In the planning of any medical X-ray facility the main priority is to ensure that persons in the vicinity of the facility are not exposed to levels of radiations which surpass the current regulatory exposure limits. Appropriate steps must be taken to ensure adequate shielding is present to meet the following requirements:
Appendix I provides a detailed description of the regulatory dose limits. For medical X-ray imaging facilities, controlled areas are typically in the immediate areas where X-ray equipment is used such as the procedure room and X-ray control booths. The workers in these areas are primarily equipment operators such as radiologists and radiation technologists who are trained in the proper use of the equipment and in radiation protection. Uncontrolled areas are those occupied by individuals such as patients, visitors to the facility, and employees who do not work routinely with or around radiation sources (NCRP 2004).
In general, attention to the basic principles of distance, time and shielding are required to determine shielding needs.
In the early stages of designing and planning a medical X-ray facility, three steps should be taken to ensure adequate shielding is in place to provide the necessary level of radiation protection:
In order to determine the shielding requirements for an X-ray facility a floor plan must be prepared, clearly identifying the following components:
When designing the layout of the X-ray facility, the following general recommendations must be considered.
The thickness of the shielding material, such as lead, concrete, or gypsum wallboard, required to reduce radiation levels to the recommended dose limits can be determined through calculations. In general, the radiation exposure to individuals depends primarily on the amount of radiation produced by the source, the distance between the exposed person and the source of the radiation, the amount of time that an individual spends in the irradiated area, and the amount of protective shielding between the individual and the radiation source.
The parameters listed below must be considered for the calculation of barrier thicknesses. Allowance should be made for possible future changes in anyone or all of these parameters, including increases in use and occupancy factors, in operating tube voltage and workload, as well as modifications in techniques that may require ancillary equipment.
The workload is a measure of the operational time or the amount of use of the X-ray equipment. A workload distribution indicates the workload across a range of operating voltages. The workload and workload spectrum can be determined by recording the operating voltage and current-time product of each irradiation taken in each X-ray suite over a set period of time (i.e., week). For irradiations made under Automatic Exposure Control, the operating voltage, procedure type and patient thickness should be recorded to be used later to estimate the current-time product. If actual workload values are not available, Table 5 presents estimated total workloads for various medical X-ray facilities (NCRP 2004).
| Total Workload per patient (mA min/patient) | Typical Number of Patients (per 40 hour week) | Total Workload per week (mA min/week) | |||
|---|---|---|---|---|---|
| Average | Busy | Average | Busy | ||
|
a. R and F is a room that contains equipment for both radiography and radioscopy (fluoroscopy). | |||||
| Radiographic Room (chest) | 0.6 | 120 | 160 | 75 | 100 |
| Radiographic Room (other) | 1.9 | 120 | 160 | 240 | 320 |
| Dedicated Chest Room | 0.22 | 200 | 400 | 50 | 100 |
| R and Fa Room (radioscopic system) | 13 | 20 | 30 | 260 | 400 |
| R and Fa Room (radiographic system) | 1.5 | 25 | 40 | 40 | 40 |
| Angiography Room (cardiac) | 160 | 20 | 30 | 3,200 | 4,800 |
| Angiography Room (other vascular) | 64 | 20 | 30 | 1,300 | 2,000 |
The Occupancy factor is the fraction of time that the area under consideration is occupied by the individual (employee or public) who spends the most time at that location while the X-ray equipment is operating. The following table present recommended occupancy factors.
The use factor, is the fraction of the workload during which the X-ray beam is pointed in the direction under consideration. The following table present recommended occupancy factors.
Shielding calculations must be made for both primary and secondary protective barriers. Primary protective barriers provide shielding from the direct X-ray beam and therefore must be placed in such an orientation as to intersect the X-ray beam. Secondary protective barriers are required to provide shielding from scattered and leakage X-rays.
Comprehensive shielding calculations for large radiological facilities should only be performed by individuals with current knowledge of structural shielding design and the acceptable methods of performing these calculations. It is recommended that shielding calculations be performed using the methodology presented in the National Council on Radiation Protection and Measurements (NCRP) Report No. 147: Structural Shielding Design for Medical X-Ray Imaging Facilities (NCRP 2004). However, it must be noted that the shielding design goals specified in NCRP Report 147 are not adopted in this Safety Code. The shielding design goal values may be lower but must not exceed the limits set out in section B1.1 for controlled and uncontrolled areas. Due to the extensiveness of the information, the methodology of NCRP 147, including equations, tables and figures, is not provided in this Safety Code. Alternatively, the methodology presented in NCRP Report No. 49 (NCRP 1976) is also acceptable and presented in Appendix III.
Under the methodology used in NCRP Report 147, the following are assumptions made in the shielding calculation:
The information outlined in sections B1.1 and B1.2 along with the final plans of the installation must be submitted for reviewed by the appropriate responsible government agency. For installations under federal jurisdiction, the responsible agency is the Consumer and Clinical Radiation Protection Bureau, Health Canada, Ottawa, Ontario K1A 1C1. Radiological facilities that fall under provincial or territorial jurisdiction should contact the responsible agency in their respective province or territory listed in Appendix V.
Film storage containers must be adequately shielded to ensure that excessive exposure of film by X- rays does not occur. Sufficient film shielding must be in place to reduce the radiation level to stored film to less than 0.1 mGy over the storage period of the film. The values presented in Appendix IV are very conservative but will protect films from radiation exposure for most circumstances. Once films are loaded into cassettes, radiation exposure levels should be less than 0.5 µGy and the resulting increase in the base-plus-fog should be less than 0.05 O.D. Refer to Appendix IV for storage guides for radiographic film.
Primary and secondary shielding must be provided for radiographic equipment where the tube can be manipulated in several directions. The walls and floor where the X-ray tube can be directed are considered primary barriers whereas the other walls and ceiling are secondary barriers. The primary barrier includes the wall behind the vertical image receptor, or "wall or chest bucky", and the floor under the radiographic table. For dedicated chest radiographic equipment, the wall behind the image receptor is considered a primary barrier.
The X-ray tube should never be directed towards the control booth. Therefore the walls of the control booth are calculated as secondary barriers. The information required for calculation of the shielding of radiographic X-ray equipment and dedicated chest radiographic equipment is found in Table AII.1 of Appendix II.
The design of radioscopic X-ray equipment is such that only secondary shielding must be provided for these types of systems. However, in systems where an X-ray tube for radiography is also present, the shielding for this X-ray tube must be evaluated independently, as in Section B1.3.2. When equipment include more than one X-ray tube, such as in cardiac systems, the shielding calculation must take into account each X-ray tube independently. The information required for calculation of the shielding of radioscopic X-ray equipment is found in Table AII.2 of Appendix II.
The design of computer tomography equipment is such that only secondary shielding must be provided. The calculation of shielding for CT rooms should not rely on workload values as defined in Section B1.2.3 and therefore it is recommended that shielding requirements be calculated using the methodology of NCRP 147 for CT equipment. The information required for calculation of the shielding of CT equipment is found in Table AII.3 of Appendix II.