there will be no vacuum applied to the suction manifold so the primary function of the affected rover canister is lost.
Trade Name
OneTouch SureStep Test Strips
Recall Posting Date
2009/12/21
Manufacturer
Lifescan Inc.
Recall Start Date
2009/09/24
Recall Number
51926
Hazard Classification
Type III
Model or Catalog #
Part # 010-665
Lot or Serial #
2749789
Reason for Recall
Expired OneTouch SureStep Test Strips (lot#274789) were shipped to four Canadian wholesalers.
Trade Name
Oxidase Touch Sticks
Recall Posting Date
2009/12/07
Manufacturer
Oxoid Limited
Recall Start Date
2009/07/07
Recall Number
51737
Hazard Classification
Type II
Model or Catalog #
BR0064A
Lot or Serial #
702079
Reason for Recall
May produce weak positive or false negative results.
Trade Name
Periarticular Locking Plate System
Recall Posting Date
2009/11/23
Manufacturer
Zimmer Inc.
Recall Start Date
2009/10/20
Recall Number
51408
Hazard Classification
Type III
Model or Catalog #
00-2357-009-06
Lot or Serial #
60594607
Reason for Recall
This action is being conducted because the part number on the locking plate (Zimmer Periarticular Locking Plate System distal lateral tibial plate) is incorrect. The part was etched as '2357-09-07', when it should have been etched '2357-09-06'.
Trade Name
Power Cord Component Of Sesmoid Plasty
Recall Posting Date
2009/12/07
Manufacturer
Zimmer CAS
Recall Start Date
2009/11/05
Recall Number
51683
Hazard Classification
Type II
Model or Catalog #
500.923 SUB COMP 200.301
Lot or Serial #
SP-122, SP-197
Reason for Recall
1. Female receptacle end: the terminal contact are not sufficiently recessed and protected from access. 2. The cable portion is not certified North America. 3. The male plug that plug to the power supply mains is missing a label.
Trade Name
pRK(Rabbit Kidney)
Recall Posting Date
2009/11/09
Manufacturer
Diagnostic Hybrids, Inc.
Recall Start Date
2009/01/21
Recall Number
51217
Hazard Classification
Type III
Model or Catalog #
48-0600
Lot or Serial #
480109
Reason for Recall
The Rabbit Kidney Culture(pRK), which consisted of a monolayer culture of rabbit kidney cells used to grow viruses from a patient specimen in order to detect or identify the virus, appeared to also contain another type of cells.
Trade Name
Pronto-Short Extraction Catheter
Recall Posting Date
2009/12/21
Manufacturer
Vascular Solutions, Inc.
Recall Start Date
2009/11/20
Recall Number
51819
Hazard Classification
Type II
Model or Catalog #
N/A
Lot or Serial #
Lot 544202, Exp. 11-2009, Lot 545693, Exp. 05-2010, Lot 546614, Exp. 09-2010
Reason for Recall
The expiration date of the sterile pouched guidewire was not considered when expiration dating was calculated for the finished device. Cannot assure sterility or function of the guidewire beyond its expiration date.
Trade Name
Quantiferon-TB Gold In-Tube
Recall Posting Date
2009/10/12
Manufacturer
Cellestis Ltd.
Recall Start Date
2009/09/01
Recall Number
50725
Hazard Classification
Type III
Model or Catalog #
N/A
Lot or Serial #
A090504Z
Reason for Recall
It was observed that there were a higher than expected number of false positive results with lot A090504Z and the presence of higher levels of endotoxins was found.
Trade Name
Rapid Eric Electronic Compendium
Recall Posting Date
2009/12/21
Manufacturer
Remel Inc.
Recall Start Date
2009/10/30
Recall Number
51669
Hazard Classification
Type I
Model or Catalog #
RE8323600
Lot or Serial #
1096277
Reason for Recall
Error limited to the Rapid Yeast Plus Database regarding differentiation of C. albicans from C. dubliniensis potentially leading to false ID of isolate.
Trade Name
Recon Plate 8 X 9mm, 3.5mm
Recall Posting Date
2009/11/23
Manufacturer
Smith & Nephew, Inc.
Recall Start Date
2009/10/14
Recall Number
51369
Hazard Classification
Type III
Model or Catalog #
7114-0108
Lot or Serial #
L06MT04447, L08ET19936
Reason for Recall
Smith & Nephew was made aware that all parts were missing the laser etching as identification information.
Trade Name
Reflex Hybrid Screw Extractor Inner Shaft
Recall Posting Date
2009/11/23
Manufacturer
Stryker Spine SAS
Recall Start Date
2009/10/13
Recall Number
51370
Hazard Classification
Type II
Model or Catalog #
48511905B, 48511905D
Lot or Serial #
All lots
Reason for Recall
The Reflex Hybrid Screw Extractor Inner Shaft was reported to fail at the threaded tip and fracture in the head of the screw. There had been reports of the threaded tip of the inner shaft being left in the screw due to the surgeon being unable to remove the screw. There had also been reports of damage to the screw and/or the plate while attempting to remove the screw.
Trade Name
Removable Magnet Hires 90K
Recall Posting Date
2009/12/07
Manufacturer
Advanced Bionics, LLC
Recall Start Date
2009/08/25
Recall Number
51652
Hazard Classification
Type II
Model or Catalog #
N/A
Lot or Serial #
N/A
Reason for Recall
Advanced Bionics (AB) is notifying all surgeons, clinicians and patients that the removable magnet component of the HiRes 90k cochlear implant (model no. CI-1400-01 and CI-1400-02H) is susceptible to corrosion. The affected units should be monitored and replaced if necessary.
Trade Name
Remstar M Series Heated Humidifier
Recall Posting Date
2009/12/07
Manufacturer
Respironics Inc.
Recall Start Date
2009/11/20
Recall Number
50202
Hazard Classification
Type II
Model or Catalog #
1049109, 1049110, 1049124, 10499110, 1051157, 1051158, R1049109, R1049110, U1049109, U1049110
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
Some humidifiers may fail to operate due to potential malfunction of the internal connection between the heated plate and the printed circuit board which results in a 'blinking blue light' on the humidifier control knob.
Trade Name
Renegade Fiber Braided Microcatheter
Recall Posting Date
2009/10/12
Manufacturer
Boston Scientific Cork Ltd.
Recall Start Date
2009/08/26
Recall Number
50941
Hazard Classification
Type II
Model or Catalog #
18-257, 18-259
Lot or Serial #
>100 Numbers, Contact Manufacturer
Reason for Recall
Sterile barrier in the packaging of affected product may be compromised. Pouch seal (top seal) may be breached.
Trade Name
RET-Search II
Recall Posting Date
2009/12/07
Manufacturer
Sysmex Reagents America
Recall Start Date
2009/11/03
Recall Number
51623
Hazard Classification
Type II
Model or Catalog #
RED-700A
Lot or Serial #
ZY9021
Reason for Recall
Several boxes of RET-Search II (Red-700A) manufactured on July 15, 2009 had an incorrect bar code label applied to the product box. The bar code label identified the product as stromatolyser-NR with the same lot number and expiration date. The box contains a reagent bottle and dye pouch, both of which were correctly labelled, and the contents of the bottle and dye pouch are of the proper chemical composition for RET-Search II and passed all in-process and final inspection test criteria.
Trade Name
Revlite
Recall Posting Date
2009/11/09
Manufacturer
Hoya Conbio
Recall Start Date
2009/03/31
Recall Number
49248
Hazard Classification
Type III
Model or Catalog #
N/A
Lot or Serial #
RL-1094
Reason for Recall
Advisory letter was sent to customers of Revlite System with Revision E1 software, declaring that if PTP keypad is pressed while the wavelength window is green, the system may not complete change to new wavelength as requested.
Trade Name
Revolution Centrifugal Blood Pump
Recall Posting Date
2009/10/26
Manufacturer
Sorin Group Italia S.R.L.
Recall Start Date
2009/09/23
Recall Number
51015
Hazard Classification
Type II
Model or Catalog #
50300700
Lot or Serial #
0905220164, 0906240184
Reason for Recall
Sorin Group Italia has recently identified a weakness in the plastic packaging tray used for the Revolution Centrifugal Blood Pump that might lead the packaging tray to crack. Cracks in the packaging tray can compromise the product sterility.
Trade Name
Sarns Disposable Centrifugal Pump
Recall Posting Date
2009/10/26
Manufacturer
Terumo Cardiovascular System Corporation
Recall Start Date
2009/07/06
Recall Number
51040
Hazard Classification
Type II
Model or Catalog #
164275
Lot or Serial #
KN10, KP08, LA12, LC16, LE15
Reason for Recall
Terumo Cardiovascular Systems has received three reports of tubing disconnecting from the pump inlet of the Sarns Disposable Centrifugal Pump during cardiopulmonary bypass.
Trade Name
Sarns Disposable Centrifugal Pump
Recall Posting Date
2009/10/26
Manufacturer
Termo Cardiovascular System Corporation
Recall Start Date
2009/07/14
Recall Number
51040
Hazard Classification
Type II
Model or Catalog #
164275
Lot or Serial #
KN10, KP08, LA12, LC16, LE15
Reason for Recall
Tubing disconnecting from the pump inlet of the disposable centrifugal pump during cardiopulmonary bypass.
Trade Name
Sequencer For Mosaiq
Recall Posting Date
2009/12/07
Manufacturer
Impac Medical Systems Inc.
Recall Start Date
2009/10/16
Recall Number
51821
Hazard Classification
Type II
Model or Catalog #
N/A
Lot or Serial #
Mosaiq 1.40 (All Builds), Mosaiq 1.41 (All Builds), Mosaiq 1.50
(All Builds), 1.60 (Up To, Including T2SP4), 2.00 (All Builds)
Reason for Recall
Consolidated field sequencing (CFS) feature not working properly as X2 jaw moved when only the X1 jaw should have been moving.
Trade Name
Signa HDE 1.5T MRI System - Magnet
Recall Posting Date
2009/12/07
Manufacturer
GE Healthcare Japan Corporation
Recall Start Date
2009/11/05
Recall Number
51766
Hazard Classification
Type II
Model or Catalog #
M3335SE
Lot or Serial #
HDMR2675, 9948YR1,10833YR2, 11827YR3, 8704YR9, 10685YR6, 7964YR0, 10023YR0
Reason for Recall
Use of CADstream with "pure" software for Dynamic MRI imaging may require a modification to the study preference settings to present colour overlays equivalent to non-pure studies.
Trade Name
Siregraph CF
Recall Posting Date
2009/12/21
Manufacturer
Siemens AG
Recall Start Date
2009/10/08
Recall Number
51376
Hazard Classification
Type III
Model or Catalog #
04 98 717
Lot or Serial #
1987, 3089
Reason for Recall
The potential hazard of fingers or hands getting caught exists when the patient holds on to the tomographic rod between the column/tube assembly and the spotfilm device.
Trade Name
Speedlink 5.5 Transverse Connector
Recall Posting Date
2009/10/12
Manufacturer
Zimmer Spine Austin, Inc.
Recall Start Date
2009/09/22
Recall Number
50964
Hazard Classification
Type III
Model or Catalog #
1302-55M
Lot or Serial #
55QJ
Reason for Recall
This action is being conducted because a small quantity of titanium alloy was discovered to have a discrepancy from the alloying process at another manufacturer. Alloy segregation can affect material strength. It has been suggested that the metal used to manufacture parts from lot 55QJ may have a similar problem, though testing has not shown this.
Trade Name
Staphytect Plus/Dryspot, Test Kits
Recall Posting Date
2009/12/07
Manufacturer
Remel Europe Ltd.
Recall Start Date
2009/07/09
Recall Number
50862
Hazard Classification
Type II
Model or Catalog #
DR0100M, DR0850M
Lot or Serial #
J577910, J055610, J539110, J610610
Reason for Recall
A potential for false positive reactions with some non staphylococcus aureus strains.
Trade Name
Steinmann Pin
Recall Posting Date
2009/11/23
Manufacturer
Zimmer Inc.
Recall Start Date
2009/10/23
Recall Number
51411
Hazard Classification
Type III
Model or Catalog #
18700259
Lot or Serial #
61211489
Reason for Recall
This action is being conducted because packages from the identified lot should contain style 5 (trocar-style three point) Steinmann pins. However, they actually contain style 6 (spade-type diamond point).
Trade Name
Steris Harmony LL500/700 Surgical Light
Recall Posting Date
2009/10/12
Manufacturer
Steris Corporation
Recall Start Date
2009/09/14
Recall Number
50864
Hazard Classification
Type III
Model or Catalog #
YG18568
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
A welded seam at the joint of the spring-loaded arm of the surgical light may brake and the light head may become disengaged and fall.
Trade Name
Streamlab Analytical Workcell
Recall Posting Date
2009/10/12
Manufacturer
Siemens Healthcare Diagnostics Products GMBH
Recall Start Date
2009/09/09
Recall Number
50779
Hazard Classification
Type III
Model or Catalog #
Workcell
Lot or Serial #
N/A
Reason for Recall
Customer complaints regarding Dimension Vista Systems connected to a Streamlab Analytical Workcell where it has been observed on rare occasions an intermittent low result.
Trade Name
Synex II System - Central Body
Recall Posting Date
2009/10/12
Manufacturer
Synthes (Canada) Ltd.
Recall Start Date
2009/09/23
Recall Number
50981
Hazard Classification
Type II
Model or Catalog #
04.808.001, 04.808.002, 04.808.003, 04.808.004, 04.808.005, 04.808.006, 04.808.007, 04.808.008, 04.808.009, 04.808.010, 04.808.011
Lot or Serial #
All Lot Numbers
Reason for Recall
Synthes (Canada), Ltd. was notified by their supplier, Synthes (USA), of six complaints reported outside the US associated with Synex II Implant (Central Body) which fully or partially collapsed in the original implant position after a period of six to fifteen months post operatively.
Trade Name
Synex II System - Central Body
Updated recall # 50981 previously posted on 2009-10-12.
Recall Posting Date
2009/11/23
Manufacturer
Synthes (Canada) Ltd.
Recall Start Date
2009/09/23
Recall Number
50981
Hazard Classification
Type I
Model or Catalog #
04.808.001, 04.808.002, 04.808.003, 04.808.004, 04.808.005, 04.808.006, 04.808.007, 04.808.008, 04.808.009, 04.808.010, 04.808.011
Lot or Serial #
All lots
Reason for Recall
Synthes (Canada), Ltd. was notified by their supplier, Synthes (USA), of six complaints reported outside the US associated with Synex II Implant (Central Body) which fully or partially collapsed in the original implant position after a period of six to fifteen months post operatively.
Trade Name
Syngo Dynamics C.S.A.N.-5.X,6.X,7.0
Recall Posting Date
2009/11/09
Manufacturer
Siemens Medical Solutions USA, Inc.
Recall Start Date
2009/04/30
Recall Number
48348
Hazard Classification
Type III
Model or Catalog #
100 90 480
Lot or Serial #
92241, 95243, 93477, 95071, 95248
Reason for Recall
To inform customers about a potential issue, when receiving the multiple PSV measurements from some ultrasound systems, the absolute value is not being used.
Trade Name
Syngo Imaging XS
Recall Posting Date
2009/11/09
Manufacturer
Siemens AG
Recall Start Date
2009/10/13
Recall Number
51220
Hazard Classification
Type III
Model or Catalog #
104 96 279
Lot or Serial #
1225, 1532, 1533
Reason for Recall
A potential of a single image of a series is displayed "distorted" after loading to a viewing application.
Trade Name
Syngo Imaging XS
Recall Posting Date
2009/11/09
Manufacturer
Siemens AG
Recall Start Date
2009/10/13
Recall Number
51221
Hazard Classification
Type III
Model or Catalog #
075 020 29
Lot or Serial #
14630
Reason for Recall
Potential loss of single image data of prior examinations when archived to network-attached storage (NAS) or hierarchical storage management (HSM).
Trade Name
Systeme PhD EIA/IFA (cat# 426-0246)
Recall Posting Date
2009/11/09
Manufacturer
Bio-Rad Laboratories Diagnostics Group
Recall Start Date
2009/10/13
Recall Number
51252
Hazard Classification
Type III
Model or Catalog #
4226-0228, 426-0246
Lot or Serial #
N/A
Reason for Recall
Results passed on integrating lab systems may be erroneous when using competitive tests, if they are located outside of the standard range.
Trade Name
TECNIS 1-Piece Posterior Chamber IOL
Recall Posting Date
2009/12/21
Manufacturer
Advanced Medical Optics, Inc.
Recall Start Date
2009/11/10
Recall Number
51829
Hazard Classification
Type II
Model or Catalog #
ZCB00
Lot or Serial #
N/A
Reason for Recall
The TECNIS 1-Piece Intraocular Lenses (IOLS) are being recalled because the haptics of some TECNIS 1-Piece IOLS (model ZCB00) have adhered to the lens optic after insertion into the eye. This situation can lead to additional manipulation to separate the haptics from the optic during surgery which could lead to an injury.
Trade Name
Tg Gen.2 (Thyroglobulin) Assay
Recall Posting Date
2009/12/07
Manufacturer
Roche Diagnostics GMBH
Recall Start Date
2009/11/12
Recall Number
51712
Hazard Classification
Type II
Model or Catalog #
5118921190
Lot or Serial #
155166
Reason for Recall
The current lot #155166 of TG Gen.2 (Thyroglobulin) Assay shows a premature ageing effect which leads to a positive bias of approximately 1ng/ml TG when used in combination with all current lots # 151605, 152737, 155488 and 157297 of elecsys thyroglobulin calibrator. The manufacturer has determined that customers should not use it for this purpose anymore.
Trade Name
Therascreen K-Ras Mutation Kit
Recall Posting Date
2009/11/23
Manufacturer
DXS Ltd.
Recall Start Date
2009/09/23
Recall Number
50937
Hazard Classification
Type III
Model or Catalog #
KR-41
Lot or Serial #
DK146-01, DK149-01
Reason for Recall
The mixed standard component of the Therascreen Mutation Kit may sporadically generate out-of-range high results when tested with the control assay.
Trade Name
TheraScreen K-Ras Mutation Kit
Recall Posting Date
2009/11/09
Manufacturer
Roche Diagnostics GMBH
Recall Start Date
2009/09/23
Recall Number
50937
Hazard Classification
Type III
Model or Catalog #
5366259190
Lot or Serial #
DK146-01, DK149-01
Reason for Recall
The mixed standard component of the Therascreen Mutation may sporadically generate out-of-range high results when tested with the control assay.
Trade Name
Titan Anchor Accessory Kits
Recall Posting Date
2009/11/23
Manufacturer
Medtronic Inc.
Recall Start Date
2009/10/27
Recall Number
51434
Hazard Classification
Type II
Model or Catalog #
3550-39
Lot or Serial #
>100 Numbers, Contact Manufacturer
Reason for Recall
Medtronic has initiated a field action to notify managing and implanting physicians of Medtronic per cutaneous leads for spinal cord stimulation of the potential for separation of the titanium insert from the silicone portion of the Titan Anchor, model 3550-39. If this occurs it may result in lead migration and compromised therapy for the patient.
Trade Name
Triage Tox Drug Screen PN 94400
Recall Posting Date
2009/12/21
Manufacturer
Biosite Incorporated
Recall Start Date
2009/05/28
Recall Number
50030
Hazard Classification
Type III
Model or Catalog #
94400
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
Biosite, Inc. initiates notice of correction for Triage Tox Drug Screen Kit, PN 94400. Boxes shipped after August 6, 2008 are misprinted on the right side of the box flap. The misprint incorrectly states "use EDTA whole blood or plasma sample only". The kit is intended for use with urine samples not blood or plasma.
Trade Name
Trinest Microplate Incubator And Shaker
Recall Posting Date
2009/11/23
Manufacturer
Wallac OY
Recall Start Date
2009/10/21
Recall Number
51375
Hazard Classification
Type II
Model or Catalog #
1296-0050
Lot or Serial #
12960024, 12960028
Reason for Recall
The protective coating around certain electrical wires inside the Trinest Incubator/Shaker may be worn out due to contact with a round aluminum part rotating during the continuous shaking movement, which would result in a warning message and the instrument would shut down.
Trade Name
Triniclot APPT S, Cat. No. T1202
Recall Posting Date
2009/12/21
Manufacturer
Trinity Biotech PLC
Recall Start Date
2009/04/23
Recall Number
49895
Hazard Classification
Type II
Model or Catalog #
T1202
Lot or Serial #
S067009
Reason for Recall
Precision and open vial stability issues with this lot may result in increased or decreased APTT times across the range of the assay. The lot is not stable for the full 30 day period when opened and stored.
Trade Name
Triniclot APTT HS, Product Code T1204
Recall Posting Date
2009/12/21
Manufacturer
Trinity Biotech PLC
Recall Start Date
2009/05/24
Recall Number
49896
Hazard Classification
Type II
Model or Catalog #
T1204
Lot or Serial #
R324001
Reason for Recall
Precision issues with this lot may result in increased APTT times across the range of the assay.
Trade Name
Triniclot PT Excel, T1106
Recall Posting Date
2009/12/07
Manufacturer
Trinity Biotech PLC
Recall Start Date
2009/08/06
Recall Number
51651
Hazard Classification
Type III
Model or Catalog #
T1106
Lot or Serial #
S172057
Reason for Recall
Quality control materials not recovering within predefined ranges. Prothrombin times are prolonged in lot S172057 by about 1.2 seconds.
Trade Name
Twist Drill 1.1x50mm 3.5mm stop
Recall Posting Date
2009/12/21
Manufacturer
Walter Lorenz Surgical, Inc. Also trading as: Biomet Microfixation
Recall Start Date
2009/11/18
Recall Number
51857
Hazard Classification
Type III
Model or Catalog #
01-7141
Lot or Serial #
537407, 536306, 535933
Reason for Recall
Three manufacturing lots of Twist Drill 1.1 x 50mm, 3.5mm stop with J notch, part number 01-7141 were packaged and distributed without the instructions for use.
Trade Name
Uric Acid Assay
Recall Posting Date
2009/10/12
Manufacturer
Randox Laboratories Ltd.
Recall Start Date
2009/09/15
Recall Number
50863
Hazard Classification
Type III
Model or Catalog #
UA233
Lot or Serial #
154107
Reason for Recall
UA233, batch 154107, was incorrectly labelled with an expiry date of August 28, 2011. The correct expiry date is November 29, 2010.
Trade Name
Varis 4D Integrated Treatment Console
Recall Posting Date
2009/12/07
Manufacturer
Varian Medical Systems, Inc.
Recall Start Date
2009/10/16
Recall Number
51774
Hazard Classification
Type II
Model or Catalog #
8.0.12, 8.0.23 SP2 , 8.1 SP2, 8.1.1.3, 8.1.1.4,
8.1.2, 8.1.3, 8.1.4, 8.3, 8.6, 8.6.1, 8.6.2,
8.8.0, H51
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
Malfunction when adding CBCT image. Present software has ability to acquire gantry angle when 'treatment approval enforced' and/or 'allow only couch parameter (vert, long, lat) editing & acquiring for approved plans' selected in treatment administration. Therapists can acquire gantry angle into plan, even without system rights to acquire gantry angle. Patient treated with wrong gantry angles one day.
Trade Name
Vertical Surgical Table
Recall Posting Date
2009/12/07
Manufacturer
Stryker Communications
Recall Start Date
2009/11/06
Recall Number
51664
Hazard Classification
Type II
Model or Catalog #
0788-100-000, 0788-100-001, 0788-100-0015
Lot or Serial #
All Lot Numbers
Reason for Recall
The table may experience involuntary movement as a result of fluid intrusion. This hazard can arise especially when the table is in the trendelenburg or anti-trendelenburg position allowing fluids, such as saline or water, to enter the table column.
Trade Name
Vitrax II (Sodium Hyaluronate)
Recall Posting Date
2009/12/21
Manufacturer
Advanced Medical Optics, Inc.
Recall Start Date
2009/11/06
Recall Number
51828
Hazard Classification
Type II
Model or Catalog #
VT465
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
The Vitrax II product is being recalled because during routine stability testing, it was discovered that a sample tested at the three month time point had a ph value of 7.70 which exceeded the ph specification range of 6.8 to 7.5 resulting in a lab investigation. The investigation revealed that some units from different lots post ETO sterilization have a ph greater than the upper ph limit of 7.5. The highest value was found to be 8.29.
Trade Name
Von Willebrand Reagent
Recall Posting Date
2009/11/09
Manufacturer
Siemens Healthcare Diagnostics Products GMBH
Recall Start Date
2009/10/13
Recall Number
51218
Hazard Classification
Type III
Model or Catalog #
OUBD23
Lot or Serial #
537675, 537682
Reason for Recall
An incorrect sensitivity limit has been assigned which is reagent specific and could cause results to be erroneously elevated between 17% and 40%.
Trade Name
Wallach LL 100 Cryosurgery Device
Recall Posting Date
2009/11/09
Manufacturer
Cooper Surgical Inc. Also Trading As Wallach Surgical Devices
Recall Start Date
2009/09/19
Recall Number
51212
Hazard Classification
Type II
Model or Catalog #
REF-900000-900002, REF-900034, REF-900509-5
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
Under certain circumstances a restriction of the flow of carbon dioxide gas within the handle may result in a less than optimal temperature delivery to the treatment site.
Trade Name
Washer Disinfectors
Recall Posting Date
2009/12/21
Manufacturer
Getinge Disinfection AB
Recall Start Date
2009/04/23
Recall Number
51707
Hazard Classification
Type II
Model or Catalog #
9027, 9100, 9120, 9125, 9128
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
The ground wire of the supply transformer exhibited signs of high resistance in a complaint and may give high eddy currents.
Trade Name
XIO Radiation Treatment Planning System
Recall Posting Date
2009/11/09
Manufacturer
Computerized Medical Systems, Inc. (DBA CMS, Inc.) The Elekta Group
Recall Start Date
2009/10/13
Recall Number
51241
Hazard Classification
Type II
Model or Catalog #
N/A
Lot or Serial #
Release 4.3.0 & Higher
Reason for Recall
If an auto port is created for a beam on focal, the plan sent to XIO and the beam edited, it is possible the graphic beam representation and dose display do not agree with the edited field on the text page.
Trade Name
XTEN, HLX 3004, HLX 3005
Recall Posting Date
2009/10/26
Manufacturer
Maquet S.A.
Recall Start Date
2009/09/24
Recall Number
50977
Hazard Classification
Type III
Model or Catalog #
ARD 567801093, ARD 567801094, ARD 567910901, ARD 567910910
Lot or Serial #
>10 Numbers, Contact Manufacturer
Reason for Recall
Excessive strain may cause impairment of the interlocking between the arm joint and surgical light. Inappropriate handling may cause a crack on the front pivot of the Acrobat 2000 spring arm.