Health Canada
Symbol of the Government of Canada
Environmental and Workplace Health

Tritium/Carbon-14 Urinalysis Intercomparison Report - June 2006

The following report summarizes the results obtained in the June 2006 sample distribution for the Tritium/Carbon-14 Intercomparison Program, which is administered by the National Calibration Reference Centre for Bioassay, Radiation Protection Bureau.

Method

Samples were prepared as described in previous reports [1,2] and were spiked with both tritium and carbon-14. The spike levels tested were:

Tritium: 2.52 x 104 Bq/L
Carbon-14: 3.34 x 103 Bq/L

and the date of spiking was May 11, 2006.

Each participant was provided with five aliquots of the dual spiked sample and of the urine blank from which the spiked sample had been prepared. Instructions for sample analysis and reporting of results were enclosed with the samples.

Results

Samples were distributed to eleven laboratories. Nine returned results as requested. These results are presented in Table 1.

The equations used to evaluate the bias and precision of results are given in the appendix. The terms used in the equations and the attached tables and figures are defined as follows:

yjkl =
the result obtained by correcting the measurement for a spiked sample by the value obtained for the appropriate urine blank.
bjkl =
the relative bias of the result obtained for each aliquot of a particular spiked sample. It represents the difference between the measurement and the true value, normalized by the true value.
bjk =
the mean relative bias, i.e. the mean of the bjkl values for a particular spiked sample.
sjk =
relative precision, calculated as the standard deviation of the relative bias of measurements of aliquots of the same spiked sample.
MDA =
Minimum detectable activity. This is an indication of the detection capability of a participant's analytical method.
MTL =
Minimum testing level. The minimum amount of radioactive material at which a dosimetry service should be tested to be in compliance with the Canadian Nuclear Safety Commission license requirements for performance in an independent testing exercise.

The Canadian Nuclear Safety Commission (CNSC), formerly the Atomic Energy Control Board, has set its performance criteria for bioassay measurements in Regulatory Standard S-106, "Technical and Quality Assurance Standards for Dosimetry Services in Canada" [3]. At levels equal to or greater than the MTL, the limits for acceptable bias and precision are:

-0.25 ≤ Mean relative bias ≤ +0.50
Relative precision ≤ 0.40

Note that the MTL is set at five or more times the MDA. In S-106, the MDA for urinalysis has been set at 70 Bq/L for carbon-14 and at 400 Bq/L for tritium. The known performance of Canadian bioassay laboratories has been taken into consideration in setting these specifications [3].

Relative Bias

The yjkl values and their means are listed in Table 2 in the same format and with the same number of significant figures given in the reports submitted to the Bureau. The yjkl values are also presented graphically in Figures 1 and 2.

The bjkl values are presented in Table 3. Values that exceeded the acceptable relative bias limits are marked with an asterisk in the table.

Table 4 lists the bjk values which are also shown in Figures 3 and 4. All values satisfied the performance criteria for mean relative bias.

Relative Precision

The sjk values are presented in Table 5 and Figures 5 and 6. All values were within the acceptable limit for relative precision.

For the participants that followed the recommended protocol (all except Laboratory 7 in this intercomparison) of only one analyst performing measurements for all the blank and spiked samples, the sjk values indicate day-to-day within-analyst variability in addition to intra-laboratory precision.

Summary

Nine laboratories participated in this intercomparison. The number of participants that obtained mean relative bias and relative precision values of 5% or less are as follows:

Spike (Bq/L)
Mean Relative Bias Relative Precision
Tritium: 2.52 x 104 7 7
Carbon-14: 3.34 x 103 7 6

References

  1. Carbon-14 in Urine Intercomparison Report, Bureau of Radiation and Medical Devices, Health and Welfare Canada, June, 1986.

  2. Carbon-14 in Urine Intercomparison Report, Bureau of Radiation and Medical Devices, Health and Welfare Canada, June, 1987.

  3. Atomic Energy Control Board [now known as the Canadian Nuclear Safety Commission], "Technical and Quality Assurance Standards for Dosimetry Services in Canada", Regulatory Standard S-106, March, 1998