The Health Products and Food Branch (HPFB) posts on the Health Canada web site safety alerts, public health advisories, press releases and other notices as a service to health professionals, consumers, and other interested parties. These advisories may be prepared with Directorates in the HPFB which includes pre-market and post-market areas as well as market authorization holders and other stakeholders. Although the HPFB grants market authorizations or licenses for therapeutic products, we do not endorse either the product or the company. Any questions regarding product information should be discussed with your health professional.
This is duplicated text of a letter from Defibtech, LLC.
Contact the company for a copy of any references, attachments or enclosures.
URGENT MEDICAL DEVICE SAFETY INFORMATION AND CORRECTIVE ACTION
Defibtech DDU-100 Series AEDs
February 16, 2007
Dear Customer:
Defibtech is notifying customers that version 2.002 and earlier versions of the software in DDU-100 Series AEDs ("Lifeline" and "ReviveR") may, under certain circumstances, clear a "Low Battery" or "Replace Battery" warning that was previously detected by an automatic self-test. If this condition occurs, the operator may be unaware of the low battery, and the device may be unable to deliver a shock.
This situation can occur only in units with low batteries. After a period of a low battery warning, the software may stop reporting the warning and the operator may be unaware of the low battery condition. A low battery may have insufficient power to deliver a shock. During a rescue, replacing the low battery pack will allow the unit to function normally.
The probability of this occurring is higher when a battery is near end-of-life. Field data indicate that battery life is typically greater than three years.
Devices with operating software version 2.002 and earlier versions are subject to this correction. Our records indicate you own at least one of the identified DDU-100 Series AEDs that was initially sold with the affected software.
Defibtech is releasing a new version of software to correct this issue. You will be able to install the new version of software in the field, where the unit is deployed. Your distributor will follow up this notification with a communication detailing the plan to provide this software upgrade. Distribution of this new software may take 12 to 16 weeks.
Follow the detailed instructions in the "Immediate Recommendations" section below and Defibtech recommends that you keep your DDU-100 Series AED in service until your unit's software is upgraded.
Immediate Recommendations
These recommendations apply only to units running version 2.002 software or earlier versions. To determine the software version of a unit, turn the unit on, wait until the unit begins speaking, and then press and hold the on/off button for approximately five seconds until the unit begins speaking again. Prior to automatically turning itself off, the unit will speak the version number; if the unit does not speak a version number, the software version is 1.201.
We advise that you perform the following procedure immediately and then once per month until the unit's software has been upgraded:
We are communicating this information to Health Canada and to international regulatory agencies.
I sincerely apologize for any inconvenience this may cause you. If you have any questions regarding this notification, please contact your distributor. Defibtech is committed to ensuring our products meet the highest quality standards and that our customers are fully supported.
Sincerely,
original signed by
Glenn W. Laub, M.D.
CEO
Defibtech, LLC