Contact: Device Licensing
Date Adopted 1999/01/06
Administrative Changes Date 2011/11/22
Effective Date 2011/04/01
Également disponible en français sous le titre: Comment compléter une nouvelle demande d'homologation pour un instrument médical
Guidance documents are meant to provide assistance to industry and health care professionals on how to comply with the policies and governing statutes and regulations. They also serve to provide review and compliance guidance to staff, thereby ensuring that mandates are implemented in a fair, consistent and effective manner.
Guidance documents are administrative instruments not having force of law and, as such, allow for flexibility in approach. Alternate approaches to the principles and practices described in this document may be acceptable provided they are supported by adequate scientific justification. Alternate approaches should be discussed in advance with the relevant program area to avoid the possible finding that applicable statutory or regulatory requirements have not been met.
As a corollary to the above, it is equally important to note that Health Canada reserves the right to request information or material, or define conditions not specifically described in this guidance, in order to allow the Department to adequately assess the safety, efficacy or quality of a therapeutic product. Health Canada is committed to ensuring that such requests are justifiable and that decisions are clearly documented.
This document should be read in conjunction with the accompanying notice and the relevant sections of other applicable guidance documents.
This document describes how to complete an application for a new medical device licence for Class II, III and IV medical devices, including payment of applicable licence fees.
The Medical Devices Regulations (MDR) have been established under the authority of the Food and Drugs Act and apply to all medical devices imported or sold in Canada. The Therapeutic Products Directorate (TPD) has the regulatory responsibility for medical devices.
The MDR set out the requirements governing the sale, importation, and advertisement of medical devices in Canada. The level of regulatory scrutiny to be applied in these areas is based on risk management principles that uses a rule-based system to classify medical devices into four classes, with Class I representing the lowest risk and Class IV the highest.
For the purpose of the MDR, the term "medical device" means a device within the meaning of the Food and Drugs Act, but does not include any device that is intended for use in relation to animals.
With the intent to ensure the safety and effectiveness of devices sold in Canada, Part 1, section 26 of the MDR prohibits the importation or sale of a Class II, III or IV medical device unless the manufacturer holds a licence for that device.
BISPHENOL A [BPA; Phenol, 4,4' -(1-methylethylidene)bis-] is an industrial raw material that was identified for screening assessment under the Canadian Environmental Protection Act (CEPA, 1999). BPA is primarily used as a raw material in the production of polycarbonates and epoxy resins. BPA or BPA-based polymers are used in the manufacture of a variety of medical devices, including resin-based dental composite restorative and prosthodontic materials, dental sealants, hemodialyzers, hemofilters and blood oxygenators. Please refer to Table 1 for the chemical identity of BPA, including its Chemical Abstracts Services (CAS) Registry Number and synonyms.
| Characteristic | Information |
|---|---|
| Chemical Abstracts Services (CAS) Registry Number | 80-05-7 |
| Domestic Substances List name | phenol, 4,4' -(1-methylethylidene)bis- |
| National Chemical Inventories (NCI) names1 | phenol, 4,4'-(1-methylethylidene)bis- (TSCA, PICCS, ASIA-PAC) 4,4'-isopropylidenediphenol (EINECS, PICCS) 2,2-Bis(4'-hydroxyphenyl) propane (ENCS) phenol, 4,4'-(1-methylethylidene)bis- (AICS, PICCS) 4,4'-(1-Methylethylidene)bisphenol (ECL) 4,4'-Bisphenol A (ECL) phenol, 4,4'-(1-methylethylidene)bis- (SWISS) bisphenol A (SWISS, PICCS) p,p'-isopropylidene diphenol (PICCS) diphenol methylethylidene (PICCS) bis[phenol], 4,4'-(1-methylethylidene)- (PICCS) bisphenol-a (PICCS) bisphenol, 4,4'-(1-methylethylidene)- (PICCS) 4,4-isopropylidene diphenyl (PICCS) 4,4'-dihydroxyphenyl-2,2-propane (PICCS) 2,2-di(4-hydroxyphenyl)propane (PICCS) 2,2-di(4-hydroxyphenyl) propane (PICCS) 2,2-bis-(4-hydroxy-phenyl)-propane (PICCS) |
| Other names | bisphenol A diphenylolpropane BPA |
| Chemical group | Discrete organics |
| Chemical subgroup | Phenols |
| Chemical formula | C15H16O2 |
| Chemical structure | ![]() |
| SMILES | Oc(ccc(c1)C(c(ccc(O)c2)c2)(C)C)c1 |
DEVICE ID refers to the device identification number assigned by Health Canada.
DI(2-ETHYLHEXYL)PHTHALATE (DEHP) is a chemical additive that is used to make polyvinyl chloride (PVC) soft, flexible and kink resistant. PVC plasticized with DEHP is currently used in a variety of medical devices, including blood bags, catheters, intravenous tubing and medical gloves. A medical device shall be considered to contain DEHP if the amount of DEHP in the device is more than or equal to 0.1% of the device's mass (that is [i.e.], ≥ 0.1% w/w). Please refer to Table 2 for the chemical identity of DEHP, including its Chemical Abstracts Services (CAS) Registry Number, synonyms and known trade names.
| Characteristic | Information | Reference |
|---|---|---|
| Chemical name | di(2-ethylhexyl) phthalate | RTECS 2000 |
| Synonyms | DEHP dioctylphthalate bis(2-ethylhexyl) phthalate |
RTECS 2000 |
| Registered trade names | Bisoflex 81 Eviplast 80 Octoil Plantinol DOP Staflex DOP |
RTECS 2000 |
| Chemical formula | C24H38O4 | RTECS 2000 |
| Chemical structure | ![]() |
Howard and Meylan, 1997 |
| Identification numbers: CAS Registry Number NIOSH RTECS EPA hazardous waste OHM/TADS DOT/UN/NA/IMCO shipping HSDB NCI |
117-81-7 TI0350000 U028 7216693 No data 334 C52733 |
HSDB 1990 |
CAS = Chemical Abstracts Services; DOT/UN/NA/IMCO = Department of Transportation/United Nations/North America/International Maritime Dangerous Goods Code; EPA = Environmental Protection Agency; HSDB = Hazardous Substances Data Bank; NCI = National Cancer Institute; NIOSH = National Institute for Occupational Safety and Health; OHM/TADS = Oil and Hazardous Materials/Technical Assistance Data System; RTECS = Registry of Toxic Effects of Chemical Substances.
IDENTIFIER means a unique series of letters or numbers or any combination of these or a bar code that is assigned to a medical device by the manufacturer and that identifies it and distinguishes it from similar devices. Examples of an identifier for a device are a catalogue, model or part number.
LICENCE APPLICATION TYPE means the application can be submitted either as a single device, a system, a test kit, a device group, a device family or a device group family. The term "test kit" applies only to in vitro diagnostic devices. For more information on licence application types, refer to the Guidance for the Interpretation of Sections 28 to 31: Licence Application Type, which is available on the website.
MANUFACTURER means a person who sells the medical device under their own name, or under a trade-mark, design, trade name or other name or mark owned or controlled by the person, and who is responsible for designing, manufacturing, assembling, processing, labelling, packaging, refurbishing or modifying the device, or for assigning to it a purpose, whether those tasks are performed by that person or on their behalf. "Person" includes a partnership, firm or association.
1. National Chemical Inventories (NCI). 2006: AICS (Australian Inventory of Chemical Substances); ASIA-PAC (Asia-Pacific Substances Lists); ECL (Korean Existing Chemicals List); EINECS (European Inventory of Existing Commercial Chemical Substances); ENCS (Japanese Existing and New Chemical Substances); PICCS (Philippine Inventory of Chemicals and Chemical Substances); SWISS (Inventory of Newly Notified Substances and Giflist 1 - List of Toxic Substances) and TSCA (Toxic Substances Control Act Chemical Substance Inventory).
Medical devices imported or sold in Canada, with the exception of Class I medical devices, are required to be licensed prior to being imported or sold in Canada. One device licence will be issued to the manufacturer for each licence application submitted, if it is determined, on the basis of the information submitted, that it meets the safety and effectiveness requirements of the MDR.
To apply for a new medical device licence, ensure that:
Device Licensing Services Division (DLSD)
Medical Devices Bureau (MDB)
Therapeutic Products Directorate (TPD)
Health Canada
2934 Baseline Road
Address Locator: 3403A
Ottawa, Ontario
K1A 0K9
Canada
Under the MDR, a device licence is a pre-market requirement for the following:
The classification of your medical device is dependent upon its intended use. If a product is a Class I device, a medical device licence is not required. The rules to classify medical devices are outlined in Schedule 1 (Parts 1 and 2) of the MDR. Part 1 of Schedule 1 addresses medical devices other than in vitro diagnostics and Part 2 addresses in vitro diagnostic devices. Refer to the documents Guidance for the Risk-based Classification System, or Guidance for the Risk-based Classification System of In Vitro Diagnostic Devices, as appropriate, for further guidance on the classification of medical devices. Once you have ascertained the class of the your device, use the appropriate application form (available on the website):
New Class II Medical Device Licence Application Form,
New Class III Medical Device Licence Application Form; or
New Class IV Medical Device Licence Application Form.
The document entitled Keyword Index to Assist Manufacturers in Verifying the Class of Medical Devices is an alphabetical listing of all the words that appear in all the short descriptors for all the devices that are entered into the Medical Devices System (MDS). The document contains synonyms and industry words that are commonly used to describe these devices, along with their respective classifications. This document is available on the website for your reference.
Item 1: Name Of The Device (as it appears on the label)
The device name indicated for a system, medical device family or a medical device group family must appear, at least in part, on the label of each of the member devices. Only one name is to be entered in Item 1. The device name on the application form will be used as the licence name unless you are applying for a family of medical devices. In this case, a generic licence name, such as urinary catheters, should be indicated to cover all the possible trade names that you sell under urinary catheters. The licence name usually reflects the types of devices that are contained within the licence and sometimes may vary from the device name.
Item 2: manufacturer Information (as it appears on the label)
This is the name and address of the manufacturer of the device and the name and address to which the licence will be issued. A complete address must include: name and title of a contact person; Company ID (if known, this number is assigned by Health Canada); telephone number, facsimile number and e-mail address of the contact person; street name and number or Post Office Box; city, province or state; postal or zip code; and country.
Item 3: Regulatory Correspondent Information
All regulatory correspondence will be sent to this address (if different from Item 2), but the licence will be issued to the Manufacturer. A medical device licence application can be submitted by a third party; the mailing address and name of this authorized Regulatory Correspondent will be entered here.
Item 4: invoicing Information
Enter the name, address and contact information of the party which will receive all invoicing and billing information; it may be the same as Item 2 or 3, or it may be a third party.
Item 5: Quality Management System Certificate
Enter the certificate number and the name of the recognized registrar that has issued the certificate. A legible copy of the certificate must accompany each medical device licence application. For more information on the content and acceptance of quality management system certificates, refer to GD207: Guidance on the content of ISO 13485 quality management system certificates issued by Health Canada recognized Registrars. The certificate must be issued by a Health Canada recognized registrar. Refer to the Health Canada website for a current list of recognized registrars.
Item 6: Attestations
Attestation of Safety and Effectiveness Requirements
As required by Sections 10 to 20 of the MDR, manufacturers of Class II medical devices must attest that they have objective evidence establishing the safety and effectiveness of their device.
Attestation of Labelling Requirements
Manufacturers of a Class II medical device must attest that their device meets the labelling requirements as required by Sections 21 to 23 of the MDR. Refer to the documents Guidance for the Labelling of Medical Devices and Guidance for the Labelling of In Vitro Diagnostic Devices, which are available on the website.
Attestation of Investigational Testing for In Vitro Diagnostic Devices (IVDDs)
Manufacturers of Class II near patient IVDDs must attest here that investigational testing of their device was conducted using human subjects representative of the intended patients and under conditions similar to the intended conditions of use of the device.
Near Patient Attestation
Manufacturers are to attest here that the device(s) is NOT a near patient IVDD.
A "near patient in vitro diagnostic device" means an in vitro diagnostic device (IVDD) that is intended for use outside a laboratory, for testing at home or at the point of care, such as a pharmacy, a health care professional's office or the bedside.
Signature
The manufacturer of the device must sign and date the application.
Evidence of safety and effectiveness
Along with the application form, a manufacturer must submit the information requested in section 32, subsection (3) or (4) of the MDR. Refer also to the document Guidance Document - Preparation of a Premarket Review Document for Class III and Class IV Device Licence Applications.
Signature
The manufacturer of the device must sign and date the application.
Item 7: purpose or Intended Use of Device
Information provided for Item 7 is crucial to establishing the appropriate device class and should include the following:
Insufficient detail to allow verification of the classification of the device may result in the application not being accepted for review.
Item 8: Licence Application Type
A manufacturer may apply for the following types of device licence:
A single medical device:
A single medical device is defined by a unique device name, is sold as a distinct packaged entity and does not meet the criteria for a medical device group, a medical device family, a medical device group family, a system, or a test kit. It may be offered in a range of package sizes, and is represented by a unique device name. Examples include: an acupuncture needle, an aneurysm clip, a larynx prosthesis or dental cement.
A medical device family:
A medical device family is a group of medical devices that are made by the same manufacturer, that differ only in shape, colour, flavour or size, that have the same design and manufacturing process and that have the same intended use. Examples include: intra vascular catheters, insulin syringes, feeding tubes or vascular access grafts.
A medical device group:
A medical device group is comprised of a collection of medical devices, such as a procedure pack or tray, that is sold under a single name. Examples include: a denture repair kit, a declotting tray, a parenteral administration kit or disposable circumcision tray.
A medical device group family:
A medical device group family is a collection of medical device groups that are made by the same manufacturer, that have the same generic name specifying their intended use, and that differ only in the number and combination of products that comprise each group. Examples include: IV administration sets, dressing trays, contact lens care kits or irrigation trays.
System:
A system is a medical device comprising a number of components or parts intended to be used together to fulfil some or all of the device's intended functions, and that is sold under a single name and are manufactured by the same manufacturer. Examples include hip prostheses, knee prostheses or an ultrasonic imaging system.
Test Kit:
A test kit is an in vitro diagnostic device that consists of reagents or articles, or any combination of these, and that is intended to be used to conduct a specific test.
For further assistance in ascertaining the appropriate licence Application type for your product, consult the Guidance for the Interpretation of Sections 28 to 31: Licence Application Type.
Item 9: place of Use
Indicate on the application form by checking the appropriate boxes.
Item 10: Medical Devices Containing Drugs
Non-IVD Devices Containing Drugs
Do not complete this item if the device is an IVDD. If the device contains a drug, which includes a pharmaceutical or biological drug, or a natural health product, specify its brand or trade name, active ingredient(s), manufacturer, and its Drug Identification Number (DIN) or Natural Product Number (NPN). If the drug does not have a DIN or a NPN, please provide evidence that the drug is sourced from a company with a valid drug establishment licence. Heparin coated catheters or wound dressings containing Zinc Oxide would be examples of devices that would need to meet this requirement. Health Canada has published a document addressing the regulation of products that are comprised of both a drug and a medical device entitled Drug/Medical Device Combination Products Policy, which is available on the Health Canada website.
IVDD Test Kits containing Controlled Substances
Please note that if the device is a test kit containing a substance listed in Schedule I, II, III or IV of the Controlled Drugs and Substances Act (CDSA), it would need to be registered with the Office of Controlled Substances. For information on how to apply for a Test Kit Number (T. K. Number), please refer to the Office of Controlled Substances Guidance Document entitled Registration of a Test Kit for Medical, Laboratory, Industrial, Educational or Research Purposes or contact the Office of Controlled Substances at (613) 952-2219 or (613) 957-1063.
Item 11: Device History
Indicate if the device has been previously authorized for sale in Canada under the investigational testing or special access provisions of the MDR. If the device has been previously authorized for sale under the investigational testing provisions of the MDR, it will have a Device Identification (ID) number. If the device had been previously authorized for sale by the special access provisions of the MDR, it will have an Authorization number. The appropriate number must be supplied.
Item 12: IDENTIFIER OF DEVICE
Only devices, components, parts and accessories listed on the application will be considered for licensing. Spare parts that do not represent medical devices on their own should not be listed. If additional space is required, photocopy the Item 12 page and attach it to the application form.
For a single device, enter the name of the device in the first column and enter the identifier for the device (bar code, catalogue, model or part number) in the second column. If the device contains ≥0.1% by mass of DEHP, check the third column. If the device is manufactured from raw materials containing or derived from BPA, check the fourth column.
For a medical device group, a medical device family, or a medical device group family, the names of the constituent members must be listed in the first column. Associated identifiers must be entered in the second column. If a constituent member contains ≥0.1% by mass of DEHP, check the associated row in the third column. If a constituent member is manufactured from raw materials containing or derived from BPA, check the associated row in the fourth column.
The fifth column is for Health Canada (HC) use only.
Refer to the definition of "BPA", "DEHP" and "identifier" in section 3 of this guidance document. Please note that it is the manufacturer's responsibility to determine whether a medical device contains ≥ 0.1% weight per weight of DEHP or is manufactured from raw materials containing or derived from BPA. The absence of a check mark in the third and fourth column in line with a specific device will be taken to indicate that the device does not contain ≥ 0.1% w/w of DEHP or is not manufactured from raw materials containing or derived from BPA.
Item 13: Compatibility Of Interdependent Devices
For a device intended to be used with another Class II, III, or IV device, a list of all medical devices that this device is intended to be used or function with (including their licence number), is required. This is intended to be for system components from the same manufacturer.
An important requirement in demonstrating safety and effectiveness of all medical devices intended to be used together is compliance with section 18 of the MDR. Section 18 requires that when medical devices are intended to be used with other medical devices, they must be compatible with every other medical device with which they interact, and don't adversely affect the performance of the combination of medical devices.
Failure to submit compatibility information for interdependent medical devices may lead to delays in the pre-market review of device licence applications while the Medical Devices Bureau requests the necessary information and manufacturers assemble and submit it for review.
Manufacturers are therefore reminded that the submission of evidence of compatibility for inter-dependent medical devices is a requirement under the MDR.
See also Notice to Industry - Licensing Requirements of Interdependent Medical Devices (April 30, 2002) available on the website.
Item 14: list Of Recognized Standards Complied With In The Manufacture Of The Device
Refer to the documents on Recognition and Use of Standards under the Medical Devices Regulations. Health Canada uses these document to identify standards that manufacturers may utilize in demonstrating that their devices comply with the safety and effectiveness requirements of sections 10 to 20 of the MDR. Under Section 35 of the MDR, the Medical Devices Bureau may also request the manufacturer to provide additional evidence of safety and effectiveness. Note that standards not contained in the Policy may also be identified by manufacturers in support of the safety and effectiveness of their device. However, these standards should not be identified in section 14 of the application form. Instead, for a Class II medical device, they should be referenced by way of a letter accompanying the licence application, and for a Class III or IV medical device, they should be listed in the safety and effectiveness information submitted.
For Class II applications, the manufacturer is to list the recognized standards complied with, or attest that they possess objective evidence that the device either meets an equivalent or better standard or has been tested and alternate evidence of safety and effectiveness exists.
For Class III and IV applications, the manufacturer must respond "YES" to one of the following statements and provide appropriate documentation:
If the manufacturer does not comply with any of these three options, a licence will not be issued.
Class II Applications
Items 15 to 22: These items pertain to the payment of fees. Instructions are provided on the form itself. The instructions given for each item must be carefully followed to avoid delays in the processing of your application. Consult also the Guidance Document - Fees in Respect of Medical Devices Regulations Licence Applications.
Class III Applications
Item 15: Review Documents
Indicate which review documents listed on the table are included as attachments to your application. For details regarding content and format of review documents, you are requested to consult the Guidance Document - Preparation of a Premarket Review Document for Class III and Class IV Device Licence Applications (available on the website).
Items 16 to 24: These items pertain to the payment of fees. Instructions are provided on the form itself. The instructions given for each item must be carefully followed to avoid delays in the processing of your application. Consult also the Guidance Document - Fees for the Review of Medical Devices License Applications.
Class IV Applications
Item 15: Review Documents
Indicate which review documents listed on the table are included as attachments to your application. For details regarding content and format of review documents, you are requested to consult the Guidance Document - Preparation of a Premarket Review Document for Class III and Class IV Device Licence Applications (available on the website).
Item 16: Devices Containing Biological Material
This section of the application form must be completed in detail. If additional space is required, photocopy the Item 16 page and attach it to the application form.
Items 17 to 25: These items pertain to the payment of fees. Instructions are provided on the form itself. The instructions given for each item must be carefully followed to avoid delays in the processing of your application. Consult also the Guidance Document - Fees for the Review of Medical Devices License Applications.
Cadogan D, Howick C. 1996. Plasticizers. In: Kroschwitz J, Howe-Grant M, eds. Kirk-Othmer encyclopedia of chemical technology. New York: John Wiley & Sons Inc., 258-290.
Canada. 1999. Canadian Environmental Protection Act, 1999. S.C., 1999, c. 33, Canada Gazette Part III. Vol. 22, No. 3. Available from:
http://www.gazette.gc.ca/archives/p3/1999/g3-02203.pdf.
Howard PH, Meylan WM. 1997. Handbook of physical properties of organic chemicals. Boca Raton, FL: Lewis Publishers.
HSDB. 1990. Hazardous Substances Data Bank. National Library of Medicine, National Toxicology Information Program, Bethesda, MD. July 18, 1990.
Montgomery JH, Welkom LM. 1990. Groundwater chemicals desk reference. Chelsea, MI: Lewis Publishers, Inc., 93-95.
RTECS. 2000. Registry of Toxic Effects of Chemical Substances. National Library of Medicine, National Toxicology Information Program, Bethesda, MD. February 2000.