Help on accessing alternative formats, such as Portable Document Format (PDF), Microsoft Word and PowerPoint (PPT) files, can be obtained in the alternate format help section.
Contact: Policy Bureau Enquiries
Medical devices have become increasingly complex as more sophisticated and less invasive surgical procedures are developed. The manufacturers' validated reprocessing instructions for these devices have not kept pace with the requirements listed in Canadian National Standards (e.g. CSA Z314.3, CSA Z314.8, CSA Z17664, CSA Z314.1, CSA Z314.22) and National or Provincial Guidelines (e.g.. Health Canada Infection Control guidelines, Provincial Infectious Diseases Advisory Committee (PIDAC) Best Practices for Cleaning, Disinfection and Sterilization - published 2006). These reprocessing issues have been compounded because medical device licensing procedures do not require pre-market review of instructions for use of Class I and Class II devices and because surgical instruments that are loaned or leased to hospitals may not be distributed with reprocessing instructions.
Concern regarding the safety of patients receiving healthcare in Canada and recommendations of the Auditor General's report of 2004 prompted Health Canada to strike a Scientific Advisory Panel on Reprocessing of Medical Devices (SAP-RMD) with the mandate of addressing both single-use and reusable medical device reprocessing issues.
The SAP-RMD has developed a series of recommendations regarding the reprocessing and reuse of Single-use Medical Devices in Canada. These recommendations are available on Health Canada's website1.
In addition, the SAP-RMD has compiled an extensive list of specific reusable medical devices currently being used in Canada that were deemed to have unique reprocessing concerns (i.e. difficult to clean and or sterilize despite compliance with manufacturers' recommended protocols) and has provided recommendations to both Health Canada and the healthcare facilities for addressing these concerns in the attached table "Reusable Medical Devices: Reprocessing concerns and recommendations of the Scientific Advisory Panel on Reprocessing of Medical Devices".
It should be noted that the information included in the table is not exhaustive and was current at the time of publication.
1 Scientific Advisory Panel on Reprocessing of Medical Devices (SAP-RMD) ( February 10-11, 2005 ) - Panel Recommendations:
http://hc-sc.gc.ca/dhp-mps/md-im/activit/sci-consult/reproc-retraite/saprmd_gcsrmm_recom_2005-02-10-eng.php
General Recommendations to Health Care Facilities:
1. Airway Introducers/Dilators
2. Esophageal Bougies
3. Electro-surgical cauteries/forceps
4. Self-retaining retractors
5. Laryngeal and silicone resuscitator masks
6. Skin Meshers, Dermabrader
7. ET tubes
8. Oral airways (e.g. two-piece)
9. Aerosol spray tips
10. Delicate clamps (vascular clamps - bulldogs, suture holders - serafins)
11. Line Access Stopcocks
12. Valvutome Scissors
13. Internal Defibrillator Paddles
14. Fallopian tube clip applicators
15. Stress incontinence devices (pessaries)
16. Vaginal speculums
17. Laser Vaginal Speculums
Dental/ENT/Facial Maxillary Instruments
18. Punch forceps; Instrument for ethmoid sinus surgery
19. Hand pieces (nasal shaver)
20. Mouth gag
21. Snares
22. Endofiles
23. Accessory devices (e.g. biopsy forceps, sphinctertomes baskets, graspers, snares, sheaths, cautery tips)
24. Flexible endoscopes
25. Fiber-optic mammary retractor
26. Brain retractor
27. Awl
28. External cranio-fixation halo device.
29. Rapid flap applier device
30. Neuro/Spinal sets
31. Aneurysm clips
32. Power Handpieces (saws, drills)
33. Handpiece Accessories (bits, burs, blades)
34. Rigid endoscope sets (includes trocar, stopcocks, etc.)
35. Accessory devices for Minimally Invasive Surgery (e.g. L-hook, biopsy forceps, baskets, snares, sheaths, cautery tips, graspers, scissors, robotic instruments)
36. Tissue fragmenters/aspirators (e.g. liposuction cannulae, cannulated handpieces)
Probes (Critical and Semi-critical)
37. Intra-operative ultrasound probe/transducer (critical device that enters sterile body cavity e.g. liver resection, brain probes, argon probes)
38. Diagnostic or interventional probes (semi-critical device that contacts intact mucus membrane e.g. ultrasonic imaging probes)
39. Needles
40. Ear suctions
41. Dental needles
42. Fine ophthalmology/intra-operative instruments (e.g. Irrigator/Aspirator (I/A) needles, Phaco tips, needles, tubing, conformers)
43. Phaco handpieces
44. Diamond knives
45. Ophthalmology Foreign Body Magnet
46. Lenses
Orthopaedic, Spinal Instruments
47. Handpieces/Drills/Saws (chucks, adapters, cordless, oscillating, reciprocating, saggital, jacobs, reamer attachment, trauma drill, shavers)
48. Handpiece Accessories (bits, burs, blades, saws)
49. Batteries
50. Acetabular reamers
51. Flexible-coiled and cannulated devices (e.g. reamers, drill guides and bits
52. Bone rasps
53. Cement bowls
54. Kerrisons, rongeurs and punches
55. Vice grips
56. External fixation components
57. Intra-medullary nailing system
58. Spinal Systems (e.g. Cable grip systems)
59. Modular fixation systems
60. Shoulder repair, replacement sets
61. Total hip repair or replacement sets
62. Total knee repair or replacement sets
63. Suctions in general
64. Tubing (e.g. Silicone, and latex)
65. Marlex Mesh
66. Aortic and Vascular grafts
1. Follow manufacturer's instructions for Extended Steam Sterilization Cycles:
Medical device manufacturers often indicate steam sterilization times that are outside the cycle parameters normally used in North American healthcare facilities. Users need to be aware that the biological indicators (BIs) and chemical indicators (CIs) that are validated for routine steam cycles cannot be used to monitor the extended steam cycles. To ensure adequate steam sterilization, facilities should contact the manufacturer to determine if standard steam cycles can be used. This confirmation should be in writing. If standard cycles are deemed inadequate, ensure the medical device manufacturer's steam sterilization cycles are followed. (For additional details see: Alfa MJ et al "User Alert: Problems with process monitors for extended steam sterilization cycles" Canadian Journal of Infection Control, Oct 2006).
2. Develop facility-specific reprocessing policy and procedures:
Every healthcare facility that offers patient procedures requiring medical device reprocessing must provide a copy of the device-specific manufacturers' instructions, but in addition there should be site-specific instructions. These should reflect the processes, equipment and products used in that facility. These should include; device handling, cleaning, wrapping, sterilization and/or disinfection instructions with appropriately detailed schematic diagrams or photographs. These need to be updated as changes occur (including site-specific processes, changes in products and equipment, and updates to the device manufacturer's instructions)
3. Plan for adequate turn-around times for device reprocessing:
When scheduling patient procedures, adequate time must be allowed for reprocessing device and instrument sets (including loaner instrument sets). Flash sterilization in the surgical suite to compensate for inadequate medical device inventory is unacceptable because the manufacturers' validated reprocessing protocols cannot be accomplished in this environment. Healthcare facilities must comply with the specific information provided in the following documents; CSA Z314.8, CSA Z314.3, CSA Z314.22 and CSA Z314.1. Failure to do so may jeopardize patient safety and expose the facility to legal liability for not exercising due diligence.
4. Provide formal training to personnel who reprocess medical devices:
Any staff involved in reprocessing of medical devices shall have formal, documented training (including theoretical and practical components) prior to being allowed independent responsibility for reprocessing of medical devices. Continuing education and regular (preferably annual) competency testing of all staff is required (PIDAC, AAMI ST79).
5. Establish procedures for Medical Device Tracking:
Life-span tracking of limited reuse devices:
Reprocessing of limited reuse medical devices must be tracked to ensure that use of the device does not exceed the manufacturers' specified lifespan, since medical devices used beyond their lifespan may fail, leading to adverse patient events. This requires a unique identifier for each instrument and a method to track use and functionality. If a tracking procedure is not provided by the device manufacturer, the healthcare facility should develop one or consider it disposable. If this is not possible, purchase of alternative devices should be considered.
Tracking to facilitate recall:
It is recommended that facilities implement an instrument tracking process to allow for: recall following sterilization failure, alerts issued by manufacturers, retrospective prion-related instrument recalls, and inventory monitoring. Implementation of a medical device tracking system ensures the ability of the facility to provide specific patient notification related to infection control issues.
6. Establish procedures for life-cycle management of medical devices and reprocessing equipment:
Medical devices and reprocessing equipment are often replaced only when they are beyond repair, long after their design has become obsolete. A plan should be made for replacement of old and obsolete devices and equipment. This should be an integral part of the initial purchase process as well as the capital equipment and medical device replacement plans.
7. Ensure notification of adverse events related to medical devices:
To ensure appropriate action, any adverse event associated with the use of a medical device must be reported to the device manufacturer as well as to Health Canada using the appropriate "Health Products and Food Branch Inspectorate Medical Devices Problem Report Form", or by calling the Inspectorate Hotline number 1-800-267-9675 to reach a Medical Device Inspector in your Region. This form along with a guidance document on Voluntary and Mandatory Problem Reporting are available on Health Canada 's website.2
Failure to comply with this may result in adverse events occurring across Canada due to lack of communication or device recall. The healthcare facility may also wish to notify the appropriate regional and provincial authorities.
8. Monitor lifespan for implantable screws, wires and plates:
Facilities should ask device manufacturers for the validated lifespan of implantable screws, wires and plates (i.e. the number of times the item can be cleaned and re-sterilized without compromising functionality).
9. Ensure device materials are compatible with sterilization process:
Facilities should ensure that their protocols consider compatibility of the medical device materials with the mode of sterilization used. Containers, wraps and devices must be validated by the manufacturer as being compatible with the sterilant to be used.
10. Disseminate and use current protocols to manage risks of Creutzfeld-Jakob Disease (CJD) and prions:
Each healthcare facility needs to ensure they have current policies and procedures relating to CJD and high-risk medical procedures that comply with Health Canada Infection Control Guidelines for Prevention of Transmission of Classic CJD. This includes the ability to track instrumentation used for neuro-surgical, spinal and intra-ocular procedures. At a minimum, the policy should include patient screening and appropriate communication regarding the risk status of patients scheduled for neuro-surgical, spinal, and intra-ocular procedures. In addition the facility should have available surgical instruments that can either be discarded or quarantined after the procedure. A protocol for handling and disposal of high-risk tissue is also recommended.
11. Develop validated reprocessing methods for Specialty Equipment:
Specialty equipment, created solely for in-house use, must be reviewed and evaluated by the reprocessing departments, risk management, legal, biomedical engineering, and infection control departments of the healthcare facility. For these devices a process for cleaning, sterilization and functionality testing shall be validated by an external laboratory. The approach outlined above can also be applied when the site decides to continue using a device no longer supported by the manufacturer, if there is no alternative device available.
2 http://www.hc-sc.gc.ca/dhp-mps/compli-conform/prob-report-rapport/md_prob_rep-rap_incident_im-eng.php
http://www.hc-sc.gc.ca/dhp-mps/compli-conform/prob-report-rapport/rep_md_prob-rap_inc_im_tc-tm-eng.php
| Item | Medical Device | *SUMD Available | Reprocessing Concerns | Recommendations |
|---|---|---|---|---|
| General Instruments | ||||
| 1 | Airway Introducers/ Dilators | yes | Porous Material
Limited life span with no expiry date to track. |
To Health Canada: Health Canada shall require manufacturers to:
SUMD recommended for those difficult to clean and track. |
| 2 | Esophageal Bougies | no | Manufacturers' do not provide instructions or validated methods for permeation of the narrow lumens with disinfectant or for rinsing. Porous material absorb liquid chemical disinfectants (LCD). Manufacturers' recommended LCD not appropriate for device reprocessing in healthcare facilities. (e.g. povidone iodine, bleach) Mercury or tungsten gel filled
|
To Health Canada: Health Canada shall require manufacturers to:
Develop a plan to remove from use the first generation devices that pose a risk to the patient. Recommend use of tungsten filled esophageal dilators. If used, do not autoclave mercury-filled esophageal bougies.4 Expiration date and integrity of device must be reviewed each time reprocessed. |
| 3 | Electro-surgical cauteries/forceps | Yes for some | Difficult to clean due to:
Sterilization instructions for some devices do not meet typical sterilization cycles used in Canadian health care facilities |
To Health Canada: Health Canada shall require manufacturers to:
SUMD recommended for those devices that are difficult to clean. SUMD recommended for limited use devices that have no unique identifier. |
| 4 | Self-retaining retractors | no | Difficult to clean due to:
|
To Health Canada: Health Canada shall require manufacturers to:
Heavy retractors should be separated into two instrument trays for sterilization. |
| 5 | Laryngeal and silicone resuscitator masks | Yes for some | Material is porous therefore absorbs detergents and/or liquid chemical disinfectants (LCD) which weakens the products. Materials are thermal labile therefore cannot be subjected to excessive heat (drying or pasteurization). Healthcare facilities often not following complex reprocessing instructions. |
To Healthcare Facilities: SUMDs are recommended if the health care facility is unable to follow the detailed instructions. Detailed and documented staff training and written procedures for reprocessing are required. |
| 6 | Skin Meshers, Dermabrader | No (Yes, single use blade available for dermabrader) |
Reprocessing instructions are:
Manufacturers advise against dismantling the ratchet handles despite the fact that debris gets trapped inside. Handling during cleaning and attempts to dismantle can cause lacerations and is an occupational health risk Regular servicing by the manufacturers is required, but is often overlooked by users because of confusing manufacturer instructions. |
To Health Canada: Health Canada shall require manufacturers to:
Ensure preventative maintenance of devices as per manufacturers' directions. Staff training required by manufacturers. Detailed competency testing of staff is required. |
| 7 | ET tubes | yes | Material is porous therefore absorbs detergents and/or liquid chemical disinfectants (LCD). Cuff is difficult to clean and deterioration from repeated reprocessing may lead to rupture which poses a patient safety issue. |
To Healthcare Facilities: SUMD recommended. |
| 8 | Oral airways (e.g. two-piece) | yes | Difficult to clean due to:
|
To Healthcare Facilities: SUMD recommended. |
| 9 | Aerosol spray tips | yes | Difficult to clean due to
|
To Health Canada: Health Canada shall require manufacturers to:
SUMD recommended. |
| 10 | Delicate clamps (vascular clamps - bulldogs, suture holders - serafins) | yes | Difficult to clean and sterilize due to:
|
To Healthcare Facilities: SUMD recommended. |
| 11 | Line Access Stopcocks | yes | Difficult to clean and sterilize due to:
|
To Healthcare Facilities: SUMD recommended. |
| 12 | Valvutome Scissors | no | First generation devices cannot be cleaned due to:
|
To Health Canada: Health Canada shall require manufacturer to:
Remove first generation devices from use. Newer models exist and should be considered to replace first generation devices. |
| 13 | Internal Defibrillator Paddles | yes | Manufacturer recommends different sterilization methods for the paddle and handles therefore increasing adverse outcomes due to delays in an emergency situation because assembly prior to use is required. Although manufacturers have recommended partial assembly before sterilization there are still difficulties because device is used in emergency situations where time for assembly is not feasible. For limited use devices, no unique identifier is provided on device components for tracking/testing purposes. Limited number of reuses but these devices do not always last as long as manufacturers' recommended number of uses. |
To Health Canada: Health Canada shall require manufacturers to:
To Healthcare Facilities: As per manufacturers' instructions:
|
| Item | Medical Device | *SUMD Available | Reprocessing Concerns | Recommendations |
|---|---|---|---|---|
| Dental/ENT/Facial Maxillary Instruments | ||||
| 18 | Punch forceps; Instrument for ethmoid sinus surgery | no | Difficult to clean due to:
|
To Health Canada: Health Canada shall require manufacturers to:
Detailed and documented staff training and written procedures for reprocessing are required. |
| 19 | Hand pieces (nasal shaver) | no | First generation devices cannot be cleaned due to:
|
To Health Canada: Health Canada shall require manufacturers to:
Develop a plan to remove from use the first generation devices that pose a risk to the patient. |
| 20 | Mouth gag | no | Difficult to clean due to:
|
To Health Canada: Health Canada shall require manufacturers to:
Detailed and documented staff training and written procedures for reprocessing are required. |
| 21 | Snares | No (except wires) | Difficult to clean due to:
|
To Health Canada: Health Canada shall require manufacturers to:
Detailed and documented staff training and written procedures for reprocessing are required. Reprocessing personnel must demonstrate competency in assembly of nasal and tonsil snares. |
| 22 | Endofiles | No | Difficult to clean due to:
|
To Healthcare Facilities: Detailed and documented staff training and written procedures for reprocessing are required. |
| 23 | Accessory devices (e.g. biopsy forceps, sphinctertomes baskets, graspers, snares, sheaths, cautery tips) | yes | Difficult to clean due to
Reprocessing instructions are:
|
To Healthcare Facilities: SUMD recommended. |
| 24 | Flexible endoscopes | no | Staff failing to clean all scope channels due to:
Problems associated with Automated Endoscope Reprocessors (AER):
Lack of appropriate storage of scopes and/or components post-reprocessing. |
To Health Canada: Address and take into consideration the issues mentioned when considering new devices for license approval. To Healthcare Facilities: Ensure problems associated with AER are reported to Health Canada. Ensure that all staff (including after-hours staff) assigned to reprocess endoscopes are appropriately trained and ongoing competency is verified for all models and makes of scopes being used. Staffing levels should be adequate to ensure complete reprocessing of endoscopes for after-hours emergency procedures (e.g. scopes not left soaking in detergent or AER for prolonged periods and are dried and put into storage). All sites should have a tracking system that allows linkage between patient, procedure, AER and scope. |
| 25 | Fiber-optic mammary retractor | no | Reprocessing instructions are:
|
To Health Canada: Health Canada shall require manufacturers to:
Detailed and documented staff training and written procedures for reprocessing are required. |
| 26 | Brain retractor | no | Difficult to clean due to:
|
To Health Canada: Health Canada shall require manufacturers to:
Must have detailed staff training (e.g. in-service). Ongoing education needed. Detailed written procedures for reprocessing required. Time for proper cleaning needs to be factored in the turnaround time. |
| 27 | Awl | No | Difficult to clean due to:
|
To Health Canada: Health Canada shall require manufacturers to:
|
| 28 | External cranio-fixation halo device | No | Difficult to clean due to:
|
To Health Canada: Health Canada shall require manufacturers to:
Detailed and documented staff training and written procedures for reprocessing are required. |
| 29 | Rapid flap applier device | Unknown | Difficult to clean due to:
|
To Health Canada: Health Canada shall require manufacturers to:
Detailed and documented staff training and written procedures for reprocessing are required. |
| 30 | Neuro/Spinal sets | No | Reprocessing instructions are:
Sterilization instructions for some devices do not meet typical sterilization cycles used in Canadian health care facilities. |
Health Canada: Health Canada shall require manufacturers to:
Must have detailed staff training (e.g. in-service). Ongoing education needed. Detailed written procedures for reprocessing required. Time for proper cleaning needs to be factored in the turnaround time. Policies and procedures should comply with HC CJD Infection Control Guidelines for reprocessing of devices used on high-risk tissues. 7, 8 |
| 31 | Aneurysm clips | yes | Clips are very fine and delicate making it hard to assess which ones are soiled. Each clip must be put into applier and spread to open position to see which is dirty: this can damage the device. |
To Healthcare Facilities: Trialed clips to be disposed of at point of use due to contamination and weakening of tensile strength. |
| 32 | Power Handpieces (saws, drills) | No | Difficult to clean due to:
Manufacturers' recommendations for preventative maintenance are unclear and often not followed by users. |
To Health Canada: Health Canada shall require manufacturers to:
Ensure that manufacturers' preventative maintenance program is followed. Detailed and documented staff training and written procedures for reprocessing are required. |
| 33 | Handpiece Accessories (bits, burs, blades) | yes | Users typically are not equipped to adequately assess device damage or cleanliness after reprocessing. | To Healthcare Facilities: SUMD recommended. If users are reprocessing these devices the following should be met: |
| Item | Medical Device | *SUMD Available | Reprocessing Concerns | Recommendations |
|---|---|---|---|---|
| Probes (Critical and Semi-critical) | ||||
| 37 | Intra-operative ultrasound probe/transducer (critical device that enters sterile body cavity e.g. liver resection, brain probes, argon probes) | no | Difficult to clean some devices due to:
Users not reprocessing device because of misconception that sheathing precludes the need to clean and sterilize. Users not doing leak test (where applicable) prior to reprocessing probe as per manufacturers' recommendations. Limited uses therefore this device requires the development of a tracking system. |
To Health Canada: Health Canada shall require manufacturers' to:
Sterilization is required for intra-operative probes; a sterile sheath does not eliminate the need to sterilize the device. Leak testing for all probes prior to cleaning as per manufacturers' instructions. Detailed written procedures for cleaning and sterilization based on manufacturers' recommendations. Extensive in-service by the manufacturer. Staff competency testing needs to be on-going. A tracking system must be utilized for limited reuse devices. |
| 38 | Diagnostic or interventional probes (semi-critical device that contacts intact mucus membrane e.g. ultrasonic imaging probes) | no | Users are inappropriately rinsing with tap water after HLD.9, 10 In some cases, users may be performing HLD in inappropriate settings (e.g. Diagnostic Imaging (DI) rooms, lack of ventilation). |
To Healthcare Facilities: Staff training required by manufacturers. Detailed competency testing of staff is required. Written procedures are required. HLD is required for diagnostic or interventional probes; a condom or sterile sheath does not eliminate the need to HLD the device. When biopsies are performed a sterile sheath is required. These probes must be rinsed with sterile distilled water after HLD. |
| 39 | Needles | yes | Reprocessing of needles poses a risk of needle-stick injury to staff. Reprocessing instructions are:
|
To Health Canada: Health Canada shall require manufacturers' to:
SUMD recommended. If reusable, must have appropriate cleaning materials (e.g. brushes, sharpening stone, water gun). |
| 40 | Ear suctions | unknown | Size 22-24 lumens are hard to clean. | To Health Canada: Health Canada shall require manufacturers' to:
Lobby manufacturers to develop SUMD. |
| 41 | Dental needles | yes | Reprocessing instructions are:
|
To Health Canada: Health Canada shall require manufacturers' to:
SUMD recommended. |
| Item | Medical Device | *SUMD Available | Reprocessing Concerns | Recommendations |
|---|---|---|---|---|
| Ophthalmologic Instruments | ||||
| 42 | Fine ophthalmology /intra-operative instruments (e.g. Irrigator/ Aspirator (I/A) needles, Phaco tips, needles, tubing, conformers) | Yes for some | Difficult to clean due to:
Manufacturers do not recommend a cleaning agent. Sterilization instructions do not meet typical sterilization cycles used in Canadian health care facilities. Impossible to track those devices that are limited reuse due to lack of unique identifier. Reprocessing poses the following risk to staff:
Users that perform manual rinsing are often not providing adequate rinse volume because it is difficult to deliver the copious volumes as recommended by manufacturer. Ultrasonic cleaning (where recommended) is not sufficient alone. |
To Health Canada: Health Canada shall require manufacturers to:
Recommend users acquire the manufacturer recommended automated rinsing equipment. Lobby manufacturers to develop and provide SUMD. If reusable, must have appropriate cleaning materials, staff training. Written procedures for reprocessing required. SUMD (where available) recommended. |
| 43 | Phaco handpieces | no | Difficult to clean due to:
Manufacturers do not recommend a cleaning agent. Sterilization instructions do not meet typical sterilization cycles used in Canadian health care facilities. Reprocessing poses the following risk to staff due to repetitive stress injury. Improper cleaning/rinsing (i.e. any inorganic or organic residuals left) poses risk of Toxic Anterior Segment Syndrome (TASS). 11 Users that perform manual rinsing are often not providing adequate rinse volume because it is difficult to deliver the copious volumes as recommended by manufacturer. |
To Health Canada: Health Canada shall require manufacturers to:
Recommend users acquire the manufacturer recommended automated rinsing equipment. Lobby manufacturers to develop and provide a reusable device for which manufacturers have validated the cleaning and sterilization protocols. If reusable, must have appropriate cleaning materials and staff training. Written procedures for reprocessing required. |
| 44 | Diamond knives | no | Extremely fragile (frequently damaged during transport) and expensive, typically owned by surgeons. Not routinely cleaned because it is not part of hospital inventory or inventory levels are inadequate for volume of cases scheduled. Users frequently flash sterilize without prior cleaning (using manufacturers validated instructions) due to lack of inventory, unwillingness to transport to reprocessing department and surgeon-owned instruments not available until just prior to surgery. |
To Healthcare Facilities: Work with surgeons to provide appropriate reprocessing that complies with national standards. Ensure devices are reprocessed using manufacturers' recommended reprocessing protocol. Ensure adequate inventory for volumes of cases scheduled. Flash sterilization does not comply with national guidelines when used for elective surgery (i.e. not an emergency situation). Recommend acquiring dedicated protective containers for transport and sterilization of device. |
| 45 | Ophthalmology Foreign Body Magnet | no | Difficult to clean due to:
Users that perform manual rinsing are often not providing adequate rinse volume because it is difficult to deliver the copious volumes as recommended by manufacturer. |
To Health Canada: Health Canada shall require manufacturers to reassess design and provide validated reprocessing instructions. To Healthcare Facilities: Users must properly inspect reprocessed devices to ensure no ingress of patient material. Users must ensure adequate volume of sterile distilled water is used for rinsing. |
| 46 | Lenses | no | Reprocessing instructions are:
Due to no company identifier on the device, users are often confused as to what company's reprocessing instructions to follow. |
To Health Canada: Health Canada shall require manufacturers to:
Staff training specific to the device. Written procedures are required. |
| Item | Medical Device | *SUMD Available | Reprocessing Concerns | Recommendations |
|---|---|---|---|---|
| Orthopaedic, Spinal Instruments | ||||
| 47 | Handpieces/ Drills/Saws (chucks, adapters, cordless, oscillating, reciprocating, saggital, jacobs, reamer attachment, trauma drill, shavers) | no | Difficult to clean due to:
Manufacturers recommendations for preventative maintenance are unclear and often not followed by users. |
To Health Canada: Health Canada shall require manufacturers to:
Ensure that manufacturers' preventative maintenance program is followed. Detailed written procedures for cleaning and sterilization based on manufacturers' recommendations. Extensive in-service by the manufacturer. On-going internal competency testing. |
| 48 | Handpiece Accessories (bits, burs, blades, saws) | yes | Users, typically, are not equipped to adequately assess device damage or cleanliness after reprocessing. | To Healthcare Facilities: SUMD recommended. If users are reprocessing these devices the following should be met:
|
| 49 | Batteries | no | Manufacturers' instructions for flash sterilization create problems with maintenance of the sterile technique. Some manufacturers recommend recharging battery immediately prior to surgery therefore potentially delaying surgery. |
To Health Canada: Health Canada shall require manufacturers to:
Ensure that manufacturers' preventative maintenance program is followed. Detailed written procedures for cleaning and sterilization based on manufacturers' recommendations. Recommend utilization of batteries that can be sterilized by traditional methods. (wrapped) |
| 50 | Acetabular reamers | no | Difficult to clean due to:
Instructions for frequency of sharpening not provided for cutting instruments. |
To Health Canada: Health Canada shall require manufacturers to:
Recommend that reamers are sharpened by qualified instrument technicians. Detailed written procedures for cleaning and sterilization based on manufacturers' recommendations. |
| *51 | Flexible-coiled and cannulated devices (e.g. reamers, drill guides and bits | (Yes for reamers and drill bits ) | Difficult to clean due to:
|
To Health Canada: Health Canada shall require manufacturers to:
SUMD recommended when available. Develop a plan to remove from use the first generation devices that pose a risk to the patient. Recommend flexible solid shaft reamers. If reusable, must have documented in-service training from the manufacturers and appropriate cleaning materials. Detailed written procedures for cleaning and sterilization based on manufacturers' recommendations. Flexible coiled reamers and guides should be placed into detergent and/or water solution immediately after use (in Operating Room). |
| 52 | Bone rasps | no | Difficult to clean due to:
|
To Health Canada: Health Canada shall require manufacturers to:
|
| 53 | Cement bowls | yes | Difficult to clean due to:
|
To Healthcare Facilities: SUMD recommended |
| 54 | Kerrisons, rongeurs and punches | No | Difficult to clean due to:
|
To Healthcare Facilities: Develop a plan to remove from use the first generation devices that pose a risk to the patient and replace with new generation that disassemble for cleaning. Detailed written procedures for cleaning and sterilization based on manufacturers' recommendations. |
| 55 | Vice grips | No | Difficult to clean due to:
|
To Health Canada: Health Canada shall require manufacturers to:
Detailed written procedures for cleaning and sterilization based on manufacturers' recommendations. |
| 56 | External fixation components | Some fixator pins and locking bolts are disposable | Impossible to track those devices that are limited reuse due to lack of unique identifier. Users cannot easily identify fixator pins that are SUMD and have been exposed to secretions/tissue and reprocessed. |
To Health Canada: Health Canada shall require manufacturers to:
Discard all SUMDs and/or implanted devices that are exposed to or trialed on a patient at time of procedure. Discard all SUMDs and/or implanted components at the time of removal from a patient (e.g. In O.R. or clinic setting). |
| 57 | Intra-medullary nailing system | No (Yes for flexible reamer ) |
Difficult to clean due to:
Sterilization instructions do not meet typical sterilization cycles used in Canadian health care facilities. Inability to assess the extent of disassembly required for multiple parts. Possible metal fatigue due to multiple reprocessing. Unable to track implant to patient. Instructions for frequency of sharpening not provided for cutting instruments. |
To Health Canada: Health Canada shall require manufacturers to:
Staff training by manufacturers is required. Detailed written procedures including:
Reusable components should be placed into detergent and/or water immediately after use (in OR). |
| 58 | Spinal Systems (e.g. Cable grip systems) | No | Difficult to clean due to:
|
To Health Canada: Health Canada shall require manufacturers to:
Must have detailed staff training (e.g. in-service). Ongoing education needed. Detailed written procedures for reprocessing required. |
| 59 | Modular fixation systems | no | Difficult to clean due to:
Inability to assess the extent of disassembly required for reprocessing for depth gauges and drill guides. End of life due to metal fatigue is unknown. Instructions for frequency of sharpening not provided for cutting instruments. Unable to track implant to patient. |
To Health Canada: Health Canada shall require manufacturers to:
Detailed written procedures for cleaning and sterilization are required with extensive in-service and competency testing. Staff training by manufacturers required. Detailed written procedures including:
Discard all implantables that have been in contact with patient secretions/tissue or trialed on patient at time of procedure. |
| 60 | Shoulder repair, replacement sets | No (Yes for drill bits ) |
Difficult to clean due to:
Inability to assess the extent of disassembly required for reprocessing for depth gauges and drill guides. The end of life for the medical device due to repeated reprocessing has not been determined. Instructions for frequency of sharpening (awl and drill bits) not provided. Unable to track implant to patient. |
To Health Canada: Health Canada shall require manufacturers to:
Detailed written procedures for cleaning and sterilization are required with extensive in-service and competency testing. Staff training by manufacturers required. Detailed written procedures including
|
| 61 | Total hip repair or replacement sets | No (Yes for guide pins and depth gauges ) |
Difficult to clean due to:
Inability to assess the extent of disassembly required for reprocessing (depth gauges, drill guides, flexible reamer, impactor, slap hammer). Handling during cleaning and attempts to dismantle can cause lacerations and is an occupational health risk. Instructions for frequency of sharpening (broaches and graters) not provided. End of life due to metal fatigue unknown. Unable to track implant (hip pinning) to patient. |
To Health Canada: Health Canada shall require manufacturers to:
Detailed written procedures for cleaning and sterilization are required with extensive in-service and competency testing. Staff training by manufacturers required. Detailed written procedures including:
To ensure adequate turnaround times (includes disassembly, cleaning, assembly, functionality testing, sterilization, and all other steps required for total reprocessing), it is critical to ensure that inventory of devices is adequate. See general r ecommendations. Reusable components should be placed into detergent and/or water immediately after use (in OR). |
| 62 | Total knee repair or replacement sets | No | Difficult to clean due to:
Instructions for frequency of sharpening (cutting block pins) not provided for cutting instruments. |
To Health Canada: Health Canada shall require manufacturers to:
Staff training by manufacturers required. Detailed written procedures including:
Reusable components should be placed into detergent and/or water immediately after use (in OR). |
References
1. CAN/CSA-Z314.8-00 (R2005): Decontamination of Reusable Medical Devices
2.Ontario Best Practice Manual: Cleaning, Disinfection and Sterilization In All Health Care Settings: http://www.health.gov.on.ca/english/providers/program/infectious/diseases/ic_cds.html
3. Health CanadaInfection Control Guidelines - Hand Washing, Cleaning, Disinfection and Sterilization in Health Care: http://www.phac-aspc.gc.ca/publicat/ccdr-rmtc/98pdf/cdr24s8e.pdf
4. Reprocessing of Mercury-Filled Esophageal Dilators: http://www.hc-sc.gc.ca/dhp-mps/medeff/advisories-avis/prof/2006/esoph-oesoph_dilat_nth-aah-eng.php
5. CAN/CSA-Z17664-06: Sterilization of Medical Devices - Information to be Provided by the Manufacturer for the Processing of Resterilizable Medical Devices (Adopted ISO 17664:2004, first edition, 2004-03-01)
6. CAN/CSA-Z314.15-03 Warehousing, Storage, and Transportation of Clean and Sterile Medical Devices:
7. Health Canada Infection Control Guidelines -. Classic Creutzfeldt-Jakob Disease in Canada: http://www.phac-aspc.gc.ca/publicat/ccdr-rmtc/02vol28/28s5/index.html
8. Health Canada Infection Control Guidelines -update.
9. MMWR Weekly, July 21, 2006 / 55(28);776-777, Pseudomonas aeruginosa Infections Associated with Transrectal Ultrasound-Guided Prostate Biopsies --- Georgia, 2005: http://www.cdc.gov/mmwr/preview/mmwrhtml/mm5528a3.htmy)
10. FDA Public Health Notification: Reprocessing of Reusable Ultrasound Transducer Assemblies Used for Biopsy Procedures,Issued: June 19, 2006 Updated: June 22, 2006: http://www.fda.gov/cdrh/safety/061906-ultrasoundtransducers.html
11. Toxic Anterior Segment Syndrome Outbreak Preliminary Report: http://www.ascrs.org/press_releases/Toxic-Anterior-Segment-Syndrome-Outbreak-Preliminary-Report.cfm
12 Inadequate Cleaning/Sterilization of EZ Clean Monobloc ACETABULAR REAMER a reuseable medical device:
http://www.hc-sc.gc.ca/dhp-mps/medeff/advisories-avis/prof/2004/acetabular_reamer_nth-ah-eng.php