Help on accessing alternative formats, such as Portable Document Format (PDF), Microsoft Word and PowerPoint (PPT) files, can be obtained in the alternate format help section.
Contact: Marketed Health Products Directorate (MHPD)
The Health Products and Food Branch (HPFB) posts on the Health Canada web site safety alerts, public health advisories, press releases and other notices as a service to health professionals, consumers, and other interested parties. These advisories may be prepared with Directorates in the HPFB which includes pre-market and post-market areas as well as market authorization holders and other stakeholders. Although the HPFB grants market authorizations or licenses for therapeutic products, we do not endorse either the product or the company. Any questions regarding product information should be discussed with your health professional.
This is duplicated text of a letter from ZOLL Medical Corporation.
Contact the company for a copy of any references, attachments or enclosures.
Health Canada Endorsed Important Safety Information on ZOLL AED Plus Defibrillator
February 12, 2009
Subject: ZOLL AED Plus Defibrillator May Not Deliver Therapy
Dear Health Care Professional,
We have received reports that following a long period (typically greater than four years) without use, a ZOLL® may prompt "change batteries" during use and fail to deliver therapy. Turning the device off completely, waiting ten seconds for the unit to re-set and then turning it back on is necessary to resume proper operation. Our investigation of this problem has determined that some batteries may develop high internal resistance that interferes with the batteries expected performance, lengthening charging time beyond specified and clinically acceptable limits.
ZOLL Medical Corporation is informing owners of all AED Plus devices with serial numbers below X_ _ _200000 to replace batteries that have been installed in devices for a period longer than three (3) years and at subsequent three (3) year intervals.
Required Corrective Actions
(1) Determine the date batteries were installed in your device.
(2) If batteries were installed more than three (3) years ago, replace the batteries as soon as possible and add the battery replacement label indicated in the Addendum to the ZOLL AED Plus Administrator's Guide showing the next scheduled replacement date. Read and incorporate the addendum information into the ZOLL AED Plus Administrator's Guide.
(3) If batteries are not more than three (3) years old, add the label completed with the next scheduled replacement date. Read and incorporate the addendum information into the ZOLL AED Plus Administrator's Guide.
Units above serial number X_ _ _200000 are not affected by this corrective action as they contain software that can detect this battery condition and identify when batteries require replacement. In lieu of adding the battery replacement label users may obtain this new software for their AEDs at no cost at
ZOLL AED Plus Battery Help.
Customer Assistance
ZOLL AED Plus Battery Help will be available on February 16th 2009 and will provide other technical information and support concerning this corrective action, such as help in determining battery age and ordering of replacement batteries and extra labels. We have arranged special pricing of $39.95 per set of ten batteries inclusive of shipping that can be ordered by credit card at this site. We are offering this special pricing for customers affected today as well as for future replacements to mitigate the cost associated with a shorter than expected battery life.
Health Canada has been notified of this corrective action and classified it as a Priority II recall.
We apologize for any inconvenience this may cause you and thank you in advance for assistance in implementing this corrective action. Avoiding this problem during clinical use is our highest priority. Our 24/7 technical support numbers 1 (800) 348-9011 or +1 (978) 421-9460 are available to assist users with any aspect of this notice.
Managing marketed health product-related adverse incidents depends on health care professionals and consumers reporting them. Reporting rates determined on the basis of spontaneously reported post-marketing adverse incidents are generally presumed to underestimate the risks associated with health product treatments. Any cases of serious specific adverse incident or other serious or unexpected adverse incidents in patients using ZOLL AED Plus Defibrillator should be reported to ZOLL Medical or Health Canada at the following addresses:
Sincerely,
original signed by
Paul Dias
VP Quality Assurance & Regulatory Affairs
Addendum to the ZOLL AED Plus Administrator's Guide
Applies to all units with SW version 5.18 and lower
Instructions for the Installation of New Batteries and Battery Replacement Label for the ZOLL AED Plus
Please Note: Replace the batteries in your AED Plus® three years from the installation date of new batteries.
Explanation of Symbols:
- Replace batteries before <date>
Place the AED Battery Replacement Label below the AED Plus Status Indicator

9652-0120-01, Rev A