The Health Products and Food Branch (HPFB) posts on the Health Canada web site safety alerts, public health advisories, press releases and other notices as a service to health professionals, consumers, and other interested parties. These advisories may be prepared with Directorates in the HPFB which includes pre-market and post-market areas as well as market authorization holders and other stakeholders. Although the HPFB grants market authorizations or licenses for therapeutic products, we do not endorse either the product or the company. Any questions regarding product information should be discussed with your health professional.
This is duplicated text of a letter from Source Medical Corporation. Contact the company for a copy of any references, attachments or enclosures.
Health Canada Endorsed Important Safety Information on the recall of Adult Respiratory Ventilator Circuits (RT110, RT111, RT210, RT211 Lots Ranging from 041101 through 050331 (inclusive))
December 28, 2005
Dear Source Medical and Cardinal Health customer,
Fisher & Paykel Healthcare has notified us of the recall of the specific circuits and lot numbers referenced above, due to potential insulation damage to the heater wires. Please review the enclosed Fisher & Paykel recall letter for more details.
The affected lot numbers were manufactured between November 1, 2004 and March 31, 2005. Our distribution records indicate that you may have received affected circuits during this time.
Therefore, please take the following steps:
Source Medical will forward the completed Customer Reply forms to Fisher & Paykel, who will ship replacement product directly to your facility if required.
If you have any questions, please contact your Cardinal Health Respiratory Therapy Specialist, or contact Kristina Buzadzija at 1-888-871-5945 ext, 5140.
Source Medical Corporation
60 International Boulevard
Toronto, ON M9W 6J2
Tel: 416-213-5000
Fax: 416-213-5199
www.sourcemedical.com
Any suspected adverse incident can also be reported to:
Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Tel: The Inspectorate Hotline 1-800-267-9675
The Medical Devices Problem Report Form and Guidelines can be obtained from the Health Canada web site.
For other inquiries related to this communication, please contact Health Canada at:
Marketed Health Products Directorate (MHPD)
Email: MHPD_DPSC@hc-sc.gc.ca
Tel: (613) 954-6522
Fax: (613) 952-7738
Sincerely,
original signed by
Angela Stanley
Q.A. Compliance Specialist
Source Medical and Cardinal Health
Encl: Customer Reply form
Fisher & Paykel Dec. 21, 2005 Recall Letter
To: Angela Stanley, Quality Assurance
Company: Source Medical
Fax: (416) 213-5019
Phone: 1-888-871-5945; (416) 213-5105
Hospital/Customer:
Customer Number:
Address:
City/Province:
Comments: Product Recall, RT110, RT111, RT210, RT211, lot range from 041101 through 050331 (inclusive), ref. 2005FCA44
I,(PLEASE PRINT), have received and acknowledge the details outlined in this letter.
TITLE :
SIGNATURE :
DATE :
PHONE#:
WE HAVE INSPECTED OUR INVENTORY AND HAVE NO AFFECTED STOCK OF THESE PRODUCTS.
WE HAVE THE FOLLOWING STOCK FROM THE AFFECTED LOT RANGE 041101 THROUGH 050331, AND HAVE DESTROYED IT AS PER INSTRUCTIONS. (Fisher & Paykel will issue replacement based on this information)
RT110
QUANTITY: EA./BX.
RT111
QUANTITY: EA./BX.
RT210
QUANTITY: EA./BX.
RT211
QUANTITY: EA./BX.
(1 BOX = 10 CIRCUITS)
2005FCA44
Fisher & Paykel Healthcare Limited
15 Maurice Paykel Place, East Tamaki
P O Box 14 348, Panmure
Auckland, New Zealand
Telephone: +64 9 574 0100
Facsimile: +64 9 574 0158
Website:
www.fphcare.com
21 December 2005
FILE REF: 2005 Adult Circuit Recall Letter Canada
Dear Sir/Madam,
Notice of Recall - Fisher & Paykel Healthcare Adult Respiratory Ventilator Circuit -
Dual Heated
Fisher & Paykel Healthcare has initiated a voluntary recall for adult breathing circuits that contain a spiral expiratory heater wire.
It has been identified that the manufacturing process of the Adult respiratory ventilator circuit - dual heated, may have had the potential to damage the insulation of the heater wire in circuits within a specified lot range. This may lead to a potential risk of fire in some circumstances.
This recall affects ONLY adult breathing circuits with a spiral wound, expiratory heater wire manufactured by Fisher & Paykel Healthcare. The following circuits are NOT affected: EVAQUA™, single-limb heated circuits, and neonatal circuits.
The affected catalog numbers are:
Canada: RT110, RT111, RT210, RT211
The affected lot codes are:
041101 to 050331 (inclusive).
Stock manufactured between 1 November 2004 and 31 March 2005.
Please check your stock of Fisher & Paykel Healthcare dual-heated, adult breathing circuits for any stock in the affected lot range, and destroy any stock you may have in that lot range. Please contact all your customers with this information and request they acknowledge this action and they destroy all stock.
In addition, Fisher & Paykel Healthcare wishes to take this opportunity to advise you that the labelling for the catalog numbers listed in this recall letter will be modified in the future to introduce a recommendation of a maximum 7 day use.
We apologise for any inconvenience this recall will cause. Please contact us if further details or clarification is required.
Yours sincerely
original signed by
Adele Bindon
Regulatory Affairs Manager - Intensive Care
Fisher & Paykel Healthcare Limited
Ph: + 64 9 574 0123, ext. 8813
Fax: + 64 9 574 0158
Email: Adele.Bindon@fphcare.co.nz
Please return this form by fax or post to the fax number listed below.
NOTE: A response is required even if you have no stock in the affected lot range.
I,PLEASE PRINT, have received and acknowledge the details outlined in this letter.
Please indicate the amount of stock in your distribution in the affected lot range (041101 to 050331, inclusive) and the amount destroyed in the table below:
Catalog Number
Number in stock
Number in stock
RT110
RT111
RT210
RT211
Please note, 1 box = 10 circuits.
Signature:
Name: PLEASE PRINT
Position:
Hospital:
Date:
New Zealand
Fisher & Paykel Healthcare Ltd
P O Box 14 348, Panmure,
Auckland 1134
New Zealand
Fax : +64 9 574 0158