Laboratory Procedure MFLP-66
June 2006
Help on accessing alternative formats, such as Portable Document Format (PDF), Microsoft Word and PowerPoint (PPT) files, can be obtained in the alternate format help section.
HEALTH PRODUCTS AND FOOD BRANCH
OTTAWA
This method is applicable to the measurement of water activity in foods and food ingredients to determine compliance with the requirements of Sections 4 and 7 of the Food and Drugs Act, Section B.27.001 of the regulations of the Food and Drugs Act and any other federal regulation.
This revised method replaces MFLP-66, dated October 2001.
Water activity plays an important role in the preservation of food and the control of microbial growth, especially pathogens. Decagon AquaLab models CX-2 and Series 3 use the chilled-mirror dewpoint technique to measure the water activity of a product.
See Appendix A of Volume 3, items 1,2,3,4.
Water activity (aw) : is the ratio of the water vapour pressure in any food system to the water vapour pressure of pure water at the same temperature.
Equilibration: When the aw values of any two readings are less than .001 apart, the instrument has reached equilibrium and the measurement process stops.
Humidity standard (slush) : is a saturated solution of a salt of a known % relative humidity at a designated temperature.
Commercial humidity standard: is an unsaturated salt solution of a specific molality representing a know Aw.
See Appendix B of Volume 3.
Keep samples sealed to prevent loss or gain of moisture during transport.