Radiation Protection in Mammography: Recommended Safety Procedures for the Use of Mammographic X-Ray Equipment - Safety Code 33
5. Equipment specifications
5.1 Newly purchased medical
X-ray equipment
All new, used, and refurbished mammographic X-ray
equipment, and accessories for such equipment, which is sold,
imported or distributed in Canada, must conform to the requirements
of the Radiation Emitting Devices Act and the Food and Drugs Act
and their promulgated regulations. These are the Radiation Emitting
Devices Regulations and the Medical Devices Regulations. The
Radiation Emitting Devices Regulations specify standards of design,
construc- tion and performance, with respect to radiation safety. The
Medical Devices Regulations encompass all other safety
considerations and the question of efficacy for all medical X-ray
equipment sold in Canada. It is the responsibility of the manufacturer
or distributor to ensure that the equipment conforms to the
requirements of these regulations.
The Radiation Emitting Devices Regulations in effect and
relevant for mammographic X-ray equipment, at the time of printing
of this Safety Code, are reproduced in Appendix VII. These regulations
may be amended, from time-to-time, to keep up-to-date with
changing technology in the field. Information on the applicability and
currency of the Radiation Emitting Devices Regulations may be
obtained by contacting the Radiation Protection Bureau, Health
Canada, Ottawa, Ontario, K1A 1C1.
In addition, mammographic X-ray equipment must meet any
applicable requirements under provincial jurisdictions for such
equipment.
5.2 Existing medical X-ray equipment
When possible, existing mammographic X-ray equipment must
be upgraded to incorporate as many as possible of the safety and
performance features required of new mammographic X-ray equipment.
It should be noted that it is a requirement of the Radiation
Emitting Devices Regulations that replacements for any component
or subassembly of an X-ray machine, for which a design, construction or performance standard has been specified in the
Regulations applicable to the class of X-ray equipment, must comply
with the standards in effect at the time of replacement.
Only equipment designed specifically for mammography shall
be used, and to ensure maximum protection for patients and staff, all
existing mammographic X-ray equipment must at least meet certain
basic requirements. These requirements are itemized in the
remainder of this Section.
5.3 General requirements for
mammographic X-ray equipment
The following requirements must be met by all mammographic
X-ray equipment. Specific requirements for film-screen mammographic
X-ray equipment are listed in Section 5.3.1, whereas
requirements for Xeromammographic X-ray equipment are listed in
Section 5.3.2.
- Warning Signs - The X-ray control panel must bear a permanent
and conspicuous sign warning that hazardous X-radiation is
emitted when the equipment is in operation and prohibit
unauthorized use.
- Markings - All controls, meters, lights and other indicators
relevant to the operation of the equipment must be readily
discernible and clearly labelled or marked as to function.
- Focal Spot Marking - The location of the focal spot must be
clearly and accurately marked on the X-ray tube housing. In the
case of dual focus X-ray tubes, the location of the mark should be
midway between the centres of the two focal spots.
- Indicator lights - There must be readily discernible, separate
indicators on the control panel that indicate:
- when the control panel is energized and the machine is
ready to produce X-rays, and
- when X-rays are being produced.
- Indication of Loading Factors - For mammographic X-ray
equipment having adjustable loading factors, the control panel
must incorporate electrical meters or other indicators that enable
determination of the X-ray tube voltage, X-ray tube current and
time, or combinations of these. For equipment having nonadjustable
loading factors, permanent marks or labels may be
used to indicate these parameters. The loading factors should be
displayed after the irradiation is complete.
If the mammographic X-ray equipment is battery-powered,
there must be a visual indicator that shows whether the battery is
adequately charged for proper operation.
- Irradiation Control - There must be an irradiation switch, or
other device to initiate and terminate X-ray production. The
irradiation switch must require continuous pressure by the
operator to produce X-rays.
- Timer - A timing device must be provided to automatically
terminate the irradiation. The timer must be designed and
constructed in such a way that it is not possible to energize the
X-ray tube without automatic or manual resetting of the timer
after each loading, that an irradiation cannot be initiated with the
timer set at its zero or OFF position and the production of X-rays
is automatically terminated after a preset time, preset
milliampere-second value, or a preset exposure or air kerma
value. The timing device must be accurate to 1/60 second or to
7 percent, whichever is greater.
A backup timer must be provided to limit the product of the
X-ray tube current and the irradiation time during an irradiation.
The product of the X-ray tube current and the irradiation time
must not exceed 2,000 milliampere-seconds per irradiation.
When possible, the product of the X-ray tube current and the
irradiation time should be set not to exceed
1,200 milliampere-seconds per irradiation.
- X-ray Tube Shielding - The X-ray tube must be enclosed in a
shielded housing. The shielding of the housing must be such that,
at every rating specified by the manufacturer, the leakage
radiation does not exceed 17.5 µGy (2 mR) per hour at 5 cmfrom
any point on the external surface of the housing.
- Beam Limiting Devices - Suitable beam limiting devices capable
of restricting the radiation beam shall be provided and shall
provide the same attenuation as the X-ray tube housing. The
beam limiting device should be designed in such a way that, for
any focal spot to image receptor distance, the radiation beam
does not extend beyond the edge of the image receptor except at
the edge adjacent to the chestwall where the X-ray field shall not
extend beyond the edge by more than 2 percent of the focal spot
to image receptor distance.
If a light localizer designed to define the outline of the X-ray
field is included, the misalignment, in the plane of the image
receptor, of the light field with respect of the X-ray field along either length or width must not exceed 2 percent of the source to
image receptor distance (SID).
- Image Receptor Support Shielding - The image receptor support
shall transmit less than 0.87 µGy (0.1 mR) per irradiation at all
operating loading factors at the minimum source to image
receptor distance.
- Breast Compression Device - A device to maintain firm breast
compression shall be provided on the mammographic X-ray
equipment. This device must provide adjustable, uniform and
constant compression of the breast during mammography. The
X-ray beam attenuation of the compression plate should be less
than that of 2.5 mm of polymethylmethacrylate (PMMA)
equivalent. There is no known optimal value of compression.
However, a minimum of 20 kg should be generated but
compression forces of 25 kg or greater should be achievable
only with manual control.
The breast compression device should not produce inhomogeneities
or artifacts which can degrade image quality.
The deformation of the breast compression device at maximal
compression force should not be greater than 15 mm at each
corner between the surface of the image receptor system and the
breast compression device when a piece of foam rubber is
compressed.
- Protective Barrier - A protective radiation barrier shall be provided.
This barrier must allow the technologist to observe the
patient during the entire procedure and should provide
attenuation equal to or greater than 0.25 mm Pb equivalent at
50 kVp. The barrier should be at least 0.6mwide and 1.85mhigh
and reach within 0.15 m above the floor.
- Mechanical Stability - The X-ray tube must be securely fixed
and correctly aligned within the tube housing. The X-ray tube
housing must maintain its required position without drift or
vibration during operation and must be balanced to provide
smooth operation.
5.3.1 Requirements for film-screen mammographic
X-ray equipment
- Target Material - Molybdenum (Mo) or Molybdenum-Tungsten
(Mo-W) alloy target X-ray tubes shall be used for film-screen
mammographic X-ray equipment. Alternate target material may
be appropriate when used with alternate filter material provided
that it produces comparable image quality at equal or reduced
dose to the breast.
- Focal Spot Size - The focal spots must be small enough not to
create excessive geometric unsharpness. The focal spots should
be measured using either a slit camera, the pinhole method, or
other method where the nominal focal spot size can be determined.
For film-screen mammographic X-ray equipment, the
nominal focal spot size should be:
equal to or less than 0.40 mm for contact or grid techniques
at 65 cm SID,
equal to or less than 0.30 mm for contact or grid techniques
at 50 cm SID,
equal to or less than 0.15 mm for 1.5 magnification,
equal to or less than 0.10 mm for 2.0 magnification
- X-ray Beam Filtration - A permanent filter of about 0.025 to
0.030 mm Mo shall be permanently installed. For magnification,
a Tungsten target microfocal-spot X-ray tube may be
used and this tube shall have at least 0.5 mm Al equivalent total
filtration. Alternate filter material may be appropriate provided
that it produces comparable image quality at equal or reduced
dose to the breast.
- Radiation Beam Quality -There must be radiation-absorbing
filters that provide a degree of attenuation such that the first
Half-Value Layer of aluminum is not less or greater than the
values shown in Table 1 for a selected X-ray tube voltage. For
other X-ray tube voltages, the Half-Value Layer of the radiation
beam must be calculated by linear interpolation from that Table.
Half-Value Layer measurements must include the attenuation of
the breast compression device if the device is of uniform thickness
and without holes.
Table 1:Acceptable Half-Value Layer forMo or Mo-W alloy target X-ray tube
X-ray Tube Voltage
(kilovolt) |
First Half-Value Layers
(minimum - maximum)
(millimetre of Al)
|
| 24 |
0.24 - 0.34 |
| 26 |
0.26 - 0.36 |
| 28 |
0.28 - 0.38 |
| 30 |
0.30 - 0.40 |
| 35 |
0.35 - 0.45 |
- X-ray Tube Voltage - The X-ray tube voltage should be
adjustable in 1 kVp increments. The lowest selectable X-ray tube
voltage shall be equal to or less than 24 kVp. The peak X-ray tube
voltage should correspond to within 5 percent of the selected or
indicated value. The X-ray tube voltage repro- ducibility should
be within 2 percent.
- X-ray Tube Radiation Output - The X-ray tube radiation output
shall be high enough to minimize irradiation time to eliminate
perceptible motion artifacts and reduce the dose to the patient
resulting from reciprocity law failure of film-screen combination.
The X-ray tube output should be at least 4.4 mGy/s
(500 mR/s) at 28 kVp.
For any combination of operating loading parameters, the coefficient
of variation of any ten consecutive radiation exposure
measurements, taken at the same source to detector distance
within a time period of one hour, is no greater than 0.05, and each
of the ten radiation exposure measurements is within 15 percent
of the mean value of the ten measurements.
- Automatic Exposure Control - An Automatic Exposure Control
(A.E.C.) system must be provided. The system should be able to
maintain a net film density of ±0.15 O.D. (Optical Density) units
within a range of 1.0 to 1.5 O.D. for a film with an average
gradient of 3.0 O.D., for the range of breast thicknesses
examined, and for all types of technique (non-grid, grid, and
magnification) and loading factors used by the facility. In addition, a film density control for the A.E.C. should be provided
with each increment increasing or decreasing film-screen
cassette dose by 20 percent.
When manual irradiation control is used, the selectable control
interval (time or mAs) shall be small enough to permit increments
smaller than 25 percent.
- Breast Support Table - The attenuation of the breast support
table should not exceed 0.3 mm Al equivalent at 30 kVp.
- Anti-scatter Grid and Bucky System - If an anti-scatter grid or
Bucky system is used, it must be designed for mammographic
purposes and must not produce inhomogeneities or artifacts
which can degrade image quality. Grid lines should not be visible
on mammograms.
5.3.2 Requirements for Xeromammographic X-ray
equipment
- Target Material - For Xeromammographic X-ray equipment, a
Tungsten (W) or Molybdenum-Tungsten (Mo-W) alloy target
X-ray tube with Aluminum (Al) filtration shall be used.
Alternate target material may be appropriate when used with
alternate filter material provided that it produces comparable
image quality at equal or reduced dose to the breast.
- Focal Spot Size - The focal spots must be small enough not to
create excessive geometric unsharpness. The focal spots should
be measured using either a slit camera, the pinhole method, or
other method where the nominal focal spot size can be determined.
For Xeromammographic X-ray equipment, the nominal
focal spot size should be:
equal to or less than 0.6 mm for contact techniques at 80 cm SID,
equal to or less than 0.5 mm for contact techniques at 65 cm SID,
equal to or less than 0.20 mm for 1.5 × magnification,
equal to or less than 0.15 mm for 2.0 × magnification.
- X-ray Beam Filtration - For Tungsten target X-ray tube, the total
filtration shall not be less than 2.0 mm Al equivalent at 50 kVp,
and for Molybdenum-Tungsten target X-ray tube, the total
filtra-tion shall not be less than 1.6 mm Al equivalent at 50 kVp.
Alternate target material may be appropriate provided that it
produces comparable image quality at equal or reduced dose to
the breast.
- Radiation Beam Quality - There must be radiation attenuating
filters that provide a degree of attenuation such that the first
Half-Value Layer of aluminum is not less or greater than the
values shown in Table 2 for a selected X-ray tube voltage. For
other X-ray tube voltages, the Half-Value Layer of the radiation
beam must be calculated by linear interpolation from that Table.
Half-Value Layer measurements must include the attenuation of
the breast compression device.
Table 2: Acceptable Half-Value Layer for W or Mo-W alloy target with Al filter
X-ray Tube Voltage
(kilovolt) |
First Half-Value Layers
(minimum - maximum)
(millimetre of Al)
|
| Tungsten target |
| 40 |
0.40 - 1.50 |
| 49 |
0.50 - 1.60 |
| Molybdenum-Tungsten target |
| 40 |
0.40 - 0.90 |
| 49 |
0.50 - 1.00 |
- X-ray Tube Voltage - The X-ray tube voltage should be
adjustable in 2 kVp increments. The lowest selectable X-ray tube
voltage shall be less than or equal to 40 kVp. The peak X-ray
tube voltage should correspond to within 5 percent of the
selected or indicated value. The X-ray tube voltage reproducibility
should be within 2 percent.
- X-ray Tube Radiation Output - The X-ray tube radiation output
shall be high enough to minimize irradiation time to eliminate
perceptible motion artifacts.
For any combination of operating loading parameters, the
coefficient of variation of any ten consecutive radiation exposure
measurements, taken at the same source to detector distance
within a time period of one hour, is no greater than 0.05, and each
of the ten radiation exposure measurements is within 15 percent
of the mean value of the ten measurements.
- Automatic Exposure Control - An Automatic Exposure Control
(A.E.C.) system should be provided. The system should be able
to maintain the dose to the cassette to within 15 percent of the
nominal value for the range of breast thicknesses examined and
for all types of technique and loading factors used by the facility.
In addition, a density control should be provided with each
increment increasing or decreasing cassette dose by 25 percent.
When manual irradiation control is used, the selectable control
interval (time or mAs) shall be small enough to permit increments
of 25 percent.