Date: 2003-07-30
Help on accessing alternative formats, such as Portable Document Format (PDF), Microsoft Word and PowerPoint (PPT) files, can be obtained in the alternate format help section.
Contact: Medical Devices Bureau Enquiries
The Marketed Health Products Directorate (MHPD), Therapeutic Products Directorate (TPD) and Biologics and Genetic Therapies Directorate
(BGTD) post safety alerts, public health advisories, press releases and other notices from industry as a service to health professionals,
consumers, and other interested parties. Although MHPD, TPD and BGTD approve therapeutic products, MHPD, TPD and BGTD do not endorse either
the product or the company. Any questions regarding product information should be discussed with your health professional.
This is duplicated text of a letter from Cordis Corporation
Contact the company for a copy of any references, attachments or enclosures.

July 23, 2003
Dear Health Care Professional:
Subacute Thrombosis in Association with CYPHERTM Coronary Stents
In Canada, it is estimated that 900 CYPHERTM Sirolimus-eluting Coronary Stents have been implanted since November 2002, when the
CYPHERTM Sirolimus-eluting Coronary Stent was granted a device licence for use in de novo coronary artery lesions with lengths that
can be appropriately covered with a stent
18 mm and with a reference vessel diameter
2.5
mm and
3.5mm. There have been no reports of subacute stent thrombosis.
Since the introduction of the CYPHERTM Sirolimus-eluting Coronary Stent in the United States on April 24, 2003, it is estimated that well over 50,000 patients have received a CYPHERTM Stent. From the more than 50,000 patient procedures an adverse event report has been received for 47 stents (34 patients) due to stent thrombosis occurring at the time of implantation or within a few days of implantation. This is less than a 0.1% incidence rate. Some centers have reported multiple events.
Thrombosis is a rare complication of any coronary stenting procedure. In completed and ongoing clinical trials of the CYPHERTM Stent worldwide, the rate of stent thrombosis appears to have been similar to that of a bare metal stent. We are carefully reviewing the adverse event reports received from U.S. centers to try to determine if the thrombosis rate in current clinical experience differs from the rate in clinical studies completed pre-approval.
Based on the reports so far, factors impacting the rate of thrombosis may include failure to achieve adequate stent apposition (due to under-deployment) or suboptimal use of antiplatelet medication. In addition, Cordis Corporation has become aware that some interventionalists have been over-expanding smaller stents for use in larger diameter vessels. Overexpansion of stents beyond their intended diameter may negatively affect performance and is not advisable. Due to the extremely high demand for the CYPHERTM Stent, Cordis Corporation concentrated its manufacturing efforts on the 2.5 mm and 3.0 mm diameter CYPHERTM Stents to serve the patient populations with the greatest perceived potential benefit. We have recently started to introduce the 3.5 mm CYPHERTM Stents.
1. Follow the Instructions for Use
To help ensure that your experience is similar to that observed in pre-approval clinical trials, we strongly advise you to use the product in accordance with the Indications for Use (IFU) and procedures contained in the package insert. Please take particular note of the following:
18 mm in native coronary arteries with reference vessel
diameters of
2.5 mm to
3.5
mm. The CYPHERTM stent is NOT indicated for treatment of the following:2. Report Your Experience
In most product development programs, rare side effects are difficult to detect, and their risks for special populations are difficult to assess. Therefore, as a matter of course, it is important for you to report any product complaints and adverse events directly to Cordis Canada, a Business unit of Johnson & Johnson Medical Products, division of Johnson & Johnson Inc by:
Please feel free to contact Cordis Canada for any questions you may have on the subject matter of this letter. We, in cooperation with Health Canada, will continue to keep you updated with the latest information.
Sincerely,
original signed by
Dennis Donohoe, M.D.
Vice President, Therapeutics and Clinical Research
Cordis Corporation
Any suspected adverse incident can also be reported to:
&Food Branch Inspectorate
HEALTH CANADA
Address Locator: 3002C
OTTAWA, Ontario, K1A 0K9
Tel: The Medical Devices Hotline 1-800-267-9675