Radiation Protection in Mammography: Recommended Safety Procedures for the Use of Mammographic X-Ray Equipment - Safety Code 33
9. Procedures for minimizing
dose to personnel
The guidelines and procedures outlined in this Section are
primarily directed toward occupational radiation protection.
Adherence to these will also, in many instances, provide protection to
visitors and other individuals in the vicinity of a facility. The
recommendations for safe work practices and procedures should be
regarded as a minimum, to be augmented with additional requirements
when warranted, to cover special circumstances in particular
facilities.
To achieve optimum safety, radiology technologists must make
every reasonable effort to keep irradiation of themselves and other
personnel as far as practicable below the limits specified in
Appendix I.
9.1 General recommendations
- A room where mammographic X-ray examinations are done
must not be used for more than one radiological investigation
simultaneously.
- Except for those persons whose presence is essential for the
investigation, all other persons must leave the room when a
mammographic X-ray examination is carried out.
- Personnel must, at all times, keep as far away from the radiation
beam as practicable. Radiation exposure of personnel by the
primary X-ray beam must never be allowed unless the beam is
adequately attenuated by protective screens or protective
clothing. Deliberate irradiation of an individual for training
purposes or equipment evaluation must never occur.
- All personnel must take full advantage of available protective
devices.
- Operation of the X-ray tube shall be controlled from the control
panel located behind a protective screen or inside a control booth.
The technologist must be shielded when exposures are made.
- If a patient escort or other person is called upon to assist, this
person must be provided with protective clothing and be
positioned so as to avoid the primary X-ray beam. No one must
regularly perform these duties.
- All technologists operating X-ray equipment and who are likely
to receive a radiation dose in excess of 5 percent of the
recommended dose limits for radiation workers specified in Appendix I, must wear personnel dosimeters.
- All entrance doors to a mammographic X-ray room, including
patient dressing room doors, must be kept closed while a patient
is in the X-ray room. There should be a light or a sign on the
outside of the patient dressing room indicating occupancy.
- Powered-on mammographic X-ray equipment must not be left
unattended.
- Where radiation doses in excess of 5 percent of the recommended
effective dose limits for radiation workers specified in
Appendix I are regularly received by any one person, appropriate
remedial steps must be taken to improve techniques and
protective measures.
- Mammographic X-ray equipment must be operated only by
individuals who are properly trained for the equipment and the
procedures being performed.
- Technologists must have a clear view of the patient during every
mammographic X-ray examination and must be able to
communicate with the patient and/or attendants.
10. Procedures for minimizing
dose to patients
The largest single contributor of man-made radiation exposure
to the population is dental and medical diagnostic radiology. In total,
such use of X-radiation accounts for more than 60 percent of the total
man-made radiation dose to the general population.
The risk to the individual patient from a single radiographic
examination is very low. However, the risk to a population is
increased by increasing the frequency of radiographic examinations
and by increasing the number of persons undergoing such
examinations. For this reason, every effort should be made to reduce
the number of radiographs and the number of persons examined
radiographically, as well as to reduce the dose involved in a particular
examination. In the case of mammographic screening programs, the
benefit from the program must outweigh the risk from an increase of
radiation dose to the group being targeted by the program.
To accomplish this reduction, it is essential that patients be
subjected to only necessary radiological examinations and, when a
radiological examination is required, it is essential that patients be
protected from excessive irradiation during the examination.
The recommendations and procedures for the protection of the
patient, outlined in this Section, are directed toward the physician,
the radiologist and the technologist. They are intended to provide
guidelines for elimination of unnecessary radiological examinations
and for minimizing doses to patients when radiological examinations
are indicated.
10.1 Guidelines for the prescription of
diagnostic mammography
The medical practitioner is in a unique position to reduce
unnecessary radiation dose to the patient by eliminating examinations
which are not clinically justified. The practitioner can achieve this by
adhering, as much as possible, to certain basic recommendations.
These recommendations are presented below.
- The request for a mammographic X-ray examination of a patient
should be based on a clinical evaluation of the patient and should
be for the purpose of obtaining diagnostic information.
- It should be determined whether there have been any previous
mammographic X-ray examinations which would make further
examination unnecessary or allow for an abbreviated examination.
The previous mammograms should be examined along
with a clinical evaluation of the patient.
- When a patient is transferred from one physician or hospital to
another, any relevant mammograms or reports must accompany
the patient and should be reviewed by the consulting practitioner.
- When prescribing a mammographic X-ray examination, the
physician should specify precisely the clinical indications and
information required.
- The number of mammographic views required in an examination
must be kept to the minimum practicable, consistent with the
clinical objectives of the examination.
- In prescribing mammographic X-ray examinations of pregnant
or possibly pregnant women, full consideration must be taken of
the consequences of foetal irradiation. While it is generally
accepted that the radiation dose to the ovaries and the foetus is
low in mammography, the radiation beam should not irradiate
the abdominal area.
- In prescribing mammographic X-ray examinations of patients
having breast implants, it should be noted that there may be a
need for the use of special compression, positioning and loading
techniques, and that the personnel of mammographic X-ray
facility should be proficient in performing such procedure.
- Repeat mammographic X-ray examinations should not be
prescribed only because a mammogram may not be of the "best"
diagnostic quality if the mammogram contains the required
diagnostic information.
- The quality of mammograms must be monitored routinely,
through a Quality Assurance program, to ensure that they satisfy
diagnostic requirements with minimal patient dose.
- A patient's clinical records must include details of all mammographic
X-ray examinations carried out.
10.2 Guidelines for screening mammography
In breast cancer screening programs, asymptomatic women
undergo mammographic X-ray examinations with the goal to reduce
breast cancer death by detecting cancer tumours at an early stage. In
such programs, it is important to minimize participant exposure to
radiation consistent with obtaining images of optimal diagnostic
quality. Therefore, a mammography screening program should not be
established unless mean glandular doses are within accepted limits
and that a Quality Assurance program is implemented.
- Selection of population groups for mammographic screening
should be based on the concept that the benefit from the program
should outweigh any risks from an increase of radiation dose to
the group being targeted by the program.
- Mammographic screening should not be done on pregnant or
possibly pregnant women because of the consequences of foetal
irradiation. The mammographic X-ray examination should be
re-scheduled at a subsequent date.
- Participants having breast implants should follow the same
mammographic screening schedule as recommended forwomen
without implants. However, for these participants, there may be
a need for the use of special compression, positioning and
loading techniques, and the personnel of mammographic X-ray
facility should be proficient in performing such procedures.
- The number of mammographic views required in an examination
must be kept to the minimum practicable, consistent with
screening program objectives.
- Repeat mammographic X-ray examinations should not be
prescribed only because a mammogram may not be of the "best"
diagnostic quality if the mammogram contains the required
diagnostic information.
- The quality of mammograms must be monitored routinely,
through a Quality Assurance program, to ensure that they satisfy
diagnostic requirements with minimal patient dose.
- For mobile mammography screening clinics, it is recommended
that image processing be performed on site so that technologists
can review their films which will reduce participant callbacks.
However, since it is often difficult to stabilize film processors in
mobile mammography screening clinics, additional care must be
taken to ensure that image processing is optimized. In the
situation where image processing cannot be optimized, batch
processing at another location is acceptable.
- Previous mammograms, including baseline mammograms, from
screening mammography programs should be available to the
radiologist for examination.
10.3 Guidelines for the carrying out of
mammographic X-ray examinations
Next to elimination of unnecessary X-ray examinations, the most
significant factor in reducing dose is ensuring that an examination is
performed with good methodology and proper X-ray tube loading. It
is possible, for example, to obtain a series of diagnostically acceptable
mammograms and have the organ dose vary widely due to the choice
of loading factors. It is the responsibility of the technologist, the
medical physicist and the radiologist to be aware of this and to know
howto carry out a mammographic X-ray examination with the lowest
possible radiation exposure to the patient or breast screening
participant.
The recommendations that follow are intended to provide
guidance to the technologist, the medical physicist and radiologist in
exercising their responsibility towards reduction of patient dose.
- The mammographic X-ray system must be designed specifically
for mammography and the image receptor must be compatible
with the system.
- A film-screen combination that provides good quality diagnostic
results must be used. Direct exposure film must never be used in
mammography.
- Except in the case of mammography performed within a
screening program, the technologist must not perform any
examination which has not been prescribed by a physician
responsible for the patient.
- The dose to the patient must be kept to the lowest practicable
value consistent with clinical objectives, and without loss of
essential diagnostic information. To achieve this, techniques
appropriate to the equipment available should be used and
evaluated from time to time in terms of effectiveness.
- Particular care in patient X-ray protection must be taken when
mammographic X-ray examinations of pregnant or possibly
pregnant women are carried out, even though the radiation dose
to the abdomen and fetus are negligible during normal mammographic
X-ray examinations.
- There may be a need to use special procedures when performing
mammographic X-ray examinations of patients with breast
implants such as extra views, modified positioning, compression
and loading techniques. Particular care of compression techniques
must be taken since excessive compression of the implants
during mammographic X-ray examination may cause rupture of
the implant.
- In mammography, it is recommended that the field normally be
the full size of the image receptor, but not larger than the image
receptor support, except at the chest wall. The amount of
unexposed areas on the films should be minimal so to avoid the
need for masking. Collimated-down views are useful in some
purposes, but should be considered as a special procedure.
- Appropriate compression must be used in all mammographic
procedures.
- Adequate X-ray beam filtration must be used in all mammographic
procedures.
- The technologist should use the maximum focal spot to skin
distance consistent with good radiographic loading. For mammographic
procedures, distances of less than 50 cm should not
be used. For mammographic procedures including magnification,
distances of less than 25 cm to the entrance surface of the
breast shall not be used.
- Full details of the mammographic procedures, including retakes,
carried out should be noted on the patient's clinical records.
- Irradiation times should be minimized to avoid unnecessary dose
increase from reciprocity law failure and to avoid motion
artifacts. This can be accomplished by the use of sufficiently
high tube current values.
- Cassettes should be loaded at least 30 minutes in advance to
allow air to escape, and thus improving film/screen contact.
- The technologist should examine the images after processing in
order to verify that the techniques being used are producing
diagnostic quality images and that the X-ray equipment is
functioning correctly
- An appropriate Quality Assurance program must be
implemented on mammographic X-ray equipment and on film
and Xerographic processing systems.
- While recommended dose limits have been defined for radiation
workers and the general population, there is no specific permissible
level recommended for patients undergoing diagnostic
X-rays procedures. However, it is possible to provide limits on
the amount of radiation breast tissues receive by setting dose
limits on the level of radiation a representative breast phantom
would receives for a given irradiation. These dose limits are
presented in Appendix II.
For patients the risk involved with exposure to radiation must
always be weighed against the medical requirement for accurate
diagnosis. However, the dose to the patient should be kept as low
as reasonably achievable.