The Health Products and Food Branch (HPFB) posts on the Health Canada web site safety alerts, public health advisories, press releases and other notices as a service to health professionals, consumers, and other interested parties. These advisories may be prepared with Directorates in the HPFB which includes pre-market and post-market areas as well as market authorization holders and other stakeholders. Although the HPFB grants market authorizations or licenses for therapeutic products, we do not endorse either the product or the company. Any questions regarding product information should be discussed with your health professional.
This is duplicated text of a letter from Cardinal Health.
Contact the company for a copy of any references, attachments or enclosures.
July 23, 2007
Subject: SAFETY ALERT NOTIFICATION - FOR ALL IV ADMINISTRATION SETS Indicated for Use with Alaris® Pump module (aka Medley™ Pump Module) and GeminiTM Infusion Pumps
Attention: Director of Nursing, Director of Materials and Director of Risk Management
This Safety Alert affects all IV administrations sets that are indicated for use with the Alaris® Pump module (aka MedleyTM Pump Module) (Model 8100) and the GeminiTM Infusion Pumps (Models 131X, 132X and 134X). A list of affected IV administration set models is attached.
Cardinal Health has determined the need to reinforce the intended use of the IV set roller clamp as the primary means to prevent over-infusions and to regulate flow to the patient upon removal from the pump.
Over-infusion may occur as a result of an open safety clamp fitment (aka Flo-Stop® fitment) on the IV administration set pumping segment, whether opened intentionally in the course of expected clinical practice, or as a result of inadvertent action by the clinician. The over-infusion occurrence is dependent on the user not closing the roller clamp first, as required by clinical practice and the IV administration set Directions for Use, and then intentionally or inadvertently opening the secondary back-up mechanism (safety clamp fitment).
RECOMMENDATIONS for YOUR FACILITY:
Per the IV administration set directions for use and acceptable clinical practices:
We want to assure you that Cardinal Health remains committed to serving your infusion product needs. Patient safety, exceptional product reliability and the highest level of customer satisfaction are our primary goals. Should you have any questions or concerns, our Customer Advocacy department is prepared to assist you at 1-800-387-8309 as well as provide any additional tip sheets you may request.
Sincerely,
original signed by
Barry Holleman
Cardinal Health, Clinical Technologies and Services
Director of Operations
Enclosures:
Note: The Safety clamp fitment on the pumping segment is received in open position (for priming)






Note: When door is opened, the administration set Safety clamp fitment will automatically close to help prevent accidental free-flow
When the white clamp is extended, it is in the "closed" position to prevent flow
2 Safety clamp fitment automatically closes
Note: Safety clamp fitment is closed (white clamp is extended) when removed from pump






For more information, please contact Customer Advocacy at 800.854.7128, ext. 7812. In Canada at 800.387.8309. Or visit our website at cardinalhealth.com.
*Safety clamp formerly known as Flo-Stop® fitment.
** Alaris® Pump module formerly known as Medley® Pump Module.
©2004-2007 Cardinal Health, Inc. or one of its subsidiaries. All rights reserved.
Alaris® is a registered trademark of Cardinal Health, Inc. or one of its subsidiaries.
3POC1474 (0607/15000)
Gemini Infusion Pump




Note: When door is opened, the administration set Safety clamp fitment will automatically close to help prevent accidental free-flow


Note: Safety clamp fitment is closed (white clamp is extended) when removed from pump


For more information, please contact Customer Advocacy at 800.854.7128, ext. 7812.
In Canada at 800.387.8309. Or visit our website at cardinalhealth.com.
* Safety clamp formerly known as Flo-Stop® fitment
©2004-2007 Cardinal Health, Inc. or one of its subsidiaries. All rights reserved.
Alaris® is a registered trademark of Cardinal Health, Inc. or one of its subsidiaries.
SSM# 3POC1560 (0607) 32M
Please assist us in making the Safety Alert follow-up process efficient and convenient for you by completing and returning this card.
(Check if Applicable)
Safety Alert Receipt Verification:
Location Verification for Alaris® Pump module (aka Medley™ Pump module) & Gemini™ Infusion Pump:
Signed: (Signature and Title)
Date:
(Please Print Name)
(Phone)
Please assist us in making the Safety Alert follow-up process efficient and convenient for you by completing and returning this card.
(Check if Applicable)
Safety Alert Receipt Verification:
Distributor Verification:
Signed: (Signature and Title)
Date:
(Please Print Name)
(Phone)
| Model Number | Model Number | Model Number | Model Number |
|---|---|---|---|
| 10010453 | 10015649 | 2241-0008 | 2455-0500 |
| 10010454 | 10015861 | 2241-0600 | 2464-0500 |
| 10010483 | 10015862 | 2247-0600 | 2477-0000 |
| 10010541 | 10015896 | 2248-0600 | 2477-0007 |
| 10010547 | 10031773 | 2255-0001 | 2478-0000 |
| 10010761 | 10061661 | 2255-0500 | 2904-0600 |
| 10010871 | 10062818 | 2260-0001 | 2906-0000 |
| 10010916 | 10074281 | 2260-0006 | 2906-0001 |
| 10011221 | 10108961 | 2260-0008 | 2907-0000 |
| 10011301 | 10137405 | 2260-0500 | 2907-0006 |
| 10011462 | 10142568 | 2264-0006 | 2920-0000 |
| 10011652 | 10178573 | 2264-0008 | 2929-0500 |
| 10012144 | 2110-0500 | 2264-0500 | 2941-0500 |
| 10012182 | 2111-0500 | 2277-0000 | 2942-0500 |
| 10012283 | 2120-0500 | 2277-0006 | 2946-0600 |
| 10012293 | 2125-0500 | 2277-0008 | 9943-0000 |
| 10012645 | 2126-0500 | 2279-0006 | 9943-0001 |
| 10012802 | 2130-0500 | 2280-0000 | 9943-0008 |
| 10013034 | 2131-0500 | 2301-0500 | 9943T |
| 10013037 | 2140-0600 | 2377-0000 | 9943T-0008 |
| 10013072 | 2141-0600 | 2401-0006 | 9950-0500 |
| 10013186 | 2147-0600 | 2401-0500 | 9952-0500 |
| 10013361 | 2177-0000 | 2403-0000 | 9953-0500 |
| 10013373 | 2200-0006 | 2403-0007 | 9966-0006 |
| 10013374 | 2200-0500 | 2406-0500 | 9971-0600 |
| 10013854 | 2202-0500 | 2410-0500 | 9974-0600 |
| 10013889 | 2203-0006 | 2411-0500 | C24101E |
| 10013890 | 2205-0000 | 2412-0500 | C24102M |
| 10013923 | 2210-0006 | 2414-0006 | C24103E |
| 10014035 | 2210-0500 | 2420-0006 | C24104E |
| 10014855 | 2211-0500 | 2420-0007 | C24105E |
| 10015012 | 2214-0001 | 2420-0500 | C24106E |
| 10015048 | 2214-0006 | 2421-0500 | C24107E |
| 10015294 | 2220-0006 | 2425-0500 | C24109E |
| 10015312 | 2220-0500 | 2426-0500 | C24111E |
| 10015362 | 2225-0500 | 2430-0500 | C24112 |
| 10015364 | 2226-0500 | 2431-0500 | C24114E |
| 10015366 | 2230-0500 | 2440-0600 | C24116E |
| 10015368 | 2231-0006 | 2441-0007 | C24117 |
| 10015370 | 2231-0500 | 2441-0600 | C24119E |
| 10015414 | 2240-0008 | 2443-0600 | C24120E |
| 10015489 | 2240-0600 | 2447-0600 | |
| 10015514 | 2241-0001 | 2448-0600 | |
| 10015645 | 2241-0006 | 2449-0600 |